Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analyzer controller

Lewis J W, Yee G G and Kliger D S 1987 Implementation of an optical multichannel analyzer controller for nanosecond flash photolysis measurements Rev. Sol. Instrum. 58 939-44... [Pg.2969]

Simulation of Dynamic Models Linear dynamic models are particularly useful for analyzing control-system behavior. The insight gained through linear analysis is invaluable. However, accurate dynamic process models can involve large sets of nonlinear equations. Analytical solution of these models is not possible. Thus, in these cases, one must turn to simulation approaches to study process dynamics and the effect of process control. Equation (8-3) will be used to illustrate the simulation of nonhnear processes. If dcjdi on the left-hand side of Eq. (8-3) is replaced with its finite difference approximation, one gets ... [Pg.720]

If produc t moisture is measured off-line, analytical results can be used to adjust K and Tb manually. If an on-hne analyzer is used, the analyzer controller would be most effective in adjusting the bias Tb, as shown in the figure. [Pg.751]

Sample Transport Transport time, the time elapsed between sample withdrawal from the process and its introduction into the analyzer, shoiild be minimized, particiilarly if the analyzer is an automatic analyzer-controller. Any sample-transport time in the analyzer-controller loop must be treated as equivalent to process dead time in determining conventional feedback controller settings or in evaluating controller performance. Reduction in transport time usually means transporting the sample in the vapor state. [Pg.767]

One of the difficulties with optimal control theory is in identifying the underlying physical mechanism, or mechanisms, leading to control. Methods [2, 7, 9, 14, 26-29], that utilize a small number of interfering pathways reveal the mechanism by construction. On the other hand, while there have been many successful experimental and theoretical demonstrations of control based on OCT, there has been little analytical work to reveal the mechanism behind the complicated optimal pulses. In addition to reducing the complexity of the pulses, the many methods for imposing explicit restrictions on the pulses, see Section II.B, can also be used to dictate the mechanisms that will be operative. However, in this section we discuss some of the analytic approaches that have been used to understand the mechanisms of optimal control or to analytically design optimal pulses. Note that we will not discuss numerical methods that have been used to analyze control mechanisms [145-150]. [Pg.71]

Analyzing control materials alongside the test samples greatly improves proficiency in mycotoxin analysis. Certified reference materials (CRMs) represent ideal control materials, due to their statement of uncertainty and traceability, and they should be routinely used as much as possible. Unfortunately, as outstanding as the improvements made in the last decade have been, even though the list of CRMs in the area of mycotoxins is rather long, it is still insufficient. A list of the available reference materials in the mycotoxins area is reported in Table 1 the issue has been reviewed by Boenke (27). [Pg.497]

EG G PAR (USA), Model 273 A, Potentiostat/Galvanostat volt-ammetric analyzer controlled by PC equipped with a data acquisition and treatment software to record the signal generated in the electrochemical cell for DPY measurements. A 25 mL glass cell at 25°C with the carbon paste biosensor, Ag/AgCl (3.0 mol L-1 KC1) reference electrode, and a platinum wire as auxiliary electrode to perform the volt-ammetric measurements. [Pg.1115]

Figure 8-54 shows a depropanizer controlled by reflux and boil-up ratios. The actual mechanism through which these ratios are manipulated is as D/(L + D) and B/(V + B), where L is reflux flow and V is vapor boil-up, which decouples the temperature loops from the liquid-level loops. Column pressure here is controlled by flooding both the condenser and accumulator however, there is no level controller on the accumulator, so this arrangement will not function with an overloaded condenser. Temperatures are used as indications of composition in this column because of the substantial difference in boiling points between propane and butanes. However, off-key components such as ethane do affect the accuracy of the relationship, so that an analyzer controller is used to set the top temperature controller (TC) in cascade. [Pg.43]

A cocurrent evaporator train with its controls is illustrated in Fig. 8-56. The control system applies equally well to countercurrent or mixed-feed evaporators, the principal difference being the tuning of the dynamic compensator ), which must be done in the field to minimize the short-term effects of changes in feed flow on product quality. Solid concentration in the product is usually measured as density feedback trim is applied by the analyzer controller AC adjusting the slope rn of the density function, which is the only term related to x . This recalibrates the system whenever v,. must move to a new set point. [Pg.45]

Analyzer controllers in a feedback configuration can only be considered when the dead time caused by analysis update is less than the response time of the process. The composition controller provides a feedback correction in response to feed composition changes, pressure variations, or changes in tower efficiencies. [Pg.242]

In part (a) of Figure 2.85, the analyzer controller (ARC) uses the chromatographic measurement to manipulate the reflux flow by adjusting the set point to the reflux flow controller (FRC). Controllability of the process is degraded by the dead time between measurement updates. [Pg.242]

Part (c) in Figure 2.85 illustrates a triple cascade loop, where a temperature controller is the slave of an analyzer controller while the reflux flow is cascaded to temperature. Because temperature is an indicator of composition at constant pressure, the analyzer controller serves only to correct for variations in feed composition. Cascade loops will work only if the slave is faster than the master, which adjusts its set point. Another important consideration in all cascade systems (not shown in Figure 2.85) is that an external reset is needed to prevent the integral mode in the master from saturating, when that output is blocked from reaching and modulating the set point of the slave (when the slave is switched to local set point). [Pg.243]

Part (d) of Figure 2.85 illustrates a limit control configuration where the analyzer controller is overruled by temperature when it reaches its high limit. The reason for this limit is energy conservation, because no additional stripping of the light component can be accomplished, once... [Pg.243]

Bypass-stream transport is a method for maintaining high sample transport velocity to minimize transportation lag. This method is used when samples are vaporized at the sampling tap and no facilities exist for returning the vapor to the process. If the sample bypass is piped to a drain or vent, this will not only waste the process material but might also pollute the environment. Therefore, the use of a fast bypass-return loop is preferred. After selecting the appropriate sample transport method, the sample time lag should be calculated and used in the tuning of the analyzer controller. [Pg.330]

On their arrival at the OPCW Laboratory, all seal numbers and the end pattern of the fiber-optic seals will be compared to the seal numbers and photographs received from the inspection site, if so requested in the presence of an ISP representative. Unpacking will be performed in a dedicated area for handling of authentic samples in the OPCW Laboratory. The vials containing the authentic sample splits will be weighted but left closed and sealed as received from the inspection site. For dispatch to each designated laboratory, one authentic sample split is packed together with a pre-analyzed control sample and the respective matrix blank in a transport container in the same manner as described above. A... [Pg.45]


See other pages where Analyzer controller is mentioned: [Pg.64]    [Pg.747]    [Pg.747]    [Pg.576]    [Pg.226]    [Pg.29]    [Pg.526]    [Pg.140]    [Pg.813]    [Pg.13]    [Pg.433]    [Pg.43]    [Pg.583]    [Pg.395]    [Pg.244]    [Pg.250]    [Pg.9]    [Pg.571]    [Pg.571]    [Pg.777]    [Pg.151]    [Pg.225]    [Pg.918]   


SEARCH



Analyzing Testing and Quality Control

Control analyzer

Control analyzer

Control analyzer/temperature

Define, measure, analyze control

Define, measure, analyze improve and control

Mass Analyzers in Routine Doping Controls

Sample Points for Bottom Product Analyzer Control

Sample Points for Top Product Analyzer Control

Time control, enzyme analyzers

© 2024 chempedia.info