Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

The determination of concentration

Spectrophotometry, the measurement of the absorption of Kght by a material, is used widely to monitor concentration. The technique is based on Beer s law (see Chapter 12), which states that the incident and transmitted intensities [Pg.219]

I and Iq, respectively, of light passing through a sample of length L are related to the molar concentration [J] of the absorbing species J by [Pg.220]

The molar absorption coefficient e (epsilon) depends on the wavelength of the radiation. Once the value of e has been measured (in a separate experiment) for an absorbing species taking part in a reaction, either as a reactant or a product, its concentration may be monitored by using eqn 6. la in the form [Pg.220]


In a titration, the volume of one solution is known, and we measure the volume of the other solution required for complete reaction. The solution being analyzed is called the analyte, and a known volume is transferred into a flask, usually with a pipet. Then a solution containing a known concentration of reactant is measured into the flask from a buret until all the analyte has reacted. The solution in the buret is called the titrant, and the difference between the initial and the final volume readings of the buret tells us the volume of titrant that has drained into the flask. The determination of concentration or amount by measuring volume is called volumetric analysis. [Pg.111]

Orion Model 95-64). In practice, one simply determines E ntot by calibration with a standard solution without the necessity of knowing the various constants mentioned. The S02 electrode allows the determination of concentrations down to 10 8 Af with a response time of a few minutes. From the above it appears that the gas-sensing electrodes show Nemstian behaviour provided that the concentrations to be measured are not high there is little or no interference by other components in the sample solution. [Pg.86]

Example The results obtained from the determination of concentration of the standard solutions and measurements of corresponding peak areas with a GC are recorded in Table 3.1 and plotted in Figure 3.3 where the former is represented along the x-axis and the latter along the y-axis. How to draw the best straight line through all these points ... [Pg.83]

Richardson, J. F. and Shabi, F. A. Trans. Inst. Chem. Eng. 38 (1960) 33. The determination of concentration distribution in a sedimenting suspension using radioactive solids. [Pg.288]

The analysis of X-ray contrast agents has not been described in too much detail in the hterature. Only scattered data for individual compounds can be found. In the following paragraphs, we will concentrate both on the determination of physicochemical characteristics, which allow for a classification of different contrast agents, e.g. into high and low-osmolar substances, and on the separation from by-products or biological material and on the determination of concentrations. Structural aspects of iodinated contrast agents have been described by Toennessen et al. [81]. [Pg.121]

The investigation of body fluids with respect to nutrient (essential) elements and toxic elements -which are challenging topics for analytical chemistry - include the determination of concentrations at the trace and ultratrace level. However, isotope variation and isotope effects (especially of lighter elements such as hydrogen, carbon, nitrogen, oxygen but also of iron and calcium) have also been studied.22 23 The most frequently applied mass spectrometric technique for the analysis of body fluids today, which fulfils all requirements and also results in accurate and precise data, is ICP-MS. [Pg.345]

Other, more limited approaches may also be employed. (2) When side reactions are unimportant and the products are stable, a pair of nucleophiles may be allowed to compete for a deficiency of alkylating agent. The relative amount of alkylated products, R, is determined when all the alkylating agent is consumed R is [MeHet,]/[MeHet2], the molar ratio of methylated products. The rate-constant ratio then is calculated with the aid of Eq. (18).62,148 This special approach allows a rate-constant ratio to be calculated using a product ratio instead of concentrations of individual products. It is conveniently applied when, say, it is hard to find an internal reference standard in NMR methods of analysis, thereby making the determination of concentrations difficult. [Pg.111]

EN (2004a) 13528-2. Diffusive Samplers for the Determination of Concentrations of Gases and Vapors Specific Requirements and Test Methods, The British Standards Institution, London, UK. [Pg.61]

Biosensors differ from bioassays mainly by the fact that in bioassays the transducer is not an integral part of the analytical system and biosensors can extract quantitative analytical information of single compounds in complex mixtures. One example is the determination of concentrations of dioxin-like compounds in the blood and environmental samples using the Calux assay, where within a complex matrix its levels are determined with great accuracy (see, e.g., Murk et al. 1997). Additionally, compounds that are difficult to detect (e.g., surfactants, chlorinated hydrocarbons, sulfophenyl carboxylates, dioxins, pesticide metabolites) can more easily be evaluated using biosensors. [Pg.146]

Another application of atomic absorption is in the determination of concentrations on steel surfaces after special sample preparation, and the analysis of steel residues (purity tests) after isolation and possible selective dissolution of the iron matrix [18, 124, 139] (Fig. 1). Atomic absorption is particularly useful for environmental analysis where dust samples can be analysed in a similar manner to steel residues water and effluents are the main examples. [Pg.213]

The three basic experimental features of gas-phase kinetic studies are temperature control, time measnrement, and the determination of concentrations. Of these, the principal problem is that of following the composition changes in the system. Perhaps the most generally applicable technique is the chemical analysis of aliqnots however, continuons methods are much more convenient. By far the easiest method is to follow the change in total pressure. This technique will be used in the present experiment. Obvionsly the pressure method is possible only for a reaction that is accompanied by a change in the niunber of moles of gas. Also the stoichiometry of the reaction should be straightforward and well understood, so that pressure changes can be related directly to extent of reaction. [Pg.291]

Ideally one would wish to measure the activities of all the reacting species in equation (4) but it is not possible to determine activities of single ionic species (at best, mean ionic activities may be determined in some cases) and the determination of activities of uncharged species is fraught with difficulties. However, modem experimental methods (see Section 3) can readily be applied to the determination of concentrations of at least some of the species concerned, leading to stoichiometric stability constants defined by... [Pg.4547]

It is also seen from Figure S-b that the extents of Ds contribution to D evaluated on LCB method almost coincides with those for DR and CMBR, implying that LCB method can be applied for the determination of concentration dependency of D . [Pg.252]

Excitation and detection geometry, filter selection and electronic settings of the PMT/gated integrator are kept the same as in the LIF measurements. For absolute calibrations, the iZi(9) and /Zi(12) transitions of CH at 387.42 and 388.15 nm were selected [see Fig. 7(a)]. The CN calibration was performed using the Pi 2(10) transition at 388.11 nm. For the determination of concentration profiles of CH, the Ri(9) line was used. For CN, relative LIF intensity profiles talmn from transitions of the unresolved P(0,0)-bandhead at 388.44 nm were compared with profiles taken with the Pi 2(10) line. This comparison showed no difference within the error limits and therefore the bandhead was chosen in order to obtain a better signal to noise ratio. [Pg.221]

Potentiometry is a method of electroanalytical measurement in which the equilibrium voltage of the cell consisting of an indicator electrode and a proper reference electrode is measured using a high-impedance voltmeter, i.e., effective at zero current. The potential of the indicator electrode is a function of particular species present in solutions and their concentration. By judicious choice of electrode material, the selectivity of the response to one of the species can be increased, and thus, interferences from other ions can be minimized. The method allows the determination of concentrations with detection limits of the order of 0.1 pmol per liter, although in some cases, as little as lOpmol differences in concentration can be measured. [Pg.1502]

In writing these expressions, it is implicitly assumed that the observations ctij are subject to random errors and that the reaction times (/y—t,o) are free from random errors. The assumption that the random errors occur solely in a is reasonable since, in the majority of kinetic studies, time can be measured with much greater precision than concentration. When the random errors in the determination of time are commensurate with the random errors in the determination of concentration, the above equations are no longer valid we shall not discuss this situation in view of the difficulties which arise but would suggest that, if possible, every effort should be made to avoid it by redesigning the experimental procedure. [Pg.371]

Many workers have used iodometric methods for the determination of concentrations of ozone in the range of several per cent by volume and higher. They have investigated the stoichiometry by comparison of the amounts of iodine liberated with the amounts of ozone determined by physical measurements of gas density or pressure change. Thus Lechner (5) found that both neutral and alkaline (0.2A potassium hydroxide) potassium iodide (0.2M) absorbed ozone efficiently and yielded the same amount of iodine, equivalent to one oxygen atom in the ozone molecule. This... [Pg.93]

When the residence time becomes shorter, this approach becomes questionable for several reasons. For example, the asymptotic state may not have been reached yet, or the peaks may be unsymmetrical. These "short-time" situations may be encountered when trying to apply chromatographic concepts to the study of dispersion in connecting tubes, or in some apphcations, such as hollow-fiber liquid chromatography. Shankar and Lenhoff [77] have derived a solution in the time domain, using series expansion. This solution can be implemented by numerical computation for the determination of concentration profiles inside a tube coated with a retentive layer, when the fluid flow is laminar. This solution is valid for systems that are either short or long after the Taylor-Aris definition. [Pg.325]

Homogeneous EIA lend themselves better to dose-response curve linearization than solid-phase EIA. This type of EIA is generally used for the determination of concentrations of small molecules (haptens, drugs). The major supplier of this type of EIA kits (Syva Co.) recommends the plotting of the response variable ( EMIT Units ) vs the log of the concentration or the presentation of the results on special non-linear graph paper provided by them, which is, in fact, logit-log paper. [Pg.413]

In the determination of concentrations ranging from I0 -10 % the error is usually... [Pg.43]

In the free volume theory (Williams et al., 1955), a binary system is characterized using the properties of pure components. In the present work a simple modification is introduced. It is based on the determination of concentration dependence of hole free volume using properties of the mixed system, rather than individual components. This modification is used to account for the effect of on the free volume in the moisturized soy flour matrix. [Pg.595]


See other pages where The determination of concentration is mentioned: [Pg.151]    [Pg.58]    [Pg.95]    [Pg.463]    [Pg.65]    [Pg.275]    [Pg.616]    [Pg.256]    [Pg.329]    [Pg.300]    [Pg.146]    [Pg.700]    [Pg.16]    [Pg.256]    [Pg.662]    [Pg.101]    [Pg.278]    [Pg.37]    [Pg.2479]    [Pg.2519]    [Pg.1511]    [Pg.2]    [Pg.93]    [Pg.43]    [Pg.276]    [Pg.224]    [Pg.380]   


SEARCH



Concentration determine

Concentration, determination

Determination of concentration

Determining concentration

© 2024 chempedia.info