Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Isothermal calorimeters

B1.27.4.1 CLASSIFICATION BY PRINCIPLE OF OPERATION ISOTHERMAL CALORIMETERS (MORE PRECISELY, QUASI-ISOTHERMAL)... [Pg.1902]

Adiabatic calorimetry. Dewar tests are carried out at atmospheric and elevated pressure. Sealed ampoules, Dewars with mixing, isothermal calorimeters, etc. can be used. Temperature and pressure are measured as a function of time. From these data rates of temperature and pressure rises as well as the adiabatic temperature ri.se may be determined. If the log p versus UT graph is a straight line, this is likely to be the vapour pressure. If the graph is curved, decomposition reactions should be considered. Typical temperature-time curves obtained from Dewar flask experiments are shown in Fig. 5.4-60. The adiabatic induction time can be evaluated as a function of the initial temperature and as a function of the temperature at which the induction time, tmi, exceeds a specified value. [Pg.368]

When the heat exchange between the inner vessel and its surroundings, maintained at a constant temperature T0, occurs at an infinitely large rate isothermal calorimeter, 2 in Fig. 1), the temperature of the inner vessel also remains constant. The heat produced or absorbed is generally evaluated from the intensity of a physical modification occuring at a constant temperature in the surrounding medium (phase transformation). [Pg.194]

Usual isothermal calorimeters also differ from the theoretical model that has been described a temperature gradient, for instance, however small it may be, must exist within the calorimeter for heat to be transferred from the inner vessel to the heat sink. [Pg.195]

It appears therefore that during the operation of all usual calorimeters, temperature gradients are developed between the inner vessel and its surroundings. The resulting thermal head must be associated, in all cases, to heat flows. In isoperibol calorimeters, heat flows (called thermal leaks in this case) are minimized. Conversely, they must be facilitated in isothermal calorimeters. All heat-measuring devices could therefore be named heat-flow calorimeters. However, it must be noted that in isoperibol or isothermal calorimeters, the consequences of the heat flow are more easily determined than the heat flow itself. The temperature decrease... [Pg.195]

Usually, isothermal calorimeters are used to measure heat flow in batch and semi-batch reactions. They can also measure the total heat generated by the reaction. With careful design, the calorimeter can simulate process variables such as addition rate, agitation, distillation and reflux. They are particularly useful for measuring the accumulation of unreacted materials in semi-batch reactions. Reaction conditions can be selected to minimize such accumulations. [Pg.99]

B. Kasting for allowing his Calvet isothermal calorimeter to be used for some preliminary measurements and Dr. D. N. Rubingh for helpful discussions. [Pg.150]

Calorimeters may also be classified with respect to the way they use the heat balance. In fact, every calorimeter is based on a heat balance (as reactors are). Here we may distinguish ideal accumulation calorimeters or adiabatic calorimeters, from ideal heat flow or isothermal calorimeters and isoperibolic11 calorimeters. [Pg.83]

In this category of calorimeters, we find the isothermal calorimeters and the dynamic calorimeters where the temperature is scanned using a constant temperature scan rate. The instrument must be designed in such a way that any departure from the set temperature is avoided and the heat of reaction must flow to the heat exchange system where it can be measured. The instrument acts as a heat sink. In this family we find the reaction calorimeters, the Calvet calorimeters [7], and the Differential scanning calorimeter (DSC) [8],... [Pg.85]

Once the melt is completed, the sample is shock-cooled by removal from the calorimeter vessel and placement on the isothermal calorimeter block. This cools the sample sufficiently rapidly that most materials will be "trapped" in their metastable supercooled liquid state. A rescan of the sample over the same temperature span will show the cold recrystallization of the material followed by the melt. For maximum time efficiency and sensitivity, a fast scanning rate is recommended. [Pg.118]

Heat-flow adsorption microcalorimetry. The most important type of isothermal calorimeter in current use is that based on the principle of the heat flowmeter, which was first applied by Tian (1923) and improved by Calvet (Calvet and Prat, 1958,... [Pg.64]

This type of isothermal calorimeter is especially suitable for the study of open systems (i.e. with the introduction and withdrawal of gas) and is therefore highly recommended for the determination of energies of adsorption. [Pg.65]

Apparatus and Procedure. The thermochemical measurements were made using a Parr fluorine combustion bomb and a National Bureau of Standards (NBS) isothermal calorimeter (No. 63090) manufactured bv the Precision Scientific Co. The bomb cylinder and all internal parts or the bomb were Monel. A Monel ampoule was fitted into the top of the bomb to retain the OF2 sample. The ampoule apparatus reduced the internal volume of the bomb from 380 to 315 cc. The ampoule screws... [Pg.222]

Benyon, J. H. and Humphreys, A. R. (1955). The enthalpy difference between a- and /3-copper phthalocyanine measured with an isothermal calorimeter. Trans. Faraday Soc., 51, 1065-71. [267]... [Pg.313]

Mention should be made here of the recently developed technique of pressure perturbation calorimetry (PPC), which measures the temperature-dependent volume change of a solute or colloidal particle in aqueous solution. PPC can also be used to detect thermotropic phase transitions in lipid model membranes and to characterize the accompanying volume changes and the kinetics of the phase transition. PPC essentially measures the heat change that results from small pressure changes at a constant temperature in a high-sensitivity isothermal calorimeter. For an excellent recent review on PPC as applied to lipid systems, the reader is referred to Heerklotz (19). [Pg.129]

The differential molar energy of adsorption can be measured by means of a closed isothermal calorimeter. This system consists of two compartments contained in a closed isothermal calorimeter. Initially one compartment is evacuated and contains a given amount of adsorbent but no adsorbate, and the other compartment contains n moles of gas at pressure p. The two compartments are then connected physically until the pressure equilibrates. [Pg.155]

Isothermal calorimeter. With the isothermal calorimeter, the condition (Tc- Tg) = constant, is attained by a total transfer of heat of the process taking place in the calorimetric vessel to the heat sink, where it will cause a partial phase transformation of the substance in the heat sink. The thermal effect of the investigated process is then determined from the volume change of the calorimetric substance. According to the phase transformation, isothermal calorimeters using transformation of the solid phase to the liquid or of the liquid to the vapor phase are known. [Pg.234]

In order to calculate the heat evolved or absorbed in the calorimeter, it is necessary to measure (except for the isothermal calorimeter) the change if temperature of the calorimetric vessel and cover, as well as the heat capacity of the calorimeter. [Pg.235]

On the other hand, for slow reactions, adiabatic and isothermal calorimeters are used and in the case of very small heat effects, heat-flow micro-calorimeters are suitable. Heat effects of thermodynamic processes lower than 1J are advantageously measured by the micro-calorimeter proposed by Tian (1923) or its modifications. For temperature measurement of the calorimetric vessel and the cover, thermoelectric batteries of thermocouples are used. At exothermic processes, the electromotive force of one battery is proportional to the heat flow between the vessel and the cover. The second battery enables us to compensate the heat evolved in the calorimetric vessel using the Peltier s effect. The endothermic heat effect is compensated using Joule heat. Calvet and Prat (1955, 1958) then improved the Tian s calorimeter, introducing the differential method of measurement using two calorimetric cells, which enabled direct determination of the reaction heat. [Pg.236]

KEY WORDS plaster, gypsum, anhydrite, vapor pressure, humidity, measure and integration. hydration ratio, plaster rocks, calcium sulfate P-hemihydraie. calcium sulfate dihydrate, reactor, water vapor molar fraction, isothermal calorimeter, thermoelectric captor... [Pg.57]

Using modem isothermal calorimeters, experiments with a reproducibility better than 20 mj can be achieved. This is therefore the value one has to compare with the expected immenion energy in order to predict the feasibiHty of an experiment and to estimate the sample mass to be used. The immersion energies range between a few mj/m (water/Teflon) and a few hundred mj/m (specially carbons in organic solvents, but also inorganic oxides in water). Up to a few hundred miUigrams of sample can be introduced in the bulb. [Pg.278]

The heat of precipitation silver selenate was measured by mixing 0.04241 moles of silver nitrate dissolved in 325 g of water with 0.02122 moles of selenic acid in an isothermal calorimeter equipped with a sensitive thermometer. The acid (7.07 M) was contained initially in a bulb that was broken in order to start the reaction. The Ag2Se04 formed was shown to be crystalline by X-ray diffraction. The water equivalent was determined after each run. The experimental results and the evaluation of the standard enthalpy of formation of Ag2Se04(cr) are shown in Table A-17. [Pg.455]

As// is easily measured in an isothermal calorimeter by monitoring the heat evolved or absorbed on successive additions of solvent to a given amount of. solute. The table below gives the integral heat-of-solution data for 1 mol of sulfuric acid in water at 25 C (the negative sign indicates that heat is evolved in the dilution process). [Pg.393]

The first difficulty in using the energy balance is that enthalpy of formation data may not be available, especially for the biochemical species that are not completely defined. Also, the substrate may be a mixture of substances, and an incompletely characterized mixture of products may be produced in a fermenter. The information that is more likely to be available are the elemental analysis and the heat of combustion (it is relatively easy to put any substance in a calorimeter and measure the heat released on combustion). As a simple example, suppose we wanted to know the enthalpy of benzene at 25 C. We could put one mole of benzene in an isothermal calorimeter with... [Pg.893]

All calorimeters consist of the calorimeter proper and its surround. This surround, which may be a jacket or a bath, is used to control the temperature of the calorimeter and the rate of heat leak to the environment. For temperatures not too far removed from room temperature, the jacket or bath usually contains a stirred liquid at a controlled temperature. For measurements at extreme temperatures, the jacket usually consists of a metal block containing a heater to control the temperature. With non-isothermal calorimeters (calorimeters where the temperature either increases or decreases as the reaction proceeds), if the jacket is kept at a constant temperature there will be some heat leak to the jacket when the temperature of the calorimeter changes. [Pg.1901]


See other pages where Isothermal calorimeters is mentioned: [Pg.195]    [Pg.196]    [Pg.62]    [Pg.99]    [Pg.196]    [Pg.79]    [Pg.326]    [Pg.274]    [Pg.89]    [Pg.135]    [Pg.520]    [Pg.116]    [Pg.286]    [Pg.208]    [Pg.209]    [Pg.58]    [Pg.232]    [Pg.59]    [Pg.289]    [Pg.295]   
See also in sourсe #XX -- [ Pg.134 ]

See also in sourсe #XX -- [ Pg.64 , Pg.65 ]

See also in sourсe #XX -- [ Pg.232 ]

See also in sourсe #XX -- [ Pg.73 , Pg.78 , Pg.134 , Pg.265 ]

See also in sourсe #XX -- [ Pg.109 , Pg.117 , Pg.345 ]




SEARCH



Calorimeter continuous , isothermal

Calorimeter discontinuous, isothermal

Calorimeter isothermal titration

Calorimeter isothermal-jacket

Calorimeters

Calorimetric methods isothermal calorimeter

Isotherm calorimeter, 3.30

Isotherm calorimeter, 3.30

Isotherm safety calorimeter

Isothermal Calorimeters with Electrical Compensation

Isothermal Flow Calorimeter TKR

Isothermal and Isoperibol Calorimeters

Isothermal flow calorimeter

Isothermal flow mixing calorimeter

Isothermal heat flow calorimeters

Isothermal jacketed calorimeter

Isothermally jacketed calorimeter

© 2024 chempedia.info