Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Internal standards estimation

For a quick estimate of total dissolved solids (TDS) in water one can run a conductivity measurement. The unit for the measurement is mhos/cm. An mho is the reciprocal of an ohm. The mho has been renamed the Sieman (S) by the International Standard Organization. Both mhos/cm and S/cm are accepted as correct terms. In water supplies (surface, well, etc.) conductivity will run about 10 S/cm or 1 iS/cm. [Pg.147]

Step 2. Add to the weighed sample an amount of reference material that contains the internal standard in an amount equal to the estimated amount of the element sought. This mixture is the diluted sample. [Pg.204]

Figure 3.36. Log-time T plots of the deconvoluted difference intensities, expressed as a percentage of the HhCO intensity, for the indicated UVRR bands. On the basis of reproducibility of the perchlorate internal standard, 1%, the uncertainty in the difference intensities is estimated to be 10%. (Reprinted from reference [156]. Copyright (2004), with permission from Elsevier.)... Figure 3.36. Log-time T plots of the deconvoluted difference intensities, expressed as a percentage of the HhCO intensity, for the indicated UVRR bands. On the basis of reproducibility of the perchlorate internal standard, 1%, the uncertainty in the difference intensities is estimated to be 10%. (Reprinted from reference [156]. Copyright (2004), with permission from Elsevier.)...
Because of difficulty in keeping vacuum pumps operating when they are exposed to significant volumes of moisture laden gas, it was decided that after flushing the sample cylinders with N2, they would be allowed to bleed down to 1 bar pressure but not evacuated. The residual 13 I of N2 in the sample cylinders acted as an internal standard, enabling an accurate estimation of dissolved gas concentration in the sample fluid (Freifeld Trautz 2006). [Pg.283]

However, it should be remembered that careful studies such as these are carried out on specimens of known composition. To estimate the atomic number of an unknown atom, intensities need to be compared under identical imaging conditions with a specimen of known composition acting as an internal standard. From the data of Isaacson et al. it appears that for elements... [Pg.372]

Because pi, mobility, and viscosity are all affected by temperature, the use of internal standards is recommended. However, the use of synthetic pi standards to estimate pis should be approached with caution because the pis of the protein sample and standards may not be affected equally by temperature. Temperature has a direct effect on viscosity and therefore all effects of viscosity on the CIEF process apply as temperature changes. [Pg.201]

The International Standard ISO 5725 establishes practical definitiorrs of repeatability r and reproducibility R and provides basic principles for the layout, organization and analysis of precision experiments designed for estimating r and R. [Pg.235]

I. 4-methoxyacetophenone (30 //moles) was added as an internal standard. The reaction was stopped after 2 hours by partitioning the mixture between methylene chloride and saturated sodium bicarbonate solution. The aqueous layer was twice extracted with methylene chloride and the extracts combined. The products were analyzed by GC after acetylation with excess 1 1 acetic anhydride/pyridine for 24 hours at room temperature. The oxidations of anisyl alcohol, in the presence of veratryl alcohol or 1,4-dimethoxybenzene, were performed as indicated in Table III and IV in 6 ml of phosphate buffer (pH 3.0). Other conditions were the same as for the oxidation of veratryl alcohol described above. TDCSPPFeCl remaining after the reaction was estimated from its Soret band absorption before and after the reaction. For the decolorization of Poly B-411 (IV) by TDCSPPFeCl and mCPBA, 25 //moles of mCPBA were added to 25 ml 0.05% Poly B-411 containing 0.01 //moles TDCSPPFeCl, 25 //moles of manganese sulfate and 1.5 mmoles of lactic acid buffered at pH 4.5. The decolorization of Poly B-411 was followed by the decrease in absorption at 596 nm. For the electrochemical decolorization of Poly B-411 in the presence of veratryl alcohol, a two-compartment cell was used. A glassy carbon plate was used as the anode, a platinum plate as the auxiliary electrode, and a silver wire as the reference electrode. The potential was controlled at 0.900 V. Poly B-411 (50 ml, 0.005%) in pH 3 buffer was added to the anode compartment and pH 3 buffer was added to the cathode compartment to the same level. The decolorization of Poly B-411 was followed by the change in absorbance at 596 nm and the simultaneous oxidation of veratryl alcohol was followed at 310 nm. The same electrochemical apparatus was used for the decolorization of Poly B-411 adsorbed onto filter paper. Tetrabutylammonium perchlorate (TBAP) was used as supporting electrolyte when methylene chloride was the solvent. [Pg.520]

The potency of interferon-a2 is estimated by comparing its effect to protect cells against a viral cytopathic effect with the same effect of the appropriate International Standard of human recombinant interferon-a2 or of a reference standard calibrated in International Units. [Pg.525]

GC provides a useful technique for estimating volatile degradation products. For example the pivalic acid release from the hydrolysis of dipivefrin in an eyedrop preparation (Fig. 11.19) used for treating glaucoma may be estimated by GC. Isovaleric acid, which is an isomer of pivalic acid, provides a suitable internal standard. Breakdown products of esters are more likely to occur in aqueous formulations such as eyedrops or injections. [Pg.229]

For quantitative analysis by either the external or internal standard methods, HPLC requires the use of calibration solutions that are injected under identical conditions. Thus to fully identify quantitative effects, calibration solutions plus standard solutions need to be analysed for each experiment in a ruggedness test. As duplicate determinations are required for the estimation of standard errors a single experiment can consist of up to six chromatographic experiments as shown below. [Pg.214]

For quantitative estimation of identified monomers, response factors were calculated, using 4-ethyl-resorcinol as an internal standard. The relative error in the determination of all the compounds was 4%. [Pg.240]

After filtration of the reaction mixture, the concentration of the nitrosated product in the solvent is estimated by NMR, using dioxane as an internal standard. Because of the extreme instability of the nitrosated product, it is never isolated and the solution is always kept as cold as possible. At —20°C, dilute solutions evidently are reasonably stable. Concentrated solutions, on the other hand, explode on warming to room temperature. [Pg.232]

For quantitative estimation, a sealed reusable capillary tube, with a known quantity of sodium salt of trimethylsilyl propionic acid (TSP) dissolved in 35 pi of D20, is inserted into the NMR tube while obtaining NMR spectra. The internal standard TSP is used as a chemical shift reference as well as a quantitative standard for the estimation of metabolites, and D20 is used as the field-frequency-lock . Spectra are acquired at room temperature. Typical spectra acquired at room temperature of human bile and standard glycine- and taurine-conjugated BAs are shown in Fig. 5.4.16. [Pg.653]

Concern for international standardization translates as the adoption of 300 g muscle, 100 g liver, 50 g kidney, 50 g fat, 100 g eggs, 1.5 L milk, and 20 g honey as daily food intake by humans. The value set for milk seems to be particularly high, but has been estimated as appropriate to ensure that infants do not consume drug residues at levels exceeding ADIs. The JECFA has considered that... [Pg.316]


See other pages where Internal standards estimation is mentioned: [Pg.21]    [Pg.426]    [Pg.427]    [Pg.413]    [Pg.254]    [Pg.155]    [Pg.85]    [Pg.88]    [Pg.114]    [Pg.239]    [Pg.37]    [Pg.145]    [Pg.180]    [Pg.181]    [Pg.483]    [Pg.492]    [Pg.541]    [Pg.278]    [Pg.674]    [Pg.218]    [Pg.158]    [Pg.350]    [Pg.296]    [Pg.149]    [Pg.313]    [Pg.690]    [Pg.203]    [Pg.722]    [Pg.117]    [Pg.42]    [Pg.419]    [Pg.10]    [Pg.690]    [Pg.112]    [Pg.694]   
See also in sourсe #XX -- [ Pg.294 ]




SEARCH



Internal standards

International Standardization

International Standards

Standardization international standards

© 2024 chempedia.info