Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Bias and

The main difference between the force-bias and the smart Monte Carlo methods is that the latter does not impose any limit on the displacement that m atom may undergo. The displacement in the force-bias method is limited to a cube of the appropriate size centred on the atom. However, in practice the two methods are very similar and there is often little to choose between them. In suitable cases they can be much more efficient at covering phase space and are better able to avoid bottlenecks in phase space than the conventional Metropolis Monte Carlo algorithm. The methods significantly enhance the acceptance rate of trial moves, thereby enabling Icirger moves to be made as well as simultaneous moves of more than one particle. However, the need to calculate the forces makes the methods much more elaborate, and comparable in complexity to molecular dynamics. [Pg.449]

The "feedback loop in the analytical approach is maintained by a quality assurance program (Figure 15.1), whose objective is to control systematic and random sources of error.The underlying assumption of a quality assurance program is that results obtained when an analytical system is in statistical control are free of bias and are characterized by well-defined confidence intervals. When used properly, a quality assurance program identifies the practices necessary to bring a system into statistical control, allows us to determine if the system remains in statistical control, and suggests a course of corrective action when the system has fallen out of statistical control. [Pg.705]

Hg Cd Te is an example of a ternary detector, in which the value of x controls the cutoff wavelength. Photoconductive detectors are generally simpler to couple to low noise amplifiers photodiodes generally have lower power consumption because these have no external bias, and better high frequency performance (15,16). [Pg.193]

Under Httle or no illumination,/ must be minimized for optimum performance. The factor B is 1.0 for pure diffusion current and approaches 2.0 as depletion and surface-mode currents become important. Generally, high crystal quality for long minority carrier lifetime and low surface-state density reduce the dark current density which is the sum of the diffusion, depletion, tunneling, and surface currents. The ZM product is typically measured at zero bias and is expressed as RM. The ideal photodiode noise current can be expressed as follows ... [Pg.426]

Other polyamides having higher moduli and T than nylon-6 and nylon-6,6 have been evaluated in an effort to reduce wrinkle resistance and eliminate flat-spotting of bias and bias-belted tires (Table 3). Nylons have also been tested extensively over the years for apparel and carpets (Table 4). [Pg.260]

In the category of industrial appHcations, nylon is the predominant fiber used in the carcass of bias tmck, racing car, and airplane tires because of its exceUent strength, adhesion to mbber, and fatigue resistance. Nylon is used less in the carcass of radial tires for automobiles and in replacement bias and bias-belted tires because of the development of temporary flat spots. For this reason, nylon has lost most of this market to polyester. [Pg.261]

This matrix will necessarily be sparse. First, not all measurements can be taken for a given stream or position (e.g., a chromatographic analysis may only measure a subset of the component compositions). Second, not all streams or positions are included. Third, some of the measurements are inadequate due to bias and are discarded. [Pg.2560]

However, other bias errors are so substantial that their presence will significantly distort any conclusions drawn from the adjusted measurements. Rectification is the detection of the presence of significant bias in a set of measurements, the isolation of the specific measurements containing bias, and the removal of those measurements from subsequent reconcihation and interpretation. Significant bias in measurements is defined as gross error in the literature. [Pg.2571]

J Cao, BJ Berne. Monte Carlo methods for accelerating hamer crossing Anti-force-bias and variable step algorithms. J Chem Phys 92 1980-1985, 1990. [Pg.89]

The Mosquito was made almost entirely of wood and for that reason was nearly canceled by the British before it was even built wood aircraft were viewed as fragile and slow based on the previous 40 years of aerospace history. Fortunately the designers of the Mosquito were able to overcome this bias and the Mosquito went on to become one of the best-performing fighter/bombers of the war. [Pg.1135]

Enantioselective processes involving chiral catalysts or reagents can provide sufficient spatial bias and transition state organization to obviate the need for control by substrate stereochemistry. Since such reactions do not require substrate spatial control, the corresponding transforms are easier to apply antithetically. The stereochemical information in the retron is used to determine which of the enantiomeric catalysts or reagents are appropriate and the transform is finally evaluated for chemical feasibility. Of course, such transforms are powerful because of their predictability and effectiveness in removing stereocenters from a target. [Pg.51]

Overall uncertainty of a measuring procedure or of an instrument The quantity used to characterize the uncertainty of results given by an apparatus or a measuring procedure, expressed on a relative basis by a combination of bias and precision, according to a formula. [Pg.1464]

By being independent of the audited activities, the auditor is unaware of the pressures, the excuses, the informal instructions handed down and can examine operations objectively without bias and without fear of reprisals. It is for this reason that it was considered appropriate for the auditor to have no direct responsibility for the work being audited i.e. audits carried out by a manager, supervisor, or foreman of his/her own department or section do not qualify as internal quality audits in ISO 9001 1994. However, they will qualify under ISO 9000 2000. [Pg.516]

FIG. 25 (a) Variation of the mean-square gyration radius Rg with the intensity of the field (bias) and with host matrix density Cob- (b) The same for the ratio of its longitudinal and transversal components (c) The ratio between the end-to-... [Pg.609]

Here the index 0 at Z) refers to the diffusion coefficient in a system at rest (no bias), and Vq denotes the zero approximation i.e., this relation is applicable for slow drift in dilute medium only. [Pg.611]

An additional limit to the size of a passive array relates to the current which flows in an OLED when it is under reverse bias [189]. When a given pixel is turned on in the array, there are many possible parallel paths for the current, each involving two diodes in reverse bias and one forward. Hence, as the number of rows and columns increases, there is a higher level of background emission from non-selected pixels that limits the contrast ratio of the array. As a result, the contrast degrades as N increases. [Pg.239]

The polarography of sulphones has been fairly extensively studied, sometimes with an analytical bias and sometimes with more theoretical interest. It is generally accepted that a two-electron reduction takes place, to sulphinate ... [Pg.108]

Although there have been many experimental and theoretical studies on the behavior of facially perturbed dienes (see below), only a few systematic experiments have been carried out to characterize facially perturbed dienophiles. Dienophiles embedded in the norbomane or norbomene motif have been rather intensively studied [146-150]. In most cases, steric effect controls selectivity, but in some cases the reactions are considered to be free from steric bias, and the selectivity has been explained in terms of other factors, such as orbital effects [151, 152]. [Pg.161]

PROS REJECT jcls Section 3.6, Fig. 1.29 In a production environment there are often several superimposed processes that yield measurement series like that depicted in the lower panel there is drift that unexpectedly changes slope, there is bias and measurement noise, and there are operators who take corrective action. The model includes the probability of drift occurring and a feed-back loop that permits both positive and negative coefficients. The operators can be ordered to react if a single value exceeds a set limit, or only if 2, 3, or more successive values do so. The program calculates the two-sided (asymmetric) total probability of a value being OOS and depicts this in the upper panel on a log(p) scale. The red horizontal is the upper limit on the total risk as set in cell B20. [Pg.398]

Dose, E. V, and Guiochon, G., Bias and Nonlinearity of Ultraviolet Calibration Curves Measured Using Diode-Array Detectors, Anal. Chem. 61, 1989,... [Pg.405]

Randomized, controlled clinical trials reduce bias and variability by a process of selection, randomization and standardization of treatment, and often take place under artificial conditions isolated from those of routine clinical practice (Freemande et al, 1993 Simon et al, 1995b). Yet it is the uncontrolled interactions of a dmg technology with patients, health-care workers and the system of health care that ultimately lead to much of the variability in outcomes and expenditures in clinical practice. Thus the value of RCTs in evaluating cost-effectiveness in clinical practice maybe limited (Reeder, 1995 Simon et al, 1995b Hotopf et al, 1996). [Pg.45]

Differences exist in primary care between the patterns of prescribing fluoxetine, paroxetine or sertraline, which may influence cost outcomes. Sertraline-treated patients are more likely to have their dose increased (Sclar et al, 1995 Donoghue, 1998), and to drop out of treatment prematurely (Donoghue, 1998). The apparent need to titrate doses upwards with sertraline may require more involvement by the clinician and may delay response to treatment, with resultant increases in direct health costs (Sclar et al, 1995). However, these economic findings are retrospective, may suffer from selection bias, and being derived from HMO patients may not be generalizable to other populations confirmation in further studies is required. [Pg.50]

The photovoltage is esentially determined by the ratio of the photo- and saturation current. Since io oomrs as a pre-exponential factor in Eq. 1 it determines also the dark current. Actually this is the main reason that it limits the photovoltage via Eq. 2, The value of io depends on the mechanism of charge transfer at the interface under forward bias and is normally different for a pn-junction and a metal-semiconductor contact. In the first case electrons are injected into the p-region and holes into the n-region. These minority carriers recombine somewhere in the bulk as illustrated in Fig. 1 c. In such a minority carrier device the forward current is essentially determined... [Pg.82]

Historical data on the indicator. Existing information on the statistical variation, bias, and other interpretational attributes of potential biological indicators should be examined and considered in the design of a sampling program for assessing trends in mercury bioaccumulation. [Pg.90]

Fig. 44.9. General structure of an MLF Network, (a) without bias and (b) with a bias neuron. Fig. 44.9. General structure of an MLF Network, (a) without bias and (b) with a bias neuron.
To account for inhomogeneity in bubble sizes, d in Eq. (20-52) should be taken as / Ln,dffLn,d, and evaluated at the top of the vertical column if coalescence is significant in the rising foam. Note that this average d for overflow differs from that employed earlier for S. Also, see Bubble Sizes regarding the correction for planar statistical sampling bias and the presence of size segregation at a wall. [Pg.34]

The result from a measurement on a RM is commonly a difference between the observed value and the certified value. This difference is called measurement bias, and can, appreciating both the uncertainty on the RM as well as the imcertainty added during the measurement, be tested for (statistical) significance. ISO Guide 33... [Pg.9]

Results from the analysis of the RM and the certified value and their uncertainties are compared using simple statistical tests (Ihnat 1993,1998a). If the measured concentration value agrees with the certified value, the analyst can deduce with some confidence that the method is applicable to the analysis of materials of similar composition. If there is disagreement, the method as applied exhibits a bias and underlying causes of error should be sought and corrected, or their effects minimized. [Pg.217]


See other pages where Bias and is mentioned: [Pg.184]    [Pg.420]    [Pg.234]    [Pg.349]    [Pg.371]    [Pg.372]    [Pg.372]    [Pg.374]    [Pg.233]    [Pg.2021]    [Pg.374]    [Pg.129]    [Pg.2]    [Pg.157]    [Pg.111]    [Pg.248]    [Pg.532]    [Pg.231]    [Pg.81]    [Pg.89]    [Pg.122]    [Pg.56]   
See also in sourсe #XX -- [ Pg.357 ]




SEARCH



A Few Words about Hindsight Bias and Examples

Bias and precision

Bias and variance analysis

Bias-Stress Instability and Hysteresis

Biases

Exponential Estimator - Issues with Sampling Error and Bias

Precision, Bias and Accuracy

Skill 1.4 Understanding procedures for collecting and interpreting data to minimize bias

Statistical error and bias

Systematic Errors and Biases

Tunnel-Diodes and Catalytic Bias

Weights and biases

© 2024 chempedia.info