Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Blanks and calibration

LOCC water blanks are run between sets of standards and samples to give an analysis baseline. If samples containing dissolved organic carbon in excess of 400 pmol/L DOC are to be analysed, a significant amount of sample carryover may be encountered, and two LOCC water samples in succession may be required to establish a true baseline. Standards are usually run after every ten samples and an example of a typical trace from an analysis is shown in Fig. 16-6. [Pg.434]


In this section, the preparation of test samples, blanks and calibration samples has to described, together with an indication of the minimum number of samples needed. If necessary, precautions should be mentioned, for instance to avoid cross-contamination of samples in the case of volatiles or to minimise chemical degradation in the case of unstable analytes etc. [Pg.312]

Here it is necessary to provide details as to how the analytical measurement of test, blank and calibration samples is executed and how the obtained data are evaluated. The measured concentration of the analyte obtained in this way may need further transformation into a different dimension and this should also be addressed in this section. [Pg.312]

In case of measured concentrations exceeding the QM limit, confirmation of the identity of carbonyl chloride is carried out by diode array detection. This is achieved by recording the spectral profiles of the samples, blanks and calibration samples over the wavelength range of 200-320 nm at the front, apex and tail of the peak identified as the carbonyl chloride derivative. Box can be identified as having an absorbance peak maximum at 272 nm and a minimum at 245 nm with an absorbance ratio of 40 (at 260 nm) 100 (at 272 nm) 70 (at 280 nm). If the peak is pure, the overlaid spectral profiles of the front, apex and tail of the peak should be identical. Therefore, if the three profiles are normalized, they should superimpose on top of each other. [Pg.327]

Some requirements cannot be fulfilled in routine analyses, because of being too costly, time expensive and large-scaled, e.g. general requirements on sample blanks and calibration curves, where among others the following procedures have to be followed and/or information must be given ... [Pg.161]

An internal standard is a substance that is added in a constant amount to all samples, blanks, and calibration standards in an analysis. Alternatively, it may be a major constituent of samples and standards that is present in a large enough amount that its concentration can be assumed to be the same in all cases. Calibration then involves plotting the ratio of the analyte signal to the internal-standard signal as a function of the analyte concentration of the standards. This ratio for the samples is then used t(t obtain their analyte concentrations from a calibration curve. [Pg.17]

Is There a Way to Couple the Blank and Calibration Approaches to Find Xloq ... [Pg.54]

Measure blank and calibration solutions in similar fashion. [Pg.387]

Treat 100 ml of the sample for investigation with 50 ml hydrochloric acid (1.17 g/ml) and boil for 15 minutes at the reflux. Treat blanks and calibration solutions in the same way. [Pg.389]

Using the exact conditions and time basis as were used in the blank and calibration, inject the sample into the chromatograph. Disregarding peaks (if any) before propane. [Pg.573]

Using the exact conditions that were used in the blank and calibration runs (see 11.1 and 11.2), and following the rigorously defined schedule (see 11.1), iryect 1 pL of the diluted crude oil plus internal standard mixture into the chromatograph. Record the area slices of each time interval through the end of the run. [Pg.874]

In the process of performing a spectrophotometric determination of Ee, an analyst prepares a calibration curve using a single-beam spectrometer, such as a Spec-20. After preparing the calibration curve, the analyst drops the cuvette used for the method blank and the standards. The analyst acquires a new cuvette, measures the absorbance of the sample, and determines the %w/w Ee in the sample. Will the change in cuvette lead to a determinate error in the analysis Explain. [Pg.450]

Procedure (ii). Make certain that the instrument is fitted with the correct burner for an acetylene-nitrous oxide flame, then set the instrument up with the calcium hollow cathode lamp, select the resonance line of wavelength 422.7 nm, and adjust the gas controls as specified in the instrument manual to give a fuel-rich flame. Take measurements with the blank, and the standard solutions, and with the test solution, all of which contain the ionisation buffer the need, mentioned under procedure (i), for adequate treatment with de-ionised water after each measurement applies with equal force in this case. Plot the calibration graph and ascertain the concentration of the unknown solution. [Pg.807]

In case of experimental calibration, the uncertainty of both the blank and the calibration coefficient, U (JBL, fi), have to be consider, e.g. according to... [Pg.229]

Currie LA (1997) Detection International update, and some emerging dilemmas involving calibration, the blank, and multiple detection decisions. Chemometrics Intell Lab Syst 37 151... [Pg.237]

El-Brashy [51] reported the determination of primaquine and other antimalarials via charge-transfer complexes. Powdered sample of primaquine phosphate was dissolved in water and the solution was adjusted to an alkaline pH with 6 M ammonia and extracted with chloroform. The extract was dried with anhydrous sodium sulfate, filtered, and evaporated to dryness under nitrogen and the residue was dissolved in acetonitrile. Portions of the solution were mixed with 0.2% 7,7,8,8-tetracyanoquinodimethane, diluted with acetonitrile, and set aside for 10 min before the absorbance was measured at 845 nm versus a reagent blank. The calibration graphs were linear from 0.4 to 3 pg/mL and recovery was 98%. [Pg.182]

In the bioanalytical studies, the basic calibration should be prepared in the same biological matrix as in the samples of the intended study, which can be achieved by spiking the matrix with known concentration of the analyte. In this case, a blank sample, a zero sample (blank and internal standard), six to eight nonzero samples covering the expected range (including the anticipated QL) should be evaluated as part of the linearity study [27]. [Pg.249]

An ideal method for the preconcentration of trace metals from natural waters should have the following characteristics it should simultaneously allow isolation of the analyte from the matrix and yield an appropriate enrichment factor it should be a simple process, requiring the introduction of few reagents in order to minimise contamination, hence producing a low sample blank and a correspondingly lower detection limit and it should produce a final solution that is readily matrix-matched with solutions of the analytical calibration method. [Pg.303]

In practical situations the absorbance of a sample is determined by making two measurements, the first to determine 70 and the second to determine I. The determination of I0 is used to cancel a large number of experimental factors that could affect the result. When measuring I0 the sample container must closely match the unknown container in all ways except for the analyte content. The cuvettes should be a matched pair if a double beam instrument is used and the same cuvette can be used for both the blank and sample with a single beam instrument. The blank solution filling the cuvette should be identical to the solvent that the sample is dissolved in, except for the sample itself. If done correctly, the least-squares line for the calibration graph will come very close to the 0,0 point on the graph. [Pg.131]

Since the time between reading the blank and reading the sample can be significant, the instrument may lose its calibration due to minor electrical fluctuations in either the source or detector. [Pg.521]

The ability to make analytical measurements depends intimately on the availability of well-defined standards and calibrants. Many measurements of analytes in seawater (such as dissolved organic carbon and dissolved organic nitrogen) cannot be compared among laboratories because of the lack of appropriate reference materials and blanks for instrument calibration and testing. Intercomparison exercises are critical (NRC, 1993, p. 75). [Pg.24]

In the text which follows we shall examine in numerical detail the decision levels and detection limits for the Fenval-erate calibration data set ( set-B ) provided by D. Kurtz (17). In order to calculate said detection limits it was necessary to assign and fit models both to the variance as a function of concentration and the response (i.e., calibration curve) as a function of concentration. No simple model (2, 3 parameter) was found that was consistent with the empirical calibration curve and the replication error, so several alternative simple functions were used to illustrate the approach for calibration curve detection limits. A more appropriate treatment would require a new design including real blanks and Fenvalerate standards spanning the region from zero to a few times the detection limit. Detailed calculations are given in the Appendix and summarized in Table V. [Pg.58]

Fenvalerate Detection Limits. To the extent that detection limits require knowledge of the calibration curve and random error (for x) as a function of concentration, all of the foregoing discussion is relevant — both for detection and estimation. However, curve shape and errors where x x, are relatively unimportant at the detection limit, in contrast to direct observations of the initial slope and the blank and its variability. (It will be seen that the initial observation in the current data set exceeded the ultimate detection limit by more than an order of magnitude )... [Pg.63]


See other pages where Blanks and calibration is mentioned: [Pg.179]    [Pg.34]    [Pg.212]    [Pg.334]    [Pg.341]    [Pg.434]    [Pg.81]    [Pg.575]    [Pg.179]    [Pg.34]    [Pg.212]    [Pg.334]    [Pg.341]    [Pg.434]    [Pg.81]    [Pg.575]    [Pg.48]    [Pg.110]    [Pg.129]    [Pg.129]    [Pg.411]    [Pg.810]    [Pg.285]    [Pg.286]    [Pg.954]    [Pg.228]    [Pg.229]    [Pg.75]    [Pg.36]    [Pg.182]    [Pg.58]   


SEARCH



Blank

Blank, blanking

Blanking

Calibration blank

Calibration, blank determination and calculation procedures

© 2024 chempedia.info