Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Peak symmetry

In exact quantitative work (integration of the peak areas), a maximum asymmetry must not be exceeded otherwise, there will be errors in determining the cut-off point of the peak with the base line. [Pg.164]

T should ideally be 1.0 for a symmetrical peak, but for a practical quality measure, it should not be greater than 2.5. If the tailing exceeds higher values, there will be errors in the quantitative area measurement because the point where the peak reaches the base line is very difficult to determine and has large impact on the final peak area. [Pg.164]


After packing and testing the column, one may find that the plate count or the peak symmetry is unsatisfactory. In this case some adjustments need to... [Pg.168]

All packing materials produced at PSS are tested for all relevant properties. This includes physical tests (e.g., pressure stability, temperature stability, permeability, particle size distribution, porosity) as well as chromatographic tests using packed columns (plate count, resolution, peak symmetry, calibration curves). PSS uses inverse SEC methodology (26,27) to determine chromatographic-active sorbent properties such as surface area, pore volume, average pore size, and pore size distribution. Table 9.10 shows details on inverse SEC tests on PSS SDV sorbent as an example. Pig. 9.10 shows the dependence... [Pg.288]

After column packing, each column is tested for theoretical plate count, peak symmetry, resolution, pore volume, and back pressure. If one of these tests fails the column is removed from the production cycle. If a PSS SEC column is kept in storage for a longer time, it is retested for theoretical plate count, peak symmetry, resolution, pore volume, and back pressure prior to shipping to the customer to prove up-to-date column performance. [Pg.289]

The efficiency of many CSPs increases dramatically when liquid eluents are replaced with sub- or supercritical fluids. During a comparison of LC and SFC performed with a Chiralcel OD CSP, Lynam and Nicolas reported that the number of theoretical plates obtained was three to five times higher in SFC than in LC [26]. The separation of metoprolol enantiomers by LC and SFC on a Chiralcel OD CSP is illustrated in Fig. 12-2. Although impressive selectivity is achieved by both techniques, resolution is higher in SFC (R = 12.7) than in LC (R = 4.8), and the higher flowrate in SFC reduces the analysis time. The increased efficiency of SFC also improves peak symmetry. [Pg.304]

Packed column SFC has also been applied to preparative-scale separations [42], In comparison to preparative LC, SFC offers reduced solvent consumption and easier product recovery [43]. Whatley [44] described the preparative-scale resolution of potassium channel blockers. Increased resolution in SFC improved peak symmetry and allowed higher sample throughput when compared to LC. The enhanced resolution obtained in SFC also increases the enantiomeric purity of the fractions collected. Currently, the major obstacle to widespread use of preparative SFC has been the cost and complexity of the instrumentation. [Pg.306]

Silica has often been modified with silver for argentation chromatography because of the additional selectivity conferred by the interactions between silver and Jt-bonds of unsaturated hydrocarbons. In a recent example, methyl linoleate was separated from methyl linolenate on silver-modified silica in a dioxane-hexane mixture.23 Bonded phases using amino or cyano groups have proved to be of great utility. In a recent application on a 250 x 1-mm Deltabond (Keystone Scientific Belief onte, PA) Cyano cyanopropyl column, carbon dioxide was dissolved under pressure into the hexane mobile phase, serving to reduce the viscosity from 6.2 to 1 MPa and improve efficiency and peak symmetry.24 It was proposed that the carbon dioxide served to suppress the effect of residual surface silanols on retention. [Pg.10]

The peak symmetry, resolution, and detector response are directly dependent on the concentration of the sample. As the concentration of a sample increases, the retention time, separation, and peak symmetry generally decrease. These phenomena are due to isotherm nonlinearity. The detector response may also be nonlinear above or below certain concentrations. In some cases, small amounts of a dilute component are irreversibly adsorbed to the column, leading to reduced recovery. Above some concentration, the response of any detector will cease to be linear. The UV-VIS is one of the most linear detectors, generally exhibiting at least three decades of linearity, while RI, electrochemical, and fluorimetric detectors have a markedly narrower range of linearity. [Pg.153]

HTLC Use of conventional HPLC Low backpressure Good peak symmetry Less organic in mobile phase Leverage of temperature for selectivity... [Pg.260]

Organic amine compounds can be separated successfully on silica gel columns with good peak symmetry using organic/aqueous mobile... [Pg.515]

The most popular and versatile bonded phase is octadecylsilane (ODS), n-C18H37, a grouping that is non-polar and used for reverse phase separations. Octylsilane, with its shorter chain length, permits faster diffusion of solutes and this results in improved peak symmetry. Other groups are attached to provide polar phases and hence perform normal phase separations. These include cyano, ether, amine and diol groups, which offer a wide range of polarities. When bonded stationary phases are used, the clear distinction between adsorption and partition chromatography is lost and the principles of separation are far more complex. [Pg.115]

For validation of columns that will be used for an official assay and to provide an unambiguous standard for qualifying future media lots, it is useful to employ more measurable comparative criteria than a simple overlay. Resolution, plate heights, and peak symmetry, as calculated by the classical formulae, should match very closely among test and reference columns (Figure 6.1, Figure 6.2). [Pg.83]

BGE dip After sample injection, dip in the capillary inlet into a BGE vial that is not the same vial as the run vial to reduce carryover and/or to increase peak symmetry. Check how often this dip vial needs to be replenished. [Pg.130]

The application of HPLC in routine environments, like pharmaceutical, food, or environmental analysis and particularly quality assurance, makes not only great demands on the robnstness of HPLC hardware, comprising pumps, column thermostats, and detection units, bnt in addition to the column reproducibility. Column reproducibility can be investigated at different levels of complexity Run-to-run reproducibility compares consecutive chromatographic runs, whereas long-term stability describes the column variance over several hundreds of injections. Column-to-column (batch-to-batch) reproducibility finally explores the match of independently fabricated chromatographic columns. Column characteristics that are routinely consulted for the determination of the robustness are retention, selectivity, column efficiency, and peak symmetry. [Pg.29]

More recently [635], a unique extraction step in supplemented foods, by using hot water and a precipitation solution, following by HPLC-ELD/UV analysis has been performed for the simultaneous determination of pyridoxine, thiamine, riboflavin, niacin, pantothenic acid, folic acid, cyanoco-balamin, and ascorbic acid. The mobile phase consisting of phosphate buffer and methanol has been modified in order to perform ion-liquid chromatography by adding l-octanesulfonic acid sodium salt. Furthermore, triethylamine has been also added to improve peak symmetry. [Pg.637]

Fig. 3.5. Parameters of a signal peak obtained in chromatography.Ve elution volume, tg elution time (retention time), h height of the peak, W width of the peak at 0.5 h, a and b indicators of peak symmetry... Fig. 3.5. Parameters of a signal peak obtained in chromatography.Ve elution volume, tg elution time (retention time), h height of the peak, W width of the peak at 0.5 h, a and b indicators of peak symmetry...
Having optimised the efficiency of a chromatographic separation the quality of the chromatography can be controlled by applying certain system suitability tests. One of these is the calculation of theoretical plates for a column and there are two other main parameters for assessing performance peak symmetry and the resolution between critical pairs of peaks. A third performance test, the peak purity parameter, can be applied where two-dimensional detectors such as diode or coulometric array or mass spectrometry detectors are being used. The reproducibility of peak retention times is also an important parameter for controlling performance. [Pg.201]

In the range of 15-30% acetonitrile, the different components of the mixture were separated according to size, in a size-exclusion mode, with very little nonspecific adsorption and good peak symmetry. It was believed that the addition of TEA to the solvent contributed to these desirable effects by neutralizing residual silanol groups. Present supports should not suffer from the presence of such uncapped silanol groups. [Pg.645]

Chromatographic results (peak symmetry, resolution, run time of first peak)... [Pg.201]

A simple estimate of peak symmetry S can be useful and is calculated as follows ... [Pg.215]

The sequence of injections from experiment 8, the original method conditions, is shown in Figure 2 from reference [19]). The calculation of peak symmetry for this design was abandoned as our calculation method could not adequately define peak start and peak end times. [Pg.221]

Carbowax 20M s U urates separated Good peak symmetry, not thermally stable... [Pg.139]

Relevant parameters for column-to-column reproducibility data are plate number, peak symmetry, selectivity, and adsorption phenomena checked with chromatography of amines and acids. For column life, the analyst must periodically check k loss, change in asymmetry factor, and plate number. [Pg.17]


See other pages where Peak symmetry is mentioned: [Pg.590]    [Pg.103]    [Pg.147]    [Pg.328]    [Pg.245]    [Pg.517]    [Pg.559]    [Pg.1]    [Pg.32]    [Pg.84]    [Pg.85]    [Pg.337]    [Pg.459]    [Pg.30]    [Pg.372]    [Pg.374]    [Pg.274]    [Pg.289]    [Pg.469]    [Pg.178]    [Pg.131]    [Pg.528]    [Pg.85]    [Pg.137]    [Pg.221]    [Pg.266]   
See also in sourсe #XX -- [ Pg.201 ]

See also in sourсe #XX -- [ Pg.24 , Pg.25 , Pg.26 ]

See also in sourсe #XX -- [ Pg.164 ]




SEARCH



© 2024 chempedia.info