Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Valves impulse

The gas quality feeding the dry faee seal should be elean and dry. Due to the possibility of eondensation of the proeess gas in the seal eavity, it was deeided to use a seal gas heater. The heater eontrol was set to provide warm gas at 15°C above the dew point to ensure no eondensate entered the seal eavity. Also, a dual filter in series with 5 and 2 p filtration elements was ehosen to provide an ideal sealing environment and maintain the optimum performanee of the seal. To reduee the risk of seal damage during reverse rotation of the turboexpander, programming logie was set to open the eompressor bypass valve whenever a shutdown impulse was initiated. [Pg.341]

Check isolating valves at process impulse line see that equalizing valve is shut. [Pg.328]

When equipment receives impulses at its own natural frequency of vibration, excessive vibration (resonance) occurs, and this can lead to rapid failure. A control valve was fitted with a new spindle with slightly different dimensions. This changed its natural frequency of vibration to that of the impulses of the liquid passing through it (the frequency of rotation of the pump times the number of passages in the impeller). The spindle failed after three months. Even a small change in the size of spindle is a modification [24]. [Pg.183]

Impulse steam traps rely on steam to maintain pressure on a piston and valve arrangement. When air and condensate enter the trap, the pressure is reduced above the piston and the valve opens. [Pg.92]

From the AV node, the electrical impulse spreads through the AV bundle or the bundle of His. This portion of the conduction system penetrates the fibrous tissue separating the atria from the ventricles and enters the interventricular septum where it divides into the left and right bundle branches. The bundle branches travel down the septum toward the apex of the heart and then reverse direction, traveling back toward the atria along the outer ventricle walls. This route of conduction of the impulse facilitates ejection of blood from the ventricles. If the impulse were to be conducted directly from the atria to the ventricles, the ventricular contraction would begin at the top of the chambers and proceed downward toward the apex. This would trap the blood at the bottom of the chambers. Instead, the wave of ventricular electrical stimulation and, therefore, contraction moves from the apex of the heart toward the top of the chambers where the semilunar valves are located and ejection takes place. [Pg.172]

Atmospheric effects of large-scale TNT expins have also been studied in depth both practically and theoretically. Factors considered include pressure and impulse effects, decay characteristics and travel and duration times, all as a function of distance, and for both free-field and reflection situations (Refs 3,9,15,16, 17,24,32, 33,34,35,36,44, 53,75,76,115 116). A distinction is made between the blast area dose to the source, comprising air and the products of expln, and that farther away involving air only (Ref 53). Double-burst conditions (fireball and shock wave interaction, and torus formation) have been studied (Ref 149), as have also the dynamics of dust formation and motion (Refs 25,26 117). Performance tests were run on a naval blast valve (Ref 92), and on aircraft wing panels (Ref 4)... [Pg.765]

Shock sensitive materials react exothermally when subject to a pressure impulse. The impulse may come from a hammer-like blow, such as used in the standard drop-weight test, or a compression, such as might be experienced in a deadheaded plant compressor or valve slamming shut. Normally, shock sensitivity increases with an increase in temperature. Materials that do not show an exotherm on the DSC are seldom shock sensitive. [Pg.233]

Gibson method. The mean velocity in a penstock leading from a reservoir can be determined by rapidly closing a valve at the lower end and recording a pressure-time diagram at a point just upstream from the valve. The principle that impulse equals change of momentum can be applied to the mass of liquid between the reservoir and the point of measurement. The impulse is given by the area under the recorded pressuretime curve, and from this the initial velocity can be determined [10]. [Pg.431]

For over 20 years the tank successfully operated with a nitrogen pad to reduce moisture intrusion and a simple hinged breather vent to provide overpressure and vacuum protection. The modification (installed and completed just three days prior to the failure) provided a vent compressor and a new state-of-the-art nearly zero leakage pressure/vacuum device and some unique piping features. See Figure 2—14, which shows impulse block valve No. 4 and the vent check valve No. 2. [Pg.32]

A small block valve is usually installed at a tank on any instrument impulse line just in case a leak develops on the small tubing. In this unit block valves are generally avoided on overpressure or vacuum relief devices. If a block valve that can defeat an overpressure device cannot be avoided, such a valve is chained and locked open or car-sealed opened and a protocol is developed to administratively control accidental closing of such a prime safety feature. [Pg.33]

The investigating team made several training and Management of Change recommendations and decided to eliminate the impulse valve (valve no. 4) connected to the... [Pg.33]

Q10 Backflow of blood into the atrium from the L ventricle through the defective mitral valve increases the volume and pressure of blood in the L atrium, leading to atrial hypertrophy. Since some of the ventricular output returns to the atrium and does not enter the aorta, the ventricle needs to pump an increased volume of blood at each beat. This increases the work of the L ventricle, causing ventricular hypertrophy. The cardiac impulse may be conducted abnormally through the hypertrophied ventricle, leading to the development of ventricular dysrhythmias and possibly to cardiac failure. [Pg.197]

VALVE 4 = IMPULSE LINE ISOLATION VALVE TO P/V AND N2 REGULATOR... [Pg.32]

With all due respect to the findings of the investigating team, the root cause appears to me to be a human factors situation. The new piping system looked simple but in fact this unique system was confusing. A check valve (valve no. 2) was installed on the vent header to prevent cross-contamination. A small block valve (valve no. 4) was installed on the vacuum relief impulse line to provide easy isolation. [5]... [Pg.32]

Tate valves, open Tlobe valves, open kngle valves, open ater meters, disk ater meters, piston Water meters, impulse wheel... [Pg.485]

A hot fan (Fi) in Figure 2 acts in a push-pull mode to suck gases from the reactor and pressurize them to move through the circuits. The volume of recirculated gas through the heat exchanger-reactor loop is set by valve V2. The off-gas valve vy operates automatically in response to a pressure impulse line in the top of the reactor above the wood feed. This line provides a pressure signal referenced to ambient barometric pressure for adjusting to accomplish the desired pressure differential. Typically this is set at zero or minus 3.0 Pa to prevent wood gas escape... [Pg.272]

The interface between the cylinder and the impulsion system consists of a connecting line, which should be as short as possible, furnished with a series of safety devices, including a bleed valve, which is specially useful to avoid the unwanted presence of gas in the pump when the connecting line is not short enough, a safety valve to avoid pressures above those the system and material used can reasonably withstand, a check valve to minimize contamination of the fluid stored in the cylinder and filters of small pore size to remove any contaminant particles from the fluid. [Pg.541]

Figure 10.8 Active control experiments to realize full tube fills in the Stanford PDE (a) indicates fuel filling duration, and (b) indicates cycle impulse for the cases of fixed valve opening duration and variable opening for active control 1 — fixed duration and 2 — active control. Figure 10.8 Active control experiments to realize full tube fills in the Stanford PDE (a) indicates fuel filling duration, and (b) indicates cycle impulse for the cases of fixed valve opening duration and variable opening for active control 1 — fixed duration and 2 — active control.

See other pages where Valves impulse is mentioned: [Pg.2495]    [Pg.126]    [Pg.195]    [Pg.462]    [Pg.462]    [Pg.69]    [Pg.124]    [Pg.600]    [Pg.184]    [Pg.33]    [Pg.34]    [Pg.34]    [Pg.237]    [Pg.117]    [Pg.485]    [Pg.2250]    [Pg.113]    [Pg.281]    [Pg.37]    [Pg.61]    [Pg.101]    [Pg.103]    [Pg.1388]    [Pg.975]    [Pg.339]   
See also in sourсe #XX -- [ Pg.33 ]




SEARCH



Impulse

Impulsive

Impulsiveness

© 2024 chempedia.info