Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytics providers

In a titrimetric method of analysis the volume of titrant reacting stoichiometrically with the analyte provides quantitative information about the amount of analyte in a sample. The volume of titrant required to achieve this stoichiometric reaction is called the equivalence point. Experimentally we determine the titration s end point using a visual indicator that changes color near the equivalence point. Alternatively, we can locate the end point by recording a titration curve showing the titration reaction s progress as a function of the titrant s volume. In either case, the end point must closely match the equivalence point if a titration is to be accurate. Knowing the shape of a titration... [Pg.357]

Another equally important consideration before development of a determinative or confirmatory method is an understanding of the chemical properties of the analyte. Such an understanding becomes the cornerstone of a successful method since the unique chemical properties of each analyte provide the basis for isolation and detection schemes. Table 1 lists some of the important chemical properties that could be considered. For example, knowing the or p/fb of an analyte could influence the choice of a liquid-liquid extraction scheme, solid-phase extraction (SPE) cartridge, mobile phase pH, or mass spectrometric ionization. Knowing the overall polarity of the analyte can be very helpful in the evaluation of an extraction or separation. Currently, computational methods are available to obtain an estimate of the logP... [Pg.301]

When F is not known, using both Eqs. 3.22 and 3.62 to compute groundwater concentrations of analytes, provides an investigator with defined limits of the range of potential analyte concentrations. [Pg.75]

The molecular ion peak directly provides valuable information on the analyte. Provided the peak being of sufficient intensity, in addition to mere molecular mass, the accurate mass can reveal the molecular formula of the analyte, and the isotopic pattern may be used to derive limits of elemental composition (Chaps. 3.2 and 3.3). Unfortunately, the peak of highest m/z in a mass spectrum must not necessarily represent the molecular ion of the analyte. This is often the case with El spectra either as a result of rapidly fragmenting molecular ions or due to thermal decomposition of the sample (Chaps. 6.9 and 6.10.3)... [Pg.263]

An ideal detector of any type (Table 25-3) is sensitive to low concentrations of every analyte, provides linear response, and does not broaden the eluted peaks. It is also insensitive to changes in temperature and solvent composition. To prevent peak broadening, the detector volume should be less than 20% of the volume of the chromatographic band. Gas bubbles in the detector create noise, so back pressure may be applied to the detector to prevent bubble formation during depressurization of eluate. [Pg.571]

The confirmation of a compound s identity is only one half of the overall confirmation procedure quantitative confirmation is the other half. Compound concentrations calculated from analyses on two columns or two detectors must be in agreement. The EPA recommends a 40 percent difference (calculated as the RPD shown in Equation 1, Table 2.2) as a threshold value for making decisions on the presence or absence of a compound (EPA, 1996a). This means that the concentrations obtained from two columns or two detectors that agree within 40 percent indicate the presence of an analyte, provided that the retention time confirmation criterion has been also met. [Pg.227]

Ion-pair extraction and IPC were combined to analyze phosphoric acid mono- and diesters originating from the microbial hydrolysis of flame retardants. Even tertiary treatment did not ensure complete removal of the studied compounds detected in municipal wastewater [107], Chlorophenols extracted from water samples as anionic chlorophenolates were studied by IPC because the anionic forms of these analytes provide better UV ultraviolet absorption than uncharged chlorophenol based on their auxochromic effects. IPC conditions yielded adequate retention of the charged analytes and good sensitivity [108]. [Pg.166]

Sirius-Analytical—provider of instruments for LogP and pKa. determination, and the Absolv program to predict physicochemical properties (123)... [Pg.389]

Despite the apparent complexity, the kinetics of this reaction can be solved analytically provided one applies the steady state approximation to each of the three biradicals The result is ... [Pg.1037]

Based on the most widely accepted meanings of leaching and digestion, the former is always desirable when possible. Partial dissolution of the sample with complete removal of the target analytes provides a less complex liquid and the possibility to avoid interferences. It should be noted, however, that complete dissolution of the analytes must be ensured if the target analytes are to be quantified with adequate precision. Special care must be exercised in applying a method developed with spiked matrices to natural samples — rather than with real samples as opposed to prepared or artificial samples since the opposite of real samples would be imaginary samples . [Pg.8]

When the extracted analytes provide no response on passing through a flow-cell located in a detector connected on-line to the continuous extractor via a dynamic manifold, the extractor outlet can also be connected to a manifold where the extract is merged with a stream of an appropriate reagent to derivatize the analytes as they are extracted, thereby enabling their subsequent on-line determination. This approach has been used to extract selenium from sand and sludge [57], and selenium, arsenic and mercury from coal. In the latter application, on-line derivatization with sodium tetrahydroborate (for selenium and arsenic) and tin chloride (for mercury) allowed the analytes to be determined in a direct manner using the atomic fluorescence technique [46]. [Pg.268]

With respect to ANIS, LC-MS does not pose additional selection rales compared to other techniques. The ANIS should be as much alike as possible. Therefore, a small difference in aliphatic substitution is generally preferred over changes in conjugation, or changes in the number or nature of the polar groups. With an ANIS, proper attention should be paid to the possibility of common fragment or adduct ions. Almost co-elution of ANIS and analyte provides the best results. [Pg.294]

Photometric linearity is tested using a set of standards with known relative transmittance or reflectance, depending on the application. Repeated measurements on stable standards give information about the long-term stability of an instrument. For analyte systems with absorbance levels below 1.0, at least four reference standards are to be used in the range of 10 to 90%. When analytes provide absorbance values that exceed 1.0, the proposed USP chapter recommends adding a 2 or 5% standard, or both. [Pg.118]

Two other types of analytical methods are based on mass. In gravimetric titrimetry, which is described in Section 13D, the mass of a reagent, of known concentration, required to react completely with the analyte provides the information needed to determine the analyte concentration. Atomic mass spectrometry uses a ma.ss. spectrometer to separate the gaseous ions formed from the elements making up a sample of matter. The concentration of the resulting ions is then determined by measuring the electrical current produced when they fall on the suiface of an ion detector. This technique is described briefly in Chapter 28. [Pg.314]

Analytics provide estabhshed tools to describe and verify the quality of drugs. Apart from pure quahty control topics, reliable analytical methods are important to assess the pharmacokinetic profile of a drug and its metaboHtes. In addition, such methods often also help in the correlation of pharmacokinetic and pharmacodynamic parameters for the prediction of a drug s therapeutic effects. [Pg.1558]

Titrations with precipitating agents are useful for determining certain analytes, provided the equilibria are rapid and a suitable means of detecting the end point is... [Pg.346]

Since the bath is made up of harmonic oscillators, it can be eliminated analytically (provided that one is only interested in the reduced density matrix of the reaction coordinate). After performing this elimination, the bath-plus-coupling part H of the Hamiltonian leads to an influence functional 0[cr] in terms of the spins cr, alone [38,39], and the correlation function takes the form... [Pg.53]

Most market participants are familiar with the concept of duration and its effect on the returns from fixed income instruments. However, the difference in the calculations of duration of a bond and a portfolio is worth noting. Most index and analytics providers (including Barclays Capital) calculate index and portfolio duration as the market value weighted duration of individual bonds that constitute the index or portfolio. The other measure, known as cash flow duration, calculates the duration based on the cash flow of the entire portfolio, using the internal rate of return (IRR) of those cash flows and then measuring the sensitivity of portfolio value to change in that IRR. [Pg.808]

Bottom Line The revision of the end of chapter questions in this edition is informed by robust data-driven analytics providing a new level of pedagogically-sound assessments for your students, all while making the time they spend working these problems even more valuable. [Pg.1241]

When the extracted analytes provide no response on passing through a flow cell located in a detector connected online to the continuous extractor via a dynamic manifold, the extractor outlet can also be connected to a manifold where the extract is merged with a stream of an appropriate reagent to derivatize the analytes as they are extracted, thereby enabling their subsequent online determination. [Pg.1199]


See other pages where Analytics providers is mentioned: [Pg.65]    [Pg.34]    [Pg.616]    [Pg.503]    [Pg.299]    [Pg.326]    [Pg.22]    [Pg.171]    [Pg.34]    [Pg.74]    [Pg.364]    [Pg.257]    [Pg.319]    [Pg.791]    [Pg.40]    [Pg.159]    [Pg.1631]    [Pg.103]    [Pg.472]    [Pg.276]    [Pg.1557]    [Pg.295]    [Pg.61]    [Pg.78]    [Pg.65]    [Pg.216]    [Pg.111]    [Pg.43]    [Pg.816]    [Pg.202]    [Pg.329]   
See also in sourсe #XX -- [ Pg.716 , Pg.717 , Pg.718 , Pg.719 ]




SEARCH



Analytical methods providing information

© 2024 chempedia.info