Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Differential-mode

Figure E-3 Common-mode and differential noise models (a) common mode (b) differential mode. Figure E-3 Common-mode and differential noise models (a) common mode (b) differential mode.
Figure E-4 A complete third-order, input EMI filter (common-mode and differential-mode). Figure E-4 A complete third-order, input EMI filter (common-mode and differential-mode).
Sometimes the high-frequency attenuation is insufficient to meet the specifications and a third pole needs to be added to the EMI filter. This filter is typically a differential-mode filter and will share the Y capacitors from the common-mode filter. Its corner frequency is typically the same as the commonmode filter. This filter is made up of a separate choke on each power line, and is placed between the input rectifiers and the common-mode filter. [Pg.248]

The differential-mode filter should have a lower damping factor than the common-mode because the combined damping response of the entire filter section would be too sluggish if higher damping factors were used. A damping factor of a minimum of 0.5 is acceptable. [Pg.248]

The addition of this stage of filtering will bring the very high-frequeney attenuation under eontrol and further attenuate any differential-mode noise on the earth ground lead. It will also produee a eombined attenuation of -36 dB at the switehing frequency of the power supply. [Pg.249]

Steady-state reactors with ideal flow pattern. In an ideal isothermal tubular pZi/g-yZovv reactor (PFR) there is no axial mixing and there are no radial concentration or velocity gradients (see also Section 5.4.3). The tubular PFR can be operated as an integral reactor or as a differential reactor. The terms integral and differential concern the observed conversions and yields. The differential mode of reactor operation can be achieved by using a shallow bed of catalyst particles. The mass-balance equation (see Table 5.4-3) can then be replaced with finite differences ... [Pg.295]

Figure 5.31. An Auger spectrum from a stainless steel surface (a) the undifferentiated N(E) mode, (b) the differentiated mode. (After Flewitt and Wild 1985.)... Figure 5.31. An Auger spectrum from a stainless steel surface (a) the undifferentiated N(E) mode, (b) the differentiated mode. (After Flewitt and Wild 1985.)...
Fig. 23. (a) X-t scan of an STM for Au(100) in 0.05 M H2SO4 + 0.4 mM CuS04 after stepping the potential from +500 mV to -250 mV vs. SCE (see arrow). The picture is shown in the differential mode (shaded) for a better contrast, (b) Cross section along the white line to-ti. The first ten copper layers (including the upd) form a smooth film on the substrate. With the deposition of the eleventh layer a striped structure appears, which is maintained during further deposition [78],... [Pg.135]

Apparatus and Procedure. The kinetic studies of the catalysts were carried out by means of the transient response method (7) and the apparatus and the procedure were the same as had been used previously (8). A flow system was employed in all the experiments and the total flow rate of the gas stream was always kept constant at 160 ml STP/min. In applying the transient response method, the concentration of a component in the inlet gas stream was changed stepwise by using helium as a balancing gas. A Pyrex glass tube microreactor having 5 mm i.d. was used in a differential mode, i.e. in no case the conversion of N2O exceeded 7 X. The reactor was immersed in a fluidized bed of sand and the reaction temperature was controlled within + 1°C. [Pg.165]

All modern heat flow calorimeters have twin cells thus, they operate in the differential mode. As mentioned earlier, this means that the thermopiles from the sample and the reference cell are connected in opposition, so that the measured output is the difference between the respective thermoelectric forces. Because the differential voltage is the only quantity to be measured, the auxiliary electronics of a heat flux instrument are fairly simple, as shown in the block diagram of figure 9.3. The main device is a nanovoltmeter interfaced to a computer for instrument control and data acquisition and handling. The remaining electronics of a microcalorimeter (not shown in figure 9.3) are related to the very accurate temperature control of the thermostat and, in some cases, with the... [Pg.141]

The system was calibrated with water and the sound path (A) was found to increase slightly as the pressure is increased (assuming t is independent of pressure) (112). Since we are interested in measuring the relative speed of sound (Au) in aqueous solutions, we use the velocimeter in a differential mode. The value of t is not a function of pressure (or temperature), as the electronics are kept under ambient pressures. The values of Au were determined from... [Pg.599]

Kinetic data are frequently acquired in continuous reactors rather than batch reactors. These data permit one to determine whether a process has come to steady state and to examine activation and deactivation processes. These data are analyzed in a similar fashion to that discussed previously for the batch reactor, but now the process variables such as reactant flow rate (mean reactor residence time) are varied, and the composition will not be a function of time after the reactor has come to steady state. Steady-state reactors can be used to obtain rates in a differential mode by maintaining conversions small. In this configuration it is particularly straightforward to vary parameters individually to find rates. One must of course wait until the reactor has come to steady state after any changes in feed or process conditions. [Pg.79]

Both types of Bourdon gauge are most suitable for use with corrosive gases and both can be used most effectively as null-point instruments. Several types of mechanical gauge are available commercially which use electrical capacitance or induction to magnify the mechanical movement of a membrane. Such gauges are easily operated in a differential mode and can be used for measuring pressure differences down to ca. 10 Torr. [Pg.50]

UV-VIS-NIR diffuse reflectance (DR) spectra were measured using a Perkin-Elmer UV-VIS-NIR spectrometer Lambda 19 equipped with a diffuse reflectance attachment with an integrating sphere coated by BaS04. Spectra of sample in 5 mm thick silica cell were recorded in a differential mode with the parent zeolite treated at the same conditions as a reference. For details see Ref. [5], The absorption intensity was calculated from the Schuster-Kubelka-Munk equation F(R ,) = (l-R< )2/2Roo, where R is the diffuse reflectance from a semi-infinite layer and F(R00) is proportional to the absorption coefficient. [Pg.237]

Several electrolytic-conductivity detectors are produced (Table 3.5). The Laboratory Data Control Model 701 Conducto Monitor (Fig.3.59) may be operated in either a differential mode or an absolute mode. It provides direct readout in units of specific conductance and differences as small as 0.01% in the differential mode between the carrier and the carrier plus solute can be measured. The dynamic range of linearity is 0.01-100,000 pSl 1 /cm. The detector can function in solvents ranging from distilled water to concentrated salt solutions without the necessity of changing the cell. The volume of the cell is 2.5 pi, and the nominal cell constant is 20 cm-1. This type of detector is of use mainly in high-speed ion-exchange chromatography for the detection of ionic species. [Pg.99]

Microadsorption detectors (Table 3.5) have not been used to any large extent for column monitoring. These detectors operate by measuring the heat of adsorption of a solute as it passes over a temperature sensor. The thermistor may be located in the analytical column or in a smaller chamber containing an adsorbent and through which the effluent flows. These detectors are operated in a differential mode and changes as small as 0.001°C can be easily detected. The main reasons for the lack of use of such detectors... [Pg.99]

A novel capillary electrophoresis method using solutions of non-crosslinked PDADMAC is reported to be effective in the separation of biomolecules [211]. Soil studies conducted with PDADMAC report the minimization of run-off and erosion of selected types of soils [212]. In similar studies, PDADMAC has found to be a good soil conditioner [213]. The use of PDADMAC for the simultaneous determination of inorganic ions and chelates in the kinetic differentiation-mode capillary electrophoresis is reported by Krokhin [214]. Protein multilayer assemblies have been reported with the alternate adsorption of oppositely charged polyions including PDADMAC. Temperature-sensitive flocculants have been prepared based on n-isopropylacrylamide and DADMAC copolymers [215]. A potentiometric titration method for the determination of anionic polyelectrolytes has been developed with the use of PDADMAC, a marker ion and a plastic membrane. The end-point is detected as a sharp potential change due to the rapid decrease in the concentration of the marker due to its association with PDADMAC [216]. [Pg.176]

Similar to the sample delivery in a reactor operated in differential mode, the sample is injected into the steady laminar carrier flow in the channel which moves at... [Pg.112]

Shapiro DA, Kristiansen K, Kroeze WK, Roth BL. Differential modes of agonist binding to 5-hydroxytryptamine(2A) serotonin receptors revealed by mutation and molecular modeling of conserved residues in transmembrane region 5. Mol Pharmacol 2000 58 877-886. [Pg.57]

Figure 3.7 Pulse polarograms for 10"4 M Cd11 ion (a) normal and (b) differential modes. Figure 3.7 Pulse polarograms for 10"4 M Cd11 ion (a) normal and (b) differential modes.

See other pages where Differential-mode is mentioned: [Pg.122]    [Pg.231]    [Pg.242]    [Pg.244]    [Pg.246]    [Pg.248]    [Pg.249]    [Pg.42]    [Pg.43]    [Pg.59]    [Pg.247]    [Pg.249]    [Pg.50]    [Pg.472]    [Pg.137]    [Pg.286]    [Pg.428]    [Pg.45]    [Pg.328]    [Pg.258]    [Pg.269]    [Pg.32]    [Pg.82]   
See also in sourсe #XX -- [ Pg.230 , Pg.231 , Pg.232 , Pg.233 , Pg.240 , Pg.245 ]

See also in sourсe #XX -- [ Pg.230 , Pg.231 , Pg.232 , Pg.233 , Pg.240 , Pg.245 ]

See also in sourсe #XX -- [ Pg.230 , Pg.231 , Pg.232 , Pg.233 , Pg.240 , Pg.245 ]




SEARCH



Basic Speed Differential between Parallel and Serial Modes

Conducted EMI Differential-mode

Differential Mode Attenuation

Differential Mode Delay

Differential pulse-mode polarography

Effective modes differential equations

Single-Mode Differential-Type Constitutive Equations

© 2024 chempedia.info