Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Diode array devices

Figure 77. A Kubelka-Munk contour-plot, measured with a diode-array device. The sample-peak at 33.9 mm separation distance can be identified as flupirtine, a centrally acting non-opioid analgesic substance. Flupirtine shows absorptions between 200 and 350 nm and an additional fluorescence quenching signal between 500 and 530 nm. The weak signal at 400 nm is the flupirtine fluorescence spectrum which would be invisible if evaluated using equation (2). A transformation of the raw-data using equation (3) would show this signal only to reveal it as a fluorescence emis-... Figure 77. A Kubelka-Munk contour-plot, measured with a diode-array device. The sample-peak at 33.9 mm separation distance can be identified as flupirtine, a centrally acting non-opioid analgesic substance. Flupirtine shows absorptions between 200 and 350 nm and an additional fluorescence quenching signal between 500 and 530 nm. The weak signal at 400 nm is the flupirtine fluorescence spectrum which would be invisible if evaluated using equation (2). A transformation of the raw-data using equation (3) would show this signal only to reveal it as a fluorescence emis-...
S. Lacorte, J. J. Vreuls, J. S. Salau, R Ventura and D. Barcelo, Monitoring of pesticides in river water using fully automated on-line solid-phase extraction and liquid cliro-matography with diode array detection with a novel filrtation device , J. Chromatogr. 795 71-82(1998). [Pg.374]

Some analytical instruments produce a table of raw data which need to be processed into the analytical result. Hyphenated measurement devices, such as HPLC linked to a diode array detector (DAD), form an important class of such instruments. In the particular case of HPLC-DAD, data tables are obtained consisting of spectra measured at several elution times. The rows represent the spectra and the columns are chromatograms detected at a particular wavelength. Consequently, rows and columns of the data table have a physical meaning. Because the data table X can be considered to be a product of a matrix C containing the concentration profiles and a matrix S containing the pure (but often unknown) spectra, we call such a table bilinear. The order of the rows in this data table corresponds to the order of the elution of the compounds from the analytical column. Each row corresponds to a particular elution time. Such bilinear data tables are therefore called ordered data tables. Trilinear data tables are obtained from LC-detectors which produce a matrix of data at any instance during the... [Pg.2]

When measuring a signal, one records the magnitude of the output or the response of a measurement device as a function of an independent variable. For instance, in chromatography the signal of a Flame Ionization Detector (FID) is measured as a function of time. In spectrometry the signal of a photomultiplier or diode array is measured as a function of the wavelength. In a potentiometric titration the current of an electrode is measured as a function of the added volume of titrant. [Pg.507]

The spectroscopy system uses a dispersive element and a detector which is either a charge-coupled device (CCD) or a diode array. A computer is required for instrument control and for intensive data processing. [Pg.52]

Homogeneous wetting, 22 111 Homogenization, 26 699 aluminum alloys, 2 329 Homogenizers, 8 703 10 127 Homoglycans, 4 697, 701 23 62-64 classification by structure, 4 723t Homo-interface, 24 71 Homo-ionic interactions, 8 77 Homojunction BJTs, 22 166 Homojunction devices, LEDs asm 173 Homojunction diode arrays, 29 163 Homojunction laser diode, 24 699 Homoleptic tetranuclear carbonyl complexes, 26 63... [Pg.441]

Microprocessor based analytical equipments is no longer an uncommon phenomenon towards the modernization, automation, and above all the ease of function and handling of sophisticated devices, for instance a microprocessor scans the array of diodes many times a second in a diode array detector a microprocessor does the temperature programming of a constant temperature chamber of HPLC unit. [Pg.466]

A diode array is a spectroscopic device ior the continuous monitoring of absorbance over a specified wavelength, range. [Pg.104]

Instrnmentation for UV-vis process analysis falls into fonr categories scanning instruments, diode-array instrnments, photometers, and fiber-optic spectrophotometers with photodiode-array (PDA) and charge-conpled device (CCD) detectors. The former two are more typically enconntered in at-line or near-line applications, whereas the latter two are better snited to actnal on-line analyses. [Pg.85]

The drawback is the measurement time, which depends on the number of pixels. Spectrometer manufacturers therefore developed line mapping, in which samples are scanned line by line, thereby reducing the acquisition time. These devices were developed for Raman, IR, or near-IR (NIR) spectroscopy (with diode array detectors). However, due to the moving stage, this kind of imaging principle is only suitable for at-line applications. [Pg.413]

Finally, the introduction of new detectors, such as diode arrays and charge-coupled devices (CCDs), has been a boon for Raman spectroscopy. CCDs permit the accumulation of light in the manner of photographic film additionally, their noise level is lower than that of the photomultiplier tube. In addition, by combining CCDs or diode arrays with optical dispersive elements, entire spectra mav be collected in fractions of a second. [Pg.61]

Schematically, two main systems can be used to collect 3D fluorescence data (time, wavelength, number of photons, see fig. 1). In a first type of system, light is directed into a monochromator connected to a photomultiplier tube and then to a fast oscilloscope (PM detection). The experimentalist thus collects luminescence decays at various wavelengths. This system is known to be very efficient for luminescence decay acquisition but is very time-consuming for the acquisition of emission spectra. In the second type of system, light is directed to a diode array detector (or CCD camera) and a subsequent electronic detection device (diode detection). The experimentalist collects emission spectra at various delay times (time zero for the pulse entering in the sample). This system is very efficient for emission data acquisition but, on the other hand, time-consuming for luminescence decay acquisitions. From this very schematic description, it appears that a system combining the two types of detections would be the optimum. Schematically, two main systems can be used to collect 3D fluorescence data (time, wavelength, number of photons, see fig. 1). In a first type of system, light is directed into a monochromator connected to a photomultiplier tube and then to a fast oscilloscope (PM detection). The experimentalist thus collects luminescence decays at various wavelengths. This system is known to be very efficient for luminescence decay acquisition but is very time-consuming for the acquisition of emission spectra. In the second type of system, light is directed to a diode array detector (or CCD camera) and a subsequent electronic detection device (diode detection). The experimentalist collects emission spectra at various delay times (time zero for the pulse entering in the sample). This system is very efficient for emission data acquisition but, on the other hand, time-consuming for luminescence decay acquisitions. From this very schematic description, it appears that a system combining the two types of detections would be the optimum.
Modifications to the experimental set-up for the acquisition of fluorescence spectra from samples within the ESR microwave cavity are described in previous work ( ). Further improvements using a fast photomultiplier/photon counting technique were made in an attempt to determine the radiative fluorescence lifetime in solution. Phosphorescence at 77 K was measured both by a conventional Varian spectrofluorimeter and a pulsed laser/cooled diode array imaging device. Radiative phosphorescence lifetimes were measured by the photon counting technique, using the Stanford Research System SR400 gated photon counter. [Pg.102]

The first detectors to be used in OMA systems were standard TV image tubes. These were silicon vidicons or the more sensitive Silicon Intensified Target (SIT) detectors, which both employed silicon targets to convert optical information into electronic form. More recently, the use of solid state detectors in the form of a diode array (Reticon) has been found to have some advantages over the vidicons and SIT tubes. Current developments in the field of charge coupled devices (CCD) will probably soon provide an even better multielement detector for use in OMA systems. [Pg.46]


See other pages where Diode array devices is mentioned: [Pg.167]    [Pg.427]    [Pg.345]    [Pg.287]    [Pg.89]    [Pg.288]    [Pg.167]    [Pg.427]    [Pg.345]    [Pg.287]    [Pg.89]    [Pg.288]    [Pg.2963]    [Pg.512]    [Pg.432]    [Pg.303]    [Pg.454]    [Pg.358]    [Pg.362]    [Pg.283]    [Pg.113]    [Pg.12]    [Pg.68]    [Pg.608]    [Pg.21]    [Pg.172]    [Pg.877]    [Pg.384]    [Pg.722]    [Pg.200]    [Pg.512]    [Pg.283]    [Pg.259]    [Pg.466]    [Pg.27]    [Pg.36]    [Pg.61]    [Pg.626]    [Pg.9]   
See also in sourсe #XX -- [ Pg.89 ]




SEARCH



Diode arrays

© 2024 chempedia.info