Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical Calibration graph

In summary, increasing Vs initially increases the sample volumetric fraction (X) and this increase becomes less pronounced for larger sample volumes or, in other words, the system tends to reach saturation, a situation where further increase in Vs no longer alters h in a significant manner. For very small injected volumes, there is an almost linear X vs Vs relationship, and exploitation of this aspect allows analytical calibration graphs to be obtained using only one standard solution [81]. [Pg.167]

Worked example 4. A confluence flow injection system with a very low confluent stream flow rate is designed for the spectrophotometric determination of phosphate in plant digests. The linearity of the analytical calibration graph is good and the recorded absorbance corresponding to a 100.0 mg L 1 P (as phosphate) standard solution is 0.21. Replacing the sample carrier stream by this standard solution (sample infinite volume) yields a steady state situation and the related absorbance is 0.68. The pump is then turned off and an asymptotic increase in absorbance towards 0.95 is observed determine the sensitivity improvement that in principle could be attained simply by increasing the sample volume and the mean sample residence time in the analytical path. [Pg.189]

Fig. 1. Analytical calibration graph for determination of hydroxylamine (conditions 1.20xl0 M hydrochloride acid, 5.00xl0 2M Na202 and 2.00 mL/min... Fig. 1. Analytical calibration graph for determination of hydroxylamine (conditions 1.20xl0 M hydrochloride acid, 5.00xl0 2M Na202 and 2.00 mL/min...
Matrix-isolation Shpol skii fluorimetry has been shown to eliminate quantitative artifacts from quenching and aggregation phenomena occurring in Shpol skii low-temperature solid matrices. The linear dynamic range for the fluorescence of PAHs in vapor-deposited alkane matrices smpasses that of solid-state Shpol skii fluorimetry. Thus, the analytical calibration graph for the fluorescence of benzo[ ]ant-hracene in -heptane is linear from 30 pg (the detection limit) to a maximmn amormt of over 35 pg. The... [Pg.1423]

Following this procedure urea can be determined with a linear calibration graph from 0.143 p.g-ml To 1.43 p.g-ml and a detection limit of 0.04 p.g-ml based on 3o criterion. Results show precision, as well as a satisfactory analytical recovery. The selectivity of the kinetic method itself is improved due to the great specificity that urease has for urea. There were no significant interferences in urea determination among the various substances tested. Method was applied for the determination of urea in semm. [Pg.371]

The following procedure has been recommended by the Analytical Methods Committee of the Society for Analytical Chemistry for the determination of small amounts of arsenic in organic matter.20 Organic matter is destroyed by wet oxidation, and the arsenic, after extraction with diethylammonium diethyldithiocarbamate in chloroform, is converted into the arsenomolybdate complex the latter is reduced by means of hydrazinium sulphate to a molybdenum blue complex and determined spectrophotometrically at 840 nm and referred to a calibration graph in the usual manner. [Pg.683]

In many cases when methods involve internal or external standards, the solutions used to construct the calibration graph are made up in pure solvents and the signal intensities obtained will not reflect any interaction of the analyte and internal standard with the matrix found in unknown samples or the effect that the matrix may have on the performance of the mass spectrometer. One way of overcoming this is to make up the calibration standards in solutions thought to reflect the matrix in which the samples are found. The major limitation of this is that the composition of the matrix may well vary widely and there can be no guarantee that the matrix effects found in the sample to be determined are identical to those in the calibration standards. [Pg.270]

Wienke D, Lucasius C, Ehrlich M, Kateman G (1993) Multicriteria target vector optimization of analytical procedures using a genetic algorithm. Part II. Polyoptimization of the photometric calibration graph of dry glucose sensors for quantitative clinical analysis. Anal Chim Acta 271 253... [Pg.148]

Gotti et al. [42] reported an analytical study of penicillamine in pharmaceuticals by capillary zone electrophoresis. Dispersions of the drug (0.4 mg/mL for the determination of (/q-penicillaminc in water containing 0.03% of the internal standard, S -met hy I - r-cystei ne, were injected at 5 kPa for 10 seconds into the capillary (48.5 cm x 50 pm i.d., 40 cm to detector). Electrophoresis was carried out at 15 °C and 30 kV, with a pH 2.5 buffer of 50 mM potassium phosphate and detection at 200 rnn. Calibration graphs were linear for 0.2-0.6 pg/mL (detection limit = 90 pM). For a more sensitive determination of penicillamine, or for the separation of its enantiomers, a derivative was prepared. Solutions (0.5 mL, final concentration 20 pg/mL) in 10 mM phosphate buffer (pH 8) were mixed with 1 mL of methanolic 0.015% 1,1 -[ethylidenebis-(sulfonyl)]bis-benzene and, after 2 min, with 0.5 mL of pH 2.5 phosphate buffer. An internal standard (0.03% tryptophan, 0.15 mL) was added and aliquots were injected. With the same pH 2.5 buffer and detection at 220 nm, calibration graphs were linear for 9.3-37.2 pg/mL, with a detection limit of 2.5 pM. For the determination of small amounts of (L)-penicillamine impurity, the final analyte concentration was 75 pg/mL, the pH 2.5 buffer contained 5 mM beta-cyclodextrin and 30 mM (+)-camphor-10-sulfonic acid, with a voltage of 20 kV, and detection at 220 nm. Calibration graphs were linear for 0.5-2% of the toxic (L)-enantiomer, with a detection limit of 0.3%. [Pg.141]

Garcia et al. [45] determined penicillamine in pharmaceutical preparations by FIA. Powdered tablets were dissolved in water, and the solution was filtered. Portions (70 pL) of the filtrate were injected into a carrier stream of water that merged with a stream of 1 mM PdCl2 in 1 M HC1 for determination of penicillamine. The mixture was passed though a reaction coil (180 cm long) and the absorbance was measured at 400 nm. Flow rates were 1.2 and 2.2 mL/min for the determination of penicillamine, the calibration graphs were linear for 0.01-0.7 mM, and the relative standard deviation (n = 10) for 0.17 mM analyte was 0.8%. The method was sufficiently selective, and there were no significant differences between the labeled contents and the obtained results. [Pg.142]

In practical situations the absorbance of a sample is determined by making two measurements, the first to determine 70 and the second to determine I. The determination of I0 is used to cancel a large number of experimental factors that could affect the result. When measuring I0 the sample container must closely match the unknown container in all ways except for the analyte content. The cuvettes should be a matched pair if a double beam instrument is used and the same cuvette can be used for both the blank and sample with a single beam instrument. The blank solution filling the cuvette should be identical to the solvent that the sample is dissolved in, except for the sample itself. If done correctly, the least-squares line for the calibration graph will come very close to the 0,0 point on the graph. [Pg.131]

Prepare a calibration graph in which the analyte of interest and the interferant are present at the same activity. The level of interference will then be constant and not affect the accuracy... [Pg.68]

The magnitude of a voltammetric current peak Ip is proportional to the concentration of analyte. In consequence, we can readily construct a calibration graph of Ip against [analyte] from known standards. We then perform our analyses by measuring Ip, and read off the concentration of analyte. This is generally quite a good method. [Pg.185]

The analytical performance of CC is demonstrated by extending the calibration graph of measured phenol with data found by CC. [Pg.108]

General Analytical Plan. A six step process is described to calculate the amount or concentration values of unknown samples using chromatographic response values and calibration graphs that were constructed by regression. The steps are ... [Pg.136]

Figure S.4 shows a calibration graph of arsenic concentrations obtained by using a Perkin Elmer 2100 atomic-absorption system bnked to a P.S. Analytical hydride/vapour generator (PSA 10.003). An electrically heated tube has been used in this work and the spectral source was an electrodeless discharge lamp. Alternatively, a flame-heated tube can be used. Figure S.4 shows a calibration graph of arsenic concentrations obtained by using a Perkin Elmer 2100 atomic-absorption system bnked to a P.S. Analytical hydride/vapour generator (PSA 10.003). An electrically heated tube has been used in this work and the spectral source was an electrodeless discharge lamp. Alternatively, a flame-heated tube can be used.
Silent-hours operation, which is commonly termed hands-off analysis, requires the automatic analysis to operate to a set protocol. For a fully automatic instrument to run in this manner, it will require a feedback system comparing the results with check cahbration standards. A calibration graph can be constructed from the analytical data, and the precision of this graph is easily evaluated. As the analyses proceed, the system can be monitored by reference to the check calibration standards. Should the performance remain within specification, the analyses can safely go on. The automatic instrument can then operate within the set protocols throughout the silent hours, taking full account of any variations in the instrument and its operating parameters. [Pg.254]

The cyanide sensor developed by the authors group is based on the formation of an addition product between cyanide ion and pyridoxal-5-phosphate, and its subsequent retention in the sensor (a fluorimetric flow-cell packed with QAE-Sephadex resin). The eluent is not injected, but merged with a stream of 0.05 M HCl after the reactor that is used both to acidify the complex and elute it after measurement. The calibration graph for the target analyte was linear from 50 ng/mL to 3.0 pg/mL, and the relative standard deviation and sample throughput were 1.4% (for 2 pg CN7mL) and... [Pg.217]

System (8) has been described for quantitation of corticosteroids as common adulterants in local drugs [156]. The sample is extracted from its matrix by methanol, and the resulting supernatant layer subjected to the HPLC analysis. The column used was an ODS-Zorbax column (25 cm x 4.6 mm), and the mobile phase 7 2 11 acetonitrile-methanol-aqueous 1% phosphoric acid. An eluent flow rate of 0.8 mL/min was used, and the analyte detection was performed using the UV absorbance at 240 nm. The calibration graph was found to be linear in the ranges of 1-15 pg/mL for betamethasone, 0.5-20 pg/mL for prednisolone, and 1-30 pg/mL for cortisone acetate. [Pg.227]


See other pages where Analytical Calibration graph is mentioned: [Pg.137]    [Pg.192]    [Pg.192]    [Pg.193]    [Pg.413]    [Pg.137]    [Pg.192]    [Pg.192]    [Pg.193]    [Pg.413]    [Pg.240]    [Pg.240]    [Pg.753]    [Pg.43]    [Pg.44]    [Pg.275]    [Pg.39]    [Pg.142]    [Pg.145]    [Pg.473]    [Pg.577]    [Pg.111]    [Pg.112]    [Pg.469]    [Pg.621]    [Pg.422]    [Pg.354]    [Pg.514]    [Pg.27]    [Pg.28]    [Pg.218]    [Pg.101]    [Pg.160]    [Pg.172]    [Pg.219]   
See also in sourсe #XX -- [ Pg.326 ]




SEARCH



Analytical Calibration

Calibration graph

© 2024 chempedia.info