Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Voltmeter, electronic

Indicator and reference electrodes potentiometer, pH meter or milli-voltmeter electronics and recorder for automated systems. [Pg.227]

Rapid-Scan Corrosion Behavior Diagram (CBD) Basically, all the same equipment used in the conductance of an ASTM G5 slow-scan polarization study is used for rapid-scan CBDs (that is, a standard test cell, potentiostat, voltmeters, log converters, X-Y recorders, and electronic potential scanning devices). The differences... [Pg.2431]

U2 = 0.1 V) and are suitable for current measurement. For smaller currents, sensitive instruments with 5 kQ per fiPc (Uj = 5 mV) are used. Small currents are usually measured by the voltage drop across a fixed resistance (calibrated shunt) using an electronic amplifier-voltmeter. This method has the advantage that the circuit does not have to be interrupted to measure the current. [Pg.85]

Fig. 1.20 Cell consisting of two reversible Ag /Ag electrodes (Ag in AgN03 solution). The rate and direction of charge transfer is indicated by the length and arrow-head as follows gain of electrons by Ag -he- Ag—> loss of electrons by Ag - Ag + e- —. (o) Both electrodes at equilibrium and (f>) electrodes polarised by an external source of e.m.f. the position of the electrodes in the vertical direction indicates the potential change. (K, high-impedance voltmeter A, ammeter R, variable resistance)... Fig. 1.20 Cell consisting of two reversible Ag /Ag electrodes (Ag in AgN03 solution). The rate and direction of charge transfer is indicated by the length and arrow-head as follows gain of electrons by Ag -he- Ag—> loss of electrons by Ag - Ag + e- —. (o) Both electrodes at equilibrium and (f>) electrodes polarised by an external source of e.m.f. the position of the electrodes in the vertical direction indicates the potential change. (K, high-impedance voltmeter A, ammeter R, variable resistance)...
Miniaturisation of electronic components has enabled the construction of a compact, portable, battery-operated recording voltmeter. The principal use of this instrument is to measure pipe/soil potential fluctuations over a period of time. The instrument can be modihed to measure current variations. [Pg.255]

Electrons generated at the anode move through the external circuit (right to left in Figure 18.2) to the copper cathode, connected through the black wire to the voltmeter. At the cathode, the electrons are consumed, reducing Cu2+ ions present in the solution around the electrode ... [Pg.482]

At all stages of the electrolysis the electronic voltmeter reads the potential difference between the cathode and the reference electrode, and the required limiting value of this potential difference is entered into the potentiostat. If the measured potential at any instant differs from the pre-set value, the potentiostat will adjust the current flowing to restore the required potential difference. [Pg.510]

If a controlled-potential determination is to be carried out, additional equipment will be required, namely an electronic voltmeter, a potentiostat and a reference electrode. The latter is most commonly a saturated calomel electrode, the construction of which is described in Chapter 14. [Pg.514]

To measure the e.m.f. the electrode system must be connected to a potentiometer or to an electronic voltmeter if the indicator electrode is a membrane electrode (e.g. a glass electrode), then a simple potentiometer is unsuitable and either a pH meter or a selective-ion meter must be employed the meter readings may give directly the varying pH (or pM) values as titration proceeds, or the meter may be used in the millivoltmeter mode, so that e.m.f. values are recorded. Used as a millivoltmeter, such meters can be used with almost any electrode assembly to record the results of many different types of potentiometric titrations, and in many cases the instruments have provision for connection to a recorder so that a continuous record of the titration results can be obtained, i.e. a titration curve is produced. [Pg.574]

This is called electrochemical shift and simply stems from the fact that the Fermi level of the reference electrode is not equal to that of the working electrode and thus to the Fermi level of the detector. Furthermore if one changes UWr via a potentiostat the core level electron binding energies of species associated with the reference electrode will shift according to Eq. (5.66), i.e. the XPS analyzer acts also as a (very expensive) voltmeter. [Pg.246]

FIGURE 12.5 The cell potential is measured with an electronic voltmeter, a device designed to draw negligible current so that the composition of the cell does not change during the measurement. The display shows a positive value when the + terminal of the meter is connected to the cathode of the galvanic cell. The salt bridge completes the electric circuit within the cell. [Pg.616]

The difference in electrical potential between two electrodes is the cell potential, designated E and measured in volts (V). The magnitude of E increases as the amount of charge imbalance between the two electrodes increases. For any galvanic cell, the value of E and the direction of electron flow can be determined experimentally by inserting a voltmeter in the external circuit. [Pg.1381]

The zinc-copper galvanic cell is under standard conditions when the concentration of each ion is 1.00 M, as shown in Figure 19-13. The cell potential under these conditions can be determined by connecting the electrodes to a voltmeter. The measured potential is 1.10 V, with the Zn electrode at the higher (more negative) potential, so Zn gives up electrons and E eii = 1.10 V ... [Pg.1382]

Adhering to the conditions given above in that at 298 K, the concentrations of Cu2+ and H+ are 1 M and the pressure of H2 is 1 atm, the voltmeter is found to read 0.34 V. The deflection direction is indicative of the fact that Cu has an inferior disposition to H2 to emit electrons. In other words, in this case, the half-reaction ... [Pg.637]

A basic electrochemical cell is depicted in Figure 9.3 and is made of a copper wire in one container with a solution of copper sulfate and a zinc rod in a different container with a zinc sulfate solution. There is a salt bridge containing a stationary saturated KC1 solution between the two containers. Electrons flow freely in the salt bridge in order to maintain electrical neutrality. A wire is connected to each rod and then to a measuring device such as a voltmeter to complete the cell. [Pg.194]

The difference is that the electrons are now flowing through a wire from the oxidation half-reaction to the reduction half-reaction. The flow of electrons through a wire is electricity. If we connect a voltmeter to the wire connecting the two electrodes, we would measure a current of 1.10 V. This galvanic cell is a Daniell cell. [Pg.269]

The battery module is a circuit board with the power source (two 9-V batteries) mounted on it and a receptacle for receiving the category 5 connector. It also has the output display terminals that the sensor requires for connection to a voltmeter to display the output voltage. The voltmeter or multimeter connects to these display terminals. The module has additional electronic components and connections with which we will not concern ourselves in this experiment. [Pg.170]

We have seen already that the root of the word potentio- means that we are looking at a potential, so the crucial piece of apparatus in a potentiometric experiment is a potentiometer (which, in practice, will almost certainly be marketed and labelled as a voltmeter ). In the previous chapter, we looked at the simplest forms of potentiometric experiment in which the two half cells were physically separated, one from the other, in order to prevent electron transfer (et) occurring. To recap, such a separation is a key requirement for ensuring that a frustrated equilibrium holds. [Pg.86]


See other pages where Voltmeter, electronic is mentioned: [Pg.93]    [Pg.942]    [Pg.467]    [Pg.414]    [Pg.84]    [Pg.96]    [Pg.257]    [Pg.258]    [Pg.565]    [Pg.72]    [Pg.615]    [Pg.615]    [Pg.615]    [Pg.35]    [Pg.195]    [Pg.406]    [Pg.634]    [Pg.637]    [Pg.202]    [Pg.188]    [Pg.1211]    [Pg.282]    [Pg.248]    [Pg.13]    [Pg.93]    [Pg.474]    [Pg.203]    [Pg.511]    [Pg.17]   
See also in sourсe #XX -- [ Pg.92 ]




SEARCH



© 2024 chempedia.info