Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Run times for

The integral of the Gaussian distribution function does not exist in closed form over an arbitrary interval, but it is a simple matter to calculate the value of p(z) for any value of z, hence numerical integration is appropriate. Like the test function, f x) = 100 — x, the accepted value (Young, 1962) of the definite integral (1-23) is approached rapidly by Simpson s rule. We have obtained four-place accuracy or better at millisecond run time. For many applications in applied probability and statistics, four significant figures are more than can be supported by the data. [Pg.16]

Several steps can be taken to maximize the run time for the reciprocating compressor. Since wear is a function of rubbing speed, the piston speed can be kept to a minimum. Chapter 3 made recommendations for piston speed. Reliability problems due to valves are reputed to account tor 40% of the maintenance cost of the compressor. Valves are the single largest cause for unplanned shutdowns. Basically, valve life can he increased by keeping the speed of the compressor as low as practical. At 360 rpm, the valves are operated six times a second. At 1,200 rpm, ihc valves operate 20 times a second or 1,728,000 times in a day. It is not difficult to understand why the valves are considered critical. To keep the reliability in mind, valve type, material selection and application considerations such as volume ratio, gas corrosiveness, and gas cleanliness need attention by the experts. One final note is that while lubrication is an asset to the rubbing parts, it is not necessarily good for valve reliability. [Pg.475]

The fact that according to (3.1) the run time for a spectrum is inversely proportional to the square of the peak absorption abSj. and that the noise improves only with the... [Pg.48]

A calibration curve for doxazosine was constructed in the range of 0 to 100 ng/mL with a correlation coefficient over 0.999. Intra-day and inter-day coefficients of variation were 5.7 and 8.0%, respectively. The limit of detection and the limit of quantification were 0.02 and 0.07 ng/mL, respectively. This validated method had a very short run time of 2 min compared with a 15-min run time for HPLC-fluorescence54 and may be used for TDM of doxazosine. [Pg.313]

In the example above, it appears that all the bases are covered and they may well be, depending upon the analyst s needs. It is easy to overlook valuable functions that we may expect without further thought. In this case, we have not specified that media should be preheated while a previous method is running. The sequence of events has not been well characterized. Here, it can cause a delay in run time for the batch is media is not heated prior to completion of the prior batch. [Pg.389]

Duverneuil and coworkers (2003) have developed a method for the determination of 11 of the most commonly prescribed non-tricyclic antidepressants and some of their metabolites these include paroxetine, fluoxetine, norfluoxetine, sertraline, citalopram, fluvoxamine mirtazapine, venlafaxine, and 0-des-methylvenlafaxine. The method involves an LLE procedure followed by an HPLC separation with photodiode-array UV detection at three different wavelengths (220, 240, and 290 nm). The total run time was 18 min. The extraction recoveries were calculated to be in the range of 74-109% and the lower limit of detection (LLOD) reported was 2.5-5 ng/ml. A method published by Tournel and associates (2001) also reported the simultaneous determination of several newer antidepressants by RP-HPLC with UV detection. The compounds were isolated from human serum using an LLE process. The LLOQ ranged from 15-50 ng/ml depending on the analyte of interest. The total run time for all compounds eluted was approximately 20 min. [Pg.32]

For very long gradient run times, for example 6h, we can reach a peak capacity of around 250 with this column. The flow rate for this gradient would be around O.hmlVmin. For more realistic conditions, for example a 45-min gradient, a peak capacity of 194 can be achieved at a flow rate of 1.4mL/min. Even for a 45-min gradient, the flow rate is... [Pg.86]

Based on the optimization criterion, SpinPro can select the most appropriate rotor. For example, suppose the investigator has a relatively large sample volume, all of which needs to be processed as soon as possible. The "minimize cumulative run time" criterion would be the appropriate choice. SpinPro would then initiate the following rotor selection procedure SpinPro determines the total sample volume based on inputs of the sample volume, the current concentration of the sample, and a correction for any pre-run dilutions of the sample. Next, consideration is made for whether tubes or bottles will be used. The program then evaluates rotors for the number of tube positions and the amount of sample per tube. At this point, SpinPro will have estimated for each rotor the number of runs required to process the sample. SpinPro then estimates the run time for each rotor to perform a single run. Based on these estimates, SpinPro selects the rotor that will give the shortest total run time when the run time is summed over the total number of runs. Similarly, the investigator can select any of the optimization criteria and initiate a variety of precise rotor selection procedures. [Pg.300]

The Optimal Plan requires 38% of the Lab Plan run time for a single run. It requires 38% of the Lab Plan run time when processing the entire sample. [Pg.303]

Rotor/run conditions SW 55 Ti rotor at 55000 rpm for approximately 6 hours. These recommendations form the core of any procedure. SpinPro usually considers more factors in the rotor selection process than does the expert. In determining the run speed, SpinPro considers every possible reason to reduce the run speed. If there are none, the rotor is run at full speed. When there are reasons (e.g., when using salt gradients, bottles, differential pelleting, or discontinuous runs), the run speed may have to be reduced dramatically, from 80,000 rpm to 40,000 rpm, for example. There are many cases of rotors being run too slow for the application or too fast for safety. Accurate determination of the run time is a complex problem based on the gradient characteristics, calculations, interpolations from numerical tables, and experience. SpinPro employs all of these methods in order to infer run times for many special cases. [Pg.304]

The theory of operation of each circuit is discussed, followed by the circuit schematic, the simulation results, and a comparison to laboratory data. Advantages and disadvantages of each circuit are added, along with any tips or hints useful in modeling the circuit accurately. We have attempted to perform each simulation using several versions of SPICE for comparison. Also included are the run times for each circuit simulation. [Pg.3]

Postpreparative Stability. The stability of processed samples, including the resident time in the autosampler, should be determined. The stability of the drug and the internal standard should be assessed over the anticipated run time for the batch size in validation samples by determining concentrations on the basis of original calibration standards. Reinjection reproducibility should be evaluated to determine if an analytical run could be reanalyzed in the case of instrument failure. [Pg.114]

Thirty minutes is sufficient for sample preparation beginning with refined vegetable oils. However, the other sample preparations take up to 120 min because of the incubation, re-extraction, and evaporation steps. Usually, a batch of samples is prepared at the same time to reduce the preparation time per sample. The running time for HPLC analysis is approximately 18 min for normal phase or 14 min for reversed phase in order to quantify all tocols. The running time for normal phase could be cut to 12 min for samples without 5-tocopherol and tocotrienol. [Pg.489]

The Basic Protocol requires 60 min of running time for HPLC analysis after each injection. The period between 55 and 60 min allows for column equilibration prior to the next injection. Analysis of the acidic fraction described in the Alternate Protocol is completed within 30 min including column equilibration time. HPLC analysis of the neutral fraction described in the Alternate Protocol needs much more time (60 min) for sample runs. It is desirable to analyze the sample on the same day as the extraction to prevent possible polyphenolic deterioration. [Pg.1265]

Stiff IVPs are best solved by the integrators of the MATLAB ode.. . suite of functions that end in the letter s. In practical terms, one need not construct a linearization to check for stiffness, but rather compare the run times for ordinary integrators (without an s in their MATLAB name) and for stiff integrators (with an s ). If the ordinary ones take too long, the IVP problem is most likely stiff and a MATLAB integrator with a name ending in s should be used to solve the problem more successfully. [Pg.546]

If the conversion of octanol is plotted over the run time for katapak-S , DX and a typical monolith experiment (Fig. 8.35), katapak-S shows clearly the highest, and DX-packings the lowest, space-time yield. [Pg.260]

NP-HPLC Normal-phase liquid chromatographic methods applying Diol-columns or common silica columns are well suited for the analysis of the total steryl ferulate content. They require very little sample preparation, as total lipid extracts can frequently be directly injected into the column without purification or fractionation. Run times for SFs are also relatively short, and a good separation from other lipid components can be obtained in less than 10 min in traditional HPLC systems. Depending on the column type and the sample, SFs elute as one or two peaks. Two peaks are obtained from the separation of SFs, which have ferulic acid both in cis- and trans- configuration (Nystrom et ah, 2008). The relative retention time (obtained with a silica column and hexane/ethyl acetate 97 3 as eluent) of the cis- form is about 0.5 smaller than that of steryl irans -ferulates (Akihisa et al., 2000). [Pg.340]

The main benefit of monolithic columns is the separation speed. As an example of the practical use of ultrafast separation, the two-dimensional HPLC separation of tryptic digest BSA is shown. The fractions of the column effluent from the first dimension (lEX) separation was collected for 2 min each and injected into the second-dimension monolithic column (RP).The run time for the second dimension could not be more than 2 min (more than fraction collection time), and the monolithic column provided fast and efficient separation (Figure 3-26). [Pg.121]

There are many HPLC methods that are developed during the process of drug development. The chromatographer needs to understand the aim of analysis in order to make judicious choices prior to the commencement of method development and the implications it may have on the final method that is developed. The following should be considered method development time, the maximum run time for analysis, the number of samples expected per week, the complexity of the mixture, the structure of the main analyte (physicochemical properties), possible degradation pathways (i.e., hydrolysis, oxidative, photolysis, dethydration, thermolytic, racemization), and whether the analyte or analytes are ionizable. [Pg.348]

Figure 6.3 Change in Friction with Running Time for a Rubbed Film of Molybdenum Disulphide (Ref. 132)... Figure 6.3 Change in Friction with Running Time for a Rubbed Film of Molybdenum Disulphide (Ref. 132)...

See other pages where Run times for is mentioned: [Pg.225]    [Pg.329]    [Pg.469]    [Pg.332]    [Pg.548]    [Pg.1047]    [Pg.76]    [Pg.30]    [Pg.189]    [Pg.121]    [Pg.259]    [Pg.31]    [Pg.419]    [Pg.421]    [Pg.289]    [Pg.491]    [Pg.497]    [Pg.173]    [Pg.88]    [Pg.271]    [Pg.221]    [Pg.221]    [Pg.384]    [Pg.55]    [Pg.398]    [Pg.87]    [Pg.73]    [Pg.167]   
See also in sourсe #XX -- [ Pg.250 , Pg.253 , Pg.325 ]




SEARCH



Run time

Running

Running time

© 2024 chempedia.info