Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Zero sample

Zero sampling error Eq. (1.6) reduces to Vreprod = V repeat-Independent individual samples/measurements in the sense of Fig. 1.5. A result that is nearer to the SL(/ and an accepted risk of 5% (alpha/2 = 0.05 for the single-sided test use the alpha = 0.1 column in the t-table). [Pg.266]

In PAMPA measurements each well is usually a one-point-in-time (single-timepoint) sample. By contrast, in the conventional multitimepoint Caco-2 assay, the acceptor solution is frequently replaced with fresh buffer solution so that the solution in contact with the membrane contains no more than a few percent of the total sample concentration at any time. This condition can be called a physically maintained sink. Under pseudo-steady state (when a practically linear solute concentration gradient is established in the membrane phase see Chapter 2), lipophilic molecules will distribute into the cell monolayer in accordance with the effective membrane-buffer partition coefficient, even when the acceptor solution contains nearly zero sample concentration (due to the physical sink). If the physical sink is maintained indefinitely, then eventually, all of the sample will be depleted from both the donor and membrane compartments, as the flux approaches zero (Chapter 2). In conventional Caco-2 data analysis, a very simple equation [Eq. (7.10) or (7.11)] is used to calculate the permeability coefficient. But when combinatorial (i.e., lipophilic) compounds are screened, this equation is often invalid, since a considerable portion of the molecules partitions into the membrane phase during the multitimepoint measurements. [Pg.138]

Similarly, for the reverse direction we can adopt a (highest) limit-perturbation xg on the high- r tail which separates the regions of perfect sampling (x < xg) and zero sampling (x > xg). The same analysis leads to... [Pg.217]

In the bioanalytical studies, the basic calibration should be prepared in the same biological matrix as in the samples of the intended study, which can be achieved by spiking the matrix with known concentration of the analyte. In this case, a blank sample, a zero sample (blank and internal standard), six to eight nonzero samples covering the expected range (including the anticipated QL) should be evaluated as part of the linearity study [27]. [Pg.249]

We also vary the sample size (from 1 to 2 y ) between runs at a given P and T. This allows extrapolation of peak areas to zero sample size as shown in Figure 4. The flow rate within the column must be constant for an isotherm. We record the following parameters ambient pressure and temperature, inlet pressure, outlet pressure, column temperature, retention time for both air and water, water peak area, ambient flow rate, regulator pressure, sample sizes, detector current and temperature, injector temperature, and attenuation. [Pg.369]

Figure 3 is the absorbance spectrum of a sample of the ambient laboratory air drawn into the cell. Here, in accord with the usual procedure, the initially determined spectrum was first corrected for radiation that had reached the detector without having passed through the sample (room temperature background radiation entering the optical path via imperfect optical components and nonoverlap of the source and detector pupils and fields), ratioed against a zero-sample spectrum, and converted to absorbance. Trace (A) shows the spectrum from 3600-600 cm l. The massive absorbances seen here truncated at 1% transmission are due to water vapor and to carbon dioxide. [Pg.165]

This sample represents the nearly zero sample (i.e., virtually no oil), which is used to establish an accurate intercept. [Pg.21]

Benzene was passed at various temperatures through a gas chromatography column packed with silica gel and the following specific retention volumes (extrapolated to zero sample size) were measured ... [Pg.282]

Add 1 pi 2N NaOH to each plus and minus sample and to the zero samples. [Pg.49]

Figure 5 Field modulation detected stochastic (FMDST) ENDOR sequence. The upper level indicates monitoring of the EPR signal prior (baseline) and (usually) after (signal) application of the rf pulse. This pulse is applied for time and at frequency v, which is a randomly selected frequency among the total munber of possible values determined by the rf range and number of spectral points. There is also a delay time, fj, that can range from zero (sample concurrent with apphcation of the rf pulse) to > fu (sample immediately after application of the rf pulse, or later) this... Figure 5 Field modulation detected stochastic (FMDST) ENDOR sequence. The upper level indicates monitoring of the EPR signal prior (baseline) and (usually) after (signal) application of the rf pulse. This pulse is applied for time and at frequency v, which is a randomly selected frequency among the total munber of possible values determined by the rf range and number of spectral points. There is also a delay time, fj, that can range from zero (sample concurrent with apphcation of the rf pulse) to > fu (sample immediately after application of the rf pulse, or later) this...
Figure 9.13 Hydrogen ion concentrations (Q. ) at intervals after a test meal (mean results are shown 2SEM) the zero samples were taken just before the meal was begun, and the 1 hour sample just before 45 cm of water. Figure 9.13 Hydrogen ion concentrations (Q. ) at intervals after a test meal (mean results are shown 2SEM) the zero samples were taken just before the meal was begun, and the 1 hour sample just before 45 cm of water.
Besides this method of extrapolation towards c = 0, n = 0 of the static isotherm, it is possible to determine the Henry constant directly from the specific retention volume for zero sample size Vm if the elution time of the substance from the column does not depend on the sample size at a constant flow rate. In this case K s c,i = Vn,. [Pg.679]

The graphical method does not take into account that there may be oxygen-consuming compounds present that give zero sample DO and consume oxygen from the dilution water, i.e., S may effectively be less than zero. [Pg.5080]

Use six lots of blank monkey plasma to prepare six lower limit of quantification quality control (LLOQ QC) samples, six zero samples (only contain IS), and six blank plasma sample. Follow sample preparation in Section 7.3.5 and LC-MS/ MS analysis in Section 7.3.6. [Pg.182]

Selectivity assessment six blank matrices, zero samples, LLOQ QC samples, and ULOQ without IS samples are used to evaluate the presence or absence of interference and lot-to-lot variation (Figures 7.2 and 7.3). [Pg.182]

Sample analysis sequence should contain a blank sample, a zero sample, two sets of calibration curves (one at the beginning and one at the end of the run), and QC samples as stated and unknown samples. QC samples should be at least 5% of the number of unknown samples or six total QCs, whichever is greater. [Pg.185]


See other pages where Zero sample is mentioned: [Pg.59]    [Pg.24]    [Pg.200]    [Pg.442]    [Pg.169]    [Pg.281]    [Pg.417]    [Pg.153]    [Pg.175]    [Pg.104]    [Pg.25]    [Pg.43]    [Pg.40]    [Pg.105]    [Pg.6546]    [Pg.96]    [Pg.37]    [Pg.324]    [Pg.111]    [Pg.6545]    [Pg.6545]    [Pg.232]    [Pg.79]    [Pg.389]    [Pg.5075]    [Pg.44]    [Pg.527]    [Pg.560]    [Pg.561]    [Pg.578]    [Pg.178]    [Pg.180]   
See also in sourсe #XX -- [ Pg.180 ]




SEARCH



Sampling zero-dimensional

Zero-contamination sampling

© 2024 chempedia.info