Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Compensation parameters

Soc. 1979, 101, 1470]. Perhaps the NMR case is a particularly extreme example, but the point is that one has two compensating parameters in temperature effects. [Pg.64]

Compensation parameters calculated from kinetic data reported by Te-tenyi et al. (184, 246, 247) for the dehydrogenation of cyclohexane and of... [Pg.293]

Compensation parameters calculated from data reported by Winter for the decomposition of nitrous oxide (263) and nitric oxide (264) on various oxides are given in Table V, B and C, respectively. The variation of data in the latter reaction is relatively small (aL = 0.795, Table V, C) and the values of B and e can be reliably determined. Interpretation of the results for nitrous oxide decomposition is, however, less straightforward since the compensation trend for reactions on the rare earth sesquioxides (B = 19.831 and e = 0.0571) was significantly different from that for all other oxides studied (B = 23.226 and e = 0.0321) and the combined data give the intermediate values of Table V, B. Thus, we are unable to discriminate between the possibilities that either there are two distinct lines or the number of points available is insufficient to characterize fully the compensation effect at the observed level of standard deviation. [Pg.298]

Compensation behavior, of an accuracy comparable with many trends so described in the literature, was found in data reported by Knozinger et al. (278) for the dehydration of alcohols over alumina. Values of A were converted to units used throughout the present review from the catalyst surface area, kindly supplied by Knozinger (279). Analyses of the results in Table 2 of (278) yielded two compensation effects one of these was for alcohols that contained a four-unit carbon chain and the other where there was a five unit carbon chain the compensation parameters are given in Table V, N and O, respectively. The accuracies of these two obediences are both significantly better than that found by the same calculation using the combined data. [Pg.302]

Compensation between instrument channels is required where the emission spectra of the fluorochromes overlap. Compensation is necessary between channels 1 and 2 and between channels 2 and 3 but is not required between channels 1 and 3 because these spectra should not overlap. To set compensation use the compensation parameter menu and cell samples of known positivity for the fluorochromes that are being used, in this case known dead cells and known infected cells, and also negative controls. [Pg.221]

Consequently, the plot of logx0 vs. AH constitutes a usual representation of the compensation rule. The origin intercept and the slope of the lines deduced from a linear regression of log x0 vs. AH are used to obtain the compensation parameters,... [Pg.365]

Having prepared an evolutionary optimization strategy, the application to earthquake resistance follows without many problems. We model the building by the 8 dof/node elements, define a goal or fitness function - a function of the building s response to the earthquake excitation - and try to minimise the goal wrt. the compensators parameters. [Pg.264]

Response function of active material is unreliable, which is different from traditional mechanism. It was figured out that most of them are reduced to the problem of delay or exceed of response. We can absorb the scatter of response by regulating switching of operators or by applying compensation parameters. The point is to include internal state changes of the materials to control laws of machines. Internal state changes are representation of activeness. [Pg.212]

Galwey AK, Brown ME (1979) Compensation parameters in heterogeneous catalysis. J Catal 60 335-338... [Pg.265]

The sensitivity of the luminescence IP s in the systems employed here decreases with increasing x-ray energy more strongly than in the case of x-ray film. Therefore, this phenomenon must be compensated by using thicker lead front and back screens. The specific contrast c,p [1,3] is an appropriate parameter for a comparison between IP s and film, since it may be measured independently of the spatial resolution. Since the absorption coefficient p remains roughly constant for constant tube voltage and the same material, it suffices to measure and compare the scatter ratio k. Fig. 2 shows k as a function of the front and back screen thickness for the IP s for 400 keV and different wall thicknesses. The corresponding measured scatter ratios for x-ray films with 0,1 mm front and back screens of lead are likewise shown. The equivalent value for the front and back screen thicknesses is found from the intersection of the curves for the IP s and the film value. [Pg.470]

Figure 5 shows the display in the measure mode. It consists of a detailed A-scan window and a number of smaller windows for display parameters and inspection parameters. The A-scan display may be used as a stand-alone tool or as a tool for measuring parameters required for a specific inspection, e.g. probe parameters, reference echoes, and depth compensation with automatic transfer to the data set. [Pg.786]

This situation, despite the fact that reliability is increasing, is very undesirable. A considerable effort will be needed to revise the shape of the potential functions such that transferability is greatly enhanced and the number of atom types can be reduced. After all, there is only one type of carbon it has mass 12 and charge 6 and that is all that matters. What is obviously most needed is to incorporate essential many-body interactions in a proper way. In all present non-polarisable force fields many-body interactions are incorporated in an average way into pair-additive terms. In general, errors in one term are compensated by parameter adjustments in other terms, and the resulting force field is only valid for a limited range of environments. [Pg.8]

When possible, quantitative analyses are best conducted using external standards. Emission intensity, however, is affected significantly by many parameters, including the temperature of the excitation source and the efficiency of atomization. An increase in temperature of 10 K, for example, results in a 4% change in the fraction of Na atoms present in the 3p excited state. The method of internal standards can be used when variations in source parameters are difficult to control. In this case an internal standard is selected that has an emission line close to that of the analyte to compensate for changes in the temperature of the excitation source. In addition, the internal standard should be subject to the same chemical interferences to compensate for changes in atomization efficiency. To accurately compensate for these errors, the analyte and internal standard emission lines must be monitored simultaneously. The method of standard additions also can be used. [Pg.438]

In addition, most devices provide operator control of settings for temperature and/or response slope, isopotential point, zero or standardization, and function (pH, mV, or monovalent—bivalent cation—anion). Microprocessors are incorporated in advanced-design meters to faciHtate caHbration, calculation of measurement parameters, and automatic temperature compensation. Furthermore, pH meters are provided with output connectors for continuous readout via a strip-chart recorder and often with binary-coded decimal output for computer interconnections or connection to a printer. Although the accuracy of the measurement is not increased by the use of a recorder, the readabiHty of the displayed pH (on analogue models) can be expanded, and recording provides a permanent record and also information on response and equiHbrium times during measurement (5). [Pg.467]

Because few scatter events are recorded, attenuation compensation is relatively easier for PET using an external positron emitting source. As a result, the technology for quantitative determinations of radioactivity distributions is significantly more advanced in PET imaging. Technology development for SPECT, however, is improving this parameter. [Pg.482]

The Smith predictor is a model-based control strategy that involves a more complicated block diagram than that for a conventional feedback controller, although a PID controller is still central to the control strategy (see Fig. 8-37). The key concept is based on better coordination of the timing of manipulated variable action. The loop configuration takes into account the facd that the current controlled variable measurement is not a result of the current manipulated variable action, but the value taken 0 time units earlier. Time-delay compensation can yield excellent performance however, if the process model parameters change (especially the time delay), the Smith predictor performance will deteriorate and is not recommended unless other precautions are taken. [Pg.733]

Adaptive Control Process control problems inevitably require on-hne tuning of the controller constants to achieve a satisfactory degree of control. If the process operating conditions or the environment changes significantly, the controller may have to be retuned. If these changes occur quite frequently, then adaptive control techniques should be considered. An adaptive control system is one in which the controller parameters are adjusted automatically to compensate for changing process conditions. [Pg.734]

The usage of the ratio of chai acteristic lines as analytical parameter in the process of formation of the calibration curve provides a significant decrease of the residual error. In Realization of this method simultaneously with the decrease of the matrix effects causes some decrease or even full compensation of the fonu and condition of the measured surface. [Pg.442]


See other pages where Compensation parameters is mentioned: [Pg.79]    [Pg.298]    [Pg.300]    [Pg.308]    [Pg.315]    [Pg.130]    [Pg.372]    [Pg.376]    [Pg.377]    [Pg.89]    [Pg.601]    [Pg.39]    [Pg.167]    [Pg.79]    [Pg.298]    [Pg.300]    [Pg.308]    [Pg.315]    [Pg.130]    [Pg.372]    [Pg.376]    [Pg.377]    [Pg.89]    [Pg.601]    [Pg.39]    [Pg.167]    [Pg.630]    [Pg.1887]    [Pg.2203]    [Pg.119]    [Pg.189]    [Pg.46]    [Pg.145]    [Pg.45]    [Pg.399]    [Pg.467]    [Pg.271]    [Pg.427]    [Pg.406]    [Pg.1136]    [Pg.1839]    [Pg.2012]    [Pg.2549]    [Pg.578]   
See also in sourсe #XX -- [ Pg.267 , Pg.315 ]




SEARCH



Compensation behavior Arrhenius parameters

© 2024 chempedia.info