Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standard of calibration

Deviations from approved standards of calibration on critical equipment/instruments should be investigated. This is to determine if these deviations affect the quality of the pharmaceutical or healthcare products manufactured using this equipment since the last successful calibration. [Pg.259]

Signals are measured using equipment or instruments that must be properly calibrated if Sjneas is to be free of determinate errors. Calibration is accomplished against a standard, adjusting S eas until it agrees with the standard s known signal. Several common examples of calibration are discussed here. [Pg.105]

Colorplate 2 shows an example of a set of standard additions and their corresponding standard additions calibration curve. [Pg.113]

Examples of calibration curves for the method of standard additions. In (a) the signal is plotted versus the volume of the added standard, and in (b) the signal is plotted versus the concentration of the added standard after dilution. [Pg.114]

Figure 5.7(b) shows the relevant relationships when Sspike is plotted versus the concentrations of the spiked standards after dilution. Standard addition calibration curves based on equation 5.8 are also possible. [Pg.115]

Since a standard additions calibration curve is constructed in the sample, it cannot be extended to the analysis of another sample. Each sample, therefore, requires its own standard additions calibration curve. This is a serious drawback to the routine application of the method of standard additions, particularly in laboratories that must handle many samples or that require a quick turnaround time. For example, suppose you need to analyze ten samples using a three-point calibration curve. For a normal calibration curve using external standards, only 13 solutions need to be analyzed (3 standards and 10 samples). Using the method of standard additions, however, requires the analysis of 30 solutions, since each of the 10 samples must be analyzed three times (once before spiking and two times after adding successive spikes). [Pg.115]

The method of standard additions can be used to check the validity of an external standardization when matrix matching is not feasible. To do this, a normal calibration curve of Sjtand versus Cs is constructed, and the value of k is determined from its slope. A standard additions calibration curve is then constructed using equation 5.6, plotting the data as shown in Figure 5.7(b). The slope of this standard additions calibration curve gives an independent determination of k. If the two values of k are identical, then any difference between the sample s matrix and that of the external standards can be ignored. When the values of k are different, a proportional determinate error is introduced if the normal calibration curve is used. [Pg.115]

An appropriate standard additions calibration curve based on equation 5.8 plots Sspi elVo + Vs) on they-axis and CsVs on the x-axis. Clearly explain why you cannot plot Sspike on the y-axis and Cs[ Vs/( Vo Vj)] on the x-axis. Derive equations for the slope and y-intercept, and explain how the amount of analyte in a sample can be determined from the calibration curve. [Pg.131]

Tor each of the pairs of calibration curves in figure 5.13 on page 132, select the calibration curve with the better set of standards. Briefly explain the reasons for your selections. The scales for the x-axes andy-axes are the same for each pair. [Pg.131]

Construct an appropriate standard additions calibration curve, and use a linear regression analysis to determine the concentration of analyte in the original sample and its 95% confidence interval. [Pg.133]

A standard addition calibration curve of emission versus the concentration of added sodium gives, by linear regression, an equation of... [Pg.440]

Samples of car exhaust are collected using a 4-L glass bottle evacuated to a level of less than 2 torr. A normal calibration curve using external standards of known Pco is used to determine the Pco in the exhaust samples. [Pg.448]

In the DPD colorimetric method for the free chlorine residual, which is reported as parts per million of CI2, the oxidizing power of free chlorine converts the colorless amine N,N-diethyl-p-phenylenediamine to a colored dye that absorbs strongly over the wavelength range of 440-580 nm. Analysis of a set of calibration standards gave the following results... [Pg.452]

A pH electrode is normally standardized using two buffers one near a pH of 7 and one that is more acidic or basic depending on the sample s expected pH. The pH electrode is immersed in the first buffer, and the standardize or calibrate control is adjusted until the meter reads the correct pH. The electrode is placed in the second buffer, and the slope or temperature control is adjusted to the-buffer s pH. Some pH meters are equipped with a temperature compensation feature, allowing the pH meter to correct the measured pH for any change in temperature. In this case a thermistor is placed in the sample and connected to the pH meter. The temperature control is set to the solution s temperature, and the pH meter is calibrated using the calibrate and slope controls. If a change in the sample s temperature is indicated by the thermistor, the pH meter adjusts the slope of the calibration based on an assumed Nerstian response of 2.303RT/F. [Pg.492]

The amount of sulfur in aromatic monomers can be determined by differential pulse polarography. Standard solutions are prepared for analysis by dissolving 1.000 mb of the purified monomer in 25.00 mb of an electrolytic solvent, adding a known amount of S, deaerating, and measuring the peak current. The following results were obtained for a set of calibration standards... [Pg.538]

The purity of a sample of K3Fe(CN)6 was determined using linear-potential scan hydrodynamic voltammetry at a glassy carbon electrode using the method of external standards. The following data were obtained for a set of calibration standards. [Pg.538]

Sittampalam and Wilson described the preparation and use of an amperometric sensor for glucose. " The sensor is calibrated by measuring the steady-state current when it is immersed in standard solutions of glucose. A typical set of calibration data is shown in the following table. [Pg.538]

Zhou and colleagues determined the %w/w H2O in methanol by GG, using a capillary column coated with a nonpolar stationary phase and a thermal conductivity detector. A series of calibration standards gave the following results. [Pg.616]

The concentration of chloride in sea water can be determined by a flow injection analysis. The analysis of a set of calibration standards gives the results in the following table. [Pg.663]

Analysis of Standards The analysis of a standard containing a known concentration of analyte also can be used to monitor a system s state of statistical control. Ideally, a standard reference material (SRM) should be used, provided that the matrix of the SRM is similar to that of the samples being analyzed. A variety of appropriate SRMs are available from the National Institute of Standards and Technology (NIST). If a suitable SRM is not available, then an independently prepared synthetic sample can be used if it is prepared from reagents of known purity. At a minimum, a standardization of the method is verified by periodically analyzing one of the calibration standards. In all cases, the analyte s experimentally determined concentration in the standard must fall within predetermined limits if the system is to be considered under statistical control. [Pg.710]

Other instrumental advantages include its high sensitivity and a linear mass scale to m/z 10,000 at full sensitivity. The linearity of the mass scale means that it is necessary to calibrate the spectrometer using a single or sometimes two known mass standards. Some calibration is necessary because the start of the mass scale is subject to some instrumental zero offset. The digitized accumulation of spectra provides a better signal-to-noise ratio than can be obtained from one spectrum alone. [Pg.167]

It is important to note that simulated distillation does not always separate hydrocarbons in the order of their boiling point. For example, high-boihng multiple-ring-type compounds may be eluted earher than normal paraffins (used as the calibration standard) of the same boiling point. Gas chromatography is also used in the ASTM D 2427 test method to determine quantitatively ethane through pentane hydrocarbons. [Pg.1326]

Robustness of methods based on broad-on-nar row approach (one of narrow standar ds is used as broad standard for correction of calibration function) is higher but some disadvantages also are observed. [Pg.345]

Internal standard of Rh with 2 ppb concentration was added to all the solutions. Three certified Multi-element solutions (CLMS-1,-2,-4, SPEX, USA) were employed for constmcting calibration graphs. [Pg.454]

U.S. Environmental Protection Agency, "Transfer Standards for Calibration of Air Monitoring Analyzers for Ozone," EPA-600/4-79-056. Office of Air Quality Planning and Standards, Research Triangle Park, NC, 1979. [Pg.213]


See other pages where Standard of calibration is mentioned: [Pg.22]    [Pg.626]    [Pg.104]    [Pg.62]    [Pg.269]    [Pg.22]    [Pg.626]    [Pg.104]    [Pg.62]    [Pg.269]    [Pg.1248]    [Pg.47]    [Pg.104]    [Pg.109]    [Pg.113]    [Pg.123]    [Pg.124]    [Pg.125]    [Pg.420]    [Pg.443]    [Pg.443]    [Pg.451]    [Pg.451]    [Pg.457]    [Pg.616]    [Pg.661]    [Pg.813]    [Pg.173]    [Pg.644]    [Pg.220]    [Pg.729]   
See also in sourсe #XX -- [ Pg.247 , Pg.253 ]




SEARCH



Calibration of Internal Standard

Calibration of Internal Standard to Acyl-ACP

External standard mode of instrument calibration

Quantification of Analytical Data via Calibration Curves in Mass Spectrometry Using Certified Reference Materials or Defined Standard Solutions

Standard addition mode of instrument calibration

Standard error of calibration

Standardization calibration standards

Standards for Calibration of Explosive Detectors

State-of-the-Art Commercial Instruments, Standards, and Calibration

© 2024 chempedia.info