Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Relative humidities, equipment

In many cases an optimized method may produce excellent results in the laboratory developing the method, but poor results in other laboratories. This is not surprising since a method is often optimized by a single analyst under an ideal set of conditions, in which the sources of reagents, equipment, and instrumentation remain the same for each trial. The procedure might also be influenced by environmental factors, such as the temperature or relative humidity in the laboratory, whose levels are not specified in the procedure and which may differ between laboratories. Finally, when optimizing a method the analyst usually takes particular care to perform the analysis in exactly the same way during every trial. [Pg.684]

Moisture measurements are important in the process industries because moisture can foul products, poison reactions, damage equipment, or cause explosions. Moisture measurements include both absolute-moisture methods and relative-humidity methods. The absolute methods are those that provide a primaiy output that can be directly calibrated in terms of dew-point temperature, molar concentration, or weight concentration. Loss of weight on heating is the most familiar of these methods. The relative-humidity methods are those that provide a primaiy output that can be more direc tly calibrated in terms of percentage of saturation of moisture. [Pg.765]

A common method of stating flow is standard cubic feet per minute where the flowing conditions are referred to an arbitrary set of standard conditions. Unfortunately, standard conditions are anything but standard. Of the many used, two are more common. The ASME standard uses 68°F and 14.7 psia. The relative humidity is given as 36%. The other standard that is used by the gas transmission industry and the API Mechanical Equipment Standards is 60°F at 14.7 psia. As can be seen from this short discussion, a flow value must be carefully evaluated before it can be used in a compressor calculation. [Pg.21]

Air Temperamre Design temperature for equipment and outdoor transformers shall be 98°F. Design Temperature for the Installation of Service Equipment - Maximum Temperamre 100°F - Minimum Temperamre 60°F Design Temperamre for Calculation of the Heat Loss - Normal Daily Average 79°F. The maximum mean relative humidity is 97%. Mean relative humidity -82%... [Pg.313]

The removal of heat within an enclosed space must be considered as a multi-step heat transfer process. Heat passes from the occupants or equipment to the air within the space, and from there to the refrigerant or chilled water. It follows that the temperature differences at each step are a reciprocal function of the air mass flow. Where there is a high latent heat load within the space, the relative humidity will also vary with the air flow - the variation being higher with low air flow. [Pg.299]

In addition, there are hand-held psychrometers for measuring air temperature and relative humidity available, and also hand-held wind meters, which measure wind speed. When using such hand-held equipment, readings should be taken each hour during the course of the day when worker exposure volunteers or re-entry worker volunteers are performing the test. [Pg.994]

The maximum capacity is about 200 persons. The room temperature should not exceed 26 ° C at a relative humidity of 5 0%. The maximum cooling load was calculated assuming 4.5 kW for lighting and 1 kW for other technical equipment. The result is a maximum latent cooling load of 22 kW (73%) and sensible of 8 kW (27%). [Pg.421]

Fig. 3.29 Resistances of the 38 sensor segments of a Sn02-microarray equipped with Si02-membrane during alternating exposure to carbon monoxide contaminated humid air and clean air with the same relative humidity... Fig. 3.29 Resistances of the 38 sensor segments of a Sn02-microarray equipped with Si02-membrane during alternating exposure to carbon monoxide contaminated humid air and clean air with the same relative humidity...
An industrial standard method has been developed to test the lightfastness of polymers in accelerated test equipment [103]. The apparatus consists of a quartz-xenon tube with a special optical filter between the light source and the specimen to produce light that resembles window glass-filtered daylight [104], Samples are mounted at a specific distance from the arc and are supported on a frame which revolves around the arc 1 to 5 times per minute for uniform exposure. A blower unit in the base provides a flow of air which makes it possible to maintain a black panel temperature of 45°C, measured by a black panel thermometer which is positioned at level with the samples. A black panel unit consists of a bimetallic thermometer mounted on a steel frame. Both faces of the frame plate and also the stem of the thermometer are coated with a heat-resistant glossy black enamel. The relative humidity level in the exposure cabinet is closely controlled. [Pg.90]

An environment should have appropriate controls for temperature, pressure, and relative humidity. For aseptic production, cleanroom conditions monitored for particles and bioburden contamination are necessary. Equipment must be validated and maintained with current calibration. Processes must be developed and validated to ensure the production of pure and consistent product. [Pg.323]

Temperature and relative humidity cause problems because they are difficult to control except when special equipment is used. The variation depends on the weather and the quality of the climate control, but they do vary. Their effect on the retention is different for different solutes. Therefore the resolution can change. [Pg.233]

Unfortunately, the design as developed could not completely be performed because of breakdown of the equipment. Therefore some experiments were performed at the temperature and relative humidity of the laboratory. The design used consists of 65 design points. [Pg.248]

Drip loss, free drip, or thaw exudate Amount of liquid expelled from a sample without the application of an external force over a specified time interval Drip loss (%) = (drip volume/initial sample weight) x 100 Advantages Simple test requiring little equipment. Commonly used for meat products. Disadvantages Test may take 24 hr to complete. If not controlled, results may be influenced by environmental conditions, such as relative humidity or temperature. Hon ikel (1987)... [Pg.295]

Use a laboratory-sized EtO sterilizer during early phases of the validation process as long as the sterilizer is equipped with devices allowing variability in vacuum, relative humidity, temperature, gas pressure, timing, and rate of gassing the chamber. Involve production sterilizer experts in these early phases of the EtO validation process. [Pg.151]

Verify the calibration of all instrumentation involved in monitoring the EtO cycle. Examples include thermocouple and pressure gauge calibration, gas leak testing equipment, relative humidity sensors, and gas chromatographic instrumentation. [Pg.151]

Critical environmental parameters that include, but are not limited to, relative humidity and temperature, must be maintained within the computer equipment manufacturer s requirements. In cases where there are a number of requirements, the most stringent should be utilized. [Pg.107]

Operation with low relative humidity of the gases at the stack inlet is preferred because it simplifies the system (humidification of reactant gases and water recovery). PEM fuel cells are operational even at room temperature, but the typical operating temperature is between 60°C and 80°C. In order to reduce the size of the heat rejection equipment there is a lot of reserch and development (R D) on high temperature membranes that would allow operation at 130-140°C. [Pg.115]


See other pages where Relative humidities, equipment is mentioned: [Pg.197]    [Pg.197]    [Pg.362]    [Pg.458]    [Pg.460]    [Pg.64]    [Pg.324]    [Pg.505]    [Pg.1160]    [Pg.79]    [Pg.158]    [Pg.201]    [Pg.293]    [Pg.774]    [Pg.366]    [Pg.626]    [Pg.7]    [Pg.181]    [Pg.59]    [Pg.357]    [Pg.353]    [Pg.64]    [Pg.79]    [Pg.246]    [Pg.517]    [Pg.637]    [Pg.262]    [Pg.214]    [Pg.103]    [Pg.181]    [Pg.222]   


SEARCH



Humidity, relative

Relative humidities, equipment ambient

© 2024 chempedia.info