Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Accuracy evaluation development

Run Acceptance Criteria. Run acceptance criteria are used to accept or reject a run because of its performance. As a consequence, no dehned run acceptance criteria are applicable during method development. Prestudy validation runs are accepted based on the standard curve acceptance criteria. No run can be rejected because of poor performance of a sample during precision and accuracy evaluation all data from prestudy validation runs are reported unless there has been a clearly recognizable error during sample preparation or measurement. Despite the fact that the standard curve must satisfy the criteria described for standard curves above, in-study runs are accepted based primarily on the performance of the QC samples. As stated above in the paragraph on precision and accuracy, the 4-6-30 rule is recommended (i.e., at least four of six QC samples must be within 30% of their theoretical values and at least 50% of the values for each level must satisfy the 30% limit. [Pg.622]

The proposed system has three functions (1) accuracy evaluation, (2) accuracy data accumulation, and (3) search and reuse of accuracy data. The objective of the study shown in this section is to build a method for identifying knowledge, knowhow and techniques in the field based on the data managed by the developed system and evaluated by the three dimensional measured data in the ship construction process. The overview of the whole system is shown in Fig. 23.10. All types of data are stored in the database and, as well, the metadata is assigned to the data. Any data stored in the system can be reachable efficiently thanks to the metadata. [Pg.689]

MAPPscan is a manual ultrasonic testing system connected to an acoustic positioning system.. The system is developed as a consequence of the increased radiation doses in nuclear plants The system has the same flexibility as manual scanning with the same accuracy and the possibilties to collect, store and evaluate the UT data as with mechanised Ultrasonic systems The positioning system is based on spatial acoustic triangulation and have an accuracy of better than 1.0 mm within its recommended range. [Pg.862]

Before a procedure can provide useful analytical information, it is necessary to demonstrate that it is capable of providing acceptable results. Validation is an evaluation of whether the precision and accuracy obtained by following the procedure are appropriate for the problem. In addition, validation ensures that the written procedure has sufficient detail so that different analysts or laboratories following the same procedure obtain comparable results. Ideally, validation uses a standard sample whose composition closely matches the samples for which the procedure was developed. The comparison of replicate analyses can be used to evaluate the procedure s precision and accuracy. Intralaboratory and interlaboratory differences in the procedure also can be evaluated. In the absence of appropriate standards, accuracy can be evaluated by comparing results obtained with a new method to those obtained using a method of known accuracy. Chapter 14 provides a more detailed discussion of validation techniques. [Pg.47]

Moreover recovery studies on commercial spiked samples were developed in order to asses the no contribution of the matrix and evaluate the accuracy of the proposed procedure. [Pg.141]

Part of the planning should include the evaluation of test uncertainty. This evaluation can be limited to a common sense approach based on available instrumentation and the locations relative to the ideal. A more sophisticated study can be made in which instrumentation accuracy and the impact of any inaccuracy on the measured parameters is evaluated. This is a complex task with the need being based on the motivation for the test. If the test is being performed to settle a dispute, a formal understanding of the uncertainty should be developed. Methods for evaluation of test uncertainty are found in ANSI/ASME PTC 19.1 [11]. [Pg.431]

Develop final Perform structural design of analysis of component acceptable accuracy Determine structural response—stresses, support reactions, deflections, and stability—based on a structural analysis of acceptable accuracy. Determine acceptable accuracy based on economic value of component, consequences of failure, state-of-the-art capability in stress and stability analysis, margin of safety, knowledge about loads and materials properties, conservatism of loads, provisions for further evaluation by prototype testing... [Pg.8]

An alternative for evaluating accuracy is spiking known amounts of standards to a food, as reported in several papers,although percent recoveries of spikes do not truly address the influence of the food matrix complexity on the extraction efficiency. Data evaluation procedures were developed as a manual system to assess the quality of analytical data for carotenoids in foods. ... [Pg.449]

Certainly, this is an extremely small value and is the reason why the determination of the gravitational constant with very high accuracy is a rather complicated experiment. During the last two hundred years there were many measurements of this constant, but still only three digits after decimal point are reliable. One can say that due to Cavendish s measurements it became possible to develop the theory of gravity and evaluate mass of the earth. In fact, determination of this mass was the main goal of this experiment. [Pg.5]

A fully automated instrumental procedure has been developed for analyzing residual corrosion inhibitors in production waters in the field. The method uses ultraviolet (UV) and fluorescence spectrophotometric techniques to characterize different types of corrosion inhibitors. Laboratory evaluations showed that fluorescence is more suitable for field application because errors from high salinity, contamination, and matrix effect are minimized in fluorescence analysis. Comparison of the automated fluorescence technique with the classic extraction-dye transfer technique showed definite advantages of the former with respect to ease, speed, accuracy, and precision [1658],... [Pg.86]

Reference methods are generally arrived at by consensus and fairly extensive testing by a number of laboratories. For example, the flame atomic absorption method for Ca in serum developed under the leadership of the agency fondly remembered as NBS, now NIST (Cali et al. 1972), was established after several inter-laboratory comparison exercises. The results were evaluated after each exercise and the procedure was changed as necessary. After five exercises, it was felt that the state-of-the-art had been reached, with the reference method being capable of measuring Ca in serum with an accuracy of 2% of the true value determined by IDMS (note that attainment of high accuracy and precision is not only a matter of the method, but is a function of both the method and analyst expertise). [Pg.54]

Currently, nutrient analytical methods development often utilizes the method of standard additions as an intrinsic aspect of the development process. Essentially, the analyte to be measured exists in the matrix to which an identical known pure standard is added. The spiked and non-spiked matrix is extracted and analysed for the nutrient of interest. By spiking at increasing levels the researcher can establish, to some degree of certainty, the recovery and linearity of the standard additions. One can also evaluate data to determine reproducibility, precision, and accuracy. Unfortunately, the method of standard additions does not allow the evaluation of the method at nutrient concentrations less than 100 % of the endogenous level. [Pg.288]


See other pages where Accuracy evaluation development is mentioned: [Pg.166]    [Pg.58]    [Pg.510]    [Pg.115]    [Pg.879]    [Pg.158]    [Pg.80]    [Pg.167]    [Pg.537]    [Pg.199]    [Pg.167]    [Pg.723]    [Pg.162]    [Pg.317]    [Pg.536]    [Pg.2575]    [Pg.24]    [Pg.405]    [Pg.288]    [Pg.186]    [Pg.2]    [Pg.80]    [Pg.387]    [Pg.527]    [Pg.749]    [Pg.335]    [Pg.122]    [Pg.775]    [Pg.79]    [Pg.100]    [Pg.148]    [Pg.142]    [Pg.448]    [Pg.296]    [Pg.104]    [Pg.606]    [Pg.1057]   
See also in sourсe #XX -- [ Pg.154 ]

See also in sourсe #XX -- [ Pg.154 ]




SEARCH



Accuracy evaluation

© 2024 chempedia.info