Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample requirements buffers

Procedure. Select a volume of sample requiring less than 15 mL of titrant to keep the analysis time under 5 min and, if necessary, dilute the sample to 50 mL with distilled water. Adjust the pH by adding 1-2 mL of a pH 10 buffer containing a small amount of Mg +-EDTA. Add 1-2 drops of indicator, and titrate with a standard solution of EDTA until the red-to-blue end point is reached. [Pg.326]

Franek et al developed an immunoassay for sulfadimidine, but they found that milk samples required dilution (1 100), and tissues required dilutions of 1 200 to 1 4000. The LOD was 0.02 pg kg from the buffer calibration curve with the IC50 at 0.15 pgkg The assay measured levels in milk from 10 to 100 pgkg with satisfactory precision. In swine muscle, kidney, and liver samples, levels from 20 to 500 pg kg could be measured when 2 g of tissue were homogenized with 20 mL of buffer and then diluted 1 20. [Pg.704]

Silverlight and Jackman developed an immunoassay for levamisole in meat and milk. The LOD in both milk and meat samples was 1 xg kg The assay was applied to milk directly, and muscle samples required only homogenization in the presence of 10-fold of buffer prior to analysis. The linear range of the assay was between 5 and 50 pg kg for meat and between 0.2 and 25 qg kg for milk. The linear range of the assay was below the MRL for milk (10 qg kg ) and meat (50 qg kg ). [Pg.707]

A key requirement is that the autosampler needs to be temperature stabilized so that the viscosity of the buffer remains consistent over time. Changes in sample or buffer temperature will directly affect the sample injected along with the migration time. Temperature stabilization can be accomplished in a variety of ways including oven and a peltier cooler. [Pg.46]

In the separation of biomolecules, sample preparation almost always involves the use of one or more pretreatment techniques. With high-performance liquid chromatography (HPLC), no one sample preparation technique can be appHed to all biological samples. Several techiques may be used to prepare the sample for injection. For example, complex samples require some form of preffactionation before analysis, samples that are too dilute for detection require concentration before analysis, samples in an inappropriate or incompatible solvent require buffer exchange before analysis, and samples that contain particulates require filtration before injection into the analytical instrument. [Pg.118]

The thermal unfolding of proteins is best measured by differential scanning calorimetry, which measures the heat absorbed by a protein as it is slowly heated through its melting transition (Figure 17.1). A solution of about 1 mg of protein in 1 mL of buffer and a separate reference sample of buffer alone are heated electrically.6 The additional current required to heat the protein solution is recorded. As the protein denatures, there is a large uptake of heat because the process is highly endothermic. The temperature at the maximum of the peak is... [Pg.268]

A highly sensitive and specific ELISA for the determination, in different types of water samples, of diclofenac, a commonly used nonsteroidal anti-inflammatory drug (NSAID), has been developed by Deng et al.357 This analyte belongs to the most frequently detected, pharmaceutically active compounds in the water cycle. The immunoassay was able to measure tap water samples directly— respective LOD and IC50 values were 6 and 60 ng E. On the other hand, surface water samples required fivefold dilution and the wastewater samples 10-fold dilution in buffer to be analyzed correctly the LODs were then 20 and 60 ng E, respectively. Recently, the development and validation of a highly sensitive and specific ELISA for the detection of pharmaceutical indomethacin in... [Pg.165]

Apart from limitations imposed by aerosol transport inefficiency, conventional analytical flame spectrometry suffers from three other significant limitations. These are the sample requirement (generally 1-2 ml per determination), the time required to add releasing agents or ionization buffers, and the limited linear and useful working ranges of calibration graphs. Sections 4-7 of this chapter consider how the effects of these restrictions may be minimized. [Pg.69]

However, it must be pointed out that most of these studies cited in Table 3 were performed in blood-perfused models. In buffer-perfused models, PBN adduct formed in coronary effluents could be extracted into toluene within seconds. On the contrary, in blood-perfused models, the whole-blood samples require centrifugation to obtain plasma fractions that are then extracted into solvents. The processing of blood samples involves a delay time of several minutes. Therefore, it can be argued that PBN/ OH would not survive the delay time for sample processing. The actual half-life of PBN/ OH in the blood is, however, not known. [Pg.351]

Protein] Sample preparation/ required buffer interference ... [Pg.22]

Acidity is determined through the acid number, which is the quantity of base, expressed in milligrams of potassium hydroxide per gram of sample, required to titrate a sample in the solvent from its initial meter reading to a meter reading corresponding to a freshly prepared nonaqueous basic buffer solution or a well-defined inflection point as specified in the test method. Test methods include potentiometric titration (ASTM D-66, IP 177) and indicator-indicator titration (ASTM D-974, IP 139) in addition to inorganic acidity (IP 182) and total acidity (IP 1) methods. [Pg.201]

For some samples, the buffer pH or gradient conditions will require modification in order to optimize the results. GE Healthcare offers a HiTrap I EX Selection Kit, which consists of 7 HiTrap 1 mL columns prepacked with different ion exchangers, i.e., strong and weak cation and anion exchangers. It provides a convenient way to find the best exchanger and buffer system for a given protein. [Pg.32]

The sample is dissolved in a solution of high density, usually containing glycerol and applied to the sample wells. To avoid blocking of the gel pores, samples should not contain any solid particles. High salt and buffer concentrations in the sample can interfere with the electrophoretic separation. Hence, these should be kept to a minimum, typically < 50 mM. The amount of sample required depends on the detection method, typical amounts are in the order of p.g. The sample volume depends on the size of the sample well, which could range from p.L to mL. [Pg.61]

After monodispersity tests, the most important physical assay is that of the sample and buffer transmission measurement. These are required for accurate sample-buffer subtractions (below) and for and matchpoint determinations (Section 2.5). Since the neutron transmissions of HjO and H20 range from about... [Pg.183]

Clearly, analysis of a sample requires that it be solubilized. Although it is not always possible to know the physicochemical properties of the sample to be analyzed, it is important to be familiar with at least some of the physical properties of the specific analyte of interest in the sample. Some physicochemical properties may ultimately have to be determined empirically. Different questions should be posed depending on whether the sample involved is lyophilized or already solubilized, as in a biological fluid. In either case, it is important to have some knowledge of the detectability, purity, and stability of the sample/component of interest. A number of questions need to be addressed. How complex is the sample Is the substance of interest a major or minor component What other substances in the sample might interfere with the detection or separation of the analyte of interest What is its X ax Is it thermally stable If structural information is available, what are the pAT values of the ionizable groups If the sample contains proteins or peptides, what pis are involved If the sample is not in solution, there are additional questions that must be considered. Is it soluble in water or low ionic strength buffer at mg/mL concentrations Does it require extremes of pH for solubility and, if so, is it stable at this pH Is solubility enhanced by buffer additives such as urea, methanol. [Pg.23]

Cytochrome was found to be reducible at the DME at an overpotential of ca. -0.26 V for 30 ixM concentration in 0.05 M phosphate buffer, pH 7.0. From 70 to 300 /xM the diffusion current was linear with concentration but the magnitude was much lower than predicted. At a vibrating nickel electrode trypsin-isolated cytochrome bs was reduced at an overpotential of ca. -0.78 V while the detergent-isolated sample required ca, -1.08 V. Under the same experimental conditions, cytochrome c was reduced at an overpotential of ca. -0.62 V. [Pg.331]

A major consideration when using mass spectrometry for analysis of DNA is the formation of Na and K adduct ions, which reduce both resolution and sensitivity. It is therefore very important to reduce the prevalence of these adduct ions during sample preparation [reviewed in ref. 14]. As PCR amplification and restriction digestion require buffered solutions containing cations, it is necessary to purify the samples prior to their introduction into the mass spectrometer. [Pg.80]


See other pages where Sample requirements buffers is mentioned: [Pg.152]    [Pg.152]    [Pg.500]    [Pg.123]    [Pg.384]    [Pg.278]    [Pg.331]    [Pg.897]    [Pg.260]    [Pg.166]    [Pg.254]    [Pg.152]    [Pg.356]    [Pg.331]    [Pg.277]    [Pg.141]    [Pg.309]    [Pg.4]    [Pg.262]    [Pg.356]    [Pg.49]    [Pg.395]    [Pg.86]    [Pg.7]    [Pg.793]    [Pg.1288]    [Pg.1382]    [Pg.143]    [Pg.246]    [Pg.25]    [Pg.127]    [Pg.325]    [Pg.142]    [Pg.2889]   
See also in sourсe #XX -- [ Pg.42 ]




SEARCH



Buffering requirements

Sample buffer

Sample requirement

Sampling requirements

Sampling samples required

© 2024 chempedia.info