Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Absorption detectors diode array

UHPLC coupled with spectrophotometric detector, diode array detector (DAD) or MS was widely used in the analysis of polyphenolic compounds in biological samples. Polyphenolic compounds are characterized by strong absorption of ultraviolet radiation therefore, the use of a UV (25-28) detector or DAD (32,35,36) allowed their determination after chromatographic separation. [Pg.208]

It is seen that the profile of the combined peaks is perfectly symmetrical and displays no hint that there are two solutes present. Obviously an absorption ratio curve from a diode array detector would quickly disclose the presence of the two components, as would an appropriate changes in mobile phase composition. However, there would be a further clue for the analyst to follow that would give warning of the "duplicity" of the peak. The double peak would be very broad and be inconsistent with the change in peak width of the other solute peaks with retention time. The peak width of a solute increases regularly with retention time but, unfortunately, the relationship is not smooth. There are good reasons for this, but they... [Pg.259]

If the wavelength of maximum absorption of the analyte (Xmax) is known, it can be monitored and the detector may be considered to be selective for that analyte(s). Since UV absorptions are, however, generally broad, this form of detection is rarely sufficiently selective. If a diode-array instrument is available, more than one wavelength may be monitored and the ratio of absorbances measured. Agreement of the ratio measured from the unknown with that measured in a reference sample provides greater confidence that the analyte of interest is being measured, although it still does not provide absolute certainty. [Pg.33]

The aim of all the foregoing methods of factor analysis is to decompose a data-set into physically meaningful factors, for instance pure spectra from a HPLC-DAD data-set. After those factors have been obtained, quantitation should be possible by calculating the contribution of each factor in the rows of the data matrix. By ITTFA (see Section 34.2.6) for example, one estimates the elution profiles of each individual compound. However, for quantitation the peak areas have to be correlated to the concentration by a calibration step. This is particularly important when using a diode array detector because the response factors (absorptivity) may considerably vary with the compound considered. Some methods of factor analysis require the presence of a pure variable for each factor. In that case quantitation becomes straightforward and does not need a multivariate approach because full selectivity is available. [Pg.298]

Polar or thermally labile compounds - many of the more modern pesticides fall into one or other of these categories - are not amenable to GC and therefore LC becomes the separation technique of choice. HPLC columns may be linked to a diode-array detector (DAD) or fluorescence detector if the target analyte(s) contain chromophores or fluorophores. When using a DAD, identification of the analyte(s) is based on the relative retention time and absorption wavelengths. Similarly, with fluorescence detection, retention time and emission and absorption wavelengths are used for identification purposes. Both can be subject to interference caused by co-extractives present in the sample extract(s) and therefore unequivocal confirmation of identity is seldom possible. [Pg.742]

UV detection, diode-array detector (DAD) and fluorescence have been the detection techniques used, coupled to HPLC for the analysis of OTC. UV detection is set at 355 nm [49-51], 350 nm [40], or at 353 nm [52], Using the diode array detector [49] offers advantages that the target peak can be identified by its retention time and absorption spectrum. Compared to UV detection, fluorescence detection is generally more specific and is less interfered by other compounds in the sample matrix [51]. A HPLC method with electrochemical detection has also been suggested recently. Zhao et al. [53] described HPLC with a coulometric electrode array system for the analysis of OTC, TC, CTC, DC, and methacycline (MC) in ovine milk. An amper-ometric detection coupled with HPLC was developed by Kazemifard and Moore [54] for the determination of tetracyclines in pharmaceutical formulations. [Pg.111]

FIGURE 13.9 The HPLC diode array UV absorbance detector. When a mixture component elutes from the column, not only the chromatography peak but the entire UV absorption spectrum for that component can be recorded. [Pg.380]

The first measurement we make when starting a fluorescence study is not usually a fluorescence measurement at all but the determination of the sample s absorption spectrum. Dual-beam differential spectrophotometers which can record up to 3 absorbance units with a spectral range of 200-1100 nm are now readily available at low cost in comparison to fluorimeters. The wide spectral response of silicon photodiode detectors has made them preeminent over photomultipliers in this area with scan speeds of a few tens of seconds over the whole spectral range being achieved, even without the use of diode array detection. [Pg.378]

The use of a fixed wavelength UV detector for liquid chromatographic separations was first described by Horvath and Lipsky in 1966 [1], and is possibly the most popular HPLC detector in general use today. Although other detection techniques are more sensitive, the UV detector provides a simple and universal answer to the majority of HPLC applications [2]. Developed in 1982, the diode array UV detector measures the full absorption spectrum of each analyte peak, and was a... [Pg.207]

Detector Technology. For copolymer composition analysis the new diode array UV/vis detectors are extremely attractive the absorption at many wavelengths are instantaneously recorded there is only a single spectrophotometer cell so that transport time delays between detectors and axial mixing in detector cells do not confound comparison of detector response at different wavelengths and for styrene copolymers, extremely low concentrations of polymer can be detected. [Pg.65]

Figure 3.14—Principle of the diode array detector. The flow cell is irradiated with a polychromatic UV/Vis light source. The light transmitted by the sample is dispersed by reflection on a grating and the reflected intensities are monitored by an array of photodiodes. Several hundred photodiodes can be used, each one monitoring the mean absorption of a narrow band of wavelengths (i.e. 1 nm). Figure 3.14—Principle of the diode array detector. The flow cell is irradiated with a polychromatic UV/Vis light source. The light transmitted by the sample is dispersed by reflection on a grating and the reflected intensities are monitored by an array of photodiodes. Several hundred photodiodes can be used, each one monitoring the mean absorption of a narrow band of wavelengths (i.e. 1 nm).
High-Performance Liquid Chromatography. A Varian 5060 delivery system was used for this work with detection by UV absorption. Either a Varian UV-50 variable wavelength detector or a Hewlett Packard 1040A scanning diode array detector was used. All HPLC columns were packed in our laboratory (10) with 5-/um particle size Spherisorb-ODS, Spherisorb-CN (Phase Separations), or 8-pm particle size Zorbax-CN (Dupont Ltd). HPLC columns (20 or 25 cm X 4.6 mm i.d.) were coupled via short lengths of stainless steel capillary tubing (5 cm X 0.25 mm i.d.). Separation conditions were as follows ... [Pg.641]

The spectral characteristics of a standard can be monitored during HPLC using a diode-array detector (unitfu). A directory of standard spectra can be stored, enabling additional identification of sample peaks. The actual absorption maxima and fine structure will be dependent on the composition of the mobile phase (see Fig. F2.2.4). Peak I may only occur as a shoulder with civ-carotenoids. while an additional peak is observed at around 340 nm (see Fig. 2.2.1). [Pg.854]

For coumarins in orange fruits (115), the HPLC used a Zorbax Rx C8 (250-mm X 4.6-mm ID, 5 fim) column maintained at 25°C, and analysis was performed by binary-gradient elution using 0.1% HOAc in acetonitrile (eluent A) and 0.1% HOAc in HzO (eluent B). In the author s lab, standard coumarins could be separated by isocratic elution on Zorbax Rx C8 column with acetonitrile-0.1% HOAc in water (35 65) at 1.0 ml/min, as presented in previous work (1). The eluate from the column was passed to a UV detector (UV 330 nm) and then into a fluorescence detector (excitation at 340 nm, emission at 425 nm). As for the specificity, some of the coumarins do not have native fluorescence. Nine coumarins are separated under UV 330 nm, and three coumarins could not be detected with fluorescence detection. Detailed conditions for coumarin analysis in foods and absorption spectra of coumarins obtained by online diode array detector with HPLC were presented by Lee and Widmer (1). Since coumarins exhibit strong absorption in the ultraviolet region, absorption at approximately 313 nm has been used to estimate the dilution of cold-pressed lemon oil with distilled oil (12). Analysis of umbelliferone (7-hydroxy-coumarin) and scopoletin (6-methoxy-7-hydroxycoumarin) in citrus fruits was performed using... [Pg.807]

Figure 1. Experimental system used in the time resolved absorption measurements. (EL=excimer laser, KrF, 248nm DG=delay generator OMA=optical multichannel analyser MC=monochromator and gated diode array detector C=cell X=xenon flash lamp L=lenses )... Figure 1. Experimental system used in the time resolved absorption measurements. (EL=excimer laser, KrF, 248nm DG=delay generator OMA=optical multichannel analyser MC=monochromator and gated diode array detector C=cell X=xenon flash lamp L=lenses )...
Rosania et al. [19] have shown this straightforward preparation to be advantageous for the transposition to a diversity-oriented, combinatorial approach to an organelle-targeted fluorescent library. Therefore, the condensation of 9 and 10 (Fig. 5.4) with pyrrolidine as a catalyst was performed in 96-well plates and the dehydration reaction was accelerated by microwave irradiation for 5 min to give 10-90% conversion. The resulting library compounds were analyzed using an LC-MS system with diode-array and fluorescence detectors and a fluorescence plate-reader to determine the absorption and emission maxima and the emission colors. [Pg.185]


See other pages where Absorption detectors diode array is mentioned: [Pg.3]    [Pg.584]    [Pg.316]    [Pg.432]    [Pg.660]    [Pg.265]    [Pg.272]    [Pg.489]    [Pg.248]    [Pg.224]    [Pg.179]    [Pg.68]    [Pg.821]    [Pg.379]    [Pg.539]    [Pg.104]    [Pg.42]    [Pg.233]    [Pg.33]    [Pg.877]    [Pg.697]    [Pg.723]    [Pg.8]    [Pg.127]    [Pg.74]    [Pg.74]    [Pg.811]    [Pg.139]    [Pg.563]    [Pg.17]    [Pg.120]    [Pg.519]    [Pg.59]   
See also in sourсe #XX -- [ Pg.459 ]




SEARCH



Absorption detectors

Absorptive Detectors

Array detectors

Detectors diode array detector

Diode array detector

Diode arrays

Diode detectors

© 2024 chempedia.info