Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration container

Stock solutions of approximately 1 mg mL were prepared by dissolving the appropriate amounts of the analytical standards in acetonitrile. Working standard solutions for fortification were prepared in volumetric flasks by appropriate dilutions of the stock solutions for each analyte or combination of analytes. During analysis, SCA is converted to DMS and HMS is derivatized therefore, the analytical standard solutions for quantitation and instrument calibration contained sulfentrazone, DMS and derivatized HMS. A measured volume of a standard solution containing sulfentrazone, DMS and HMS (prepared from stock solutions) was derivatized simultaneously with the samples. [Pg.573]

The volume of water to be purged from the well must be calculated. Methods for determining purge volume include the use of a flow meter on the discharge line, the discharging of purge water into a calibrated container, and/or purging at a known flow rate for a specific period of time. [Pg.802]

Pipet 1 mL of the stock solution aliquot and transfer into the calibrated container. Add sufficient quantity of D5W to fill to the required volume of 120 cc. [Pg.351]

Weigh out 120 mg of gentian violet and dissolve in water to make 20 mL stock solution. Transfer 5 mL of the stock solution to calibrated container and qs with water to 120 mL. [Pg.369]

Underground mains should be tested before joints are covered. Water leakage through the joints should be measured by pumping at the specified test pressure from a calibrated container into the section of pipe being tested. Care must be taken to expel all entrapped air and ensure the main is completely full of water. Permissible leakage per 100 joints per hour is 2 quarts (1.9 I) (NFPA 24). [Pg.336]

TPX has a coefficient of linear expansion of 11.7 X 10 s cm/cm C, which is similar to that of water, and hence it is useful for calibrating containers such as laboratory graduated cylinders. TPX is resistant to nonoxidizing acids, alkalis, and salts but is not resistant to oxidizing acids, such as nitric acid. Because of the presence of tertiary carbon atoms, TPX, like PP, is readily oxidized and cannot be used outdoors in the absence of antioxidants. [Pg.141]

A calibration curve with aqueous calibrators containing 0, 0.1, 0.5,1.0, and 5.0 nmol homocarnosine are measured. The curve is plotted and results determined. The calibration curve is linear to at least 5.0 nmol. [Pg.123]

Chromium and Sb have been determined in cachaca by ET-AAS using matrix matching calibration and Ru as a permanent modifier [128], Calibrants containing both elements were prepared in a 40 percent v/v ethanol solution and spiked samples were used to check accuracy. LoDs of 2 and 43 pg were achieved for Cr and Sb, respectively. The results obtained from the analysis of 52 samples varied from nondetectable to 39.1 xg l-1 for Sb and between 0.64 and 1.53 p,g l-1 for Cr. [Pg.482]

The requirement for calibrated laboratory-ware for containing and dispensing solutions extends to all facets of the analytical procedure. It rests with the analyst to ascertain accuracy suitable to the application, of all devices used, including micropipets used in conjunction with electrothermal atomizers. The temptation to use other than appropriately calibrated containers, and chemical reagents of suitably high purity, and not to follow good analytical procedures as practised by those with only peripheral contact with analytical chemistry, should be avoided. [Pg.160]

Calibrators are used to show deficiencies in the equipment, for example, chain looseness, tilting of stirrer motor, excessive vibration, or misalignment of the flasks with the stirrers. Non-disintegrating calibrators are composed of 300 mg salicylic acid, whereas disintegrating calibrators contain 10 mg prednisone. The dissolution rates obtained from the calibrants should fall within established ranges at both 50 and 100 rpm. [Pg.920]

Calibration is effected by means of six calibrators contained in the kits. The relevant graph is stored in the apparatus and is re-plotted whenever the reagent batch is changed or if a technical change is made in the instrument. The sample rotor is filled with 2 vessels each per calibrator. The calibrators consist of a human serum base into which the appropriate calibrator substance has been weighed with the exception of free T4. [Pg.558]

The approaches for estimation of the LOD are basically the same as those described for LLOQ under the headings LLOQ based on signal to noise ratio (S/N) - LOQ based on a specific calibration curve in the range of LLOQ . However, for LOD a S/N or k-factor equal to or greater than three is usually chosen [1, 6, 9,14], If the calibration curve approach is used for determination of the LOD, only calibrators containing the analyte in the range of LOD must be used. [Pg.6]

Enzyme-Linked Immunosorbent Assay. LlSAis a heterogeneous EIA technique that is widely used in clinical analyses. In this type of assay, one of the reaction components is nonspecifically adsorbed or covalently bound to the surface of a solid phase, such as that of a microtiter well, a magnetic particle, or a plastic bead. This attachment facilitates separation of bound- and free-labeled reactants. In the most common approach to using the ELISA technique, an aliquot of sample or calibrator containing the antigen to be quantitated is added to and allowed to bind with a solid phase antibody. After washing, enzyme-labeled antibody is... [Pg.235]

A set of calibrants containing the seven CB congeners was supplied to each participating laboratory in the form of pure, crystalline, certified materials from BCR (CRMs 291 -298). Each laboratory was requested to prepare separate calibration solutions of the appropriate concentration in iso-octane, to determine the linearity of the electron capture detector and to calibrate the gas chromatograph prior to the analysis of the fish oils. [Pg.279]

An alternative is to fit a calibrated container, referred to as a Kalibottle (Figure 5.6), to the nozzle and to collect the spray while treating 25 m2. The volume application rate is then read from the side of the bottle. This technique has the advantage when training teams of spray operators, as they can see the effects of walking speed or other changes in application technique. [Pg.80]

Prepare calibrators containing 0, 1, and lOng/mL cotinine (see Note 13). [Pg.226]

FIGURE 9.2 A comparison of two calibration curves in cynomolgus monkey plasma. One curve was prepared in K3 EDTA plasma and the other in lithium heparin plasma. Each calibrator contained 2% of the corresponding matrix prepared from the same stock solution of drug, spiked directly into 100% plasma and then diluted 50 fold in assay buffer. The pH of the diluted calibrators was 7.4 for each matrix. [Pg.259]

For most analytical processes, a measured signal of the sample is compared with the signals of calibrators in order to calculate the concentration present in the sample. An ideal calibrator contains a known concentration of the analyte and is by all means comparable to the sample to be analysed. As serum contains various protein bound cobalamins and cobalamin analogues in a very complex matrix, it is not possible to use an ideal calibrator when analysing cobalamin in serum. A suitable approximation is a solution of cyanocobalamin prepared as described in Box 26.1. For the final dilution of the calibrator, a buffer comparable to the one used for dilution of the samples to be analysed should be employed. [Pg.456]


See other pages where Calibration container is mentioned: [Pg.814]    [Pg.42]    [Pg.51]    [Pg.128]    [Pg.225]    [Pg.401]    [Pg.1257]    [Pg.468]    [Pg.205]    [Pg.153]    [Pg.205]    [Pg.126]    [Pg.335]    [Pg.19]    [Pg.32]    [Pg.283]   
See also in sourсe #XX -- [ Pg.144 ]




SEARCH



Permeation devices, containing dangerous goods, for calibrating air quality monitoring

© 2024 chempedia.info