Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analysis repeatability

Analysis repeated with the assumption that the hub intermediate H (see Figure 4.36) is an available starting material. [Pg.142]

They were also typical when the regression model chosen was first order. Mean-level bandwidths greater than 20-30% are probably indicative that errors have been made in the analysis process that should not be tolerated. In this case techniques would be carefully scrutinized to find errors, outliers, or changing chromatographic conditions. These should be remedied and the analysis repeated whenever possible. Certain manipulation can be done to reduce the bandwidth values. For example, they would be... [Pg.158]

For early phase methods, the precision tests only include injection repeatability (also referred to as system repeatability) and method repeatability (also referred to as analysis repeatability). The former is demonstrated by repeating injections of a standard solution and the latter by preparing multiple samples over multiple concentration levels (usually at 80%, 100%, and 120% of the nominal concentration) from the same lot of a composite sample of the dosage form. [Pg.163]

Repeatability can be divided into two areas injection repeatability and analysis repeatability. Injection repeatability is measured by analyzing multiple injections of the same solution preparation. It is an indicator of the performance of the HPLC system under the specified conditions and at the time of the analyses. This information is included as part of the validation package and is also used during routine analysis in the form of a system suitability. During the validation the specification for % RSD will be set, which will determine the variation limit for the analysis. If the value is low... [Pg.277]

The area of the internal standard peaks must remain constant throughout a series. If not, the cause must be determined and corrected and, if necessary, the analysis repeated. One possible cause is quenching due to incorrect injection. If the signal of the internal standard is more than 50% reduced, the sample can be re-injected as soon as possible after dilution. Note that the AdoMet-D3 signal in samples is usually 1.5 times higher than that in standard runs. [Pg.110]

B. Fast (about 10 minutes per analysis). Repeat analysis on weighed sample requires only 1 minute). [Pg.80]

The standard and docked (n = 202) aligned databases were analyzed using CoMFA/pls. The results indicated three compounds having especially high residual values (differences between predicted and actual activities). These outliers were removed, and the pis analysis repeated (n = 199). [Pg.202]

Soil, sediment (2,3,7,8-TCDD) Soxhlet extraction of sample volume reduction clean-up on basic silica/acidic, silica/alumina, elution with CH2CI2 in hexane analysis clean-up on silver nitrate silica or 2,3,7,8-TCDD-specific alumina, elution with CH2CI2 in hexane analysis repeating of clean-up or extraction if needed HRGC/LRMS (SIM) HRGC/MS/MS (SIM) 1 ng/g <1 ng/g 40-90 57-102 Simon et al. 1989... [Pg.553]

In the context of such interim analysis—repeated analysis to see if the drug response, on average, differs from the placebo response—the same hypothesis is tested five times if the trial is not terminated early and proceeds to its final stage. While the chance of erroneously obtaining a significant result at the 5% level of significance is 1 in 20 if only one analysis is conducted, this chance increases every time more than one test is performed. [Pg.183]

Repeatability represents the simplest situation and involves analysis of replicates by the same analyst, generally one injection after the other. Repeat-abihty tests are mandatory for all tests dehvering numerical data. Repeatability is divided into two parts injection repeatability and analysis repeatability (multiple preparations) [24]. [Pg.471]

Too high concentrations (activities) are indicated by an error report or by a maximal value. The samples must then be diluted further and the analysis repeated. Pressing the DIL button causes automatic multiplication in the instrument, but the prescribed dilution must always be employed by the user. Samples of very low concentration (activity) will also cause an error report. These samples must then be measured by a different method. [Pg.444]

Finally, having performed a cluster analysis, statistical tests can be employed to assess the contribution of each variable to the clustering process. Variables found to contribute little may be omitted and the cluster analysis repeated. [Pg.95]

Following the establishment of specificity, the method(s) should be validated to allow for use in release and stability testing. Such validation is typically less stringent than for final methods (sec Chapter 12), but should demonstrate specificity, linearity, range, accuracy, and analysis repeatability for the API. For related substances, specificity should be demonstrated and the limit of detection (LOD) and limit of quantitation (LOQ) should be established for the API to serve as surrogate values for the LOD and LOQ of impurities for which authentic substances are not available. To achieve a sufficient LOD and simultaneously keep the API in the linear dynamic range of the detector, it may be necessary to use different sample concentrations for the analyses of the API and related substances. It is additionally beneficial to repeat the separation on new columns from different batches to ascertain that the separation obtained can be maintained column to column. [Pg.357]

Analysis Repeatability. Analysis repeatability expresses the precision under the same operating conditions over a short interval of time. It typically consists of multiple preparations and measurements of a homogeneous sample by the same analyst on the same day. The analysis repeatability can be determined by assaying 15 individual sample preparations covering the specified range for the procedure (5 concentration levels/3 replicates each). [Pg.434]

New pyrolysis vessels require conditioning before analysis. Standards should be loaded, pyrolysed and the resulting gas released for at least two runs before CO2 is collected for analysis. Repeated runs with water and cellulose standards should then be undertaken until consistent reproducibility is obtained and until accurate calibration of results can be developed. After extended periods of non-use, standards should be run in each vessel to re-establish the catalytic activity of the nickel. A water or cellulose standard should be routinely run with each sample batch, rotating the vessel in which the standard is loaded to monitor vessel performance. Analytical uncertainties of water, lake sediment and terrestrial cellulose are generally 0.5 to 1.0%o. [Pg.384]

This equation is used alternately with the flow equation, to update the melt temperature distribution. If the injection pressure is below the limit of the machine, the new position of the melt front is computed. The programmes output the melt front shape and melt pressure isobars at various times, hence predict whether a mould can be filled satisfactorily. If not, modifications can be made to the CAD file for the mould geometry, and the analysis repeated, before the mould cavity is machined. Figure 5.27 shows the predicted flow fronts for an instrument panel moulding. [Pg.168]

Step 4 - Repeat the Analysis Repeat steps 1 to 3 using the revised PFDa, g value of the proposed instrumented modification, to recalculate the overall incident frequency meeting the risk target... [Pg.86]

MDP, as all the N-acetyl-muramyl-peptides, can be precipitated by addition of ether from its ethanol or methanol solution, but it occludes a considerable amount of solvent judging from elemental analysis. Repeated lyophilization allows the elimination of all traces of solvent 33). [Pg.35]

The proof proceeds in principle as a natural generalization of that established for the finite-range case. It presents an ingenious combination of soft analysis (repeated use of Schauder-Tychonov theorem) and hard analysis (tricky majorizations, etc.), which we do not want to reproduce here. It should, however, be noticed that we can show explicitly for these (one-dimensional) models that the effect of the boundary conditions adopted for the computation of the Gibbs state does disappear in the thermodjmamic limit. This is a reflection of the fact that these models do not exhibit phase transitions. [Pg.349]

Laboratory and/or field data were analyzed using SAS systems (SAS Institute Inc. 1999) utilizing analysis of variance, regression analysis, response surface analysis, univariate analysis, repeated measures analysis (multivariate profile analysis), covariance analysis and/or principle components analysis. Good statistical practices were used to verify that the data satisfied the assumptions underlying the various analyses. Significant differences between means were determined by Tukey s Studentized Range Test, the Tukey-Kramer HSD test, or the Bonferroni t test. Alpha was set at 0.05. [Pg.97]

Rodriguez-Otero et al. (46) tried to measure cheese composition by NIR in cheese. They tried to analyze cheese without any prior sample manipulation (not even grating) on the basis of increased knowledge in calibration techniques based on multivariate analysis. Repeatability of NIR moisture determination was approximately double in comparison with the reference method. Repeatability of determination of protein by reference (Kjeldahl) and NIR methods was higher for the NIR spectroscopy, probably because of the large sample size rather than to a lack of method precision. The repeatability of fat determination by reference (gravimetric extraction) and NIR methods was 0.31% for reference and 0.40% for NIRS. [Pg.329]


See other pages where Analysis repeatability is mentioned: [Pg.125]    [Pg.153]    [Pg.105]    [Pg.54]    [Pg.184]    [Pg.278]    [Pg.717]    [Pg.178]    [Pg.192]    [Pg.164]    [Pg.169]    [Pg.2850]    [Pg.142]    [Pg.444]    [Pg.613]    [Pg.147]    [Pg.435]    [Pg.154]    [Pg.181]    [Pg.418]    [Pg.92]    [Pg.33]    [Pg.56]    [Pg.128]    [Pg.2868]    [Pg.1043]   
See also in sourсe #XX -- [ Pg.163 , Pg.278 ]




SEARCH



© 2024 chempedia.info