Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standard deviation of blank

Lowest amount of the analyte in the sample which can be detected but not necessarily quantified as an exact value. LOD is expressed as a concentration or a quantity derived from the smallest signal xLi which can be detected with a reasonable certainty for an analytical procedure. The LOD can be determined using the equation xL = xbl +ksbl with xbl and sbl as the mean value and the standard deviation of blank, respectively. LOD of an element or element species is often used as 3sbl (3cr) of blank.3... [Pg.474]

LLOQ based on standard deviation of the response from blank samples [14], Another definition of LLOQ is the concentration that corresponds to a detector response that is k times greater than the estimated standard deviation of blank samples (SDbl). From the detector signal, the LLOQ can be calculated using the slope of the calibration curve (S) with following formula LLOQ= k-SDbl/S (for blank corrected signals). [Pg.5]

Table 7.3 Results of mean and standard deviation of blank measurements using ICP-OES-FIA... Table 7.3 Results of mean and standard deviation of blank measurements using ICP-OES-FIA...
A widely used detection limit technique is the 3ct approach that is mandated for EPA testing. The standard deviation in concentration units is calculated by computing the standard deviation of blank replicates (s7) and dividing by the slope of the calibration curve. The number is multiplied by the appropriate value of the Student s t with n — 1 degrees of freedom for the chosen a. [Pg.179]

ICP-SFMS (Thermo Finnigan, Flement) with cold vapour generation was developed with a guard electrode and a gold amalgamation device using an Au-sorbent for sample pre-concentration to improve the sensitivity. Instrumental parameters of ICP-SFMS such as take-up time, heating temperature of Au-sorbent, additional gas flow, and sample gas flow were optimized. Detection limit calculated as 3 times the standard deviation of 10 blanks was 0,05 ng/1, RSD = 7-9 %. [Pg.171]

The detection limit is another value which is often quoted, and this may be defined in a variety of ways. The most widely accepted definition is that the detection limit is the smallest concentration of a solution of an element that can be detected with 95 per cent certainty. This is the quantity of the element that gives a reading equal to twice the standard deviation of a series of at least ten determinations taken with solutions of concentrations which are close to the level of the blank. [Pg.804]

As discussed earlier, a concentration that would produce a signal of approximately 10-12 times the standard deviation of the blank (or in this case xeloq) is considered to be the limit of quantification. Therefore, if the LOQ was set at 10 times 5eloq, for 7 replicates (6 degrees of freedom) of the fortified control samples... [Pg.71]

Net recoveries of cyfluthrin from matrices fortified at 0.01-5.05 mg kg ranged from 77 to 119%. The limit of detection (LOD) is defined as the lowest concentration that can be determined to be statistically different from a blank or control. Calculate the value by taking the standard deviation of the residue values from the analysis of the recovery samples at the limit of quantification (LOQ) and using the equation... [Pg.1286]

Assuming that the standard deviation of the critical value is approximately equal to the standard deviation of the blank, scv Srl> it results for a = fi in... [Pg.230]

The DL and QL for chromatographic analytical methods can be defined in terms of the signal-to-noise ratio, with values of 2 1-3 1 defining the DL and a value of 10 1 defining the QL. Alternatively, in terms of the ratio of the standard deviation of the blank response, the residual standard deviation of the calibration line, or the standard deviation of intercept (s) and slope (5) can be used [40, 42], where ... [Pg.255]

The limit of detection (LoD) has already been mentioned in Section 4.3.1. This is the minimum concentration of analyte that can be detected with statistical confidence, based on the concept of an adequately low risk of failure to detect a determinand. Only one value is indicated in Figure 4.9 but there are many ways of estimating the value of the LoD and the choice depends on how well the level needs to be defined. It is determined by repeat analysis of a blank test portion or a test portion containing a very small amount of analyte. A measured signal of three times the standard deviation of the blank signal (3sbi) is unlikely to happen by chance and is commonly taken as an approximate estimation of the LoD. This approach is usually adequate if all of the analytical results are well above this value. The value of Sbi used should be the standard deviation of the results obtained from a large number of batches of blank or low-level spike solutions. In addition, the approximation only applies to results that are normally distributed and are quoted with a level of confidence of 95%. [Pg.87]

The limit of quantitation (LoQ) is the lowest concentration of analyte that can be determined with an acceptable level of uncertainty. This should be established by using an appropriate reference material or sample. It should not be determined by extrapolation. Various conventions take the approximate limit to be 5, 6 or 10 times the standard deviation of a number of measurements made on a blank or a low-level spiked solution. [Pg.88]

The determination of iron at trace levels can be accomplished by the spectrophotometric measurement of its complex with the reagent o-phenanthroline. The sensitivity of a particular method is 53 ppm per unit absorbance and the standard deviation of the blank estimated from 25 measurements is 0.002 of an absorbance unit. The practical detection limit is therefore 1.64aB or 0.0033 of an absorbance unit which corresponds to 0.17 ppm, and the theoretical detection limit is 3.28aB or 0.0066 of an absorbance unit which corresponds to 0.35 ppm. (The value of pB in this case is assumed to be zero.) Hence if a result is less than 0.17 ppm, the conclusion is that iron is not detected. If the value lies between 0.17 ppm and 0.35 ppm, the iron content should be reported as less than 0.35 ppm. ... [Pg.14]

Residue was incorporated to a volumetric flask. Determinations of REE were carried out with an ICP mass spectrometer VG Elemental model PQ3. Detection limits were calculated as the concentration equivalent to three times the standard deviation of five replicates of the blank... [Pg.462]

Quantification of the limits of detection (LOD), or minimum detectable levels (MDL statistically defined in Section 13.4), is an important part of any analysis. They are used to describe the smallest concentration of each element which can be determined, and will vary from element to element, from matrix to matrix, and from day to day. Any element in a sample which has a value below, or similar to, the limits of detection should be excluded from subsequent interpretation. A generally accepted definition of detection limit is the concentration equal to a signal of twice (95% confidence level) or three times (99% confidence) the standard deviation of the signal produced by the background noise at the position of the peak. In practice, detection limits in ICP-MS are usually based on ten runs of a matrix matched blank and a standard. In this case ... [Pg.204]

Depending on method type or intent, the LOD or LOQ may need to be determined. ICH guidelines describe several approaches and allow alternative approaches, if scientifically justifiable. Suggested approaches include calculation based on signal-to-noise ratio (typically set at 3 1 for LOD and 10 1 for LOQ) standard deviation of the response and slope standard deviation of the response of a blank, or a calibration curve. [Pg.389]

The procedure for measuring the solvent removal efficiencies has been described elsewhere [4]. All of the detection hmits in the work described here were estimated as the concentration giving a signal equal to three times the standard deviation of the signal at the analytical wavelength when aspirating a blank. The lowest value reported for the results of the determinations is take to equal twice the detection Hmit. [Pg.141]

The simple approach is just using the mean value of several determinations of blank samples plus a multiple (factor k) of the standard deviation of these measnrements. With the choice of k we define the level of confidence. [Pg.194]

The Limit of Detection is expressed as the concentration or the qnantity derived from the smallest signal that can be detected with reasonable certainty for a given analytical procedure. One possibility to express the LoD is as the mean blank valne pins three times the standard deviation of 10 measurements of the blank sample (see also chapter 9). [Pg.227]

It is an indicative valne and should not normally be nsed for decision-making pnrposes. It shonld be established nsing an appropriate measurement standard or sample and should not be determined by extrapolation. The LoQ is calculated as the analyte concentration corresponding to the sample blank value plus 10 standard deviations of the blank measurement. If measurements are made under repeatability conditions, a measure of the repeatability precision at this concentration is also obtained. [Pg.228]


See other pages where Standard deviation of blank is mentioned: [Pg.194]    [Pg.194]    [Pg.113]    [Pg.267]    [Pg.2338]    [Pg.284]    [Pg.207]    [Pg.194]    [Pg.194]    [Pg.113]    [Pg.267]    [Pg.2338]    [Pg.284]    [Pg.207]    [Pg.99]    [Pg.48]    [Pg.67]    [Pg.320]    [Pg.834]    [Pg.51]    [Pg.571]    [Pg.655]    [Pg.149]    [Pg.233]    [Pg.56]    [Pg.87]    [Pg.205]    [Pg.409]    [Pg.204]    [Pg.209]    [Pg.302]    [Pg.44]    [Pg.130]    [Pg.138]    [Pg.8]    [Pg.9]   
See also in sourсe #XX -- [ Pg.239 , Pg.240 , Pg.241 ]




SEARCH



Blank

Blank, blanking

Blanking

Standard deviation

Standard deviation blank

Standard deviation of the blank

Standard deviation standardization

© 2024 chempedia.info