Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Reference measurements direct

Wlien an electrical coimection is made between two metal surfaces, a contact potential difference arises from the transfer of electrons from the metal of lower work function to the second metal until their Femii levels line up. The difference in contact potential between the two metals is just equal to the difference in their respective work fiinctions. In the absence of an applied emf, there is electric field between two parallel metal plates arranged as a capacitor. If a potential is applied, the field can be eliminated and at this point tire potential equals the contact potential difference of tlie two metal plates. If one plate of known work fiinction is used as a reference electrode, the work function of the second plate can be detennined by measuring tliis applied potential between the plates [ ]. One can detemiine the zero-electric-field condition between the two parallel plates by measuring directly the tendency for charge to flow through the external circuit. This is called the static capacitor method [59]. [Pg.1894]

Capacity. Pumps deHver a certain capacity, Q, sometimes referred to as flow, which can be measured directly by venturi, orifice plate (11), or magnetic meters (12) (see Flow measurement). The indirect way to determine capacity is often used. Whereas this method is less accurate than applying a flow meter, it often is the only method available in the field. The total head is measured and the capacity found from the pump head—capacity (H— curve (Fig. 2). More recently, sonic flow meters (13) have been used, which can be installed on the piping without the need for pipe disassembly. These meters are simple to use, but require relatively clean single-phase Hquid for reHable measurements. [Pg.288]

Here the voltage is measured directly between reference electrodes at the positions a and a,. [Pg.136]

The precursive anisotropy in the thermal expansion for an In-26.5at%Tl alloy is shown in Figure 1 taken from reference 7, where for curve I the measurement direction becomes a c axis in the transformed crystal and for curve II it becomes an a axis. [Pg.335]

Also, a chromatographic profile or fingerprint of trace unknowns can be established and monitored, so that if product performance unexpectedly changes, there will be a starting point for troubleshooting. The effects of experimental variables on sample recoveries should be measured directly by controlled variation of an experimental factor, using the reference standard, or suitable external standards, or spiked addition of an external standard to the reference standard. A detailed example of the use of internal and external standards is presented in Chapter 4. [Pg.30]

Hence, a plot of log (pAR /(I -pAR)) against log [A] should give a straight line with a slope of one. Such a graph is described as a Hill plot, again after A. V. Hill, who was the first to employ it, and it is often used whenpAR is measured directly with a radiolabeled ligand (see Chapter 5). In practice, the slope of the line is not always unity, or even constant, as will be discussed. It is referred to as the Hill coefficient (%) the term Hill slope is also used. [Pg.9]

Direct reference measurements by the relation of measured value to content (concentration) of a reference material (RM)... [Pg.62]

In the simplest case, experimental calibration can be carried out by direct reference measurements where the sensitivity factor b is given by the relation of measured value to concentration of a reference material (RM), b = yRvi/xRv,. Direct reference calibration is frequently used in NAA and X-ray analytical techniques (XRF, EPMA, TXRF). [Pg.150]

In order to ameliorate the sharply sloping background obtained in an STS spectrum, the data are often presented as di,/dFh vs. Vb, i.e. the data are either numerically differentiated after collection or Vb has a small modulation applied on top of the ramp, and the differential di,/d Vb is measured directly as a function of Vb. The ripples due to the presence of LDOS are now manifest as clear peaks in the differential plot. dt,/dFb vs. Vb curves are often referred to as conductance plots and directly reflect the spatial distribution of the surface electronic states they may be used to identify the energy of a state and its associated width. If V is the bias potential at which the onset of a ripple in the ijV plot occurs, or the onset of the corresponding peak in the dt/dF plot, then the energy of the localised surface state is e0 x F. Some caution must be exercised in interpreting the differential plots, however, since... [Pg.83]

Saturation properties such as solubility in water and vapor pressure can be measured directly for solids and liquids. For certain purposes it is useful to estimate the solubility that a solid substance would have if it were liquid at a temperature below the melting point. For example, naphthalene melts at 80°C and at 25°C the solid has a solubility in water of 33 g/m3 and a vapor pressure of 10.9 Pa. If naphthalene was a liquid at 25°C it is estimated that its solubility would be 115 g/m3 and its vapor pressure 38.1 Pa, both a factor of 3.5 greater. This ratio of solid to liquid solubilities or vapor pressures is referred to as the fugacity ratio. It is 1.0 at the melting point and falls, in this case at lower temperatures to 0.286 at 25°C. [Pg.9]

Air-dry soil is mixed with 0.02 M calcium chloride solution (1 2 ratio, for instance, 10-g soil 20 mL 0.02 M CaCl2 solution) and mixed for 1 hour. The pH of the suspension can be measured directly. In addition, the solution can be filtered for the determination of aluminum or magnesium by atomic absorption spectroscopy (AAS) or inductively coupled plasma (ICP) spectroscopy (adapted from Reference 5). [Pg.232]

The stepwise formation of fluoro complexes in aqueous solution has been extensively studied in recent years by standard techniques (192). The introduction of the LaF3 electrode has enabled fluoride activities to be measured directly, even with the "insoluble fluorides such as CaF2 or PbF2 (34, 81). Values obtained depend on the salts added to maintain constant ionic strength, and care must be taken in making comparisons since not all results are extrapolated to zero strength (15). Key references to work in this field can be gleaned from standard compilations of stability constants. [Pg.29]

An efficient way of overcoming this difficulty is to use a reference fluorophore (instead of a scattering solution) (i) whose fluorescence decay is a single exponential, (ii) which is excitable at the same wavelength as the sample, and (iii) which emits fluorescence at the observation wavelength of the sample. In pulse fluorometry, the deconvolution of the fluorescence response can be carried out against that of the reference fluorophore. In phase-modulation fluorometry, the phase shift and the relative modulation can be measured directly against the reference fluorophore. [Pg.181]

DR. SUTIN I think that this is always a problem with exchange reactions involving aquometal ions. The manganese system bears out what Dr. Taube said we have to be sure of the numbers which we are attempting to interpret. In this regard, I believe that the manganese self-exchange rate constant which you referred to was not measured directly. [Pg.132]

References. Because the detected fluorescence signal is a direct response of the dye-analyte complex formed, no reference measurement is required. Also no calibration of the probe is required, although the response function of the probe may be needed. [Pg.196]

The maximum closedloop log modulus does not have these pioblems since it measures directly the closeness of the G B curve to the (—1,0) point at all frequencies. The closedloop log modulus refers to the closedloop servo transfer function ... [Pg.474]

Provide a reference or direct experimental proof that the conditions chosen do provide initial rate measurements. At a minimum, the percentage of substrate consumed during the course of an initial rate determination should be specified. One should also show that under initial rate conditions a doubling of enzyme concentration should exactly produce a doubling in the observed initial rate. Likewise, if an auxiliary enzyme assay is used to monitor the primary enzyme s activity the observed rate at low or high substrate concentration should not depend on the concentration of additional enzyme(s), substrate(s), or factors used for the coupled assay. 2. Describe all assay conditions (eg., concentrations of substrates, products, inhibitors, and/or activators enzyme concentration temperature pH and buffer composition ... [Pg.364]

E is the standard equilibrium potential, i. e. the potential corresponding to unit activity and RTF. The dissolution reaction leads to the development of an electrical double layer at the iron-solution interface. The potential difference of the Fe/Fe " half cell cannot be measured directly, but if the iron electrode is coupled with a reference electrode (usually the standard hydrogen electrode, SHE), a relative potential difference, E, can be measured. This potential is termed the single potential of the Fe/Fe electrode on the scale of the standard hydrogen couple H2/H, the standard potential of which is taken as zero. The value of the equilibrium potential of an electrochemical cell depends upon the concentrations of the species involved. [Pg.491]

Here either the C.P.D. between a surface A and a reference surface R is measured directly, or else some property dependent on the C.P.D. is measured. In each case, when the surface A is changed to A by the process of adsorption, a varying potential is applied to A until it behaves as A alternatively, the new C.P.D. is measured directly. This change in potential is the required C.P.D. between A and A. Thus, if Var and Va r are the C.P. differences between the reference electrode R and the clean and covered surfaces A and A, respectively,... [Pg.87]

This AE can be measured directly since there is no ohmic drop at t > tx. The coulostatic impulse method is, therefore, particularly suitable for kinetic studies in low-conducting media. For further details, the reader is referred to the extensive reviews of van Leeuwen [39]. [Pg.227]

If there is a bottleneck at r=30, the system is much more likely to find it and suddenly leak through if not, one has a least measured the equilibrium distribution of r in a region where it would be too low to measure directly. The normalizing factor Q/Q, necessary to make the connection between p and p, can found be from the histograms via eq. 20 or, more accurately, by eqs. 12a and 12b of reference 17. [Pg.97]

ACTIVITY SERIES- Also referred to as the electromotive series or the displacement series, this is an arrangement of the metals (other elements can be included) in the order of their tendency to react with water and acids, so that each metal displaces from solution those below itiu the series and is displaced by those above it. See Table 1. Since the electrode potential of a metal in equilibrium with a solution of its ions cannot be measured directly, the values in the activity series are, in each case, the difference between the electrode potential of the given metal tor element) in equilibrium with a solution of its ions, and that of hydrogen in equilibrium with a solution of its ions. Thus in the table, it will be noted that hydrogen lias a value of 0.000. In experimental procedure, the hydrogen electrode is used as the standard with which the electrode potentials of other substances are compared. The theory of displacement plays a major role in electrochemistry and corrosion engineering. See also Corrosion and Electrochemistry. [Pg.31]

The term property refers to a characteristic of a material and can be measured. Examples are pressure, temperature and volume. Properties may also be computed, such as, for example, internal energy, which cannot be measured directly. An extensive property is one whose value is the sum of each of the subsystems comprising the entire system. An example is a gas mixture, in which each constituent (or subsystem) has masses or volumes different from the original system. Thus, mass or volume is an extensive property. [Pg.20]


See other pages where Reference measurements direct is mentioned: [Pg.20]    [Pg.503]    [Pg.369]    [Pg.285]    [Pg.57]    [Pg.367]    [Pg.29]    [Pg.549]    [Pg.345]    [Pg.278]    [Pg.97]    [Pg.144]    [Pg.78]    [Pg.91]    [Pg.152]    [Pg.320]    [Pg.176]    [Pg.557]    [Pg.173]    [Pg.425]    [Pg.369]    [Pg.94]    [Pg.107]    [Pg.631]    [Pg.286]    [Pg.340]   
See also in sourсe #XX -- [ Pg.124 ]

See also in sourсe #XX -- [ Pg.124 ]




SEARCH



Direct measure

Direct measurement

Directly measured

Reference measurements

© 2024 chempedia.info