Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Transmittance, with spectrophotometer

The final optically clear solution is evaluated as to intensity of color at 555 m/x, and with slit width at 0.02 to 0.04 mm., on a Beckman spectrophotometer adjusted to 100% transmittance with an untreated sample freshly processed in parallel with the unknown sample. From a standard calibration curve (see Figure 3), the concentration of parathion may readily be ascertained, and an appropriate factor converts this value to micrograms of parathion present in the original field sample. [Pg.75]

T adjustment Adjustment of a spectrophotometer to register 100% transmittance with a blank in the light path. [Pg.1119]

Set the spectrophotometer to 0% transmittance with the tube chamber empty by turning. [Pg.187]

When a spectrophotometer is used it is unnecessary to make comparison with solutions of known concentration. With such an instrument the intensity of the transmitted light or, better, the ratio I,/I0 (the transmittance) is found directly at a known thickness /. By varying / and c the validity of the Beer-Lambert Law, equation (9), can be tested and the value of may be evaluated. When the latter is known, the concentration cx of an unknown solution can be calculated from the formula ... [Pg.650]

All infrared spectrophotometers are provided with chart recorders which will present the complete infrared spectrum on a single continuous sheet, usually with wavelength and wavenumber scales shown for the abscissa and with absorbance and percentage transmittance as the ordinates. More advanced instruments also possess visual display units on which the spectra can be displayed as they are recorded and on which they can be compared with earlier spectra previously obtained or with spectra drawn from an extensive library held in a computer memory. These modern developments have all led to quantitative infrared spectrophotometry being a much more viable and useful analytical procedure than it was just a few years ago. [Pg.747]

Scenario A student determined that the optimal wavelength for the absorbance of FeSCN2+ experiment was 445 nm. Then the student prepared samples of known concentrations of FeSCN2+ ranging from 4.0 x 10 5 M to 1.4 x 10 4 M. The samples were then examined by means of a spectrophotometer and their transmittances recorded. From the transmittance, the absorbance was calculated and graphed. Next, he mixed 5.0 mL of 2.0 x 10 3 M Fe(N03)3 with 5.0 mL of 2.0 x 10 3 M KSCN. This solution was then analyzed in the spectrophotometer and through extrapolation, he was able to determine that the concentration of FeSCN2+ at equilibrium was 1.3 x 10" M. [Pg.288]

Before the concentration could be determined through colorimetry, the student needed to know the wavelength of light that was most absorbed by the complex ion in order to set the spectrophotometer properly. The student calibrated the spectrophotometer by setting the transmittance to 100% with the FeCl3-KCl-HCl solution as a reference. The optimal wavelength was found to be 525 nm (see Figure 1). [Pg.333]

When using a spectrophotometer for a colorimetric analysis, both the 0% and 100% transmittance (oo and 0 absorbance) readings must be set. Once the instrument has warmed up, with the light beam blocked and with nothing in the sample compartment, the readout is set to 0% transmittance (oo abs.). Again, this measurement is done to set / in the absorbance equation shown earlier. A blank, a solution containing all the components used in the analysis except the analyte being measured, is placed in a cuvette, placed in the sample... [Pg.310]

The transmittance values for the calibration and experiment solutions can be measured either with the spectrophotometer or with the pyroelectric probe. In the first case, this reading can obviously be done at any time after the calorimeter is filled with the solution. In the latter case, the value is computed by using equation 13.21. [Pg.201]

Iodine in water also may be determined by the Leucocrystal violet colorimetric method. An aqueous sample is treated with mercuric chloride followed by Leucocrystal violet reagent [4,4 ,4 —methylidynetris(N,N-dimethylani-hne)] in the pH range 3.5 to 4.0. A violet color is produced. The absorbance or transmittance is measured at 592 nm by a spectrophotometer or filter photometer. Iodine concentration is calculated from a standard calibration curve. [Pg.401]

Photometric accuracy is determined by comparing the difference between the measured absorbance of the reference standard materials and the established standard value. Many solid and liquid standards are commonly used to verify the photometric accuracy of a spectrophotometer. An optically neutral material with little wavelength dependency for its transmittance/absorbance is desirable because it eliminates the spectral bandwidth dependency of measurements. The advantages and disadvantages of various commonly used photometric accuracy standards are summarized in Table 10.6. Even for a relatively stable reference standard, the intrinsic optical properties may change over time. Recertification at regular intervals is required to ensure that the certified values of the standards are meaningful and accurate for the intended use. [Pg.165]

Neutral-Density Filters. The empty reference filter holder (air reference) and then filters of various transmittance values at 440, 465, 546.1, 590, and 635 nm are scanned (these wavelengths are selected to minimize interaction between the absorbance and wavelength scales of the spectrophotometer being tested [12]). Compare the results with the values in the certificates. The certified values for the National Institute of Standards and Technology (NIST) SRM 930-e series glass... [Pg.165]

The films contained between 0.1 and 1% residual monomer according to gas chromatographic analysis. The transmittance and fluorescence spectra were measured with a Spectronic 505 spectrophotometer by Bausch and Lomb. To obtain correct fluorescence spectra with respect to energy, the above mentioned apparatus was calibrated using a wolfram-band lamp and an aperture driven by a wavelength-dependent correction... [Pg.592]

In the presence of cadmium, nitrate (N03 ) is reduced to nitrite (N03). The nitrite produced is diazotized with sulfanilamide. This is followed by coupling with N-(l-naphthyl)-ethylenediamine to form a highly colored azo dye. The intensity of the color developed is measured by a spectrophotometer or a filter photometer at 540 nm. The concentration of oxidized N/L (N03-N plus NO AN) is read from a standard curve prepared by plotting absorbance (or transmittance) of standard against N03-N concentrations. [Pg.179]

It has proven feasible to take the electrical output from photocells or phototubes and either with or without amplification record the magnitude and duration of the output. The recording may be made either in the form of a line tracing on a moving chart or may be converted to numerical values and printed by a read-out device. Further refinements can be supplied in which the instrument converts optical densities (or transmittance units) to concentration values. More intricate recording colorimeters or spectrophotometers are also available for continuous scanning and recording of complete spectra from ultraviolet to infrared. No further discussion of these will be attempted since they do not serve a normal function in routine clinical chemistry laboratories. [Pg.331]

The transmittance of a 1 in 100 solution of the test article dissolved in 2 N HC1, and determined in a 1-cm cell at 430 nm with a suitable spectrophotometer, is not less than 0.95 (corresponding to an absorbance of not more than about 0.022). The test solution is prepared by means of sonication. [Pg.37]

Turn on the spectrophotometer and let it warm up for a few minutes. Turn the wavelength control knob to read 340 nm. With no sample tube in the sample compartment, adjust the amplifier control knob so that 0% transmittance or infinite absorbance is read. [Pg.498]

Insert the sample tube with the Blank solution into the spectrophotometer. Adjust the reading to 100% transmittance (or 0 absorbance). This zeroing must be performed every... [Pg.498]

A Mutiple Internal Reflection Accessory for use with Perkin-Elmer infrared spectrophotometers was used to introduce the sample to the IR beam. To correct for losses in energy transmittance through the sample beam resulting from the use of the MIR Accessory, a comb-type reference beam attenuator was employed. A LFE Corpora-... [Pg.68]


See other pages where Transmittance, with spectrophotometer is mentioned: [Pg.1122]    [Pg.1136]    [Pg.389]    [Pg.410]    [Pg.416]    [Pg.101]    [Pg.420]    [Pg.386]    [Pg.1136]    [Pg.489]    [Pg.199]    [Pg.2]    [Pg.505]    [Pg.765]    [Pg.101]    [Pg.109]    [Pg.382]    [Pg.170]    [Pg.1136]    [Pg.424]    [Pg.425]    [Pg.194]    [Pg.173]    [Pg.144]    [Pg.248]    [Pg.253]    [Pg.254]    [Pg.287]    [Pg.749]   
See also in sourсe #XX -- [ Pg.956 ]




SEARCH



Spectrophotometers

Transmittance

Transmittancy

Transmittivity

© 2024 chempedia.info