Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

ICP measurement

Patients with massive cerebral infarction may require ICP monitoring, as this may help to guide therapy and predict outcome. Schwab et al. ° evaluated 48 patients with massive hemispheric infarctions and clinical signs of elevated ICP. They found that ICP measurements correlated well with the patient s clinical status, CT findings and outcome, although they did not find a significant effect of their therapies for elevated ICP on patient outcomes. Multiple methods of monitoring ICP are avail-... [Pg.185]

ICP) measurements. The catalytic performance of the nanocatalysts was finally tested in the Fischer-Tropsch synthesis carried out in a fixed bed reactor. The obtained results were compared with literature data of commercially used Fischer-Tropsch catalysts. [Pg.18]

The cobalt content of the catalyst was analyzed by ICP measurements (Perkin Elmer, Plasma 400) at the Institute of Chemical Engineering of the University of Erlangen-Nuremberg. [Pg.19]

To repeat a statement in our opening paragraph of this section, ICP measures the emission of light from the atomization/ionization/excitation source (ICP torch) rather than absorption of light by atoms... [Pg.263]

ICP measurements were determined by the group of Prof. Dr. A. Behr, Universitat Dortmund. [Pg.106]

Iridium may be analyzed by x-ray. Also, AA or ICP measurements may be performed after digestion in hot acid mixture. The metal is practically insoluble in all mineral acids and aquaregia, even on heating. A 20 1 mixture of HChHNOs may be used to dissolve it with strong heating. [Pg.410]

Fig. 5 Aquaporin-4 deletion increases brain swelling in vasogenic edema, a Increased ICP in AQP4-null mice in response to intraparenchymal fluid infusion. Left representative ICP traces from two wildtype and AQP4 null mice. Right, increased ICP at 60 min in response to isotonic fluid infusion (0.5 p.l/min). Data shown for individual mice and mean SE. b Increased ICP and in AQP4 null mice with melanoma brain tumor. Top, site of injection of melanoma cells. Bottom., tumor size at 4 and 7 days after implantation showing similar-sized tumors in wildtype and AQP4 null mice, c ICP measured 7 days after tumor implantation. From Papadopoulos et al. (2004)... Fig. 5 Aquaporin-4 deletion increases brain swelling in vasogenic edema, a Increased ICP in AQP4-null mice in response to intraparenchymal fluid infusion. Left representative ICP traces from two wildtype and AQP4 null mice. Right, increased ICP at 60 min in response to isotonic fluid infusion (0.5 p.l/min). Data shown for individual mice and mean SE. b Increased ICP and in AQP4 null mice with melanoma brain tumor. Top, site of injection of melanoma cells. Bottom., tumor size at 4 and 7 days after implantation showing similar-sized tumors in wildtype and AQP4 null mice, c ICP measured 7 days after tumor implantation. From Papadopoulos et al. (2004)...
Subsequently, we found that a zeroth-order QWP could be employed successfully in the laser beam for ICP measurements, and from that point it was relatively straightforward to carry out DCP ROA measurements [35]. If desired, linear polarization modulation can be set up by using half waveplates in place of the QWPs. In the following sections we provide details associated with present operation of our dual QWP ROA spectrometer. [Pg.75]

P. Schramel, Improvement of ICP-measurements by using a water-cooled spray chamber, Fresenius Z. Anal. Chem., 320 (1985), 233-236. [Pg.434]

Claisse Fluxer Analysis - Lithium tetraborate or metaborate fusion for the dissolution of rocks has been in use for many years. The Claisse Fluxer fusion device simply makes this fusion automated. We have used the method in the past for the fusion of coal and fly ashes (10,13). Oil shales can be dissolved by this method without pre-ashing. Once the solution is prepared, it may be analyzed for the most part by ICPES or by AAS. Analysis of U.S.G.S. Devonian Ohio shale SDO-1 by fusion followed by AAS or ICPES measurements is illustrated in Table III. [Pg.483]

Combining the fusion technique with ICPES measurements gives a rapid and accurate method for the ash elemental analysis. The total analysis time needed is 20-25 minutes per sample. However, although the fusion procedure is excellent for the determination of all major elements, it is not suitable for the determination of trace elements, because the final solution (1 L) is too dilute for detection of trace elements. If the solution volume is kept small, extremely high concentrations of lithium and boron in the solution give an undesirable high background spectrum for trace element measurements. Hence, it is necessary to resort to a separate procedure where both trace and major elements can be simultaneously determined. [Pg.484]

Fig 4 compares kinetics in the presence of the catalyst and the support treated separately with HNOj 0.25 M with the curve obtained under the same experimental conditions with the catalyst directly treated with HNO3 0.25 M. However, this nitric acid treatment of the catalysts could result in a dissolution of the metal, and a modification of the metallic dispersion. ICP measurement shows that the metal loading remains unchanged after treatment, while dispersion decreased from 12% to 8%. As shown in fig 4, this lower dispersion is surprisingly accompanied by an increase in activity. The easiest explanation of this experimental fact, confirmed by the results obtained with a mechanical mixture of carbon and catalyst, consists in the catalytic role of the support. [Pg.267]

We very much appreciated the generous gift of catalyst samples from Johnson Matthey. We acknowledge Perkin Elmer for ICP measurement, J. Van Cautenberg and P. Uliana for then-valuable help in preparing the manuscript. [Pg.271]

Lower results (factor 2) were observed for ICP in comparison to ETAAS. It was assumed that interferences were taken into account in the calibration in the ICP measurement. However, when standard additions were used or Ca was added to the calibrants, ICP results agreed with those of ETAAS, which demonstrated the need to perform a matrix matching calibration or a standard addition procedure. The scatter of results observed in step 2 (0.73 to 1.66 mg kg ) was found to be high but not surprising considering the low Cr content in the extract. [Pg.189]

Figure 7 illustrates the kinetics of Ni sorption on pyrophyllite, along with dissolved Si data from the Ni-treated pyrophyllite and from an untreated pyrophyllite (16). The release of Si into solution shows a similar kinetic behavior as Ni sorption on pyrophyllite. When one compares the Si release rate with the dissolution rate of the clay alone, the Si release rate in the Ni-treated system is strongly enhanced as long as Ni removal from solution is pronounced (Figure 7). Although not shown, a similar correlation between Ni sorption and Si release was observed for the Ni/kaolinite system but not for the Ni/montmorillonite system. The dissolution rate of the Ni/gibbsite system was not determined since the [Al] in solution was too low (<50 ppb) to produce reliable ICP measurements. Figure 7 illustrates the kinetics of Ni sorption on pyrophyllite, along with dissolved Si data from the Ni-treated pyrophyllite and from an untreated pyrophyllite (16). The release of Si into solution shows a similar kinetic behavior as Ni sorption on pyrophyllite. When one compares the Si release rate with the dissolution rate of the clay alone, the Si release rate in the Ni-treated system is strongly enhanced as long as Ni removal from solution is pronounced (Figure 7). Although not shown, a similar correlation between Ni sorption and Si release was observed for the Ni/kaolinite system but not for the Ni/montmorillonite system. The dissolution rate of the Ni/gibbsite system was not determined since the [Al] in solution was too low (<50 ppb) to produce reliable ICP measurements.
Ray et al. [24] treated Cloisite 20A (montmorillonite modified with dimethyl-ditallow-containing approximately 65% Cig, 30% Cis, and 5% Ci4-ammonium cation chains) with a MAO solution, after vacuo-drying at 100 °C. The resulting MAO-treated clay was subsequently used for ethylene polymerizahon in the presence of a late transition metal catalyst (2,6-bis[l-(2,6-diisopropylphenylimino)ethyl] pyridine iron(ll) dichloride) and additional MAO in a glass reactor. They compared the result with homogeneous polymerization with the same catalyst in the presence of Cloisite 20A and observed that the supported catalyst was more efficiently exfoliated than when only a mixture of catalyst and clay was used. This comparison led them to conclude that at least some of the active centers resided within the clay galleries. Inductively coupled plasma (ICP) measurements showed that all MAO and catalyst remained in the solid catalyst after drying. [Pg.73]

ICP-MS and ion chromatographic method with suppressed conductivity detection were used to measure bromine species in water samples. ICP measures the total bromine concentration while the IC detects the bromide ions. In river water samples [70] no significant differences were obtained between concentration values determined by the two different methods. This shows that the bromine exists mostly as bromide ion. [Pg.190]

Different spectroseopic methods ean be used to support the speeies expheitly ineluded in the SCMs. Johnston et al. [72] have reviewed various types of speetroseopie method apphed to environmental particles. Adsorption measurements alone and even the advaneed modeling based on only the analysis of adsorption data may lead to misinterpretation of proeesses. Independent methods such as electrophoresis, surface and bulk speetroseopy, the isotope teehnique, and multiple quantitative analysis (e.g., total dissolved eoneentration of a metal from ICP measurements in parallel with its free-activity data by using ion-seleetive eleetrode) are required to support the evaluation of adsorption measurements. [Pg.737]

Chapters on sample introduction and hyphenated sample treatment and ICP systems have also been further updated since the last edition. No doubt that chromatographic, electrophoresis, flow injection and field flow fraction separations have extended ICP-MS (and AES) measurements as the mainstay of elanental specia-tion measurements in biological and environmental fields. Without the combination of these separation techniques and ICP measurements, elemental speciation applications would be severely hampered... if not impossible (Chapter 18). The ability to measure P and S with high sensitivity has opened up new opportunities in proteomics, for example. Species-specific and unspecific isotopic dilution (ID-MS) has been critical in quantifying speciation analysis and revealing recovery errors (Chapter 13). Species-specific techniques have been applied to identify species transformations, resulting in the development of multi-species methods whereas, hyphenated species-unspecific ICP-ID-MS determinations of heteroatoms such as sulfur have become a common quantification technique in proteomics. [Pg.430]

An extensive study has focused on terpenes with structures related to pinane, camphor [8], and limonene [9]. The in-phase dual circular polarization (DCPj) ROA spectra and the normalized CIDs are presented for fourteen compounds. Correlations between ROA features and structural elements of the molecules are discussed. The study clearly demonstrates the advantage of the backscattering measurements, which for theoretical reasons should be about three times as intensive as the normal right-angle incident circular polarization (ICP) measurements. [Pg.793]


See other pages where ICP measurement is mentioned: [Pg.307]    [Pg.397]    [Pg.17]    [Pg.19]    [Pg.21]    [Pg.256]    [Pg.157]    [Pg.84]    [Pg.84]    [Pg.125]    [Pg.149]    [Pg.492]    [Pg.111]    [Pg.846]    [Pg.574]    [Pg.80]    [Pg.64]    [Pg.504]    [Pg.149]    [Pg.301]    [Pg.561]    [Pg.715]    [Pg.136]    [Pg.607]    [Pg.428]    [Pg.112]    [Pg.187]   
See also in sourсe #XX -- [ Pg.319 , Pg.320 ]




SEARCH



Isotope Ratio Measurements by Laser Ablation ICP-MS

Isotope Ratio Measurements by Quadrupole based ICP-MS

Isotope ratio measurements by LA-ICP

Measuring Isotope Ratios with Single-Collector ICP-MS

© 2024 chempedia.info