Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analysis of the sample

Weigh accurately 10 mL of the seawater sample in an Erlenmeyer flask covered with a watch glass and quantitatively transfer into the titration vessel. Add 5 mL of buffer solution, 1 mL of EBT indicator and dilute with distilled water to 350 mL, When applying the conventional spectrophotometer method proceed as follows. Position the titration vessel in the cell compartment and cover it with the light-proof lid accommodating the stirring rod and the burette tip. Draw out the capillary end of the burette to a sufficient length so that the tip is always below the surface of the solution in the titration vessel. [Pg.235]

Stir vigorously and adjust the spectrophotometer slit so that the instrument reads zero absorbance at 640 nm. Add EDTA solution continuously until the first faint colour change occurs. As the endpoint is approached, add aliquots of 0.005 mL, wait for equilibrium, and record the absorbance. [Pg.235]

Plot the absorbance versus added mL of EDTA (see Fig. 11-1) and determine the volume of EDTA needed to reach the endpoint from the resulting titration curve. The intersection of the straight lines indicates the endpoint. [Pg.235]

In the case where an automatic titration system is applied (see Section 11.2.2.2), the volume of EDTA needed to reach the endpoint is reported in a print-out together with registration of the titration curve. It is important with this system to pre-select the increments of volume with which the titration should proceed (e.g., by 0.1 mL steps in this case). [Pg.236]


A standard sample was prepared containing 10.0 ppm of an analyte and 15.0 ppm of an internal standard. Analysis of the sample gave signals for the analyte and internal standard of 0.155 and 0.233 (arbitrary units), respectively. Sufficient internal standard was added to a sample to make it 15.0 ppm in the internal standard. Analysis of the sample yielded signals for the analyte and internal standard of 0.274 and 0.198, respectively. Report the concentration of analyte in the sample. [Pg.131]

After adding a 10.00-mL portion of the internal standard, the solution was filtered. Analysis of the sample gave a peak height ratio of 23.2 for ASA and 17.9 for CAE. [Pg.618]

A spike recovery for the analysis of chloride in well water was performed by adding 5.00 mb of a 25,000-ppm solution of Ck to a 500-mL volumetric flask and diluting to volume with the sample. Analysis of the sample and the spiked sample resulted in chloride concentrations of 183 ppm and 409 ppm, respectively. Determine the percent recovery of the spike. [Pg.711]

Analysis of the sample before and after the spike gave signals of0.456 for B and f.03 for sample Bsl- Considering these data, what is the most likely source of the systematic error ... [Pg.723]

On-Hne Procedures The growing trend toward automation in industry has resiilted in many studies of rapid procedures for generating size information so that feedback loops can be instituted as an integral part of a process. Many of these techniques are modifications of more traditional methods. The problems associated with on-line methods include allocation and preparation of a representative sample analysis of the sample evaluation of the results. The interface between the measuring apparatus and the process has the potential of high complexity, and consequently, high costs [Leschonsld, Paiticle Cha racterization, 1, 1 (July 1984)]. [Pg.1828]

To answer questions regarding dislocation multiplication in Mg-doped LiF single crystals, Vorthman and Duvall [19] describe soft-recovery experiments on <100)-oriented crystals shock loaded above the critical shear stress necessary for rapid precursor decay. Postshock analysis of the samples indicate that the dislocation density in recovered samples is not significantly greater than the preshock value. The predicted dislocation density (using precursor-decay analysis) is not observed. It is found, however, that the critical shear stress, above which the precursor amplitude decays rapidly, corresponds to the shear stress required to disturb grown-in dislocations which make up subgrain boundaries. [Pg.229]

Sampling for Gases and Vapors Many gases and vapors can be sampled by deviees whieh indicate the eoncentration of the substance during sampling or shortly thereafter, without the necessity for chemical analysis. These direct reading devices are convenient and useful when properly calibrated. Other substances cannot be sampled by this method, because no appropriate instrument is available, and indirect methods which require laboratory analysis of the sample must therefore be used. Such analyses are often delayed by days or weeks, depending upon laboratory schedules. [Pg.265]

Laboratory blanks Sample media that is not sampled on, but is analyzed by the laboratory to detect contamination or other problems associated with preparation and analysis of the samples. See also Field blanks. [Pg.1454]

Running a blank determination. This consists in carrying out a separate determination, the sample being omitted, under exactly the same experimental conditions as are employed in the actual analysis of the sample. The object is to find out the effect of the impurities introduced through the reagents and vessels, or to determine the excess of standard solution necessary to establish the end-point under the conditions met with in the titration of the unknown sample. A large blank correction is undesirable, because the exact value then becomes uncertain and the precision of the analysis is reduced. [Pg.131]

X-ray Photoelectron Spectroscopy analysis of the samples was performed with a Surface Science Instruments spectrometer (SSI 100) with a resolution (FWHM Au 4f7/2) of 1.0 eV. The X-ray beam was a monochromatised AlKa radiation (1486.6 eV). A charge neutraliser (flood gun) was adjusted at an energy of 6 eV. As the Cls spectra of these compounds were very complex, the binding energies were referenced to the binding energy of Ols, considered experimentally to be at 531.8 eV [8). [Pg.78]

The analysis of the sample coals used in the experiments are shown in Table 1. [Pg.308]

For example, for the iron oxide dust considered in the previous case study, Table 2 suggested Vfmm = 18 to 20 m s1 (i.e., assuming an average industrial dust ) On analysis of the sample, it was found dp50 80 pm, which appeared to support this classification. However, upon further examination of the actual distribution of size, a significant proportion of the material was found > 1000 pm (e.g., large flakes). A minimum conveying velocity of at least Vjmm 25 m s 1 was estimated for this dust. This explains why the iron oxide material built up and eventually blocked branch II-IV, which was sized/balanced mainly for air distribution purposes and produced transport velocities < Vfi r... [Pg.758]

Because the reproducibility of a-particle counting with the Pylon sources was quite good, this study was conducted as follows. Four identical samples, i.e., same material, were placed in the decay product standard sources for a period of 24 h at a time. The four samples were then removed and measured. Gross a-particle count and a-spectrometry analysis of the samples were done. [Pg.277]

The metal content analysis of the samples was effected by Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES Varian Liberty II Instrument) after microwaves assisted mineralisation in hydrofluoric/hydrochloric acid mixture. Ultraviolet and visible diffuse reflectance spectroscopy (UV-Vis DRS) was carried out in the 200-900 nm range with a Lambda 40 Perkin Elmer spectrophotometer with a BaS04 reflection sphere. HF was used as a reference. Data processing was carried out with Microcal Origin 7.1 software. [Pg.286]

In recent years, several groups have proposed the use of Laser Induced Breakdown Spectroscopy as a technique capable of giving information on the pigment compositions with minimal damage of the artwork. However, until the development of quantitative methods for accurate elemental analysis, the LIBS technique was hardly competitive with other methods for quantitative analysis of the samples. [Pg.515]

Principle Biosensors consist of paper matrixes and tissue enzyme preparations, often the pure enzyme AChE or AChE-containing cells. As seen in Fig.l, main scheme of the preparation technology and procedure includes (i) the preparation of same kinds of biotests-biosensors, which are paper matrixes impregnated with tissue preparation of AChE and covered by polymer film. (ii) biochemical reactions of the AChE activity with and without inhibitors tested and (iii) the photometric analysis of the samples for quantitative estimation of the biochemical reactions. [Pg.150]

The total variance in the final result (,v2ita ) is made up of two contributions. One is from variation in the composition of the laboratory samples due to the nature of the bulk material and the sampling procedures used ( ample). The other (Tanalysis) is from the analysis of the sample carried out in the laboratory ... [Pg.36]

Evolved gas analysis (EGA) is based on the study of gases or volatile breakdown products emitted by a sample on heating. The identity and properties of the volatile materials emitted serve as a basis for the analysis of the sample. One particular technique of EGA which has attracted sub-... [Pg.497]

The actual analysis of the samples started with a qualitative survey to identify the principal metals present. ICP-AES was selected as being most suitable for this because it provided a permanent record, a multi-element analysis, and good sensitivity. Solutions of the shellfish tissue were prepared for analysis by wet oxidation, and the sediments were digested with 40% v/v... [Pg.512]

Method development in SPE begins with an analysis of the sample matrix as well as the sample analytes. One of the most efficient mechanisms of SPE is to use mixed-mode ion exchange sorbents. If the analytes are bases, cation exchange is the most selective SPE mode. If the analytes are acids, anion exhange is the most efficient. However, it is sometimes easier to capture the matrix components and allow the analytes to pass through the sorbent unretained. [Pg.43]

Unfortunately, analysis of the sample headspace (the gas above the liquid in the drum) by GC does not show any ETO. The easy hypothesis has been disproven You need another hypothesis of what is happening—and time is of the essence ... [Pg.800]

Analysis of the samples by pyrolysis-GC/MS and solution 29Si NMR indicated that some randomization occurred due to cleavage of the original Si-Si bonds. The extent of randomization was significantly greater in the higher molecular weight fractions. [Pg.556]


See other pages where Analysis of the sample is mentioned: [Pg.616]    [Pg.91]    [Pg.106]    [Pg.2201]    [Pg.156]    [Pg.314]    [Pg.516]    [Pg.237]    [Pg.66]    [Pg.314]    [Pg.92]    [Pg.163]    [Pg.183]    [Pg.206]    [Pg.956]    [Pg.395]    [Pg.233]    [Pg.106]    [Pg.46]    [Pg.297]    [Pg.255]    [Pg.277]    [Pg.277]    [Pg.146]    [Pg.174]    [Pg.156]    [Pg.383]    [Pg.62]    [Pg.401]   


SEARCH



Analysis of Polymer Samples in which all Chains Possess the Same Backbone

Environmental Sampling and Analysis to Verify the Completeness of State Declarations

Experiment 31 The Analysis of Soil Samples for Iron Using Atomic Absorption

Mossbauer Analyses of the As-Mined Samples

Permethylated Sample, and Analysis of the Partially Methylated Sugars

Reference Materials Currently Available for the Analysis of Sediment and Particulate Samples

Sampling for chemical analysis of the

The Collection of Drainage Samples for Environmental Analyses from Active Stream Channels

The Sample

© 2024 chempedia.info