Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Response function transmission

For each EA spectrum, the transmission T was measured with the mechanical chopper in place and the electric field off. The differential transmission AT was subsequently measured without the chopper, with the electric field on, and with the lock-in amplifier set to detect signals at twice the electric-field modulation frequency. The 2/ dependency of the EA signal is due to the quadratic nature of EA in materials with definite parity. AT was then normalized to AT/T, which was free of the spectral response function. To a good approximation [18], the EA signal is related to the imaginary part of the optical third-order susceptibility ... [Pg.114]

In a more detailed description of the analyser transmission, one has to distinguish the nominal values of voltages or energies, indicated, for convenience, by a superscript zero, from actual values (without a superscript), because maximum transmission occurs only for matched nominal values U°p and °in. In the more general case of arbitrary values, one has to consider the full response function of the analyser. This instrumental function is shown schematically in Fig. 1.14. [Pg.39]

Figure 1.14 Instrumental response function of an electrostatic energy analyser shown as a function of the spectrometer voltage l/sp. Maximum transmission is achieved at the nominal voltage l/°p, and this maximum value is equal to the luminosity L (left-hand scale) or set to unity (right-hand scale), respectively. For values other than l/°p the response function decreases, and the characteristic fwhm value is indicated. For the relation fU% = °, see... Figure 1.14 Instrumental response function of an electrostatic energy analyser shown as a function of the spectrometer voltage l/sp. Maximum transmission is achieved at the nominal voltage l/°p, and this maximum value is equal to the luminosity L (left-hand scale) or set to unity (right-hand scale), respectively. For values other than l/°p the response function decreases, and the characteristic fwhm value is indicated. For the relation fU% = °, see...
ICP-MS can provide semiquantitative analysis for about 70 elements by using element response functions built into the instrument software and calibration of only a few elements [205,206]. Most elements are more than 90% ionized in the ICP (with the exception of elements with ionization potentials greater than about 8 eV). Ion transmission efficiency is a smooth function of mass. The natural isotopic abundances of the elements are well known. Therefore, it is possible to predict the relative sensitivities of the elements and any isobaric overlaps. [Pg.122]

Nicotinic acetylcholine (ACh) receptors are responsible for transmission of nerve impulses from motor nerves to muscle fibers (muscle types) and for synaptic transmission in autonomic ganglia (neuronal types). They are also present in the brain, where they are presumed to be responsible for nicotine addiction, although little is known about their normal physiological function there. Nicotinic receptors form cation-selective ion channels. When a pulse of ACh is released at the nerve-muscle synapse, the channels in the postsynaptic membrane of the muscle cell open, and the initial electrochemical driving force is mainly for sodium ions to pass from the extracellular space into the interior of the cell. However, as the membrane depolarizes, the driving force increases for potassium ions to go in the opposite direction. Nicotinic channels (particularly some of the neuronal type) are also permeable to divalent cations, such as calcium. [Pg.358]

Thus, in principle if the atmospheric abundance of the absorbing constituents, the response function of the instrument and the surface temperature are known, the temperature profile can be retreived by inverting equation (4.71). Similarly, the concentration of the absorbing species can be derived for a known temperature profile. If the absorption results from a uniformly mixed compound such as CO2 (constant mixing ratio X), the monochromatic transmission can be expressed as... [Pg.196]

Having determined = the remaining task is to try to learn the extent of vibrational excitation in the NO. When the NO/I fluorescence amplitude ratios are (1) measured as a function of NO pressure (2) corrected for detector response, filter transmission, and radiative lifetime and (3) extrapolated to zero NO pressure to correct for any NO selfabsorption, a ratio of 3.4 0.7 is obtained. Generalization of the kinetic scheme to the case where = 4 shows that this fluorescence ratio should be equal to the average number of NO quanta excited per... [Pg.395]

The intensity/energy response function (lERF) of a spectrometer is the product of the area from which photoelectrons are collected, of the transmission function (T), and of the detector efficiency D). With a microfocused X-ray beam, such as in SSX 100/206 spectrometer, the area of collection is defined by the X-ray spot size and is smaller than the acceptance area of the analyzer. The relative intensity of the peaks is sensitive to the position of the sample on the vertical axis. When a broad X-ray source is used, the lERF includes also the variation of the area of collection as the latter may depend on the energy of the photoelectrons detected (cf. the section on Basic Equations). [Pg.206]

If the reflectance of a sample is low, as it is with gaseous samples, e(v), is approximately equal to 1 — r(. Thus, for any sample for which a transmittance spectrum with discrete absorption bands can be measured, the emittance spectmm should yield equivalent information. As a result, qualitative analysis of the components of hot gases by infrared emission spectroscopy can be as easy as it is by transmission spectrometry. The problem of obtaining quantitative information by infrared emission spectroscopy is more difficult, since not only must the temperature of the sample be known if the radiant power from the blackbody is to be calculated, but the instrument response function must also be taken into account [1]. [Pg.364]

Limitations in the model used for the detector response function and absorber transmission. The response of the detector varies with X-ray energy (and also depends on any absorbers used). This can be simulated using a simple physical model of the detector and parameterized expressions for X-ray absorption coefficient. However, for the highest accuracy it is necessary to analyse standard samples and derive a correction factor for each X-ray line of interest. [Pg.754]

The detector characteristic may very well be included in the filter design. For example, an indium arsenide photovoltaic detector, operating at 195 K, has a very sharp cut-off at 3.6 m. In combination with a thin germanium window, a well-defined 1.9-3.6 m response function is obtained. However, with a limited number of substances available for the design of filters based on intrinsic absorption and reflection phenomena other methods must be found to constmct filters where the transmission limits can be set by the scientific objectives and not so much by the absorption properties of available substances such methods are based on the interference principle, to be discussed in Section 5.6, but first we deal with prism spectrometers, gas filters, and pressure modulation. [Pg.190]

FIGURE 5.10 Effects of co-expressed G-protein (G ) on neuropeptide NPY4 receptor responses (NPY-4). (a) Dose-response curves for NPY-4. Ordinates Xenopus laevis melanophore responses (increases light transmission). Ordinates logarithms of molar concentrations of neuropeptide Y peptide agonist PYY. Curves obtained after no co-transfection (labeled 0 jig) and co-transfection with cDNA for Gai6. Numbers next to the curves indicate jig of cDNA of Ga]g used for co-transfection, (b) Maximal response to neuropeptide Y (filled circles) and constitutive activity (open circles) as a function of pg cDNA of co-transfected G g. [Pg.86]


See other pages where Response function transmission is mentioned: [Pg.173]    [Pg.173]    [Pg.77]    [Pg.172]    [Pg.335]    [Pg.40]    [Pg.107]    [Pg.393]    [Pg.40]    [Pg.107]    [Pg.393]    [Pg.172]    [Pg.237]    [Pg.9]    [Pg.269]    [Pg.218]    [Pg.144]    [Pg.78]    [Pg.275]    [Pg.237]    [Pg.1755]    [Pg.485]    [Pg.1088]    [Pg.499]    [Pg.407]    [Pg.296]    [Pg.438]    [Pg.1204]    [Pg.153]    [Pg.519]    [Pg.251]    [Pg.281]    [Pg.387]    [Pg.25]    [Pg.60]    [Pg.68]   
See also in sourсe #XX -- [ Pg.125 ]




SEARCH



Response functions

Transmission function

© 2024 chempedia.info