Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Least count, measurements

When we look at the residuals carefully, we are likely to find unexpected appearances. This seems to be true in our example here. Table 6 gives us a rather hard look at two sets of residuals, both of which seem to show some tendency toward relatively regular clumping. Inquiry of Dr. Lide as to whether there was a least count of 0.04 cm"1 to 0.05 cm"1 (40 or 50 in the units of Table 5) in the wave numbers either in measurement or calculation brought a response that this was not to be ruled out. While the discovery of such an effect at the time when the data were being analyzed might have been quite helpful, neither the suggestion nor a possible proof of its existence today can be of importance. [Pg.33]

Background QC At specified intervals, in many instances daily or for each batch, the background count rate for each system must be measured. The count rate is recorded and plotted on control charts, either by hand or by computer. The mean value of the background is found by averaging at least 20 measurements. The 2(7 and 3ageviations are calculated from the individual and mean values, and these multiple-standard-deviation lines are plotted on the control chart (see Section 10.5.1). Once the control chart is established, each newly measured value is recorded. The measurement should be repeated if it falls outside the 2cr band to distinguish between a random event and an instrumental problem. Remedial action with the detector or its environment is necessary if the repeated measurement is beyond the 3a band. [Pg.231]

The most successful technique for the stable species has been microprobe sampling followed by mass spectral analysis (alternatively by gas chromatography or other microanalytical techniques). The most successful microprobes have been pencil-like quartz tubes drawn down to orifices a thousandth of an inch in diameter or less. Because of the strong pressure drop (samples are pumped off at 10 atm), the probes act as miniature supersonic nozzles and the gas residence time in the probe is very short. The reactions are quenched by the rapid pressure drop (10- second and temperature drop (lO °K/second) due to adiabatic expansion in the nozzle. Reactions with half-lives as short as a few microseconds will be quenched. It is not necessary to cool the probes because of the short residence times. The effect which such a probe has on a flame is minor because of the small size of sample withdrawn (1-2 jag/sec). The aerodsmamic disturbance is minimized because the probe sucks off its own bow wave. The thermal disturbance is small because the probe wall temperature is only slightly below that of the flame. With such probes, it has been possible in favorable cases to make reproducible and reliable composition measurements with a precision of 2% and a least count of 10 mole fraction with a spatial reproducibility of 10 inches and resoution of 2 X 10 inches (see Figs. 3 and 5a). The scope and limitations of such composition measurements are discussed in Chapter IX of Fristom and Westenberg (1965). [Pg.83]

It has not yet been made use of the fact that c/ = for nuclear spectra. Before including this relationship in the least squares method, a very simple nuclear spectrum will be considered consisting of the counts measured with a long-lived radionuclide for equal periods of time. The model function is obviously very simple in this case. It only contains one single parameter representing the common expected value of the measured counts. The minimization problem therefore can be expressed like this... [Pg.451]

A reading is the single determination of a value at one point on a measuring scale. Generally, a reading can be estimated to one-half of the smallest I I I I I I I I division (least count) on a measuring scale. In the I case of the ruler in Figure 11.6, half of the smallest... [Pg.380]

Fig. 7 Time-correlated single photon counting measurements of chromatophores containing wild type and mutant reaction centers. The measurements were performed in phosphate buffer containing 10 mM sodium dithionite. Excitation was at 855 nm and fluorescence was detected at 920 nm through a monochromator with a 16 nm bandwidth. The excitation intensity was approximately 20 mW per cm at 3.6 Mhz repetition rate. is the reduced chi squared from the nonlinear least squares fit of the data to three (wild type) or four (mutant) exponentials A(i) are the resulting initial amplitudes of the exponential terms and t(i) are the exponential decay constants. The weighted residuals from the fit are shown above each decay curve. Fig. 7 Time-correlated single photon counting measurements of chromatophores containing wild type and mutant reaction centers. The measurements were performed in phosphate buffer containing 10 mM sodium dithionite. Excitation was at 855 nm and fluorescence was detected at 920 nm through a monochromator with a 16 nm bandwidth. The excitation intensity was approximately 20 mW per cm at 3.6 Mhz repetition rate. is the reduced chi squared from the nonlinear least squares fit of the data to three (wild type) or four (mutant) exponentials A(i) are the resulting initial amplitudes of the exponential terms and t(i) are the exponential decay constants. The weighted residuals from the fit are shown above each decay curve.
This technique is invasive however, the particle can be designed to be neutrally buoyant so that it well represents the flow of the phase of interest. An array of detectors is positioned around the reactor vessel. Calibration must be performed by positioning the particle in the vessel at a number of known locations and recording each of the detector counts. During actual measurements, the y-ray emissions from the particle are monitored over many hours as it moves freely in the system maintained at steady state. Least-squares regression methods can be applied to evaluate the temporal position of the particle and thus velocity field [13, 14]. This technique offers modest spatial resolutions of 2-5 mm and sampling frequencies up to 25 Hz. [Pg.337]

Having chosen the test mixture and mobile diase composition, the chromatogram is run, usually at a fairly fast chart speed to reduce errors associated with the measurement of peak widths, etc.. Figure 4.10. The parameters calculated from the chromatogram are the retention volume and capacity factor of each component, the plate count for the unretained peak and at least one of the retained peaks, the peak asymmetry factor for each component, and the separation factor for at least one pair of solutes. The pressure drop for the column at the optimum test flow rate should also be noted. This data is then used to determine two types of performance criteria. These are kinetic parameters, which indicate how well the column is physically packed, and thermodynamic parameters, which indicate whether the column packing material meets the manufacturer s specifications. Examples of such thermodynamic parameters are whether the percentage oi bonded... [Pg.184]

The random nature of the ionizing events recorded by the detector must also be borne in mind. To achieve measurements with standard deviations of 1% it is necessary to record at least 104 counts. For signals of low intensity this may take several hours to accumulate. This point is discussed further in Chapter 10. [Pg.350]

Continuous variables are those that can at least theoretically assume any of an infinite number of values between any two fixed points (such as measurements of body weight between 2.0 and 3.0 kilograms). Discontinuous variables, meanwhile, are those which can have only certain fixed values, with no possible intermediate values (such as counts of five and six dead animals, respectively). [Pg.870]

A third source of error is associated with the fragmentation pattern caused by dissociation of the molecular ions formed in the source region of the spectrometer. Under severe conditions these processes may proceed with substantial isotopic fractionation, and this obscures the measurements of isotopic composition at the collector. To some extent careful standardization of the instrumental conditions may ensure that errors from fragmentation are systematic, and thus cancel (at least to some extent). Alternatively, softer ionization methods can be used to prevent most or all of the fragmentation. The bottom spectrum in Fig. 7.7 illustrates this approach it shows the mass spectrum of chlorobenzene obtained by photoionization. Only the parent molecular ions are observed. It should be kept in mind, however, that softer ionization usually yields smaller ion currents and consequently statistical counting errors increase. [Pg.218]

Another interesting consequence arises from the fact that diffusion and viscosity both depend on temperature. As the viscosity decreases, the diffusivity increases. As a consequence of this, the column performance as measured by the plate count is exclusively determined by the column backpressure. While this is not 100% accurate, it is at least a good rule of thumb. To get about the same column performance, you should increase the flow rate when you increase the temperature such that the column... [Pg.90]

Comparison of the kinetic features of different LMWPs revealed that all LMWPs tested so far (such as lysozyme, cytochrome-c and aprotinin) are quickly cleared from the circulation and accumulate rapidly in the kidney [38]. The fractions of the injected LMWP that are reported to be taken up by the kidney vary between 40-80 % of the injected dose. In our studies, using external counting of radioactivity, at least 80 % of the intravenously injected LMWPs was finally taken up by the kidneys, which is in agreement with renal extraction studies [69,70]. However, studies in which the actual amount of LMWP in the kidney was measured directly in the tissue, indicated a lower, but still substantial accumulation of 40% of the injected dose [71,72]. Apart from the kidney, LMWPs do not seem to accumulate elsewhere in the body (Figure 5.8). [Pg.137]


See other pages where Least count, measurements is mentioned: [Pg.195]    [Pg.1101]    [Pg.226]    [Pg.356]    [Pg.253]    [Pg.139]    [Pg.314]    [Pg.118]    [Pg.287]    [Pg.2493]    [Pg.3060]    [Pg.6]    [Pg.321]    [Pg.153]    [Pg.1]    [Pg.85]    [Pg.343]    [Pg.462]    [Pg.631]    [Pg.877]    [Pg.361]    [Pg.122]    [Pg.101]    [Pg.102]    [Pg.396]    [Pg.471]    [Pg.55]    [Pg.166]    [Pg.523]    [Pg.30]    [Pg.476]    [Pg.263]    [Pg.124]    [Pg.15]    [Pg.123]    [Pg.148]    [Pg.155]   
See also in sourсe #XX -- [ Pg.380 ]




SEARCH



Least count

© 2024 chempedia.info