Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Separated zone detection

The principle of MS/MS for direct analysis of a multicomponent system is shown in Figure 6.18, in which the first mass spectrometer (MS I) operates with soft ionisation (FI, FD, Cl, LD), and thus produces an ensemble of molecular ions (M + H+, M — H+, or adducts). For identification of molecule ABC only ABC+ is allowed to enter an interface or fragmentation zone for excitation by collisional activation, laser radiation or surface-induced dissociation. Within the time of one vibration (10-13s), ABC+ dissociates into fragments characterising the original molecule. They are separated and detected by MS II [226]. Soft ionisation with FI/FD produces low ion yields, which may be insufficient for MS/MS LVEI (typically at 20 V) can be an alternative. Complete analysis of a multicomponent system is carried out in some 20 min. [Pg.399]

Otsuka, K., Smith, C. J., Grainger, J., Barr, J. R., Patterson, J., Tanaka, N., and Terabe, S. (1998). Stereoselective separation and detection of phenoxy acid herbicide enantiomers by cyclodextrin-modified capillary zone electrophoresis-electrospray ionization mass spectrometry. /. Chromatogr. A 817, 75-81. [Pg.510]

A main control and annunciator panel should be installed when the fire alarm system requires more than a single alarm zone. The panel should be installed in the control room or other continuously staffed location. Separate detection zones should be provided for each distinct fire area and identified by a permanent label. A detailed map of the area should also be provided at the annunciator that identifies which zone relates to which annunciator lamp. Systems with more than ten separate zones should be provided with an electric or electroniczone "mimic" panel showingthe location of all alarms on the graphic display of the platform. Basic arrangements of equipment and system design should be in accordance with NFPA 72. A locked main fire panel and control cabinet should be provided. [Pg.187]

FIGURE 6.12 Schematic view of the CITP separation mechanism. The sample is introduced into the capillary between two electrolyte systems a leading electrolyte (L), having electrophoretic mobility higher than any of the sample components to be separated and a terminating electrolyte (T), having electrophoretic mobility lower than any of the sample components (A). The sample components are separated according to the order of their individual mobility into distinct zones, which are sandwiched between T and L (B). The separated zones move with the same velocity toward the capillary end where they are detected as bands (C). [Pg.200]

Berzas Nevado et al. [138] developed a new capillary zone electrophoresis method for the separation of omeprazole enantiomers. Methyl-/ -cyclodextrin was chosen as the chiral selector, and several parameters, such as cyclodextrin structure and concentration, buffer concentration, pH, and capillary temperature were investigated to optimize separation and run times. Analysis time, shorter than 8 min was found using a background electrolyte solution consisting of 40 mM phosphate buffer adjusted to pH 2.2, 30 mM /1-cyclodextrin and 5 mM sodium disulfide, hydrodynamic injection, and 15 kV separation voltage. Detection limits were evaluated on the basis of baseline noise and were established 0.31 mg/1 for the omeprazole enantiomers. The method was applied to pharmaceutical preparations with recoveries between 84% and 104% of the labeled contents. [Pg.238]

The first two points represent a general motivation for miniaturization in separation science independent of the actual fabrication technology. The benefit of a reduction of the consumption of sample, reagents, and mobile phase in chemical and biochemical analysis is self-evident and does not need to be discussed further (reduced consumption of precious samples and reagents, reduced amounts of waste, environmental aspects). This advantage is, however, sharply contrasted by its severe implications on the detection side, as discussed elsewhere in this volume in detail. The detection of the separated zones of very small sample volumes critically depends on the availability of highly sensitive detection methods. It is not surprising that extremely sensitive laser-induced-fluorescence (LIF) has been the mostly used detection principle for chip-based separation systems so far. [Pg.54]

Altria et al. reported the CE separation and detection of radiopharmaceuticals containing mTc, a 7 emitter with a 6-hour half-life (2, see also 10). Their design involved passing a capillary tube (= 2 cm long) through a solid block of scintillator material and detecting the light emitted as technetium-labeled sample zones traversed the detection volume. Unfortunately, detection limits and detector efficiency were not reported. [Pg.61]

In conclusion, capillary electrophoresis in carbohydrate analysis has advantages in both separation and detection over other techniques of electrophoresis, as well as chromatography. It allows high efficiency (up to a few million plate numbers) and very good sensitivities (up to femtomolar). In addition, CE permits analysis by a variety of separation modes simply by changing the electrolyte (capillary zone electrophoresis, MEKC, CGE). [Pg.306]

Following an electrophoretic run, the band from the tracking dye is often the only visible band. The detection of separated proteins and nucleic acids requires subsequent treatment of the separation pattern for visualization. This treatment may be performed directly on the gel, or may require a blotting step in which the entire separation pattern is transferred onto a thin membrane material. The choice of detection method depends on the concentrations of analytes in the separated zones and whether recovery of the purified sample is required. [Pg.180]

With modem columns and dilute solutions of a strong acid as the eluent, cations may be separated and detected with excellent sensitivity by direct conductivity as well as by suppressed conductivity [2]. The basis for direct conductivity detection is that the highly conductive (equivalent conductance = 350 S cm equi r ) in the eluent is partially replaced by a cation of lower conductance when a sample zone passes through the detector. For example, the equivalent conductance of Li+, Na+, and K+ is 39,50 and 74 S cm e-quiv", respectively. The decrease in conductance on an equivalent basis can be calculated as follows Background + NO3" = 350 + 71 = 421 (S cm equitr ). Sample peaks Li+ +... [Pg.146]

Electrophoresis is the most powerful method available for separation and analysis of complex mixtures of charged biopolmers. This chapter provides an overview of modern electophoresis as a general introduction to the chapters which follow. The basic electrophoretic operating modes and formats for these modes are described. Means for detection of separated zones are reviewed. Finally, an approach to fully instrumental electrophoresis is discussed. [Pg.1]

It should be emphasized that, within the separator, three different regions are present and each has its own regulating behaviour. The regulating functions (equations 10 and 13) are the mathematical expression of this regulating behaviour and locally they cannot be overruled by the electrophoretic process. All changes in electrophoretic parameters, e.g. concentration (conductance), pH and temperature, will be in agreement with the local regulating function. It is obvious to use these parameters for a universal detection of the zones of the various constituents. Photometric and radiometric detectors can be used for specific zone detection. [Pg.204]

Detection of separated zones by radioactivity measurement gained considerable... [Pg.473]


See other pages where Separated zone detection is mentioned: [Pg.202]    [Pg.4]    [Pg.500]    [Pg.351]    [Pg.773]    [Pg.860]    [Pg.533]    [Pg.369]    [Pg.145]    [Pg.239]    [Pg.189]    [Pg.676]    [Pg.172]    [Pg.202]    [Pg.260]    [Pg.184]    [Pg.147]    [Pg.53]    [Pg.437]    [Pg.270]    [Pg.434]    [Pg.538]    [Pg.286]    [Pg.478]    [Pg.761]    [Pg.1494]    [Pg.142]    [Pg.887]    [Pg.268]    [Pg.269]    [Pg.1576]    [Pg.3]    [Pg.17]    [Pg.9]    [Pg.31]    [Pg.868]   


SEARCH



Detection separation

Detection zones

Separated Zones

Separated zone detection electrophoresis

Zone separation

© 2024 chempedia.info