Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Buffer standardizing aqueous measurements

Prescriptions for seawater scale buffers are available for a range of salinities. Reliable estuarine pH measurements can be made by calibrating with a buffer of the same salinity as the sample. However, these buffers are difficult to prepare and their use presumes prior knowledge of salinity of the sample. Interpretable measurements of estuarine pH can be made by calibration with lUPAC aqueous RVS or PS standards if the electrode pair is additionally calibrated using a 20%o salinity seawater buffer. The difference between the assigned pH of the seawater buffer and its measured pH(X) value using RVS or PS standards is... [Pg.1232]

The pH meter cell must be standardised with buffers of known pH. Aqueous buffers are sometimes used when the solvent under investigation is a methanol-water mixture. One must then assume that liquid junction potentials thereby introduced are approximately constant from one test solution to another. 2 There is also a risk that transfer of a glass electrode from aqueous to methanolic solution may cause ir-reproducible changes in its asymmetry potential. A better procedure is to prepare standard buffers in methanol. Three such buffers were assigned pH values by the same procedure used for standard aqueous buffers. The assigned values, designated pH (5), should closely approximate pC %). If an unknown solution X is now measured in the standardised cell, the operational definition of its pH is... [Pg.354]

A standard calibration curve is prepared by subsequent addition of small volumes of standard aqueous saturated NO solution (concentration 1.74 mM) into a constant volume of freshly boiled and deoxygenated buffer (pH 7.4 at 37 C). -The current generated after each addition of NO is measured, and the resulting plot of current versus concentration should be linear. The detection limit of a correctly prepared sensor should be about 10 nM for a single-fiber sensor, and 1 to 5 nM for multifiber catheter-protected sensors for in vivo measurement. The detection limit of die cell-culture macrosensor is 5 nM. The response time (time for signal increase from 10-75%) of porphyrinic. sensors is 0.1 ms for micromolar NO concentrations, and 10 ms for nanomolar concentrations. [Pg.246]

From the data in Fig. 4.8b, estimate the shift factors required to displace the data at 0 = 0.5 (consider only this point) so that all runs superimpose on the experiment conducted at 128 C at 0 = 0.5. Either a ruler or proportional dividers can be used to measure displacements. Criticize or defend the following proposition Whether a buffered aqueous solution of H2O2 and 1. containing small amounts of S2O3 and starch, appears blue or colorless depends on both the time and the temperature. This standard general chemistry experiment could be used to demonstrate the equivalency of time and temperature. The pertinent reactions for the iodine clock are... [Pg.266]

Oumada et al. [148] described a new chromatographic method for determining the aqueous pKa of dmg compounds that are sparingly soluble in water. The method uses a rigorous intersolvent pH scale in a mobile phase consisting of a mixture of aqueous buffer and methanol. A glass electrode, previously standardized with common aqueous buffers, was used to measure pH online. The apparent ionization constants were corrected to a zero-cosolvent pH scale. Six sparingly soluble nonsteroidal antiinflammatory weak acids (diclofenac, flurbiprofen, naproxen, ibu-profen, butibufen, fenbufen) were used successfully to illustrate the new technique. [Pg.33]

Gelsema, W. J. de Ligny, C. L. Remijnse, A. G. Blijleven, H. A., pH-measurements in alcohol-water mixtures, using aqueous standard buffer solutions for calibration, Reel. Trav. Chim. Pays-Bas 85, 647-660 (1966). [Pg.261]

Fletsch and Richards [51] determined fluoride in seawater spectrophotometri-cally as the cerium alizarin complex. The cerium alizarin complex and chelate was formed in 20% aqueous acetone at pH 4.35 (sodium acetate buffer) and, after 20-60 min, the extinction measured at 625 nm (2.5 cm cell) against water. The calibration graph was rectilinear for 8-200 ig/l fluoride the mean standard deviation was 10 xg/l at a concentration of 1100 ig/l fluoride. [Pg.72]

Dissolution indicates the rate-limiting step for compound absorption when drugs are administered orally. The solubility of a pharmaceutical compound represents its maximum concentration in an aqueous buffer. Additional compound will not dissolve above this concentration. The solubility value is often heavily dependent upon pH and temperature and is typically measured at physiologically important pH levels and body temperature. The standards for dissolution testing are determined by the United States Pharmacopoeia (USP). Testing typically requires sampling of a solution at 15, 30, 45, and 60 min for immediate-release products. /./Pl.C is ideally suited for use in conjunction with USP apparatus types I or II and can rapidly analyze multiple time points or replicate samples. [Pg.185]

Another CE method was developed and employed for the separation of the components of FD C Red No. 3 (erythrosine). The separations were also carried out by RP-HPLC and the efficacy of the methods was compared. The chemical structures of the main components of the dye are shown in Fig. 3.167. The components of erythrosine were separated in a fused-silica capillary (43 cm effective length X 75 /xm i.d.). The running buffer was 50 mM sodium tetraborate, 25 mM SDS (pH 9.3). Analytes were detected at 516 nm. HPLC measurements were realized in an octylsilica column (150 X 4.6 mm i.d. particle size 5 /xm) at 35°C. Solvent A was 0.1 M aqueous ammonium acetate and solvent B consisted of methanol. The gradient programme was 0 min, 55 per cent A 20 min 35 per cent A 21 min, 100 per cent B, final hold, 4 min. The flow rate was 1 ml/min. The separations of the components of the standard mixture (left) and those of a real sample (right) by CE are shown in Fig. 3.168. The electropherograms clearly illustrate that the method allows the baseline separation of the dye components even in real commercial samples. The main... [Pg.548]

Elemental composition Cr 52.00%, 0 48.00%. The compound may he identified from its dark red color. Other color phases are noted above. Chromium may he measured in the aqueous phase hy AA, ICP or x-ray techniques, or in the solid phase hy x-ray methods. Hexavalent chromium (Cr6+) may he analyzed hy ion chromatography. For this, the aqueous sample is adjusted to pH 9 to 9.5 with a concentrated buffer (ammonium sulfate and ammonium hydroxide mixture) and mixed into the eluent stream of the buffer. Cr " is separated from Cr + on a column, and derivatized with an azide dye as a colored product measured at 530 nm, which is identified from its retention time. (APHA, AWWA, and WEF. 1999. Standard Methods for The Examination of Water and Wastewater, 20th ed., Washington, DC American Public Health Association.)... [Pg.228]

Hydrazine reduces iodine to hydrogen iodide. Thus, an excess of standard solution of iodine is added to a measured volume of aqueous hydrazine solution and the excess iodine is back titrated at pH 7.0 to 7.2 (buffered by sodium bicarbonate) against a standard solution of sodium thiosulfate using starch indicator. [Pg.348]

The above conceptual and operational pH definitions for solutions in non-aqueous and mixed solvents are very similar to those for aqueous solutions [16]. At present, pH values are available for the RVS and some primary standards in the mixtures between water and eight organic solvents (see 5 in Section 6.2) [17]. If a reliable pH standard is available for the solvent under study, the pH can be determined with a pH meter and a glass electrode, just as in aqueous solutions. However, in order to apply the IUPAC method to the solutions in neat organic solvents or water-poor mixed solvents, there are still some problems to be solved. One of them is that it is difficult to get the RVS in such solvents, because (i) the solubility of KHPh is not enough and (ii) the buffer action of KHPh is too low in solutions of an aprotic nature [18].8) Another problem is that the response of the glass electrode is often very slow in non-aqueous solvents,9 although this has been considerably improved by the use of pH-ISFETs [19]. Practical pH measurements in non-aqueous solutions and their applications are discussed in Chapter 6. [Pg.79]

Henrickson and Selmer-Olson [18] applied an autoanalyser to the determination of nitrate and nitrite in soil extracts. In an autoanalyser, the water sample, buffered to pH 8.6 with aqueous ammonia-ammonium chloride, is passed through a copperised cadmium reductor column. The nitrite formed is reacted with sulfuric acid and N-l-naphthylethylenediamine, and the extinction of the azo dye is measured at 520 nm. For soil extracts, the range and standard deviation are 0.5-1.0 and 0.007mg/1, respectively. [Pg.159]

The primary method is applied by national metrology institutes to assign conventional pH values to a limited number of primary standard (PS) buffer solutions in dilute aqueous solutions. The experimental details are given in [8, 9] where national standard measurement devices for pH in Denmark and Germany are described. [Pg.208]

Aqueous invertase solutions (soluble and immobilized enzyme) were incubated in acetate buffer (0.010 M, pH5.5) for 20,40,60,80, and 120 min at 30,37,40,45,50, and 55°C. After incubation at each specified time, both forms were assayed for residual activity according to the standard procedure (see Standard Assay for Measuring Invertase Activity). [Pg.148]

Before using the pH electrode, it should be calibrated using two (or more) buffers of known pH. Many standard buffers are commercially available, with an accuracy of 0.01 pH unit. Calibration must be performed at the same temperature at which the measurement will be made care must be taken to match the temperature of samples and standards. The exact procedure depends on the model of pH meter used. Modern pH meters, such as the one shown in Figure 5.8, are microcomputer-controlled, and allow double-point calibration, slope calculation, temperature adjustment, and accuracy to +0.001 pH unit, all with few basic steps. The electrode must be stored in an aqueous solution when not in use, so that the hydrated gel layer of the glass does not dry out. A highly stable response can thus be obtained over long time periods. As with other ion-selective electrodes, the operator should consult the manufacturer s instructions for proper use. Commercial glass electrodes are remarkably... [Pg.176]


See other pages where Buffer standardizing aqueous measurements is mentioned: [Pg.266]    [Pg.276]    [Pg.20]    [Pg.120]    [Pg.2378]    [Pg.704]    [Pg.15]    [Pg.27]    [Pg.3]    [Pg.58]    [Pg.59]    [Pg.305]    [Pg.93]    [Pg.331]    [Pg.99]    [Pg.28]    [Pg.138]    [Pg.134]    [Pg.166]    [Pg.132]    [Pg.208]    [Pg.40]    [Pg.8]    [Pg.259]    [Pg.108]    [Pg.111]   


SEARCH



Standard measures

Standards aqueous

© 2024 chempedia.info