Trapping-mass spectrometer

The most widely used type of trap for the study of ion-molecule reactivity is the ion-cyclotron-resonance (ICR) [99] mass spectrometer and its successor, the Fourier-transfomi mass spectrometer (FTMS) [100, 101]. Figure A3.5.8 shows the cubic trapping cell used in many FTMS instmments [101]. Ions are created in or injected into a cubic cell in a vacuum of 10 Pa or lower. A magnetic field, B, confines the motion in the x-y... [Pg.810]

In essence, a guided-ion beam is a double mass spectrometer. Figure A3.5.9 shows a schematic diagram of a griided-ion beam apparatus [104]. Ions are created and extracted from an ion source. Many types of source have been used and the choice depends upon the application. Combining a flow tube such as that described in this chapter has proven to be versatile and it ensures the ions are thennalized [105]. After extraction, the ions are mass selected. Many types of mass spectrometer can be used a Wien ExB filter is shown. The ions are then injected into an octopole ion trap. The octopole consists of eight parallel rods arranged on a circle. An RF... [Pg.811]

Another approach to mass analysis is based on stable ion trajectories in quadnipole fields. The two most prominent members of this family of mass spectrometers are the quadnipole mass filter and the quadnipole ion trap. Quadnipole mass filters are one of the most connnon mass spectrometers, being extensively used as detectors in analytical instnunents, especially gas clnomatographs. The quadnipole ion trap (which also goes by the name quadnipole ion store, QUISTOR , Paul trap, or just ion trap) is fairly new to the physical chemistry laboratory. Its early development was due to its use as an inexpensive alternative to tandem magnetic sector and quadnipole filter instnunents for analytical analysis. It has, however, staned to be used more in die chemical physics and physical chemistry domains, and so it will be described in some detail in this section. [Pg.1339]

Figure Bl.7.14. Schematic cross-sectional diagram of a quadnipole ion trap mass spectrometer. The distance between the two endcap electrodes is 2zq, while the radius of the ring electrode is (reproduced with pennission of Professor R March, Trent University, Peterborough, ON, Canada). Figure Bl.7.14. Schematic cross-sectional diagram of a quadnipole ion trap mass spectrometer. The distance between the two endcap electrodes is 2zq, while the radius of the ring electrode is (reproduced with pennission of Professor R March, Trent University, Peterborough, ON, Canada).
Figure Bl.7.16. Mass spectra obtained with a Finnigan GCQ quadnipole ion trap mass spectrometer, (a)... Figure Bl.7.16. Mass spectra obtained with a Finnigan GCQ quadnipole ion trap mass spectrometer, (a)...
Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s). Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s).
In the other types of mass spectrometer discussed in this chapter, ions are detected by having them hit a detector such as an electron multiplier. In early ICR instruments, the same approach was taken, but FT-ICR uses a very different teclmique. If an RF potential is applied to the excitation plates of the trapping cell (figure B 1.7.18(b)) equal to the cyclotron frequency of a particular ion m/z ratio, resonant excitation of the ion trajectories takes place (without changing the cyclotron frequency). The result is ion trajectories of higher... [Pg.1356]

Brodbelt J, Liou C-C and Donovan T 1991 Selective adduct formation by dimethyl ether chemical ionization is a quadrupole ion trap mass spectrometer and a conventional ion source Ana/. Chem. 63 1205-9... [Pg.1359]

Other types of mass spectrometer may use point, array, or both types of collector. The time-of-flight (TOF) instrument uses a special multichannel plate collector an ion trap can record ion arrivals either sequentially in time or all at once a Fourier-transform ion cyclotron resonance (FTICR) instrument can record ion arrivals in either time or frequency domains which are interconvertible (by the Fourier-transform technique). [Pg.201]

There are two common occasions when rapid measurement is preferable. The first is with ionization sources using laser desorption or radionuclides. A pulse of ions is produced in a very short interval of time, often of the order of a few nanoseconds. If the mass spectrometer takes 1 sec to attempt to scan the range of ions produced, then clearly there will be no ions left by the time the scan has completed more than a few nanoseconds (ion traps excluded). If a point ion detector were to be used for this type of pulsed ionization, then after the beginning of the scan no more ions would reach the collector because there would not be any left The array collector overcomes this difficulty by detecting the ions produced all at the same instant. [Pg.209]

Other types of mass spectrometer can use point, array, or both types of ion detection. Ion trap mass spectrometers can detect ions sequentially or simultaneously and in some cases, as with ion cyclotron resonance (ICR), may not use a formal electron multiplier type of ion collector at all the ions can be detected by their different electric field frequencies in flight. [Pg.212]

Linked Scanning, Ion Traps, and Hybrid Mass Spectrometers... [Pg.243]

Instruments are available that can perform MS/MS type experiments using a single analyzer. These instruments trap and manipulate ions in a trapping cell, which also serves as the mass analyzer. The ion trap and fourier transform ion cyclotron resonance (FT-ICR) mass spectrometers are examples. [Pg.14]

Tandem quadrupole and magnetic-sector mass spectrometers as well as FT-ICR and ion trap instruments have been employed in MS/MS experiments involving precursor/product/neutral relationships. Fragmentation can be the result of a metastable decomposition or collision-induced dissociation (CID). The purpose of this type of instrumentation is to identify, qualitatively or quantitatively, specific compounds contained in complex mixtures. This method provides high sensitivity and high specificity. The instrumentation commonly applied in GC/MS is discussed under the MS/MS Instrumentation heading, which appears earlier in this chapter. [Pg.17]

Electrospray ionization mass spectrometry (ESI-MS) is an analytical method for mass determination of ionized molecules. It is a commonly used method for soft ionization of peptides and proteins in quadmpole, ion-trap, or time-of-flight mass spectrometers. The ionization is performed by application of a high voltage to a stream of liquid emitted from a capillaty. The highly charged droplets are shrunk and the resulting peptide or protein ions are sampled and separated by the mass spectrometer. [Pg.458]

Tandem mass spectrometry (MS/MS) is a method for obtaining sequence and structural information by measurement of the mass-to-charge ratios of ionized molecules before and after dissociation reactions within a mass spectrometer which consists essentially of two mass spectrometers in tandem. In the first step, precursor ions are selected for further fragmentation by energy impact and interaction with a collision gas. The generated product ions can be analyzed by a second scan step. MS/MS measurements of peptides can be performed using electrospray or matrix-assisted laser desorption/ionization in combination with triple quadruple, ion trap, quadrupole-TOF (time-of-flight), TOF-TOF or ion cyclotron resonance MS. Tandem... [Pg.1191]

GC = gas chromatography ECD = electron capture detector EIA = enzyme-immunoassay GPC = gel permeation chromatography HPLC = high-performance liquid chromatography ITMS = ion trap mass spectrometer LSE = liquid solid extraction MS = mass spectrometry RSD = relative standard deviation SPE = solid phase extraction... [Pg.259]

A triple-quadrupole mass spectrometer with an electrospray interface is recommended for achieving the best sensitivity and selectivity in the quantitative determination of sulfonylurea herbicides. Ion trap mass spectrometers may also be used, but reduced sensitivity may be observed, in addition to more severe matrix suppression due to the increased need for sample concentration or to the space charge effect. Also, we have observed that two parent to daughter transitions cannot be obtained for some of the sulfonylurea compounds when ion traps are used in the MS/MS mode. Most electrospray LC/MS and LC/MS/MS analyses of sulfonylureas have been done in the positive ion mode with acidic HPLC mobile phases. The formation of (M - - H)+ ions in solution and in the gas phase under these conditions is favorable, and fragmentation or formation of undesirable adducts can easily be minimized. Owing to the acid-base nature of these molecules, negative ionization can also be used, with the formation of (M - H) ions at mobile phase pH values of approximately 5-7, but the sensitivity is often reduced as compared with the positive ion mode. [Pg.402]

Specificity is unsurpassed. Traditionally, MS was performed on very large and expensive high-resolution sector instruments operated by experienced specialists. The introduction of low-resolution (1 amu), low-cost, bench-top mass spectrometers in the early 1980s provided analysts with a robust analytical tool with a more universal range of application. Two types of bench-top mass spectrometers have predominated the quadrupole or mass-selective detector (MSD) and the ion-trap detector (ITD). These instruments do not have to be operated by specialists and can be utilized routinely by residue analysts after limited training. The MSD is normally operated in the SIM mode to increase detection sensitivity, whereas the ITD is more suited to operate in the full-scan mode, as little or no increase in sensitivity is gained by using SIM. Both MSDs and ITDs are widely used in many laboratories for pesticide residue analyses, and the preferred choice of instrument can only be made after assessment of the performance for a particular application. [Pg.740]


See other pages where Trapping-mass spectrometer is mentioned: [Pg.399]    [Pg.399]    [Pg.813]    [Pg.1355]    [Pg.1355]    [Pg.1358]    [Pg.189]    [Pg.195]    [Pg.205]    [Pg.277]    [Pg.281]    [Pg.282]    [Pg.415]    [Pg.541]    [Pg.541]    [Pg.547]    [Pg.402]    [Pg.142]    [Pg.251]    [Pg.545]    [Pg.205]    [Pg.1029]    [Pg.1030]    [Pg.52]    [Pg.71]    [Pg.285]    [Pg.284]    [Pg.55]    [Pg.57]    [Pg.401]   
See also in sourсe #XX -- [ Pg.176 , Pg.177 , Pg.177 , Pg.178 ]




SEARCH



3D Quadrupole ion trap mass spectrometer

Instrumentation trapping-mass spectrometer

Ion trap mass spectrometer, ITMS

Ion trap mass spectrometers

Linear ion trap mass spectrometer

Linear trap quadrupole mass spectrometer

Linear-trap quadrupole Orbitrap mass spectrometer

Mass spectrometers orbital trap

Mass trapping

Orbital trap mass spectromete

Plasma source ion trap mass spectrometer

Quadrupole ion trap -time of flight mass spectrometer

Quadrupole ion trap mass spectrometers

Quadrupole linear ion trap mass spectrometers

Scanning, Ion Traps, and Hybrid Mass Spectrometers

Trap-triple quadrupole mass spectrometer