Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Flow-controlled ventilators

If flow is measured and used to deliver a preset volume, then a ventilator is considered to be a flow controller (1). In most of the cases, volume preset ventilation is provided by ventilators that actually measure flow and use flow over time to deliver a preset volume. These machines maintain an approximately constant volume in the face of varying lung mechanics. The most common flow-controlled ventilators are listed in Table 2. As indicated in the table, most of these machines can also provide pressure ventilation. [Pg.232]

At high pressure experiments the reactor should be installed in a pressure cell. All check valves before it, and the filter with the flow controller after it, can be kept in the vented operating room. As a minimum, the bypass valve and the flow controller must be accessible to the operator. This can be done by extended valve stems that reach through the protecting wall. Both the operating room and the pressure cell should be well ventilated and equipped by CO alarm instruments. [Pg.86]

Crankcase emissions in the United States have been effectively controlled since 1963 by positive crankcase ventilation systems which take the gases from the crankcase, through a flow control valve, and into the intake... [Pg.523]

Other Applications Very small, very low-flow, and relatively high-velocity exhaust inlets, similar to LVHV nozzles, have been used successfully to control fumes from electric soldering irons." " Some investigations have been made into small, point-control exhaust ventilation for aerosols generated by high-speed dental tools. However, such low-volume point-control ventilation systems have not seen widespread use. [Pg.854]

In thermal models, the ventilation airflow rates normally arc input parameters, to be defined by the user or to be calculated by the program on the basis of a nominal air exchange or flow rate) and some control parameters (demand-controlled ventilation, variable air volume flow ventilation systems), in airflow models, on the other hand, room air temperatures must be defined in the input (see Fig. 11.49). [Pg.1095]

Low-velocity ATD An air terminal device which is designed for thermally controlled ventilation, e.g., displacement flow applications. See also Air terminal device. [Pg.1456]

The mixing behavior inside the test chamber can easily be checked hy the tracer gas method , in which an inert tracer gas (e.g. SFg) is continuously added to the inlet air stream of the test chamber by means of the mass flow control. After a certain period of time the tracer gas supply is cut off. At the exit of the chamber the concentration of tracer gas is determined continuously. The mixing should be checked inside the gas chamber including a sample to be tested. Figure 2.1-3 shows a typical concentra-tion/time profile for this test set-up. Should the measured values of the increasing or decreasing tracer gas concentration differ more than 10% from the theoretical curve, the mixing of the chamber atmosphere is unsatisfactory. This happens, e.g., in the case of a technical ventilation problem, when the air inlet and the air outlet of the test chamber are short-circuited . [Pg.135]

Figure 18.2 shows the flow and pressure waveforms for volume-controlled ventilation. In this illustration, the inspiratory flow waveform is chosen to be a half sine wave. In Figure 18.2a, f is the inspiration duration, t the exhalation period, and Q the amplitude of inspiratory flow. The ventilator delivers a tidal volume equal to the area under the flow waveform in Figure 18.2a at regular intervals (t + Q set by the therapist. The resulting pressure waveform is shown in Figure 18.2b. It is noted that during volume-controlled ventilation, the ventilator attempts to deliver the desired volume of breath, irrespective of the patient s respiratory mechanics. However, the resulting pressure waveform, such as the one shown in Figure 18.2b, will be different depending on the patient s respiratory mechanics. Of course, for safety... Figure 18.2 shows the flow and pressure waveforms for volume-controlled ventilation. In this illustration, the inspiratory flow waveform is chosen to be a half sine wave. In Figure 18.2a, f is the inspiration duration, t the exhalation period, and Q the amplitude of inspiratory flow. The ventilator delivers a tidal volume equal to the area under the flow waveform in Figure 18.2a at regular intervals (t + Q set by the therapist. The resulting pressure waveform is shown in Figure 18.2b. It is noted that during volume-controlled ventilation, the ventilator attempts to deliver the desired volume of breath, irrespective of the patient s respiratory mechanics. However, the resulting pressure waveform, such as the one shown in Figure 18.2b, will be different depending on the patient s respiratory mechanics. Of course, for safety...
FIGURE 18.2 (a) Inspiratory flow for a mandatory volume-controlled ventilation breath and (b) airway pressure resulting from the breath delivery with a nonzero PEEP. [Pg.271]

Figure 18.3 shows a plot of the pressure and flow during a mandatory pressure-controlled ventilation. In this case, the respirator raises the airway pressure and maintains it at the desired level, Pj, which is set by the therapist, independent of the patient s respiratory mechanics. Although the ventilator maintains the same pressure trajectory for patients with different respiratory mechanics, the resulting flow trajectory, shown in Figure 18.3b, will depend on the respiratory mechanics of each patient. As in the case of mandatory volume-controlled ventilation, the total volume of delivered breaths is monitored to ensure that patients receive adequate ventilation. [Pg.272]

This new mode is a form of pressure-controlled ventilation that simultaneously keeps track of the delivered tidal volume. Figure 18.4 shows characteristic changes of pressure, volume, and flow of inspiratory flow in this mode. As shown in the top panel of Figure 18.4, for each breath, (a) through (e), the ventilator controls the inspiratory pressure to a level that may vary from breath to breath. Specifically, the ventilator controls the pressure, but also monitors the delivered tidal volume and compares it with the desired tidal volume. If the actual delivered tidal volume matches the desired level, such as in (a), then the level of controlled pressure for the next breath will be the same. However, if the next breath produced a larger than desired tidal volume, such as in (b), then the controlled pressure will be reduced in the next breath (c). Similarly, if the tidal volume falls short, such as in (d), then the controlled pressure in the next breath, (e), will be raised to... [Pg.272]

FIGURE 18.4 Patterns respiratory desired pressure and resulting volume and flow for adaptive pressvue control ventilation mode. Breath illustration (a) through (e) shows how the applied inspiratory pressure is automatically adjusted to achieve the desired tidal volume. [Pg.273]

In a microprocessor-controlled ventilator (Figure 18.8), the electronically actuated valves open from a closed position to allow the flow of blended gases to the patient. The control of flow through each valve depends on the therapist s specification for the mandatory breath. That is, the clinician must specify the following parameters for the delivery of volume-controlled mandatory ventilation breaths (1) respiration rate (2) flow waveform (3) tidal volume (4) oxygen concentration (of the delivered breath) (5) peak flow and (6) PEEP, as shown in the lower left side of Figure 18.8. It is noted that the PEEP selected by the therapist in the mandatory mode is only used for the control of exhalation flow this... [Pg.276]

The therapist entry for pressure-controlled ventilation is shown in Figure 18.8 (lower left-hand side). In contrast to the volume-controlled ventilation, where Qj(t) was computed directly from operators entry (Equations 18.1 through 18.3), the total desired flow is generated by the closed-loop airway pressure controller shown in Figure 18.8. This controller uses the therapist-selected inspiratory pressure, respiration rate, and the 1 E ratio to compute the desired inspiratory pressure trajectory. The trajectory serves as the controller reference input. The controller then computes the flow necessary to make the actual airway pressure track the reference input. Assuming a proportional-plus-integral controller, the governing equations are... [Pg.278]

Gassy operations shall be provided with controls located above ground for reversing the air flow of ventilation systems. [Pg.687]

PPMV modes that permit spontaneous ventilatory activity are termed interactive modes, in that patients can affect various aspects of the mechanical ventilator s functions. These interactions can range from simple triggering of mechanical breaths to more complex processes affecting delivered flow patterns and breath timing. Interactive modes allow for inspiratory muscle activity which, when done at nonfatiguing or physiologic levels, may prevent muscle atrophy and facilitate recovery (31-34). Spontaneous patient ventilatory activity and comfortable interactive modes may improve ventilation and reduce the need for the sedation or neuromuscular blockers that may be required to prevent patients from fighting machine-controlled ventilation (27,35-37). [Pg.17]

During interactive modes, insufficient unloading from either inadequate support levels or from dys-synchronous flow can produce or perpetuate muscle dysfunction, from imposed loading (35-38). Mechanical ventilation can also produce muscle dysfunction if only controlled ventilation is used for prolonged periods. This ventilator-induced diaphragmatic dysfunction is akin to muscle atrophy in other skeletal muscles (33,34). [Pg.19]

When providing support, a ventilator can control four primary variables during inspiration pressure, volume, flow, and time. If a ventilator controls a given variable, then the waveform of this variable during inspiration will ideally remain unchanged from breath to breath regardless of how the load (compliance and resistance) changes (1). Most modem home ventilators are either pressure or flow controllers. [Pg.231]

Fires are ventilation-controlled therefore, control ventilation rather than base mine designs on incubation time. Where there is a potential for spon com, make every effort to minimize pressure differentials across the gob, and maintain ventilation controls that prevent air from getting behind the shields and chocks. Equally important is constant monitoring of the atmosphere flowing into and out of the gob. Remember, in the analysis of those data, specific numbers are not relevant foUow trends only. This too is discussed in the book Mine Fires, particularly on pages 31 and 32. The procedures described have never yet failed. [Pg.373]

This delivery module is an enclosed ventilated cabinet that contains the reagent flow control valves and reagent reservoirs. There are four glass reservoirs of 220-mL capacity for normal active nucleoside... [Pg.441]


See other pages where Flow-controlled ventilators is mentioned: [Pg.236]    [Pg.242]    [Pg.248]    [Pg.236]    [Pg.242]    [Pg.248]    [Pg.105]    [Pg.495]    [Pg.242]    [Pg.302]    [Pg.495]    [Pg.72]    [Pg.214]    [Pg.334]    [Pg.329]    [Pg.103]    [Pg.271]    [Pg.278]    [Pg.279]    [Pg.328]    [Pg.177]    [Pg.324]    [Pg.287]    [Pg.137]    [Pg.243]    [Pg.104]    [Pg.2361]    [Pg.2404]    [Pg.406]    [Pg.241]    [Pg.94]   
See also in sourсe #XX -- [ Pg.232 , Pg.233 , Pg.236 , Pg.242 , Pg.243 ]




SEARCH



Flow control

Flow controllers

Ventilation control

© 2024 chempedia.info