Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calorimeter continuous , isothermal

The principles of titration calorimetry will now be introduced using isoperibol continuous titration calorimetry as an example. These principles, with slight modifications, can be adapted to the incremental method and to techniques based on other types of calorimeters, such as heat flow isothermal titration calorimetry. This method, which has gained increasing importance, is covered in section 11.2. [Pg.158]

Figure 11.5 Typical curve for a continuous titration calorimetry study of an exothermic reaction, using the calorimeter of Figure 11.1 in the heat flow isothermal mode of measurement./ is the frequency of the constant energy pulses supplied to the heater C in Figure 11.1 b. Adapted from [196,197],... Figure 11.5 Typical curve for a continuous titration calorimetry study of an exothermic reaction, using the calorimeter of Figure 11.1 in the heat flow isothermal mode of measurement./ is the frequency of the constant energy pulses supplied to the heater C in Figure 11.1 b. Adapted from [196,197],...
The results can be presented in the form of a continuous curve of differential enthalpies of adsorption A versus na, as shown in Figure 3.16b, with a resolution which is much higher than that obtained by the discontinuous procedure (Figure 3.16a). If the adsorption calorimeter cannot be easily connected to a well-calibrated and well-temperature-controlled adsorption sonic nozzle set-up, or when the adsorption isotherm is difficult to determine (e.g. if very small amounts are adsorbed), there remains the possibility of determining, separately, the adsorption isotherm by any of the discontinuous or continuous procedures described in Sections 3.3.1 or 3.3.2. A simple procedure can be applied which does not require the gas flow rate calibration ... [Pg.74]

Conduction calorimeters typically, the Tian-Calvet calorimeter. They comment that, in spite of its good external thermal insulation, this calorimeter is not adiabatic, because the calories produced are continuously eliminated from the calorimetric vessel. They also consider that, in spite of the very small temperature variations of the sample cell, this calorimeter is not strictly isothermal , which justifies a separate group, except when a Peltier compensation totally cancels the temperature variations in the sample. [Pg.38]

Stokes has used an isothermal dilution calorimeter (to be described in detail later) as an adiabatic calorimeter to follow the equilibrium melting curve of a mixture until the last trace of solid disappears. At this point the composition of the mixture is precisely known from data on its preparation. A brief summary of the method follows pure solvent is thoroughly degassed, in situ, in the calorimeter and the vapour space is removed by the addition of a known volume of mercury. The calorimeter is placed in a thermostat controlled at about 0.5 K below the melting temperature of the solvent. It is stirred continuously. After supercooling the solvent freezes as shown by the reversal of the time-... [Pg.26]

The measurement can be done both in isothermal or adiabatic calorimeters, the latter being preferred. For isothermal measurement (see ASTM D3286), the temperature of the calorimeter jacket is held constant and a correction for heat transfer from the calorimeter is applied, while in the adiabatic measurement (see ISO 1928 and ASTM D2015), the temperature of the calorimeter jacket is continuously adjusted to approximate that of the calorimeter itself. [Pg.762]

Applications The ITC measures the heat production or adsorption when the molecules of the two components come together. When a certain amount of one component is injected into the calorimeter vessel containing the other component, the respective heat is compensated for electrically, and the power needed to maintain isothermal conditions between the two vessels is measured. The injection can be done stepwise or continuously. From the raw values, all interesting thermodynamic data (AH, AS, AG) of the reaction as well as other reaction parameters can be determined using the software of the respective instrument. [Pg.156]

The essentia] point for classification of this type of calorimeter is that the local temperature difference, which invariably exists wherever a heat exchange takes place, is measured. The heat released or consumed in the calorimeter (measuring system) initially causes a change of temperature with regard to the surroundings. This causes a relaxation process the heat exchange with the surroundings continues until the reestablishment of isothermal or steady-state conditions. The possible mechanisms of heat transport - thermal conduction, convection, and radiation - are discussed in Chapter 4. [Pg.166]

A heat-flow calorimeter is a variation of an isothermal-jacket calorimeter. It uses a thermopile (Fig. 2.7) to continuously measure the temperature difference between the reaction vessel and an outer jacket acting as a constant-temperature heat sink. The heat transfer takes place mostly through the thermocouple wires, and to a high degree of accuracy is proportional to the temperature difference integrated over time. This is the best method for an extremely slow reaction, and it can also be used for rapid reactions. [Pg.341]

Another development in calorimetry, at least in retrospect, was the construction of adiabatic calorimeters operating at constant heating rate. In such an instrument the heating was carried out continuously (scanning calorimeter), i.e., the measurement was not interrupted every 20 K or so to check the isothermal condition, but was carried out in one, continuous run. In such operation, the heat losses were minimized since the experiment could be completed faster, but the accuracy of such scanning calorimeters was considerably less than that of the standard adiabatic calorimeters. The reason for the lesser accuracy is the fact that the heat could not be distributed nearly as uniformly in the sample as in the adiabatic calorimeter. In addition, the loss calibration was also less accurate. [Pg.235]

Device for Continuous Reaction 2.2.1 Isothermal Flow Calorimeter TKR... [Pg.45]

One possibility proposes a isothermal, calorimetric investigation of continuous runs, in which the measuring kettle of the calorimeter functions as a totally mixed throughput reactor" the inlet stream in the measuring kettle is nearly instantaneously mixed by a quickly agitating mixer. There are differences in either... [Pg.146]

Chapter 2 presents calorimeters for measuring accurately the rate of heat release during discontinuous and continuous reactions versus time under isothermal and nonisothermal conditions. In addition, the chapter contains a description of an apparatus that can be used to record online the rate of heat release within a stirred tank reactor during a reaction. [Pg.253]


See other pages where Calorimeter continuous , isothermal is mentioned: [Pg.129]    [Pg.45]    [Pg.167]    [Pg.116]    [Pg.286]    [Pg.325]    [Pg.60]    [Pg.308]    [Pg.162]    [Pg.31]    [Pg.398]    [Pg.263]    [Pg.26]    [Pg.103]    [Pg.433]    [Pg.337]    [Pg.124]    [Pg.125]    [Pg.131]    [Pg.316]    [Pg.43]    [Pg.390]    [Pg.274]    [Pg.417]    [Pg.43]   
See also in sourсe #XX -- [ Pg.45 ]




SEARCH



Calorimeters

Continuous isotherms

Isotherm calorimeter, 3.30

Isothermal calorimeter

© 2024 chempedia.info