Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Selective calibration

A quantitative procedure should be validated for selectivity, calibration model, stability, accuracy (bias, precision), linearity, and limit of quantification (LOQ). Additional... [Pg.318]

Here is a list of common strategies in selecting calibration procedure and adopting acceptance criteria ... [Pg.293]

Select calibration hierarchy. By selecting a working calibrator, its calibration hierarchy is determined by the available documentation. Attention should also be paid to the calibration and metrological traceability of measurement results for input quantities to a measurement function measured by accessory equipment such as balances, thermometers, and volumetric ware. [Pg.223]

Identification and analysis of a particular form of a substance Selectivity Calibration, MU, ruggedness, (LOD)... [Pg.234]

Selected Calibration Procedures in Flow Analysis In recent decades there have been great developments in various fields of flow analysis, including a number of interesting and specific proposals for the flow mode of analysis calibration. These allow implementation of the various methods of calibration in accordance with procedures that are often completely different from those that analysts use in their everyday work [4, 5]. The advantage of these approaches compared with the procedures existing in traditional analysis relies not only on more efficient, automated implementation of full calibration (and thus also the complete analytical procedure), but also on the opportunities for more efficient use of the registered analytical signals to obtain richer measurement information. [Pg.38]

There is a general agreement that at least the following validation parameters should be evaluated for quantitative procedures selectivity, calibration model (linearity), stability, accuracy (bias, precision) and limit of quantification. Additional parameters which might have to be evaluated include limit of detection, recovery, reproducibility and ruggedness (robustness) [2,4-10,12],... [Pg.2]

Commercial spectrometers now come with powerful computers and software. Most of the newer ICP emission systems provide software that can assist in wavelength selection, calibration, background correction, interelement correction, spectral deconvolution, standard additions calibration, quality control charts, and report generation. [Pg.855]

Ultimately, those sensors are cheaper whose signals follow thermodynamic relations and whose reliability is guaranteed to be long-term stable by their construction and by the materials selections (calibration-free-method). Devices with such sensors are far-reachingly free of maintenance and are checked only temporarily or in the case of striking disturbances. [Pg.423]

Calibration refers to the procedures used for correlating test method ontpnt or response to an amount of analyte (concentration or other quantity). The characteristics of a calibration fnnction and justification for a selected calibration model should be demonstrated dnring SLV and ILS stndies. The performance of a calibration technique and the choice of calibration model (e.g., first-order linear, cnrvifinear, or nonlinear mathematical function) are critical for minimizing method bias and optimizing precision. The parameters of the model are nsnally estimated from the responses of known, pnre materials. Calibration errors can result from failure to identify the best calibration model inaccnrate estimates of the parameters of the model errors in the composition of calibration materials or inadeqnately smdied, systematic effects from matrix components. This section focnses on the critical issne of the traceability and supply of materials used for calibration of marine biotoxin methods. [Pg.36]

GMP regulations encompass every phase of product development (manufacturing, testing, acceptance, etc.), from the purity of the raw materials to the quality of the finished product. Established procedures include staffing and training, quality control, facilities and maintenance, selection, calibration and maintenance of equipment, procurement, testing and storage of raw materials,... [Pg.87]

Liquid junction problems can be avoided by using as reference a solid-state lanthanum fluoride membrane electrode in conjunction with spiked fluoride samples [9j. On details, Moore [46,53] has highlighted the need for careful sampling and subsequent processing by his studies on technique, selectivity, calibration, sample freezing and pH. [Pg.63]

Method validation makes use of a set of tests that both test any assumptions on which the analytical method is based and establish and document the performance characteristics of a method, thereby demonstrating whether the method is fit for a particular analytical purpose. Typical performance characteristics of analytical methods are applicability, selectivity, calibration, trueness, precision, recovery, operating range, limit of quantification, limit of detection, sensitivity. [Pg.539]

Another related problem is that of sample purity. If the unknowns contain a constituent that was not present in the original calibration set, and the contaminant has spectral bands in the region of the selected calibration peak, the accuracy of the predicted quantity will be significantly diminished. [Pg.97]

The majority of all work carried out to date makes use of arrival time analysis, and subsequently a wave velocity value for thickness measurement and for defect location. The wave velocity in a composite material is difficult to use precisely because of the anisotropic characteristics of a composite material. A typical wave or phase velocity profile in polar coordinates is illustrated in Figure 8, showing the nonspherical nature of the wave velocity distribution. The wave velocity distribution can even be more complex for a realistic composite material structure. Despite a possible weakness in the understanding of basic physics and wave propagation in a composite material, thickness measurement can still be carried out by using carefully selected calibration specimens and procedures. Calibration of a test instrument can therefore be carried out for subsequent thickness measurement and defect location analysis in a structure. [Pg.433]

Select calibrations with lowest noise, wavelength shift sensitivity, and offset sensitivity Identify and transfer actual (not nominal) wavelengths and corresponding regression coefficients... [Pg.132]

Selecting calibration equations for routine or research analysis requires familiarity with the NIR forage literature. Of all the applications of NIR analysis, by far the greatest number of papers to date have been generated in regard to forage and grain analyses [47,80-95]. As seen in the references of this chapter, the main theme of the literature has been the analysis or the development of the... [Pg.373]

J. Workman, Near Infrared Spectroscopy A Comparison of Random Versus Spectrally Selected Calibration Sets for Use in Forage Quality Evaluation, in Proc. Forage and Grasslands Conf, Athens, Georgia, Am. For. Grass. Conn., Lexington, KY, pp. 132-136 (1986). [Pg.384]

The use of generously sized, carefully selected calibration sets together with high-grade reference data is mandatory. Elaborate computational procedures appear to add little. [Pg.474]

Tables 2.4 to 2.6 present the confidence intervals (expressed as Cl) obtained when the three selected calibrators suffer from some alteration in their signals and the extrapolation procedure is applied. Figures 2.17 and 2.18 compare the effects of an increment or depletion in the signal on the Cl obtained by extrapolation. The first conclusion is that extrapolation leads to higher Cl than interpolation (see next section for a more comprehensive comparison). Tables 2.4 to 2.6 present the confidence intervals (expressed as Cl) obtained when the three selected calibrators suffer from some alteration in their signals and the extrapolation procedure is applied. Figures 2.17 and 2.18 compare the effects of an increment or depletion in the signal on the Cl obtained by extrapolation. The first conclusion is that extrapolation leads to higher Cl than interpolation (see next section for a more comprehensive comparison).
The accuracy and speed of response of the pH measurement depends on the condition of a hydrated gel layer on the surface of the glass electrode that is only 10 mm thick and the condition of a junction of the reference electrode that provides a physical contact between its internal electrolyte and the process solution. The need for the gel layer and the process junction of the reference to be structurally consistent, clean, and free from contamination has profound implications for the proper selection, calibration, and maintenance of pH meastirements. llie fact that pH sensors perform as well as they do despite exposure to a wide variety of chemicals, temperatures, and pressures is truly remarkable and testimony to the ingenuity of various electrode designs. [Pg.78]

The most basic cabbration is to introduce a selectivity calibration factor for classes of the reactions in kinetic network. [Pg.174]

The last step in the calibration procedure is to export calibration factors back into the main flowsheet. To do this, we select Calibration factors section Analysis Tab. Then we click the Save for Simulation... button to save current calibration factors as Set-1 as shown in Figure 4.87. [Pg.228]


See other pages where Selective calibration is mentioned: [Pg.105]    [Pg.271]    [Pg.152]    [Pg.338]    [Pg.301]    [Pg.29]    [Pg.359]    [Pg.2919]    [Pg.140]    [Pg.867]    [Pg.131]    [Pg.133]    [Pg.269]    [Pg.348]    [Pg.366]    [Pg.373]    [Pg.381]    [Pg.330]   
See also in sourсe #XX -- [ Pg.11 ]




SEARCH



© 2024 chempedia.info