Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standards, in calibration

We can summarise some other ideas for evaluating the predictive ability of the PLS model. First, you can compare the average error (RMSEP) with the concentration levels of the standards (in calibration) and evaluate whether you (or your client) can accept the magnitude of this error (fit-for-purpose). Then, it is interesting to calculate the so-called "ratio of prediction to deviation", which is just RPD=SD/SEP, where, SD is the standard deviation of the concentrations of the validation samples and SEP is the bias-corrected standard error of prediction (for SEP, see Section 4.6 for more details). As a rule of thumb, an RPD ratio lower than 3 suggests the model has poor predictive capabilities [54]. [Pg.222]

The success of studies on ocean chemistry usually hinges on the availability of adequate analytical methodology and expertise. The nature of the chemical problems involved the size and heterogeneity of the ocean the extremes of concentration, composition, temperature, and pressure and the dynamic interactions at its interfaces with the lithosphere, atmosphere, and biosphere create challenging and often unique analytical problems. Current analytical methods are limited, and improved capability in referee, routine, monitoring, and in situ analyses is needed. Valid sampling, the use of reliable standards in calibration, and the collaborative evaluation of methods under real-world conditions are important. [Pg.1]

Samples with concentrations higher than the highest standard in calibration curve should be diluted and reanalyzed. Otherwise, a higher standard concentration should be included in the calibration method and the samples should be requantified. [Pg.638]

Figure 10.7 Calibration curve for internal standard method where A is analyte and IS is internal standard in calibration samples. Figure 10.7 Calibration curve for internal standard method where A is analyte and IS is internal standard in calibration samples.
Standardless analysis, 323 Standards, in calibration, 11, 17 States, virtual, 495,606 in laser media, 171 in Raman spectroscopy, 482,484 Stationary phase, 762 chiral, 806, 837... [Pg.527]

C = Peak height of internal standard in sample jar D = Peak height of solvent a in calibration jar E = a content of retained solvent mg square metre F = Peak height of internal standard in calibration jar G = Area of sample in square cm. [Pg.213]

Q = weight percent internal standard in calibration mixture, and... [Pg.798]

RFi = response factor relative to the internal standard, A = area of internal standard in calibration mixture, Ai = area of impurity peak in calibration mixture. [Pg.798]

Since a standard additions calibration curve is constructed in the sample, it cannot be extended to the analysis of another sample. Each sample, therefore, requires its own standard additions calibration curve. This is a serious drawback to the routine application of the method of standard additions, particularly in laboratories that must handle many samples or that require a quick turnaround time. For example, suppose you need to analyze ten samples using a three-point calibration curve. For a normal calibration curve using external standards, only 13 solutions need to be analyzed (3 standards and 10 samples). Using the method of standard additions, however, requires the analysis of 30 solutions, since each of the 10 samples must be analyzed three times (once before spiking and two times after adding successive spikes). [Pg.115]

The method of standard additions can be used to check the validity of an external standardization when matrix matching is not feasible. To do this, a normal calibration curve of Sjtand versus Cs is constructed, and the value of k is determined from its slope. A standard additions calibration curve is then constructed using equation 5.6, plotting the data as shown in Figure 5.7(b). The slope of this standard additions calibration curve gives an independent determination of k. If the two values of k are identical, then any difference between the sample s matrix and that of the external standards can be ignored. When the values of k are different, a proportional determinate error is introduced if the normal calibration curve is used. [Pg.115]

An appropriate standard additions calibration curve based on equation 5.8 plots Sspi elVo + Vs) on they-axis and CsVs on the x-axis. Clearly explain why you cannot plot Sspike on the y-axis and Cs[ Vs/( Vo Vj)] on the x-axis. Derive equations for the slope and y-intercept, and explain how the amount of analyte in a sample can be determined from the calibration curve. [Pg.131]

Construct an appropriate standard additions calibration curve, and use a linear regression analysis to determine the concentration of analyte in the original sample and its 95% confidence interval. [Pg.133]

Sensitivity Sensitivity in flame atomic emission is strongly influenced by the temperature of the excitation source and the composition of the sample matrix. Normally, sensitivity is optimized by aspirating a standard solution and adjusting the flame s composition and the height from which emission is monitored until the emission intensity is maximized. Chemical interferences, when present, decrease the sensitivity of the analysis. With plasma emission, sensitivity is less influenced by the sample matrix. In some cases, for example, a plasma calibration curve prepared using standards in a matrix of distilled water can be used for samples with more complex matrices. [Pg.440]

A pH electrode is normally standardized using two buffers one near a pH of 7 and one that is more acidic or basic depending on the sample s expected pH. The pH electrode is immersed in the first buffer, and the standardize or calibrate control is adjusted until the meter reads the correct pH. The electrode is placed in the second buffer, and the slope or temperature control is adjusted to the-buffer s pH. Some pH meters are equipped with a temperature compensation feature, allowing the pH meter to correct the measured pH for any change in temperature. In this case a thermistor is placed in the sample and connected to the pH meter. The temperature control is set to the solution s temperature, and the pH meter is calibrated using the calibrate and slope controls. If a change in the sample s temperature is indicated by the thermistor, the pH meter adjusts the slope of the calibration based on an assumed Nerstian response of 2.303RT/F. [Pg.492]

Analysis of Standards The analysis of a standard containing a known concentration of analyte also can be used to monitor a system s state of statistical control. Ideally, a standard reference material (SRM) should be used, provided that the matrix of the SRM is similar to that of the samples being analyzed. A variety of appropriate SRMs are available from the National Institute of Standards and Technology (NIST). If a suitable SRM is not available, then an independently prepared synthetic sample can be used if it is prepared from reagents of known purity. At a minimum, a standardization of the method is verified by periodically analyzing one of the calibration standards. In all cases, the analyte s experimentally determined concentration in the standard must fall within predetermined limits if the system is to be considered under statistical control. [Pg.710]

Standard. These solution are made in calibrated tubes, and are compared with a set of standard colour disks. [Pg.615]

From the ventilation point of view, the fixed points -38.83 °C (triple-point of mercury), 0.010 °C (triple-point of water), 29.76 °C (melting point of gallium), and 156.60 °C (freezing point of indium) are of relevance. The triple-point of water is relatively simple to achieve and maintain with a triple-point apparatus. Some freezing point cells are covered in standards. In practical temperature calibration of measuring instruments, the lTS-90 fixed points are not used directly. [Pg.1140]

BS ISO TR 7066-1 1997. Assessment of Uncertainty in Calibration and Use of Flow Measurement Devices Linear Calibration Relationships. British Standards Institution, 1997. [Pg.1175]

The intent is that wherever the calibration is performed, the same standards apply. Calibrating equipment in-house should not absolve you from complying with the same requirements that you would need to impose on an external test house. [Pg.393]

Due to the limitation of space, only the linear calibration curves for TSK, PSS, APSC, and PL PEO standards in water/methanol for the Shodex SB-806M... [Pg.506]

Campbell (6) was perhaps the first to investigate the relationship between tenderometer values for raw peas and the texture characteristics of collateral samples after freezing and cooking. A high degree of correlation was observed. The investigations of Nielsen and co-workers (31) indicate the usefulness of the instrument for estimation of maturity in peas intended for freezing preservation. In order to secure consistent and reproducible results, the instrument requires standardization and calibration, and the measurements must be made under standardized conditions of temperature and interval after harvest. [Pg.31]

In many cases when methods involve internal or external standards, the solutions used to construct the calibration graph are made up in pure solvents and the signal intensities obtained will not reflect any interaction of the analyte and internal standard with the matrix found in unknown samples or the effect that the matrix may have on the performance of the mass spectrometer. One way of overcoming this is to make up the calibration standards in solutions thought to reflect the matrix in which the samples are found. The major limitation of this is that the composition of the matrix may well vary widely and there can be no guarantee that the matrix effects found in the sample to be determined are identical to those in the calibration standards. [Pg.270]

Definition and Uses of Standards. In the context of this paper, the term "standard" denotes a well-characterized material for which a physical parameter or concentration of chemical constituent has been determined with a known precision and accuracy. These standards can be used to check or determine (a) instrumental parameters such as wavelength accuracy, detection-system spectral responsivity, and stability (b) the instrument response to specific fluorescent species and (c) the accuracy of measurements made by specific Instruments or measurement procedures (assess whether the analytical measurement process is in statistical control and whether it exhibits bias). Once the luminescence instrumentation has been calibrated, it can be used to measure the luminescence characteristics of chemical systems, including corrected excitation and emission spectra, quantum yields, decay times, emission anisotropies, energy transfer, and, with appropriate standards, the concentrations of chemical constituents in complex S2unples. [Pg.99]


See other pages where Standards, in calibration is mentioned: [Pg.74]    [Pg.474]    [Pg.54]    [Pg.196]    [Pg.88]    [Pg.83]    [Pg.74]    [Pg.474]    [Pg.54]    [Pg.196]    [Pg.88]    [Pg.83]    [Pg.1248]    [Pg.109]    [Pg.113]    [Pg.124]    [Pg.420]    [Pg.431]    [Pg.586]    [Pg.813]    [Pg.173]    [Pg.56]    [Pg.220]    [Pg.729]    [Pg.1125]    [Pg.413]    [Pg.362]    [Pg.500]    [Pg.565]    [Pg.71]    [Pg.182]   
See also in sourсe #XX -- [ Pg.11 , Pg.17 ]




SEARCH



Standardization calibration standards

© 2024 chempedia.info