Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical methods standard addition

In Section 8.2.8 we have discussed the standard addition method as a means to quantitate an analyte in the presence of unknown matrix effects cf. Section 13.9). While the matrix effect is corrected for, the presence of other emalytes may still interfere with the analysis. The method can be generalized, however, to the simultaneous analysis of p analytes. Multiple standard additions are applied in order to determine the analytes of interest using many q > p) analytical sensors. It... [Pg.367]

The method for chloroacetanilide soil metabolites in water determines concentrations of ethanesulfonic acid (ESA) and oxanilic acid (OXA) metabolites of alachlor, acetochlor, and metolachlor in surface water and groundwater samples by direct aqueous injection LC/MS/MS. After injection, compounds are separated by reversed-phase HPLC and introduced into the mass spectrometer with a TurboIonSpray atmospheric pressure ionization (API) interface. Using direct aqueous injection without prior SPE and/or concentration minimizes losses and greatly simplifies the analytical procedure. Standard addition experiments can be used to check for matrix effects. With multiple-reaction monitoring in the negative electrospray ionization mode, LC/MS/MS provides superior specificity and sensitivity compared with conventional liquid chromatography/mass spectrometry (LC/MS) or liquid chromatography/ultraviolet detection (LC/UV), and the need for a confirmatory method is eliminated. In summary,... [Pg.349]

Analyte dilution sacrifices sensitivity. Matrix matching can only be applied for simple matrices, but is clearly not applicable for complex matrices of varying composition. Accurate correction for matrix effect is possible only if the IS is chosen with a mass number as close as possible to that of the analyte elements). Standard addition of a known amount of the element(s) of interest is a safe method for samples of unknown composition and thus unknown matrix effect. Chemical separations avoid spectral interference and allow preconcentration of the analyte elements. Sampling and sample preparation have recently been reviewed [4]. [Pg.589]

The internal standard method uses an internal standard substance added in a constant amount to all standards and the sample. Area ratio of analyte peak to internal standard peak is plotted vs. concentration of analyte. The standard additions method uses the addition of the analyte in increasing amounts to the sample. Peak area is plotted vs. concentration added and the line is extrapolated to zero peak area to get the sample concentration. [Pg.535]

This quotient is the critical quantity for the practical application of an analytical method. In addition, the relative standard deviation of an analyte result sometimes must not exceed a specified maximum. These two requirements lead to the. .. pragmatic setting of so-called limits of quantitation. Since such limits depend on the particular value observed, they cannot be regarded as meaningful characteristics of the method [MULLER et al., 1994], In brief conclusion, the analyst is well advised not to report determinations below this limit, because they will probably be biased owing to unacceptably high random errors. [Pg.68]

The standard addition method is often used in cases when it is not possible to obtain suitable blank matrices one example is the analysis of endogenous compounds in body fluids. The approach is to add different weights of analyte to the unknown sample, which initially contains an unknown concentration of the analyte. After the chromatographic analysis, peak areas (or heights) are plotted versus the added concentration. Extrapolation of the calibration plot provides the original unknown concentration of the analyte. A standard addition method that possesses even greater accuracy and precision is obtained if one incorporates an internal standard [48],... [Pg.27]

Illustration showing the method of standard additions in which separate aliquots of sample are diluted to the same final volume. One aliquot of sample is spiked with a known volume of a standard solution of analyte before diluting to the final volume. [Pg.111]

The successful application of an external standardization or the method of standard additions, depends on the analyst s ability to handle samples and standards repro-ducibly. When a procedure cannot be controlled to the extent that all samples and standards are treated equally, the accuracy and precision of the standardization may suffer. For example, if an analyte is present in a volatile solvent, its concentration will increase if some solvent is lost to evaporation. Suppose that you have a sample and a standard with identical concentrations of analyte and identical signals. If both experience the same loss of solvent their concentrations of analyte and signals will continue to be identical. In effect, we can ignore changes in concentration due to evaporation provided that the samples and standards experience an equivalent loss of solvent. If an identical standard and sample experience different losses of solvent. [Pg.115]

An analytical method is standardized by determining its sensitivity. There are several approaches to standardization, including the use of external standards, the method of standard addition. [Pg.130]

Quantitative Analysis for a Single Analyte The concentration of a single analyte is determined by measuring the absorbance of the sample and applying Beer s law (equation 10.5) using any of the standardization methods described in Chapter 5. The most common methods are the normal calibration curve and the method of standard additions. Single-point standardizations also can be used, provided that the validity of Beer s law has been demonstrated. [Pg.400]

The generalized standard addition method (GSAM) extends the analysis of mixtures to situations in which matrix effects prevent the determination of 8x and 8y using external standards.When adding a known concentration of analyte to a solution containing an unknown concentration of analyte, the concentrations usually are not additive (see question 9 in Chapter 5). Conservation of mass, however, is always obeyed. Equation 10.11 can be written in terms of moles, n, by using the relationship... [Pg.402]

When possible, quantitative analyses are best conducted using external standards. Emission intensity, however, is affected significantly by many parameters, including the temperature of the excitation source and the efficiency of atomization. An increase in temperature of 10 K, for example, results in a 4% change in the fraction of Na atoms present in the 3p excited state. The method of internal standards can be used when variations in source parameters are difficult to control. In this case an internal standard is selected that has an emission line close to that of the analyte to compensate for changes in the temperature of the excitation source. In addition, the internal standard should be subject to the same chemical interferences to compensate for changes in atomization efficiency. To accurately compensate for these errors, the analyte and internal standard emission lines must be monitored simultaneously. The method of standard additions also can be used. [Pg.438]

Quantitative Analysis Using the Method of Standard Additions Because of the difficulty of maintaining a constant matrix for samples and standards, many quantitative potentiometric methods use the method of standard additions. A sample of volume, Vx) and analyte concentration, Cx, is transferred to a sample cell, and the potential, (ficell)x) measured. A standard addition is made by adding a small volume, Vs) of a standard containing a known concentration of analyte, Cs, to the sample, and the potential, (ficell)s) measured. Provided that Vs is significantly smaller than Vx, the change in sample matrix is ignored, and the analyte s activity coefficient remains constant. Example 11.7 shows how a one-point standard addition can be used to determine the concentration of an analyte. [Pg.488]

Quantitative analytical methods using FIA have been developed for cationic, anionic, and molecular pollutants in wastewater, fresh waters, groundwaters, and marine waters, several examples of which were described in the previous section. Table 13.2 provides a partial listing of other analytes that have been determined using FIA, many of which are modifications of conventional standard spectropho-tometric and potentiometric methods. An additional advantage of FIA for environmental analysis is its ability to provide for the continuous, in situ monitoring of pollutants in the field. ... [Pg.655]

Spike recoveries for samples are used to detect systematic errors due to the sample matrix or the stability of the sample after its collection. Ideally, samples should be spiked in the field at a concentration between 1 and 10 times the expected concentration of the analyte or 5 to 50 times the method s detection limit, whichever is larger. If the recovery for a field spike is unacceptable, then a sample is spiked in the laboratory and analyzed immediately. If the recovery for the laboratory spike is acceptable, then the poor recovery for the field spike may be due to the sample s deterioration during storage. When the recovery for the laboratory spike also is unacceptable, the most probable cause is a matrix-dependent relationship between the analytical signal and the concentration of the analyte. In this case the samples should be analyzed by the method of standard additions. Typical limits for acceptable spike recoveries for the analysis of waters and wastewaters are shown in Table 15.1. ... [Pg.711]

Analytical Methods. A classical and stiU widely employed analytical method is iodimetric titration. This is suitable for determination of sodium sulfite, for example, in boiler water. Standard potassium iodate—potassium iodide solution is commonly used as the titrant with a starch or starch-substitute indicator. Sodium bisulfite occurring as an impurity in sodium sulfite can be determined by addition of hydrogen peroxide to oxidize the bisulfite to bisulfate, followed by titration with standard sodium hydroxide (279). [Pg.149]

Standards used to constmct a cahbration curve must be prepared such that the matrix of the standard is identical to the sample s matrix because the values of the parameters k and b associated with a linear cahbration curve are matrix dependent. Many areas of chemical analysis are plagued by matrix effects, and it is often difficult to duphcate the sample matrix when preparing external standards. Because it is desirable to eliminate matrix effects, cahbration in the sample matrix itself can be performed. This approach is called the standard addition method (SAM) (14). In this method, the standards are added to the sample matrix and the response of the analyte plus the standard is monitored as a function of the added amount of the standard. The initial response is assumed to be Rq, and the relationship between the response and the concentration of the analyte is... [Pg.427]

The complex of the following destmctive and nondestmctive analytical methods was used for studying the composition of sponges inductively coupled plasma mass-spectrometry (ICP-MS), X-ray fluorescence (XRF), electron probe microanalysis (EPMA), and atomic absorption spectrometry (AAS). Techniques of sample preparation were developed for each method and their metrological characteristics were defined. Relative standard deviations for all the elements did not exceed 0.25 within detection limit. The accuracy of techniques elaborated was checked with the method of additions and control methods of analysis. [Pg.223]

Standard addition. A known amount of the constituent being determined is added to the sample, which is then analysed for the total amount of constituent present. The difference between the analytical results for samples with and without the added constituent gives the recovery of the amount of added constituent. If the recovery is satisfactory our confidence in the accuracy of the procedure is enhanced. The method is usually applied to physico-chemical procedures such as polarography and spectrophotometry. [Pg.132]

Adequate precision and accuracy are only likely to be achieved if some standardization procedure is employed and the nature of this, internal or external standards or the method of standard additions, needs to be chosen carefully. If internal standardization procedures are adopted then appropriate compound(s) must be chosen and their effect on the chromatographic and mass spectrometry methods assessed. The ideal internal standard is an isotopically labelled analogue of the analyte but, although there are a number of commercial companies who produce a range of such molecules, these are not always readily available. An analytical laboratory is then faced with the choice of carrying out the synthesis of the internal standard themselves or choosing a less appropriate alternative with implications on the accuracy and precision of the method to be developed. [Pg.270]

The fact that APCl and electrospray are soft ionization techniques is often advantageous because the molecular ion alone, in conjunction with HPLC separation, often provides adequate selectivity and sensitivity to allow an analytical method to be developed. Again, method development is important, particularly when more than one analyte is to be determined, when the effect of experimental parameters, such as pH, flow rate, etc., is not likely to be the same for each. Electrospray, in particular, is susceptible to matrix effects and the method of standard additions is often required to provide adequate accuracy and precision. [Pg.290]

Standard additions A method of relating the intensity of signal from an analyte measured in an unknown to the amount of analyte present. This technique is designed to take matrix effects into account. [Pg.311]

The purpose of this chapter is to describe the analytical methods that are available for detecting, measuring, and/or monitoring methyl parathion, its metabolites, and other biomarkers of exposure and effect to methyl parathion. The intent is not to provide an exhaustive list of analytical methods. Rather, the intention is to identify well-established methods that are used as the standard methods of analysis. Many of the analytical methods used for environmental samples are the methods approved by federal agencies and organizations such as EPA and the National Institute for Occupational Safety and Health (NIOSH). Other methods presented in this chapter are those that are approved by groups such as the Association of Official Analytical Chemists (AOAC) and the American Public Health Association (APHA). Additionally, analytical methods are included that modify previously used methods to obtain lower detection limits and/or to improve accuracy and precision. [Pg.175]

The trade-offs between direct calibration and standard addition are treated in Ref 103. The same recovery as is found for the native analyte has to be obtained for the spiked analyte (see Section 3.2). The application of spiking to potentiometry is reviewed in Refs. 104 and 105. A worked example for the application of standard addition methodology to FIA/AAS is found in Ref 106. Reference 70 discusses the optimization of the standard addition method. [Pg.122]

The "method of standard additions" has been employed as a technique for standardization of atomic absorption analyses of metals In biological fluids (13,21) In this procedure, several concentrations of standard analyte are added to samples of the biological fluid to be analyzed The calibration curve which Is obtained after additions of the standard analyte to the biological fluid should parallel that obtained when aqueous standards are analyzed Extrapolation of the standard additions curve back to a negative Intercept on the abscissa furnishes an estimate of the concentration of the analyte In the original sample (21) This technique Is helpful In assessing the validity of methods of trace metal analysis (11,13,58) However, In the author s opinion, the "method of standard additions" Is neither practical nor reliable as a routine method for standardization... [Pg.255]

Currently, nutrient analytical methods development often utilizes the method of standard additions as an intrinsic aspect of the development process. Essentially, the analyte to be measured exists in the matrix to which an identical known pure standard is added. The spiked and non-spiked matrix is extracted and analysed for the nutrient of interest. By spiking at increasing levels the researcher can establish, to some degree of certainty, the recovery and linearity of the standard additions. One can also evaluate data to determine reproducibility, precision, and accuracy. Unfortunately, the method of standard additions does not allow the evaluation of the method at nutrient concentrations less than 100 % of the endogenous level. [Pg.288]


See other pages where Analytical methods standard addition is mentioned: [Pg.202]    [Pg.175]    [Pg.31]    [Pg.115]    [Pg.329]    [Pg.388]    [Pg.124]    [Pg.130]    [Pg.611]    [Pg.775]    [Pg.811]    [Pg.83]    [Pg.428]    [Pg.753]    [Pg.118]    [Pg.44]    [Pg.277]    [Pg.278]    [Pg.54]    [Pg.7]    [Pg.113]    [Pg.619]   
See also in sourсe #XX -- [ Pg.230 , Pg.233 ]




SEARCH



Additional methods

Additive method

Additivity methods

Analyte addition method

Analyte standard

Analytical method standardized

Analytical standards

Method standardization

Standard addition

Standard addition method

Standard method

Standardizing method

© 2024 chempedia.info