Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Validation tests

The CamuS system is currently in the form of a laboratory prototype and is undergoing a series of validation tests using an extensive set of test-pieces covering a range of geometries and classes of defect which has been manufactured for the purpose. [Pg.772]

Dyad fractions are more sensitive, but must be examined over a wide range of compositions to provide a valid test. [Pg.500]

If you conduct trials on parts and materials to prove reliability or durability, these can be considered to be verification tests. For example, you may test metals for corrosion resistance or hinges for reliability in the laboratory and then conduct validation tests under actual operating conditions when these items are installed in the final product. [Pg.263]

If the answer in both cases is "yes," then consider the locd climate. Local cooperation—or lack of it—may he the single most important determinant of site selection, since without it you cannot field a valid test. However if you have kept facility managers informed or involved as the PSM process has gone forward, you should have some ideas about which of them would be par-ticulariy enthusiastic—or otherwise. In addition, having identified site-specific benefits may help you win the support of a manager who might otherwise be reluctant to participate. [Pg.148]

We have said that every time the calibration analyzes a new unknown sample, this amounts to an additional validation test of the calibration. It can be a major mistake to believe that, just because a calibration worked well when it was being developed, it will continue to produce reliable results from that point on. When we discussed the requirements for a training set, we said that collection of samples in the training set must, as a group, be representative in all ways of the unknowns that will be analyzed by the calibration. If this condition is not met, then the calibration is invalid and cannot be expected to produce reliable results. Any change in the process, the instrument, or the measurement procedure which introduces changes into the data measured on an unknown will violate this condition and invalidate the method If this occurs, the concentration values that the calibration predicts for unknown samples are completely unreliable We must therefore have a plan and procedures in place that will insure that we are alerted if such a condition should arise. [Pg.24]

Design Validation Testing of the device under intended use conditions, field trials, clinical trials... [Pg.181]

Table 4.42. Raw Data from Method Validation Tests... Table 4.42. Raw Data from Method Validation Tests...
Once agreement and consensus have been reached by all infimal decision units, for X p to be declared an active decision policy, it is necessary to bring it to the attention of DUq. The effects of a possible implementation of Xdp on the system as a whole are examined, to check that it also translates into improved performance from a global perspective. If this final validation test is passed, X p represents an active decision policy, and one can proceed to its implementation. [Pg.146]

All the controls may be conducted either before, or in parallel with, the test itself, providing that the same batches of media are used for both. If the controls are carried out in parallel with the tests and one ofthe controls gives an unexpected result, the test for sterility attempt is recorded as invalid, and, when the problem is resolved, the test is recommenced as if for the first time. It is important to recognize that the terms recommenced and retest have different meanings. A retest may, under certain circumstances, be performed when the first (and, exceptionally, even the second) valid test shows signs of product contamination. [Pg.449]

Cross-validation test The values of q for these QSAR models are from 0.549 to 0.972. The high values of q validate the QSAR models. From the literature, it must be greater than 0.50 [73,74]. [Pg.69]

Validate the current version of the system according to a formal validation test plan. [Pg.1061]

Model Predictions vs. Field Observations The Model Validation/Testing Process... [Pg.151]

The acceptance criterion for recovery data is 98-102% or 95-105% for drug preparations. In biological samples, the recovery should be 10%, and the range of the investigated concentrations is 20% of the target concentrations. For trace level analysis, the acceptance criteria are 70-120% (for below 1 ppm), 80-120% (for above 100 ppb), and 60-100% (for below 100 ppb) [2]. For impurities, the acceptance criteria are 20% (for impurity levels <0.5%) and 10% (for impurity levels >0.5%) [30], The AOAC (cited in Ref. [11]) described the recovery acceptance criteria at different concentrations, as detailed in Table 2. A statistically valid test, such as a /-test, the Doerffel-test, or the Wilcoxon-test, can be used to prove whether there is no significant difference between the result of accuracy study with the true value [29],... [Pg.252]

But to pose the question more directly how can we tell if any set of samples constitute a valid test set Even if they were chosen in a proper random manner, are there any independent tests for their validity What characteristics should the criteria for deciding be based on, and what are the criteria to use ... [Pg.136]

Control limit, n - for validation tests, the maximum difference allowed between a valid analytical result, and a reference method result for the same sample. A measured value that exceeds a control limit requires that action be taken to correct the process. Control limits are statistically determined. [Pg.510]

Validation test, n - a test performed on a validation sample that demonstrates that the result produced by the instrument or analytical method and the result produced by the reference method are equivalent to within statistical tests. [Pg.512]

If the validated test method requires 1 g of material but only 100 mg is available, you must find out if the method is sufficiently robust to stand this amount of scaling down. This has to be checked before the analysis starts, i.e. the method must be validated for analysis of 100 mg of material. Even if the method of analysis is found to be robust, scaling down is only a viable option if the smaller test portion size remains representative, within acceptable limits. This will depend on the homogeneity of the material. [Pg.35]

Several classic pharmacogenetic traits are listed in Table 8.1 and discussed below. For a variety of reasons, mostly reflecting the slow penetration of genetic thinking into clinical medicine and the limited predictive value of even the best-validated tests, these tests have been significantly underutilized. [Pg.142]

An analytically sensitive test always detects analytes when they are present in specimens. An analytically specific test does not detect analytes when they are absent. The Task Force also recommended that laboratories providing a test for routine clinical use (after it had been developed) demonstrate their ability to provide analytically valid tests. These laboratories are regulated under the Clinical Laboratories Improvement Amendments of 1988 (Holtzman, 2000). [Pg.182]

If the scope of mass spectrometry is limitless, why are the applications of clinical MS almost completely small molecules The answer is that most clinical tests analyze small molecules, biomarkers that are either metabolites or steroids and, hence, mass spectrometers would target those first. Perhaps a more complete answer would also include that methods must be very robust, easily reproduced in different labs, reliable, and subjected to an extensive array of validation tests. Although peptide and protein analysis is increasing rapidly in clinical labs, the MS approaches to these assays is lagging behind somewhat. MS techniques targeting these peptides and proteins exist, but they are primarily in the research stage, with few systems and methods subjected to the clinical rigors of validation. Once the necessary validations occur and methods simplified, it will only be a short time before MS is used routinely in clinical proteomics. [Pg.289]

Various agencies are currently validating tests designed to test for the smallpox virus in the environment. [Pg.355]

In order to utilize TLC as a reliable technique, the sample must be readily soluble in organic solvents such as methanol, isopropanol, or mixtures of methanol and other additives such as ammonium hydroxide and hydrochloric acid. For any type of quantitative testing to be done on a pharmaceutical sample, it must be soluble at a concentration of = 25 mg/ml. If sonication is necessary to disperse the sample, it may be used. The sample must also be stable in organic solvents for at least 1 h ideally. Validation testing should be conducted throughout the duration of this test to ensure the sample stays in solution, and does not degrade in the solvent of choice. [Pg.422]


See other pages where Validation tests is mentioned: [Pg.57]    [Pg.19]    [Pg.23]    [Pg.268]    [Pg.48]    [Pg.983]    [Pg.265]    [Pg.265]    [Pg.207]    [Pg.183]    [Pg.227]    [Pg.229]    [Pg.237]    [Pg.369]    [Pg.74]    [Pg.746]    [Pg.433]    [Pg.460]    [Pg.97]    [Pg.439]    [Pg.509]    [Pg.229]    [Pg.229]    [Pg.476]    [Pg.507]   
See also in sourсe #XX -- [ Pg.199 ]




SEARCH



Test validity

© 2024 chempedia.info