MALDI desorption/ionisation

MALDI-TOF MS Matrix-Assisted Laser Desorption Ionisation... [Pg.351]

Fast atom bombardment (FAB) Plasma desorption (PD) Liquid secondary-ion mass spectrometry (LSIMS) Thermospray (TSP)/plasmaspray (PSP) Electrohydrodynamic ionisation (EHI) Multiphoton ionisation (MPI) Atmospheric pressure chemical ionisation (APCI) Electrospray ionisation (ESI) Ion spray (ISP) Matrix-assisted laser desorption/ionisation (MALDI) Atmospheric pressure photoionisation (APPI) Triple quadrupole (QQQ) Four sector (EBEB) Hybrid (EBQQ) Hybrid (EB-ToF, Q-ToF) Tandem ToF-ToF Photomultiplier... [Pg.352]

Matrix-assisted laser desorption ionisation (MALDI) MH+ (M - H) ToF, IT, FTMS Polar and some nonpolar biopolymers, synthetic polymers >250000... [Pg.358]

As evident from Scheme 7.13, most modern ionisation techniques have been used for TLC-MS, and no single ionisation method is used exclusively with TLC-MS. Various ionisation methods may be applied that avoid the need to evaporate the sample into an El or Cl source these are based in particular on sputtering (FAB, SIMS) or laser desorption. Several sputtering methods of ionisation do not require the use of a liquid matrix, e.g. TLC-SIMS [797], Recent developments include the use of matrix-assisted laser desorption ionisation (MALDI) and surface-assisted laser desorption ionisation (SALDI). It is obvious that TLC-MS is complemented with TLC-MS11 [800] and TLC-HRMS techniques. Table 7.82 lists the general characteristics of TLC-MS. [Pg.539]

Meyer-Dulheuer [55] has analysed the pure additives (phenolic antioxidants, benzotriazole UV stabilisers and HALS compounds) of Table 9.8 in THF solutions by means of MALDI-ToFMS. As it turns out, polar molecules in the mass range of below 800 Da, which have a high absorption coefficient at the laser wavelength used, can often be measured without any matrix [55,56]. In this case, there is no matrix-assisted laser desorption and ionisation (MALDI) process any more. It is a simple laser desorption/ionisation (LDI) process. The advantage of this method is a matrix-free mass spectrum with the same mass resolution as in the MALDI case,... [Pg.703]

Even HALS compounds which absorb weakly at 337 nm can be analysed directly without matrix assistance, with the exception of the high-MW Hostavin N 30 (ca. 1500 Da), which fragments by direct laser desorption ionisation of intact molecules occurs only in the presence of a (dithranol) matrix. Direct laser desorption leads only to noncharacteristic, low-MW fragments. Hostavin N 20 leads to [M + H]+, [M + Na]+, [M + K]+ and some fragmentation peaks. MALDI-ToFMS of Tinuvin 765, which consists of a mono- and bifunctional sterically hindered amine, only shows the adduct peaks of the bifunctional amine apparently, the monofunctional amine is not ionisable. [Pg.704]

Experimental considerations Sample preparation and data evaluation are similar to membrane osmometry. Since there is no lower cut-off as in membrane osmometry, the method is very sensitive to low molar mass impurities like residual solvent and monomers. As a consequence, the method is more suitable for oligomers and short polymers with molar masses up to (M)n 50kg/mol. Today, vapour pressure osmometry faces strong competition from mass spectrometry techniques such as matrix-assisted laser desorption ionisation mass spectrometry (MALDI-MS) [20,21]. Nevertheless, vapour pressure osmometry still has advantages in cases where fragmentation issues or molar mass-dependent desorption and ionization probabilities come into play. [Pg.217]

Other types of atmospheric pressure ionisation detectors are available including APPI (photon ionisation) and MALDI (matrix-assisted laser desorption ionisation). More conventional detection systems for LC include UV absorption, fluorescence and evaporative light scattering. Developments in column packings,... [Pg.570]

Mass spectrometry is used to identify unknown compounds by means of their fragmentation pattern after electron impact. This pattern provides structural information. Mixtures of compounds must be separated by chromatography beforehand, e.g. gas chromatography/mass spectrometry (GC-MS) because fragments of different compounds may be superposed, thus making spectral interpretation complicated or impossible. To obtain complementary information about complex mixtures as a whole, it may be advantageous to have only one peak for each compound that corresponds to its molecular mass ([M]+). Even for thermally labile, nonvolatile compounds, this can be achieved by so-called soft desorption/ionisation techniques that evaporate and ionise the analytes without fragmentation, e.g. matrix-assisted laser desorption/ionisation mass spectrometry (MALDI-MS). [Pg.131]

Two slightly different laser desorption/ionisation methods were developed simultaneously by Karas and Hillenkamp [1] and Tanaka et al. [2]. Whereas Karas and Hillenkamp used small organic matrix molecules to assist and facilitate the desorption and ionisation of analytes (MALDI), Tanaka et al. used ultra-fine metal powders and glycerol. Zumbiihl et al. first analysed natural triterpenoid resins, dammar and mastic, both... [Pg.131]

For the quick characterisation of polydisperse surfactants with relative high molecular weight distributions matrix-assisted laser desorption/ionisation (MALDI)-time of flight (TOF)-MS represented an interesting alternative since low mass compounds did not interfere with the mass spectrometric detection of the compounds of interest. For example, the mass spectrum of C12-APG (Fig. 2.7.8) exhibited equally spaced signals with Am/z 162 corresponding to sodiated adduct ions of the mono- (m/z 371) to heptaglucosides (m/z 1343) [7]. [Pg.228]

Matrix assisted laser desorption ionisation (MALDI) and ESI-MS spectra of non-ionic surfactant blends of AE obtained after positive ionisation were compared [28]. Both the ionisation procedures, which produce [M + Na]+ ion clusters, were very useful for this purpose, but the ESI spectra generated were more complex, whereas MALDI ionisation led to simpler spectra that can be interpreted more easily [28]. [Pg.263]

For the characterisation of the biodegradation intermediates of C12-LAS, metabolised in pure culture by an a-proteobacterium, Cook and co-workers [23] used matrix-assisted laser desorption/ionisation (MALDI)-time of flight (TOF)-MS as a complementary tool to HPLC with diode array detection and 1H-nuclear magnetic resonance. The dominating signal in the spectrum at m/z 271 and 293 were assigned to the ions [M - H] and [M - 2H + Na]- of C6-SPC. Of minor intensity were the ions with m/z 285 and 299, interpreted to be the deprotonated molecular ions of C7- and C8-SPC, respectively. [Pg.332]

While fast atom bombardment (FAB) [66] and TSI [25] built up the basis for a substance-specific analysis of the low-volatile surfactants within the late 1980s and early 1990s, these techniques nowadays have been replaced successfully by the API methods [22], ESI and APCI, and matrix assisted laser desorption ionisation (MALDI). In the analyses of anionic surfactants, the negative ionisation mode can be applied in FIA-MS and LC-MS providing a more selective determination for these types of compounds than other analytical approaches. Application of positive ionisation to anionics of ethoxylate type compounds led to the abstraction of the anionic moiety in the molecule while the alkyl or alkylaryl ethoxylate moiety is ionised in the form of AE or APEO ions. Identification of most anionic surfactants by MS-MS was observed to be more complicated than the identification of non-ionic surfactants. Product ion spectra often suffer from a reduced number of negative product ions and, in addition, product ions that are observed are less characteristic than positively generated product ions of non-ionics. The most important obstacle in the identification and quantification of surfactants and their metabolites, however, is the lack of commercially available standards. The problems with identification will be aggravated by an absence of universally applicable product ion libraries. [Pg.376]

Below we report methodological studies based upon HPLC, GC/FID, GC-MS, LC-MS, matrix-assisted laser desorption ionisation coupled with time-of-flight mass spectrometry (MALDI-ToF/MS), CE, proton nuclear magnetic resonance ( I INMR), RIA and enzymatic colorimetric techniques. [Pg.612]

Matrix assisted laser desorption ionisation (MALDI)... [Pg.310]

This desorption ionisation technique leads to weak fragmentation. The analyte is incorporated into a solid organic matrix (such as hydroxybenzoic acid) and the mixture is placed on a sample holder that is irradiated with UV laser pulses (e.g. N2 laser, A = 337 nm, pulse width = 5 ns). The laser energy is absorbed by the matrix and transferred to the analyte, which becomes desorbed and ionised (Fig. 16.18c). Although MALDI is considered to be a soft ionisation technique, a substantial amount of energy is involved. Because the technique involves pulsed ionisation, it is well suited for time-of-flight mass analysis of biomolecules. The analysis of small molecules (M < 500 Da) is limited because the matrix decomposes upon absorption of the laser radiation. However, solid supports such as silicone can be used as the matrix to overcome this disadvantage. [Pg.310]

Figure 16.18—FAB and MALDI techniques, a) Principle of fast-atom generation using xenon b) formation of a fast-atom beam of argon in a collision chamber and bombardment of the sample (using a FAB gun) c) MALDI. The impact of a photon leads to a similar result as with a fast atom. The mechanism for desorption ionisation is not entirely known. These ionisation modes are particularly well suited for the study of medium to high molecular weight species. They are mostly used in biomedical sciences but not for routine determinations. Figure 16.18—FAB and MALDI techniques, a) Principle of fast-atom generation using xenon b) formation of a fast-atom beam of argon in a collision chamber and bombardment of the sample (using a FAB gun) c) MALDI. The impact of a photon leads to a similar result as with a fast atom. The mechanism for desorption ionisation is not entirely known. These ionisation modes are particularly well suited for the study of medium to high molecular weight species. They are mostly used in biomedical sciences but not for routine determinations.
MALDI Matrix-assisted Laser desorption/ionisation... [Pg.436]

MALDI-TOF matrix-assisted laser desorption/ionisation time-of-flight... [Pg.226]

The mass spectrometry analysis was performed by the matrix assisted laser desorption/ionisation time-of-flight (S8-MALDI) technique using a Voyager-DE PRO Biospectrometry Workstation (Applied Biosystems, USA). Radiation pulses of 0.5 ns and 3 Hz frequency from N2 laser operating at 337 nm were used to desorb the species and negative/positive ions formed were detected in reflectron mode. Sulfur used as a matrix material was also dissolved in toluene and mixed with the samples solution prior to deposition onto a target. [Pg.244]

The two additives have heen identified hy matrix assisted laser desorption ionisation mass spectrometry (MALDI-MS) and attenuated total reflection infrared spectroscopy (ATR). [Pg.186]

Two techniques that have become preffered for ionisation of proteins/peptides is electrospray ionisation (ESI) and matrix-assisted laser desorption/ionisation (MALDI). Although different combinations of ionisation techniques and mass analyser exist, MALDI usually is coupled with a time-of-flight (TOF) (Figure 7) tube as a mass analyser while ESI is tradionally combined with quadrupole mass analysers. Instruments capable of MS/MS have the ability to select ions of particular m/z ratio from a mixture, to fragment selected ions and to record the precise masses of the resulting fragment ions. If this process is applied to the analysis of peptide ions, in principle the amino acid sequence of the peptide can be deduced. [Pg.862]


See other pages where MALDI desorption/ionisation is mentioned: [Pg.384]    [Pg.384]    [Pg.396]    [Pg.396]    [Pg.529]    [Pg.541]    [Pg.379]    [Pg.173]    [Pg.715]    [Pg.738]    [Pg.132]    [Pg.571]    [Pg.458]    [Pg.22]    [Pg.103]    [Pg.401]    [Pg.899]    [Pg.237]    [Pg.869]    [Pg.283]    [Pg.237]   


SEARCH



Desorption/ionisation

Ionisation

Ionised

MALDI

Mass spectrometry desorption/ionisation (MALDI

Matrix associated laser desorption ionisation MALDI)

Matrix-Assisted Laser Desorption Ionisation Mass Spectrometry (MALDI MS)

Matrix-assisted laser desorption ionisation MALDI) mass spectrometry

Matrix-assisted laser desorption/ionisation MALDI)