Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Ultra violet UV detection

Without doubt UV detectors are the most popular detectors currently used in combination with HPLC and are favoured because of their versatility, reliability, sensitivity and relatively low cost. Detection depends on the presence of a suitable chromophore and is completely non-destructive. The relationship between sample concentration and absorbance can be defined by the Beer-Lambert equation  [Pg.31]

When monitoring at low wavelengths care should be taken in the selection of the solvents used in the mobile phase as several popular solvents have UV cut-offs at relatively high wavelengths. [Pg.33]

It should be emphasised that generally HPLC-grade solvents should be used, especially where detectors are being used at high sensitivity. Moreover, the use of purified solvents may help to extend the hfetime of analytical columns (Chapter 10). [Pg.33]

The sensitivity of UV detectors is generally in the low nanogram range although this can be increased by sample derivatisation. The sensitivity of detection may also be increased by alterations in the design of the detector flow cell and it is important to emphasise that the volume of the flow cell should be chosen to match the HPLC system thus a very low volume flow ceU (e.g. 1-2 /il) should be used with microbore, or very small particle columns where very high column efficiencies are anticipated, whereas larger flow cells (e.g. 5 jttl) should be used with more conventional systems. [Pg.33]


Whereas carbonyl chloride itself is very moisture sensitive and requires the corresponding precautions such as efficiently dried glassware and solvents, the Box derivative is very stable and can be analysed by high performance liquid chromatography (HPLC) with ultra violet (UV) detection at 270 nm. Quantification is achieved by the standard addition procedure spiking carbonyl chloride into the test polymer solution. However, since Box is a commercially available chemical, it is advisable to work also with Box standards, especially when the method is used for the first time and when problems are experienced in the HPLC determination or the derivation procedure. The standards of the carbonyl chloride derivative are particularly useful to establish the analytical system and to check linearity of detector response as well as for the recovery check. [Pg.326]

FSCE can also show impurities present in solid-phase synthesis products that are not clearly evident by conventional purification methods. RP-HPLC with ultra violet (UV) detection is traditionally used to analyze solid-phase synthesis products such as peptoids and peptides as well as to separate them preparatively away from any deletion products. Since each synthetic monomer addition cycle is less than 100% efficientj products that are one or two monomer units shorter than the desired product must be purified out to ensure monodispersity of the final polypeptide or polypeptoid. The separating power of FSCE was compared with analytical RP-HPLC traces after RP-HPLC purification (Figure 12.7). The one peak in the RP-HPLC trace actually separates into one large peak... [Pg.394]

The detection of material in a capillary column has proven to be interesting and dynamic over the years. Ultimately, the ability to detect a given analyte comes down to the amount of material in the cross-sectional area of the capillary at the detector location. For example in ultra violet (UV) detection, the absorbance of light. A, is related to three terms in the Beer-Lambert law, ... [Pg.413]

Solute property detectors measure a characteristic of the solute alone. These detectors are generally more sensitive yielding a detectable signal for nanogram quantities of solute. Representative detectors of this type include, for example, ultra-violet (UV), solute transport, fluorescence, and conductivity monitors. Other less frequently employed detectors of this nature are those based on radioactivity, polarography, and... [Pg.91]

Most optical detection methods for biosensors are based on ultra-violet (UV) absorption spectrometry, emission spectroscopic measurement of fluorescence and luminescence, and Raman spectroscopy. However, surface plasmon resonance (SPR) has quickly been widely adopted as a nonlabeling technique that provides attractive advantages. Fueled by numerous new nanomateiials, their unique, SPR-based or related detection techniques are increasingly being investigated [28-31]. [Pg.120]

Since its introduction at the end of the 1950s, gas chromatography (GC) has developed into a versatile tool in the analysis of natural products. Clear advantages of GC are the high resolution and high sensitivity of the most common detection method, the flame ionization detector (FID), and the fact that the detector response of similar compounds will be about the same, i.e., peak areas may be directly compared for quantification. This is in contrast to high performance liquid chromatography (HPLC), in which the detector response in the most common detection mode, ultra violet (UV)"absorption, may vary widely for different compounds since the molar extinction coefficients can be very different. [Pg.115]

The reactants are then mixed in various proportions, and the sedimentation ratio (scompiex/smucin)—the ratio of the sedimentation coefficient of the complex to that of the pure mucin itself—is used as the measure for mucoad-hesion. The ultra-violet absorption optics on the XL-A or XL-1 ultracentrifuge have been used as the main optical detection system. Although the polysaccharide is generally invisible in the near UV ( 280 nm), at the concentrations normally employed the mucin—in uncomplexed and complexed form—is detectable. [Pg.244]

In modern times, most analyses are performed on an analytical instrument for, e.g., gas chromatography (GC), high-performance liquid chromatography (HPLC), ultra-violet/visible (UV) or infrared (IR) spectrophotometry, atomic absorption spectrometry, inductively coupled plasma mass spectrometry (ICP-MS), mass spectrometry. Each of these instruments has a limitation on the amount of an analyte that they can detect. This limitation can be expressed as the IDL, which may be defined as the smallest amount of an analyte that can be reliably detected or differentiated from the background on an instrument. [Pg.63]

Detection is also frequently a key issue in polymer analysis, so much so that a section below is devoted to detectors. Only two detectors, the ultra-violet-visible spectrophotometer (UV-VIS) and the differential refractive index (DRI), are commonly in use as concentration-sensitive detectors in GPC. Many of the common polymer solvents absorb in the UV, so UV detection is the exception rather than the rule. Refractive index detectors have improved markedly in the last decade, but the limit of detection remains a common problem. Also, it is quite common that one component may have a positive RI response, while a second has a zero or negative response. This can be particularly problematic in co-polymer analysis. Although such problems can often be solved by changing or blending solvents, a third detector, the evaporative light-scattering detector, has found some favor. [Pg.333]

Because of the use of TATP by terrorists, a reliable and fast detection of this material is desirable. In addition to conventional analytical methods such as mass spectrometry and UV (ultra violet) spectroscopy specially trained explosive detec-... [Pg.21]

In the polluted atmosphere, the ultra-violet optical absorption technique (UV-photometer) does not always provides quality ozone data. A possible influence of various interferences on the measuring procedure can be responsible for this. The chemiluminescence method of O3 detection is considerably fi eer from interferences. ... [Pg.357]

Procyanidins can usually not be completely separated from each other, from phenolic acids and from accompanying flavonoids by sample clean up and/or chromatographic procedures. The differentiation between these individual compounds can therefore only by accomplished by the selectivity of a suitable detection method. UV-DAD (ultra violet diode array detection), electrochemical and mass spectrometry detection have been used for the identification and quantitative determination of procyanidins. [Pg.543]

LC-UV Liquid chromatography with ultra-violet detection... [Pg.1050]

Because of the use of TATP by terrorists, a rehable and fast detection of this material is desirable. In addition to conventional analytical methods such as mass spectrometry and UV (ultra violet) spectroscopy specially trained explosive detection dogs (FDD) play an important role in the detection of organic peroxides. Although the high vapor pressure helps the dogs to detect the material, it is also a disadvantage because of the limited time-span in which the dog is able to find it (traces may sublime and disappear forever). Matrices in which the compounds can... [Pg.16]


See other pages where Ultra violet UV detection is mentioned: [Pg.31]    [Pg.1786]    [Pg.296]    [Pg.1759]    [Pg.31]    [Pg.1786]    [Pg.296]    [Pg.1759]    [Pg.678]    [Pg.630]    [Pg.353]    [Pg.102]    [Pg.528]    [Pg.389]    [Pg.96]    [Pg.139]    [Pg.175]    [Pg.142]    [Pg.79]    [Pg.261]    [Pg.103]    [Pg.854]    [Pg.34]    [Pg.13]    [Pg.171]    [Pg.472]    [Pg.320]   


SEARCH



UV detection

Ultra-violet

Violets

© 2024 chempedia.info