Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Potentiostatic measuring system

Fig. 4.3 A potentiostatic measuring system WE, working electrode RE. reference electrode CE, counter electrode... Fig. 4.3 A potentiostatic measuring system WE, working electrode RE. reference electrode CE, counter electrode...
Cahan, Nagy and Genshaw examine design criteria for an electrochemical measuring system to be used for potentiostatic transient investigation of fast electrode reactions. They emphasise the importance of co-design of the experimental cell and electronics. [Pg.1125]

B. Prasad and R. Lai, A capacitative immunosensor measurement system with a lock-in amplifier and potentiostatic control by software. Meas. Sci. Technol. 10,1097—1104 (1999). [Pg.280]

A Gamry electrochemical measurements system and a Pine Bi-Potentiostat were used to study the experimental decomposition potential and current response to the applied voltage. The experimental variables were electrolyte flow rate and temperature. Linear sweep voltammetry (LSV) technique was the main method used to study the electrolytic processes. [Pg.252]

Four-electrode system — For electrochemical measurements with the -> interface between two immiscible electrolyte solutions (or the oil/water interface), the four-electrode system is used. In potentiostatic measurements such as cyclic voltammetry, two reference elec-... [Pg.277]

Three-electrode system — A measurement system with a - potentiostat that uses three electrodes - working, -r counter, and - reference. The systems work in such a way that a desired potential is imposed to the working electrode vs. the reference electrode. The current in the cell flows only between the working and counter electrodes. The reference electrode is not loaded with the current, therefore it preserves its potential even under conditions of high current flowing in the cell. The application of the three-electrode system allows also the elimination of -> salt-bridge resistance and consequently the ohmic potential drop which influences the recorded -> voltammograms. Three-electrode systems do not compensate the entire resistance in the cell. See also - electrochemical cell, -> IRU potential drop. [Pg.673]

There are two options for controlling the perturbation to the measured system one is to control the current perturbation then record the voltage response from the system the other is to control the voltage perturbation then record the current response. For a current control measurement, when the leads are all connected the electrical load is set to a DC constant current. The FRA will generate an AC current perturbation and interrupt the fuel cell through a load bank or a potentiostat. The response to the interruption from the fuel cell will enter into the FRA for analysis to obtain the AC impedance spectra. [Pg.131]

Lorenz et al. [3.92-3.94, 3.98, 3.101] demonstrated that potentiostatically measured non-monotonous current transients represent a necessary, but not a sufficient criterion for 2D nucleation and growth. For example, such transients observed in the systems Ag(lll)/Pb, H, CIO4 and Ag(lll)/TT, If, S04 were quantitatively explained by Mefj g charge transfer and bulk diffusion as rate determining steps as illustrated in Fig. 3.48 [3.92, 3.93]. [Pg.121]

Kolb et al. [3.155, 3.224, 3.225] studied experimentally the system Au(lll)/Cu, H, so/. Along with observed discontinuous-like q E) isotherms (Fig. 3.50), non-monotonous current transients were potentiostatically measured (Fig. 3.51). These findings were explained in terms of first order phase transitions related to the peaks in the cyclic voltammograms. [Pg.124]

In a PEC system, IPCE is usually obtained from a chronoamperometry (potentiostatic) measurement. In this system, a bias can be applied between the sample/working electrode versus a counter electrode (2-electrode experiment) or a reference electrode (3-electrode experiment) while measuring the current that arises from subjecting the PEC electrode to monochromatic light at various wavelengths. The difference between the steady state current under monochromatic illumination... [Pg.11]

Cyclic voltammetrlc measurements were made with a BAS CV-27 potentiostat (Bioanalytlcal Systems, West Lafayette, IN). A Ag/AgCl electrode (saturated with KCl) served as the reference. [Pg.290]

A standard potential vs. pH diagram correlates the thermodynamics of the aquocobalamin (4+)-Bi2r (23)-Bi2s (40") system (see Fig. 5). The interconversion between the different oxidation states of B -derivatives can usually be monitored effectively by UV-vis spectroscopy, and the relevant data were obtained from potentiostatic measurements, which were followed by UV-vis spectroscopy [90,94], Within the pH range - 1 to 11 and applied potentials = 0.5 V and -1.2 V vs. SCE, seven solution cobalamins are thermodynamically predominant spanning a range of the three formal oxidation states of Bi2 [90]. [Pg.15]

The polarization curve is measured well, when one can use the polarization measurement system. A typical constituent of the system is shown in Fig. 14. A potentiostat is used for the system. It is the electronic hardware to control a... [Pg.35]

Electrochemical impedance measurement systems used for the analysis of the ac properties of electrochemical cells typically consist of a potentiostat (sometimes called an electrochemical interface) together with a frequency response analyzer (FRA) or a spectrum analyzer, or even a combination of the two. The potentiostat provides buffered connections to the cell under investigation together with circuitry for applying a controlled voltage or current stimulus and for the measurement of the dc properties of the cell. The FRA is connected through the potentiostat to the cell and therefore the bandwidth of the potentiostat is a very important consideration for accurate high frequency analysis. [Pg.168]

There are many apphcations where multichannel impedance measurement systems are particularly useful. The throughput of testing in a laboratory can be increased either by the use of multiplexed systems where multiple cells are connected and are automatically tested in sequence by, for example, a potentiostat and FRA, or by the use of true parallel measurement systems where each cell has access to its own potentiostat and FRA. The parallel system, of course, is the more efficient method since aU cells can be tested simultaneously however the equipment required for this is more expensive since there are separate potentiostats and FRAs on every channel. The introduction of multichannel systems, however, has seen a reduction... [Pg.181]

A potentiostat-galvanostat system with high accuracy potential and current measurement capability. [Pg.569]

Currents obtained at microelectrodes are generally in the nanoamp or picoamp range. These measurement conditions place severe demands on the measuring system. Namely, high gain and wide bandwidth at a reasonable signal-to-noise ratio potentiostats are needed. The instrument proposed in [132] can work in the staircase mode at 1 /iA full scale current over 2 MHz frequency. [Pg.117]

The CV curves were recorded by using a measuring system consisting of potentiostat, generator and x-y recorder. For rapid scan rates the curves were recorded on the screen of the monitor of an IBM PC-XT computer using a programme written in our laboratory. In kinetic experiments the positive feedback technique was used for cell resistance iR compensation. [Pg.424]

Impedance measurement system usually integrates an AC measurement unit such as frequency response analyzer (FRA), a potentiostat or galvanostat of suitably high bandwidth, and the electrochemical cell composed of 2, 3, or 4 electrodes in contact with an investigated sample (Fig. 6). The analyzed electrochemical interface is located between the sample and the working electrode (WE). A counter electrode (CE) is used to supply a current through the cell. Where there is a need to control the potential difference across the interface, one or two reference electrodes... [Pg.496]

Because of the great amplification factors of modem potentiostats, pick-up of stray signals must be avoided. Provisions for shielded cables and connections are, therefore, highly desirable. Finally, a built-in precision resistor circuit provides a convenient means of checking the accuracy of the entire measuring system. [Pg.25]

The complex transient r vs t, or equivalently r vs 0Na or r vs Uwr behaviour of Fig. 4.15 parallels the steady-state rvs UWr behaviour shown in Fig. 4.16, where for each point UWr has been imposed potentiostatically, until the current I has vanished and the corresponding rate value, r, has been measured. This shows that the catalyst surface readjusts fairly fast to the galvanostatically imposed transient 0Na values (Fig. 4.15). The dashed and dotted line transients on the same figure were obtained with the same gaseous composition but with initial Uwr values of 0 and -0.3 V respectively. It is noteworthy that the three transients are practically identical which shows the reversibility of the system. [Pg.133]


See other pages where Potentiostatic measuring system is mentioned: [Pg.1148]    [Pg.1148]    [Pg.1686]    [Pg.244]    [Pg.244]    [Pg.2140]    [Pg.678]    [Pg.456]    [Pg.1686]    [Pg.60]    [Pg.220]    [Pg.1235]    [Pg.180]    [Pg.524]    [Pg.21]    [Pg.130]    [Pg.641]    [Pg.279]    [Pg.63]    [Pg.335]    [Pg.22]    [Pg.221]    [Pg.105]   
See also in sourсe #XX -- [ Pg.63 ]

See also in sourсe #XX -- [ Pg.67 ]




SEARCH



Measurement systems

Measures Systems

Potentiostat

Potentiostatic

Potentiostatic measurements

Potentiostats

© 2024 chempedia.info