Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Linear absorber

It should be noted that the calculation is based upon an assumption of a linear absorbance/concentration relationship and this may only apply over short concentration ranges. [Pg.753]

A) The use of a calibration graph. This overcomes any problems created due to non-linear absorbance/concentration features and means that any unknown concentration run under the same conditions as the series of standards can be determined directly from the graph. The procedure requires that all standards and samples are measured in the same fixed-path-length cell, although the dimensions of the cell and the molar absorption coefficient for the chosen absorption band are not needed as these are constant throughout all the measurements. [Pg.753]

The protein analyzer tests response linearity (absorbance/fluorescence), wavelength accuracy and UV linearity, dynamic noise, drift, and the zero offset. [Pg.59]

Bischoff and coworkers showed that there is a clear parhcle size dependent shift of the linearly absorbed CO stretching vibrational frequency on Pt supported in a zeolite [144]. They observed shift from 2055 cm" for CO chemisorbed on 1-2 nm particles to 2070 cm" for CO on 4—5nm particles. They correlated the frequency shift with the occurrence of higher indexed crystal planes for the smaller particles. [Pg.139]

Before kinetic constants can be evaluated, it is critical to find the correct concentration of enzyme to use for the assays. If too little enzyme is used, the overall absorbance change for a reaction time period will be so small that it is difficult to detect differences due to substrate concentration changes or inhibitor action. On the other hand, too much enzyme will allow the reaction to proceed too rapidly, and the leveling off of the time course curve as shown in Figure E5.7 will occur very early because of the rapid disappearance of substrate. A rate that is intermediate between these two extremes is best. For the dopachrome assay, it is desirable to use the level of tyrosinase that gives a linear absorbance change at 475 nm for 2 minutes. [Pg.293]

Two UV detectors are also available from Laboratory Data Control, the UV Monitor and the Duo Monitor. The UV Monitor (Fig.3.45) consists of an optical unit anda control unit. The optical unit contains the UV source (low-pressure mercury lamp), sample, reference cells and photodetector. The control unit is connected by cable to the optical unit and may be located at a distance of up to 25 ft. The dual quartz flow cells (path-length, 10 mm diameter, 1 mm) each have a capacity of 8 (i 1. Double-beam linear-absorbance measurements may be made at either 254 nm or 280 nm. The absorbance ranges vary from 0.01 to 0.64 optical density units full scale (ODFS). The minimum detectable absorbance (equivalent to the noise) is 0.001 optical density units (OD). The drift of the photometer is usually less than 0.002 OD/h. With this system, it is possible to monitor continuously and quantitatively the absorbance at 254 or 280 nm of one liquid stream or the differential absorbance between two streams. The absorbance readout is linear and is directly related to the concentration in accordance with Beer s law. In the 280 nm mode, the 254-nm light is converted by a phosphor into a band with a maximum at 280 nm. This light is then passed to a photodetector which is sensitized for a response at 280 nm. The Duo Monitor (Fig.3.46) is a dual-wavelength continuous-flow detector with which effluents can be monitored simultaneously at 254 nm and 280 nm. The system consists of two modules, and the principle of operation is based on a modification of the 280-nm conversion kit for the UV Monitor. Light of 254-nm wavelength from a low-pressure mercury lamp is partially converted by the phosphor into a band at 280 nm. [Pg.89]

A mixed quantum classical description of EET does not represent a unique approach. On the one hand side, as already indicated, one may solve the time-dependent Schrodinger equation responsible for the electronic states of the system and couple it to the classical nuclear dynamics. Alternatively, one may also start from the full quantum theory and derive rate equations where, in a second step, the transfer rates are transformed in a mixed description (this is the standard procedure when considering linear or nonlinear optical response functions). Such alternative ways have been already studied in discussing the linear absorbance of a CC in [9] and the computation of the Forster-rate in [10]. [Pg.40]

The paper is organized as follows. The next section quotes details of the Frenkel exciton model necessary for the later discussion. Comments on a full quantum dynamical description of all those quantities which are of interest in the mixed description are shortly introduced in Section 3. The used mixed quantum classical methodology is introduced in Section 4. Its application to EET processes is given in 5, to the computation of linear absorbance in Section 3.2, and to the determination of emission spectra in Section 7. The paper ends with some concluding remarks in Section 8. [Pg.40]

Continue monitoring the reaction absorbance change with the chart-recorder. A lag phase of up to 30 s may occur. The reaction should be monitored for 5-10 min or until a linear absorbance change of >0.2 occurs. [Pg.72]

The absorbance A of a solution is related to the transmittance in a logarithmic manner, as shown in Equation 24-5. Notice that as the absorbance of a solution increa.ses, the transmittance decreases. The relationship between transmittance and absorbance is illustrated by the conversion spreadsheet shown in Figure 24-8. The scales on earlier instruments were linear in transmittance modern instruments have linear absorbance scales or a computer that calculates absorbance from measured quantities. [Pg.718]

Fig. 5.6. Linear absorbance difference diagram for the photoisomerisation of azobcnzene,... Fig. 5.6. Linear absorbance difference diagram for the photoisomerisation of azobcnzene,...
A prerequisite for the evaluation mentioned is knowledge about the reaction mechanism. Linear absorbance diagrams proved the photoisomerisation taking place as in solutions. However, the siloxane matrix has to be fresh. Different types of siloxanes were tested, some photochemically polymerised, others fabricated by a catalyst induced process. In the latter case the Pt-catalyst must not overcome a concentration limit otherwise it influences the azobenzene photoreaction. Approximate evaluations at low absorption (assuming a irradiation intensity independent of the volume element) do not offer appropriate results because of measurement problems. Therefore a transformation of the time scale has been used, discussed in Section 5.7.3. [Pg.465]

The output of the detector is determined by the energy falling on it, and is directly proportional to the transmittance. Transmittance values are then converted to absorbance values by calculation, by a non-linear scale on the instrument meter, or electrically in instruments with linear absorbance readout. [Pg.46]

Commercial instruments have been modified to make possible the simultaneous recording of linear absorption and CD data and to facilitate work on turbid preparations (Kfivacic et ai, 1971). Also, a prototype spectrophotometer for simultaneous measurement of ORD, CD, and linear absorbance has been described (Amato and Ewing, 1974). A Roussel-Jouan Dichrograph was modified to permit the direct reading of A and A /2 (Snatzke and Lohr, 1968). The sensitivity of a commercial CD instrument was improved by a factor of 10-12 by the use of a time-averaging system (Myer and MacDonald, 1967 Horwitz et al, 1968). Fourier transform spectrometry was applied to CD (Stewart, 1971). The fundamental aspects of its application were described, but no information was given about the performance. [Pg.170]

Three y-lactones (parthenolide, marrubiin, artemisinin) from plant extracts were quantitated on a silica column (A = 210nm or 225nm) using either an 85/15 or a 90/10 hexane/dioxane mobile phase [683]. Good resolution of the analytes of interest from other extracted components was obtained and linear absorbance vs. concentration plots were obtained for the 0.2-5 mg/mL range. Analysis time was 15 min. [Pg.245]


See other pages where Linear absorber is mentioned: [Pg.302]    [Pg.153]    [Pg.536]    [Pg.57]    [Pg.1505]    [Pg.83]    [Pg.96]    [Pg.90]    [Pg.35]    [Pg.36]    [Pg.54]    [Pg.67]    [Pg.68]    [Pg.402]    [Pg.553]    [Pg.215]    [Pg.325]    [Pg.271]    [Pg.153]    [Pg.104]    [Pg.158]    [Pg.272]    [Pg.777]    [Pg.349]    [Pg.423]    [Pg.436]    [Pg.373]    [Pg.388]    [Pg.737]    [Pg.35]    [Pg.1921]    [Pg.304]   
See also in sourсe #XX -- [ Pg.553 ]




SEARCH



Detectors absorbance linearity

Linear absorbance coefficient

© 2024 chempedia.info