Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Computerized mass spectrometers

In many cases, solvent mixtures are used. Their composition must be identified. Rodofos is one example of such solvent used in Poland. Its composition was determined by gas chromatography. Analyses were performed using a Hewlett-Packard gas chromatograph model 5890 coupled with computerized mass spectrometer instrument, model 5970. [Pg.1228]

Modern mass spectrometers are interfaced with computerized data handling sys terns capable of displaying the mass spectrum according to a number of different for mats Bar graphs on which relative intensity is plotted versus m z are the most common Figure 13 40 shows the mass spectrum of benzene m bar graph form... [Pg.568]

The DBMS setup and experimental procedures used in this study were the same as described in more detail elsewhere [Jusys et al., 2001]. Briefly, the DBMS setup consisted of two differentially pumped chambers, a Balzers QMS 112 quadrupole mass spectrometer (MS), a Pine Instruments potentiostat, and a computerized data acquisition system. [Pg.415]

GC-IR is becoming more widely used because FT spectrometers (p. 281) have virtually replaced the older dispersive types and even with computerized enhancements are much cheaper than mass spectrometers. [Pg.117]

The most straightforward tool for the introduction of a sample into a mass spectrometer is called the direct inlet system. It consists of a metal probe (sample rod) with a heater on its tip. The sample is inserted into a cmcible made of glass, metal, or silica, which is secured at the heated tip. The probe is introduced into the ion source through a vacuum lock. Since the pressure in the ion source is 10-5 to 10-6 torr, while the sample may be heated up to 400°C, quite a lot of organic compounds may be vaporized and analyzed. Very often there is no need to heat the sample, as the vapor pressure of an analyte in a vacuum is sufficient to record a reasonable mass spectrum. If an analyte is too volatile the cmcible may be cooled rather than heated. There are two main disadvantages of this system. If a sample contains more than one compound with close volatilities, the recorded spectrum will be a superposition of spectra of individual compounds. This phenomenon may significantly complicate the identification (both manual and computerized). Another drawback deals with the possibility of introducing too much sample. This may lead to a drop in pressure, ion-molecule reactions, poor quality of spectra, and source contamination. [Pg.121]

Mass-spectral (MS) Analysis. Mass spectral analyses were obtained using a computerized Finnigan Model 1015 gas chromatograph-mass spectrometer (GC-MS) operated at 70 eV. Samples were introduced via direct insertion probe or by utilizing the GC-MS combination, both operated over a programmed temperature range. [Pg.97]

A modem mass spectrometer is constmcted from elements which approach the state-of-the-art in solid-state electronics, vacuum systems, magnet design, precision machining, and computerized data acquisition and processing. [1] This is and has ever been a fully valid statement about mass spectrometers. [Pg.111]

Provided El spectra have been measured under some sort of standard conditions (70 eV, ion source at 150-250 °C, pressure in the order of 10 " Pa), they exhibit very good reproducibility. This is not only the case for repeated measurements on the same instrument, but also between mass spectrometers having different types of mass analyzers, and/or coming from different manufacturers. This property soon led to the collection of large El mass spectral libraries, either printed [76-78] or computerized. [79] The best established El mass spectral databases are the NIST/EPA/NIH Mass Spectral Database and the Wiley/NBS Mass Spectral Database, each of them giving access to about 120,000 evaluated spectra. [80-83]... [Pg.218]

Mass spectrometry is a sensitive analytical technique which is able to quantify known analytes and to identify unknown molecules at the picomoles or femto-moles level. A fundamental requirement is that atoms or molecules are ionized and analyzed as gas phase ions which are characterized by their mass (m) and charge (z). A mass spectrometer is an instrument which measures precisely the abundance of molecules which have been converted to ions. In a mass spectrum m/z is used as the dimensionless quantity that is an independent variable. There is still some ambiguity how the x-axis of the mass spectrum should be defined. Mass to charge ratio should not lo longer be used because the quantity measured is not the quotient of the ion s mass to its electric charge. Also, the use of the Thomson unit (Th) is considered obsolete [15, 16]. Typically, a mass spectrometer is formed by the following components (i) a sample introduction device (direct probe inlet, liquid interface), (ii) a source to produce ions, (iii) one or several mass analyzers, (iv) a detector to measure the abundance of ions, (v) a computerized system for data treatment (Fig. 1.1). [Pg.4]

There are several mass spectrometer manufacturers which have begun to offer commercial, fully computerized ms/ms instruments (16). These instruments are capable of generating and searching... [Pg.269]

The mass spectrometer used for these studies was a Type 21-103-C (Consolidated Electrodynamics Corp.) operated with a 250°C Isotron and 70-V ionization potential. Results were derived by using a computerized matrix developed for pyrolysis product gas mixtures. [Pg.30]

Figure 3. Flow sheet of a computerized combination gas chromatograph ((X) and mass spectrometer (MS) DAS = data acquisition system, s separator, i = interface, d.r. = data reduction... Figure 3. Flow sheet of a computerized combination gas chromatograph ((X) and mass spectrometer (MS) DAS = data acquisition system, s separator, i = interface, d.r. = data reduction...
The mass spectrometer is a highly sophisticated and computerized instrument, which basically consists of five parts sample introduction, ionization, mass analysis, ion detection, and data handhng. In principle, hquid chromatography is just one of the possible analyte techniques, or the mass spectrometer just another detector for LC. However, on-line chromatography-MS systems offer additional value, especially in terms of selectivity. [Pg.24]

Mass spectrometer (MS) [S] sales have always been high, especially since MS is the principal detector in a number of hyphenated techniques such as GC-MS, MS-MS, and LC-MS. The GC — MS combination accounts for about 60% of MS sales since it is used widely in drug and environmental testing. Innovations in interface technology such as inductively coupled plasma (ICP)-MS, FC-MS and thermospray or particle beam interfaces for LC-MS have both advanced the technology and expanded the interest in applications. Recent introductions of lower cost MS instruments with automated sampling and computerized data analysis have added to the attraction of the technique for first-time users. [Pg.128]

Figure 20-24 is a block diagram of the computerized control and dala-acquisition system of a triple quadrupole mass spectrometer. This figure shows two features encountered in any modern instrument. I hc first is a computer that serves as the main instrument controller. The operator communicates via a keyboard with the spectrometer by selening operating parameters and conditions via easy-to-use interactive software. The computer also controls the programs responsible for data manipulations and output. The second rcature common to almost all instruments is a set of microprocessors (often as many as six) that are responsible for specific aspects of instrument control and the transmission of information between the computer and spectrometer. [Pg.576]


See other pages where Computerized mass spectrometers is mentioned: [Pg.576]    [Pg.577]    [Pg.296]    [Pg.823]    [Pg.576]    [Pg.577]    [Pg.296]    [Pg.823]    [Pg.28]    [Pg.536]    [Pg.367]    [Pg.175]    [Pg.71]    [Pg.41]    [Pg.37]    [Pg.536]    [Pg.219]    [Pg.34]    [Pg.102]    [Pg.237]    [Pg.402]    [Pg.1809]    [Pg.3800]    [Pg.7]    [Pg.197]    [Pg.108]    [Pg.113]    [Pg.71]    [Pg.149]    [Pg.44]   
See also in sourсe #XX -- [ Pg.576 , Pg.577 ]




SEARCH



Computerization

Computerized

© 2024 chempedia.info