Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analysers clinical

The example given in Table 7.2 is taken from a study to verify the trueness of clinical analyses (Streck [2004]). Recovery rates have been used as the criterion to accept a good agreement between the measured results and the reference values as it is frequently done by analysts. [Pg.210]

In addition to automated analysers for general use, sophisticated single purpose instruments have been developed and marketed, chiefly for clinical analyses (for example Astra 4 and Astra 8 from Beckman for the determination of sodium and potassium in blood or Orion Space-Stat SS-20 or SS-30 for the determination of calcium in blood). [Pg.116]

An interesting cell construction, described in [128], is suitable especially for clinical analysers, which limits the dead space and thus optimizes solution flow (fig. 5.8). The ISE is placed in a channel and the elevation in the channel opposite the electrode ensures that the flow-rate is highest at the ISE membrane. [Pg.121]

Performance trials and evaluation tests on the technique indicate that it is both rehable and accurate, and, in addition, that the specificity is sufficient to cope with most chnical requirements. An evaluation was made by Haeckel et al. [19]. If this approach is successful, the dispensers and tubes in laboratories will become redundant. It may well become possible for a clinical test to be undertaken close to the patient rather than in the laboratory. Whilst the techniques have as yet been used only for clinical analyses, there are many other potential applications, for example in the water industry. However, the very nature of the technique necessitates development by Eastman Kodak. Very few users will be able to influence the choice of analytical problems to be tackled by this unique approach. [Pg.46]

Light production in some biochemical reactions (bioluminescence) has been recognized as a powerful tool for biochemical and clinical analyses. Emitted light can be measured with a high sensitivity, so very low detection levels can indeed be achieved. [Pg.90]

Cling control agents Clinical analyses Clinical chemistry Clinical performance Clinical trials... [Pg.227]

Immobilized Enzymes in Diagnostic Reagents. The use of immobilized. instead of soluble, enzymes for measurement of analytes lias received considerable attemion. especially lor clinical analyses. Use of immobilized enzymes offers the advantages of greater accuracy, stability, and convenience. Only a few meihods utilizing immobilized enzymes have become commercially available. [Pg.975]

Pseudocumene is used as a component in liquid scintillation cocktails for clinical analyses. Pseudocumene and durene are oxidized to trimel-litic anhydride and pyromellitic dianhydride, respectively. Mesitylene is a key building block for important antioxidants and agricultural chemicals. Prehnitene, isodurene, pentamethylbenzene, and hexamethylbenzene... [Pg.1352]

RPLC-ICP-MS has been used for clinical analyses by a number of workers, particularly with regard to speciation studies. [Pg.1233]

Clinical analyses confirmed that topic glucocorticoids as well as most H, antihistamines, especially of the 3rd generation, are able to inhibit the upreg-ulation of ICAM-1 on epithelial cells during early and late phases of allergic inflammation [11, 12],... [Pg.47]

Analyses are performed in accordance with standardized methods issued under the responsibility of a Technical Committee within the Health Ministry. Usually such measurements rely on a comparison of the measured quantity in the unknown sample with the same quantity in a standard , i.e. an RM, according to a specific measurement equation [6], after calibrating the instrument. Calibration of a photometric system for clinical analyses usually means the set of operations that establish, under specific conditions, the relationship, within a specified range, between values indicated by the instrument and the corresponding values assigned to the RMs at the stated uncertainty. Calibration of the photometer itself implies the calibration of wavelength and absorbance scale by means of proper wavelength and absorbance RMs [5], traceable to national standards. A calibration of the instrument is still needed in concentration units to check the indicated provided value. The measurement result is then verified by application of that method of measurement to a certified reference material (CRM). Both the comparator - a photometric device with narrow or wide bandwidth, and the RMs should thus be validated. [Pg.186]

Evaluation of uncertainty components in photometric measurement specific to clinical analyses, performed in INM, follows the ISO guide Expression and Quantification of Uncertainty Measurements [7], using RMs and experimental quantification. [Pg.188]

The steps considered when evaluating uncertainty measurement components in clinical analyses are illustrated in Fig. 5. [Pg.188]

Flexibility. An instrument which can analyze a sample only for a fixed combination of elements is useful for routine analyses, such as clinical analyses, but becomes useless if the combination of elements is to be varied. The ideal multielement spectroscopic system would be able to determine any combination of elements desired, if those elements are amenable to analysis by spectroscopic techniques. Since real samples contain various elements in differing relative concentrations, it is necessary that an ideal multielement spectrometer be able to accept such samples without the need for varying dilutions to accomodate all elements present. Thus the instrument should have a wide dynamic range and be capable of adjustment so that different elements present in major, minor, and trace quantities in the same sample can be simultaneously monitored. [Pg.29]

Peptides and proteins by comparison have stronger rotatory powers with obvious potentials in clinical analyses, either as the analyte or the auxiliary reagent. However, typical of the general trends in the study of oligomers and macromolecules, the major research interests were focussed on abstracting structural information [75], (and see the Chapter by Manning and Towell). The... [Pg.262]

Since ISEs can be used in continuous flow systems or in flow systems with sample injection (flow injection analysis, FIA)21 their application is wide, not limited to discrete samples. Analysis time becomes shorter, with faster recycling. Additionally, in flow systems the experimental assembly and data analysis can be controlled automatically by microcomputer, including periodic calibration. Another development is the use of sensors for the detection of eluents of chromatographic columns in high-pressure liquid chromatography (HPLC). Miniaturization has permitted an increase in the use of sensors in foods, biological tissues, and clinical analyses in general. [Pg.308]

Mass spectroscopic analysis of proteins and peptide fragments has become an often-utilized technique over the past few years, and a number of microdevices have been reported for use with this analysis method. Though not yet used routinely in clinical analyses, detecting expression patterns and specific mutations in proteins may well be the best way to detect diseases and propose treatments specific for a given mutation. Transfer of this type of analysis to the clinical laboratory may await the development of microdevices for the sample preparation steps, which can interface directly with MS detectors. [Pg.438]

The field of immunoassays is large in terms of variety of compounds to be analyzed and the concentrations involved such as millimoles to subpicomoles. One of the optimum requirements is high sensitivity. Immunoassays involving labeling with radio isotopes have been popular in clinical analyses [181]. Some of the advantages and disadvantages of radioimmunoassays are listed in Table 12.32. [Pg.963]

Amino acid analysis remains an indispensable tool in a variety of biological research and development fields, e.g. the biochemical study of proteins, quality control in biotechnology and nutrition, and in clinical analyses. The classic chromatographic technologies will be for the foreseeable future the major quantitative tools for amino acid analyses. The techniques are deceptively difficult and there remains a need to standardize techniques for good quantitation. [Pg.185]

The precision required for clinical analyses has been expressed as the tolerable analytic variability (Y2), and for many tests it is equivalent to a CV of less than 1%. The precision required from an instrument will therefore be less than this, depending on its contribution to the overall variability. [Pg.292]


See other pages where Analysers clinical is mentioned: [Pg.38]    [Pg.40]    [Pg.509]    [Pg.69]    [Pg.587]    [Pg.64]    [Pg.275]    [Pg.233]    [Pg.132]    [Pg.225]    [Pg.373]    [Pg.519]    [Pg.509]    [Pg.975]    [Pg.48]    [Pg.888]    [Pg.186]    [Pg.252]    [Pg.110]    [Pg.260]    [Pg.272]    [Pg.326]    [Pg.427]    [Pg.427]    [Pg.428]    [Pg.428]    [Pg.432]    [Pg.439]    [Pg.462]    [Pg.252]   
See also in sourсe #XX -- [ Pg.252 ]




SEARCH



Analyse

Analyser

Classification of clinical analysers

Clinical analysers batch

© 2024 chempedia.info