Analysis standard addition

Since a standard additions calibration curve is constructed in the sample, it cannot be extended to the analysis of another sample. Each sample, therefore, requires its own standard additions calibration curve. This is a serious drawback to the routine application of the method of standard additions, particularly in laboratories that must handle many samples or that require a quick turnaround time. For example, suppose you need to analyze ten samples using a three-point calibration curve. For a normal calibration curve using external standards, only 13 solutions need to be analyzed (3 standards and 10 samples). Using the method of standard additions, however, requires the analysis of 30 solutions, since each of the 10 samples must be analyzed three times (once before spiking and two times after adding successive spikes). [Pg.115]

Standardization—External standards, standard additions, and internal standards are a common feature of many quantitative analyses. Suggested experiments using these standardization methods are found in later chapters. A good project experiment for introducing external standardization, standard additions, and the importance of the sample s matrix is to explore the effect of pH on the quantitative analysis of an acid-base indicator. Using bromothymol blue as an example, external standards can be prepared in a pH 9 buffer and used to analyze samples buffered to different pHs in the range of 6-10. Results can be compared with those obtained using a standard addition. [Pg.130]

Construct an appropriate standard additions calibration curve, and use a linear regression analysis to determine the concentration of analyte in the original sample and its 95% confidence interval. [Pg.133]

Bader, M. A Systematic Approach to Standard Addition Methods in Instrumental Analysis, /. Chem. Educ. 1980, 57, 703-706. [Pg.133]

Ophardt, G. E. Acid Rain Analysis by Standard Addition Titration, /. Chem. Educ. 1985, 62, 257-258. [Pg.358]

This experiment describes a method for determining the acidity, reported as an equivalent molarity of H2SO4, of rain water. Because the volume of standard base needed to titrate a sample of rain water is small, the analysis is done by a standard addition. A 10.00-mL sample of nominally 0.005 M H2SO4 is diluted with 100.0 mL of distilled water and standardized by titrating with 0.0100 M NaOH. A second 10.00-mL sample of the sulfuric acid is mixed with 100.0 mL of rain water and titrated with the same solution of NaOH. The difference between the two equivalence point volumes... [Pg.358]

Quantitative Analysis for a Single Analyte The concentration of a single analyte is determined by measuring the absorbance of the sample and applying Beer s law (equation 10.5) using any of the standardization methods described in Chapter 5. The most common methods are the normal calibration curve and the method of standard additions. Single-point standardizations also can be used, provided that the validity of Beer s law has been demonstrated. [Pg.400]

The generalized standard addition method (GSAM) extends the analysis of mixtures to situations in which matrix effects prevent the determination of 8x and 8y using external standards.When adding a known concentration of analyte to a solution containing an unknown concentration of analyte, the concentrations usually are not additive (see question 9 in Chapter 5). Conservation of mass, however, is always obeyed. Equation 10.11 can be written in terms of moles, n, by using the relationship... [Pg.402]

When possible, a quantitative analysis is best conducted using external standards. Unfortunately, matrix interferences are a frequent problem, particularly when using electrothermal atomization. Eor this reason the method of standard additions is often used. One limitation to this method of standardization, however, is the requirement that there be a linear relationship between absorbance and concentration. [Pg.420]

Description of Method. Salt substitutes, which are used in place of table salt for individuals on a low-sodium diet, contain KCI. Depending on the brand, fumaric acid, calcium hydrogen phosphate, or potassium tartrate also may be present. Typically, the concentration of sodium in a salt substitute is about 100 ppm. The concentration of sodium is easily determined by flame atomic emission. Because it is difficult to match the matrix of the standards to that of the sample, the analysis is accomplished by the method of standard additions. [Pg.439]

Another nice example of a multicomponent analysis based on current research is presented in this experiment. Although the H-point standard addition is not discussed in this text, this paper provides adequate theory and references to the original literature. [Pg.448]

Raymond, M. Jochum, C. Kowalski, B. R. Optimal Multicomponent Analysis Using the Generalized Standard Addition Method, /. Chem. Educ. 1983, 60, 1072-1073. [Pg.448]

This experiment demonstrates the application of the generalized standard additions method for the analysis of mixtures of K2Cr207 and KMn04. [Pg.448]

Quantitative Analysis Using the Method of Standard Additions Because of the difficulty of maintaining a constant matrix for samples and standards, many quantitative potentiometric methods use the method of standard additions. A sample of volume, Vx) and analyte concentration, Cx, is transferred to a sample cell, and the potential, (ficell)x) measured. A standard addition is made by adding a small volume, Vs) of a standard containing a known concentration of analyte, Cs, to the sample, and the potential, (ficell)s) measured. Provided that Vs is significantly smaller than Vx, the change in sample matrix is ignored, and the analyte s activity coefficient remains constant. Example 11.7 shows how a one-point standard addition can be used to determine the concentration of an analyte. [Pg.488]

The technique of hydrodynamic modulation voltammetry (HMV), in which the rate of stirring is pulsed between high and low values, is demonstrated in this experiment. The application of HMV for the quantitative analysis of ascorbic acid in vitamin C tablets using the method of standard additions also is outlined. [Pg.535]

Bohman and colleagues described a reverse-phase HPLC method for the quantitative analysis of vitamin A in food using the method of standard additions. In a typical example, a 10.067-g sample of cereal is placed in a 250-mL Erlenmeyer flask along with 1 g of sodium ascorbate,... [Pg.618]

Spike recoveries for samples are used to detect systematic errors due to the sample matrix or the stability of the sample after its collection. Ideally, samples should be spiked in the field at a concentration between 1 and 10 times the expected concentration of the analyte or 5 to 50 times the method s detection limit, whichever is larger. If the recovery for a field spike is unacceptable, then a sample is spiked in the laboratory and analyzed immediately. If the recovery for the laboratory spike is acceptable, then the poor recovery for the field spike may be due to the sample s deterioration during storage. When the recovery for the laboratory spike also is unacceptable, the most probable cause is a matrix-dependent relationship between the analytical signal and the concentration of the analyte. In this case the samples should be analyzed by the method of standard additions. Typical limits for acceptable spike recoveries for the analysis of waters and wastewaters are shown in Table 15.1. ... [Pg.711]

Standards used to constmct a cahbration curve must be prepared such that the matrix of the standard is identical to the sample s matrix because the values of the parameters k and b associated with a linear cahbration curve are matrix dependent. Many areas of chemical analysis are plagued by matrix effects, and it is often difficult to duphcate the sample matrix when preparing external standards. Because it is desirable to eliminate matrix effects, cahbration in the sample matrix itself can be performed. This approach is called the standard addition method (SAM) (14). In this method, the standards are added to the sample matrix and the response of the analyte plus the standard is monitored as a function of the added amount of the standard. The initial response is assumed to be Rq, and the relationship between the response and the concentration of the analyte is... [Pg.427]

Standard addition. The sample is chromatographed before and after the addition of an accurately known amount of the pure component to be determined, and its weight in the sample is then derived from the ratio of its peak areas in the two chromatograms. Standard addition is particularly useful in the analysis of complex mixtures where it may be difficult to find an internal standard which meets the necessary requirements. [Pg.247]

Figure 5.59 Molecular structures of the diarrhetic shellfish poisons (a) pectenotoxin-6 (PTX6) (b) okadaic acid (OA) (c) dinophysistoxin-1 (DTXl) (d) yessotoxin (YTX). Reprinted from J. Chromatogr., A, 943, Matrix effect and correction by standard addition in quantitative liquid chromatographic-mass spectrometric analysis of diarrhetic shellfish poisoning toxins , Ito, S. and Tsukada, K., 39-46, Copyright (2002), with permission from Elsevier Science. Figure 5.59 Molecular structures of the diarrhetic shellfish poisons (a) pectenotoxin-6 (PTX6) (b) okadaic acid (OA) (c) dinophysistoxin-1 (DTXl) (d) yessotoxin (YTX). Reprinted from J. Chromatogr., A, 943, Matrix effect and correction by standard addition in quantitative liquid chromatographic-mass spectrometric analysis of diarrhetic shellfish poisoning toxins , Ito, S. and Tsukada, K., 39-46, Copyright (2002), with permission from Elsevier Science.
Haghighi, B., Maleki, N., and Safavi, A., Standard Additions in Flow-Injection Analysis with Atomic Absorption Spectrometry, Analytica Chimica Acta 357, 1997, 151-156. [Pg.410]

The "method of standard additions" has been employed as a technique for standardization of atomic absorption analyses of metals In biological fluids (13,21) In this procedure, several concentrations of standard analyte are added to samples of the biological fluid to be analyzed The calibration curve which Is obtained after additions of the standard analyte to the biological fluid should parallel that obtained when aqueous standards are analyzed Extrapolation of the standard additions curve back to a negative Intercept on the abscissa furnishes an estimate of the concentration of the analyte In the original sample (21) This technique Is helpful In assessing the validity of methods of trace metal analysis (11,13,58) However, In the author s opinion, the "method of standard additions" Is neither practical nor reliable as a routine method for standardization... [Pg.255]

In Section 8.2.8 we have discussed the standard addition method as a means to quantitate an analyte in the presence of unknown matrix effects cf. Section 13.9). While the matrix effect is corrected for, the presence of other emalytes may still interfere with the analysis. The method can be generalized, however, to the simultaneous analysis of p analytes. Multiple standard additions are applied in order to determine the analytes of interest using many q > p) analytical sensors. It... [Pg.367]

Internal rather than external standards are recommended for quantitative analysis. If standard additions are used for calibration all the parameters listed in this table must be held constant. [Pg.128]

Brandt [200] has extracted tri(nonylphenyl) phosphite (TNPP) from a styrene-butadiene polymer using iso-octane. Brown [211] has reported US extraction of acrylic acid monomer from polyacrylates. Ultrasonication was also shown to be a fast and efficient extraction method for organophosphate ester flame retardants and plasticisers [212]. Greenpeace [213] has recently reported the concentration of phthalate esters in 72 toys (mostly made in China) using shaking and sonication extraction methods. Extraction and analytical procedures were carefully quality controlled. QC procedures and acceptance criteria were based on USEPA method 606 for the analysis of phthalates in water samples [214]. Extraction efficiency was tested by spiking blank matrix and by standard addition to phthalate-containing samples. For removal of fatty acids from the surface of EVA pellets a lmin ultrasonic bath treatment in isopropanol is sufficient [215]. It has been noticed that the experimental ultrasonic extraction conditions are often ill defined and do not allow independent verification. [Pg.80]

Table 4.15 fists the many possibilities for solid sampling for GC analysis. In general, sample preparation should be considered in close conjunction with injection. Robotic sample processors have been introduced for automatic preparation, solvent extraction and injection of samples for GC and GC-MS analyses. Usually, facilities are included for solvent, reagent, and standard additions and for derivatisation of samples. [Pg.182]

RPLC-ELSD has been used for analysis of polyolefin additives [678], Improvements in sensitivity by use of narrow-bore columns were reported. MAE-RPLC-ELSD/UV (273 nm) has been proposed as a universal one-step and/or two-step procedure for the analysis of additives in polyolefins (Section 3.4.5.1), with the reference mixture of standard additives for... [Pg.249]

Table 8.76 shows the main characteristics of voltammetry. Trace-element analysis by electrochemical methods is attractive due to the low limits of detection that can be achieved at relatively low cost. The advantage of using standard addition as a means of calibration and quantification is that matrix effects in the sample are taken into consideration. Analytical responses in voltammetry sometimes lack the predictability of techniques such as optical spectrometry, mostly because interactions at electrode/solution interfaces can be extremely complex. The role of the electrolyte and additional solutions in voltammetry are crucial. Many determinations are pH dependent, and the electrolyte can increase both the conductivity and selectivity of the solution. Voltammetry offers some advantages over atomic absorption. It allows the determination of an element under different oxidation states (e.g. Fe2+/Fe3+). [Pg.670]


See other pages where Analysis standard addition is mentioned: [Pg.117]    [Pg.130]    [Pg.454]    [Pg.456]    [Pg.616]    [Pg.616]    [Pg.811]    [Pg.811]    [Pg.428]    [Pg.393]    [Pg.18]    [Pg.130]    [Pg.151]    [Pg.202]    [Pg.243]    [Pg.309]    [Pg.513]    [Pg.609]   
See also in sourсe #XX -- [ Pg.85 , Pg.88 , Pg.108 , Pg.115 , Pg.116 , Pg.117 , Pg.121 ]