Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sensitization level

Sensitivity levels more typical of kinetic studies are of the order of lO molecules cm . A schematic diagram of an apparatus for kinetic LIF measurements is shown in figure C3.I.8. A limitation of this approach is that only relative concentrations are easily measured, in contrast to absorjDtion measurements, which yield absolute concentrations. Another important limitation is that not all molecules have measurable fluorescence, as radiationless transitions can be the dominant decay route for electronic excitation in polyatomic molecules. However, the latter situation can also be an advantage in complex molecules, such as proteins, where a lack of background fluorescence allow s the selective introduction of fluorescent chromophores as probes for kinetic studies. (Tryptophan is the only strongly fluorescent amino acid naturally present in proteins, for instance.)... [Pg.2958]

Since it determines the sensitivity level of the protection scheme, it must be kept as low as possible to detect even a small fault. To achieve a high degree of sensitivity it is therefore essential... [Pg.484]

When the scheme is required to detect the ground fault as well as the phase faults, a triple-pole relay is used, eaeh pole of which is connected between the shorted terminals of the two same phase CTs and the neutral formed by shorting the other terminals of all the CTs, as shown in Figure 15.22. The setting of all the poles is kept the same. In other words, the sensitivity level remains the same for all types of faults. [Pg.484]

The important question, then, is not whether a substance is pure but whether a given sample is sufficiently pure for some intended purpose. That is, are the contaminants likely to interfere in the process or measurement that is to be studied. By suitable manipulation it is often possible to reduce levels of impurities to acceptable limits, but absolute purity is an ideal which, no matter how closely approached, can never be attained. A negative physical or chemical test indicates only that the amount of an impurity in a substance lies below a certain sensitivity level no test can demonstrate that a specified impurity is entirely al ent. [Pg.1]

Balance machines are limited to the 25 to 50 pin sensitivity level, indicating that Equation 9.3 for the example will take the balance close to the limit. [Pg.373]

Group Al, 80% of the cardiac sensitization level for R-11, R-12, R-22, R-113, R-134a, R-500, and R-502 others are limited by levels in which oxygen... [Pg.317]

Table 8.80 shows the present status of speciation methodology. For trace-metal speciation, atomic absorption detectors feature a relatively high absolute detection limit (10 pg level), as compared to the 0.1 to 1 pg sensitivity level for molecular ion MS techniques as well as for MIP-AES. The detection limit of LEI-ToFMS is in the attogram range. Speciation has been reviewed [550]. Various monographs deal with speciation analysis [542,551,552]. [Pg.676]

The on-line coupling of HPLC and NMR can either be performed in the stopped-flow or in the continuous-flow mode (Krucker et al. 2004, Grynbaum et al. 2005, Putzbach et al. 2005, Albert et al. 2006, Hentschel et al. 2006, Rehbein et al. 2007). Current sensitivity levels are in the lower nanogram range for ID II NMR spectra and in the microgram range for 2D spectra. [Pg.64]

The main drawback of the galvanometer-spot follower system is that the sensitivity level of the line is defined ab initio and cannot be modified except by increasing or decreasing the number of thermoelements in the circuit. A more versatile, but more expensive, system is provided by the addition of a dc amplifier to a recording voltmeter. [Pg.216]

In the various sections of this article, it has been attempted to show that heat-flow calorimetry does not present some of the theoretical or practical limitations which restrain the use of other calorimetric techniques in adsorption or heterogeneous catalysis studies. Provided that some relatively simple calibration tests and preliminary experiments, which have been described, are carefully made, the heat evolved during fast or slow adsorptions or surface interactions may be measured with precision in heat-flow calorimeters which are, moreover, particularly suitable for investigating surface phenomena on solids with a poor heat conductivity, as most industrial catalysts indeed are. The excellent stability of the zero reading, the high sensitivity level, and the remarkable fidelity which characterize many heat-flow microcalorimeters, and especially the Calvet microcalorimeters, permit, in most cases, the correct determination of the Q-0 curve—the energy spectrum of the adsorbent surface with respect to... [Pg.259]

Microprobe laser desorption laser ionisation mass spectrometry (/xL2MS) is used to provide spatial resolution and identification of organic molecules across a meteorite sample. Tracking the chemical composition across the surface of the meteorite requires a full mass spectrum to be measured every 10 p,m across the surface. The molecules must be desorbed from the surface with minimal disruption to their chemical structure to prevent fragmentation so that the mass spectrum consists principally of parent ions. Ideally, the conventional electron bombardment ionisation technique can be replaced with an ionisation that is selective to the carbonaceous species of interest to simplify the mass spectrum. Most information will be obtained if small samples are used so that sensitivity levels should be lower than attomolar (10—18 M) fewer than 1000 molecules can be detected and above all it must be certain that the molecules came from the sample and are not introduced by the instrument itself. [Pg.169]

We have only found one example of the application of an immunoassay kit to the analysis of fluoroquinolones in environmental samples [84]. The assay is able to detect enrofloxcin as standard analyte with sensitivity levels of 5 pg L1. [Pg.216]

An HPLC detector is often a modified spectrophotometer equipped with a small flow cell, which monitors the concentration (or mass) of eluting analytes.Common detectors in the pharmaceutical laboratory are listed in Table 2 with their respective attributes and sensitivity levels. A recent survey found that 85% of pharmaceutical applications use absorbance detectors such as UVA/ is or photodiode array detectors (PDA). These two detectors are covered in more detail in this section. [Pg.62]

Lueking et al. (1999) arrayed recombinant proteins on NC membranes and screened them with different antibodies. Joos and coworkers (2000) printed down autoantigens onto NC membranes and compared performance relative to silylated (aldehyde) and PLL glass slides. Protein arrays could be stored at room temperature for a month without significant loss in activity. Huang (2001) hand spotted down IgC species and antibodies directed toward various cytokines onto membranes. The properties of various commercial membranes were assessed in terms of absorption, background, and sensitivity levels based upon detection by enhanced chemiluminescence (ECL). [Pg.60]

In neutral medium, and in the presence of air and cupric acetate, the ketol group is first oxidized into a ketoaldehyde that then condenses with o-phenylenediamine to produce a quinoxaline derivative. This derivative absorbs strongly in the near-UV range [82], with all of the corticosteroids affording equivalent sensitivity levels [72]. [Pg.209]

NTO or NTO/RDX or HMX have been reported as fillings to get insensitive munitions (IM) [155]. A comparison of PBXs based on TATB and NTO further confirms the same sensitivity levels while VOD of the NTO-based PBX is slightly higher. [Pg.113]

Two main models are usually discussed for the mechanism of the spectral sensitization. The excitation of the sensitizer by absorbed light and electron transfer from the excited sensitizer to the semiconductor is the first model. The alternative mechanism consists of the transfer of the excitation energy from the sensitizer to the semiconductor. This energy is used for photogeneration of the charge carriers in the sensitized photoconductor. In the first case the excited singlet level of the sensitizers has to be located above the conduction band of the semiconductor for realization of the electron transfer. For hole transfer the basic sensitizer level has to be located lower than the valence band of the sensitized photoconductor. The energy transfer mechanism does not need a special mutual location of the semiconductor and sensitizer levels. [Pg.13]

The glass floats, which control the liquid levels in the distillation chambers, leaked and sank and thus caused problems with liquid level control in the distillation chambers. New floats were made, but minor leakage persisted. The float has since been eliminated, and a more sensitive level detecting device is now used. [Pg.561]

Because of environmental concerns, a new generation of biodegradable penetrants having sensitivity levels ranging from 1 to 4 has been developed. The new penetrants are water washable and. in most instances, can be directly discharged into sewers. They are free of petroleum-based solutions. [Pg.1093]

Figure 27-10 Photoelectron spectra of ethene, ethyne, and benzene induced by 58.4-nm radiation from a helium-discharge lamp. For ethyne, the left part of the spectrum is shown with three different sensitivity levels. The horizontal scale here is in units of electron volts (eV), which can be converted to kcal mole-1 by multiplying by 23.06. (Published by permission of A. D. Baker and D. W. Turner, and of Accounts of Chemical Research.)... Figure 27-10 Photoelectron spectra of ethene, ethyne, and benzene induced by 58.4-nm radiation from a helium-discharge lamp. For ethyne, the left part of the spectrum is shown with three different sensitivity levels. The horizontal scale here is in units of electron volts (eV), which can be converted to kcal mole-1 by multiplying by 23.06. (Published by permission of A. D. Baker and D. W. Turner, and of Accounts of Chemical Research.)...

See other pages where Sensitization level is mentioned: [Pg.444]    [Pg.175]    [Pg.46]    [Pg.206]    [Pg.905]    [Pg.401]    [Pg.387]    [Pg.106]    [Pg.184]    [Pg.15]    [Pg.444]    [Pg.214]    [Pg.86]    [Pg.277]    [Pg.164]    [Pg.489]    [Pg.215]    [Pg.215]    [Pg.45]    [Pg.66]    [Pg.179]    [Pg.236]    [Pg.364]    [Pg.109]    [Pg.360]    [Pg.347]    [Pg.350]    [Pg.358]    [Pg.785]    [Pg.8]   
See also in sourсe #XX -- [ Pg.1073 ]




SEARCH



Inoculum levels, sensitivity

Laboratory rats sensitivity levels

Level sensitive scan design

Level-sensitive storage element

Moisture sensitivity level

Sensitivity at the Cellular Level

Sensitization energy level, triplet excited

Sensitizing dyes energy levels

© 2024 chempedia.info