Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

MIM analysis

MIM analysis was difficult to develop, as it required special HPLC and GPC conditions to determine the purity and mass spectroscopy to confirm the structures. [Pg.162]

The newest approach to measuring denitrification uses a mass spectrometer that can rapidly measure the N2/Ar ratio in water using membrane inlet mass spectrometry (MIMS, Kana et al., 1994, 1998). In this approach, sediment cores are incubated under continuous flow conditions. The N2/Ar ratio is then measured at the inlet and outlet of the core over time. Advantages of the technique are the short incubation times needed (usually less than 12 h), and the high throughput and small sample size ( 5 ml) of MIMS analysis. [Pg.1255]

From the data shown in Table 16.2 concerning the numbers of operating days and the number of analyses carried out each day at each of the four sites, it is found that averages of 2825 analyses were carried out with each filament and 1883 analyses were carried out between successive calibrations. Comparing these data with those of a GC/MS instrument used presently in our laboratory and which has shown good instrumental stability and reliability, it was found that the GC/MS instrument performed 305 analyses per filament and 78 analyses per calibration. When it is borne in mind that the duration of a GC/MS analysis was 50 min while that of a MIMS analysis was 5 min, it is clear that both systems performed comparably in terms of operation time per filament. [Pg.505]

For iron(III) eomplexes, uic venues /vlh [Fe(aepa)2]BPh4 H2O and k = 6.7 x 10 s for [Fe(mim)2(salacen)]PF6 have been obtained [156, 166]. The rate constants derived from the line shape analysis of Mossbauer spectra thus vary between 2.1 x 10 and 2.3 x 10 s at room temperature, no significant difference between iron(II) and iron(III) being apparent. In addition, it is evident that the rates of spin-state conversion in solution and in the crystalline solid are almost the same. For iron(II) eomplexes, for example, the solution rates vary between /cjjl = 5 x 10 and 2 x 10 s , whereas in solid compounds values between kjjL = 6.6 x 10 and 2.3 x 10 s have been obtained. Rates resulting from the relaxation of thermally quenched spin transition systems are considerably slower, since they have been measured only over a small range of relatively low temperatures. Extrapolation of the kinetic data to room temperature is, however, of uncertain validity. [Pg.147]

Combined synthesis/analysis operations were performed to characterize the MIMs and identify the impurities. The principal impurities were synthesized and isolated using HPLC. The intermediates and final products were purified on macroreticular resins. [Pg.162]

The combination of enantio-MDGC with high-resolution MS or mass-selective detectors, both used in full scan or (at least) in the multiple ion monitoring (MIM) mode is currently the most potent analytical tool in enantioselective analysis of chiral compounds from complex mixtures. [Pg.384]

The PPP configuration analysis technique and the mo ecules-in-molecules/localized configuration interaction (MIM/LCI) method can provide an explanation of the fact that some cyclic sulfur diimides are colorless, while others are not. Thus, while 1,2,5-thiadiazoles 231 and 232 are colorless, naphtho[l,8- /][l 4,2,6]thiadiazine 229 is colored (low intensity absorption at 642 nm). The energy level of the color band appears to be largely determined by frontier orbital interactions. By considering the overlap between the two fragments, (i) the NSN group and (ii) the... [Pg.308]

ILs are also attractive candidates for replacing volatile organic compounds (VOCs) as solvents, because they have practically no vapor pressure [165]. However, the environmental impact of ILs and VOCs should be compared on the basis of life-cycle analysis, and for that we are still missing many data on the toxicity and environmental effects of I Ls [ 166,167]. Another point is that the current prices of I Ls are much higher than those of VOCs. Handy et al. recently demonstrated a handy synthesis of mim-type ILs starting from fructose, which could eventually lead to truly eco-friendly IL solvents [168]. [Pg.164]

Ibrahim MIM, Wertherimer AL 1998. Management leadership styles of effectiveness of community pharmacists A descriptive analysis. / Sue Admin Pharm 15 57. [Pg.146]

The core development of MIMS has centered upon MS ionization techniques for the analysis of aromatic contaminants in water,47 improving membrane extraction selectivity48 and enhancing MIMS sensitivity by cooling and heating the membrane (trap and release MIMS).49 50 A dedicated review has been written on the applications of MIMS in environmental analysis.51... [Pg.76]

The use of silicone membranes as an interface in MIMS for direct extraction and analysis by MS has fostered their implementation for extraction purposes that can be combined off-line or on-line with other analytical instrumentation, such as GC. The technique of membrane extraction with sorbent interface (MESI) (Figure 4.2) employs the pervaporation principle in a nonporous polymeric membrane unit, where the membrane is used as a selective barrier for the extraction of VOCs and SVOCs in gaseous or liquid samples. [Pg.76]

In an attempt to overcome the significant difficulties that the presence of water vapor poses to the analysis of very volatile compounds, purge-and-membrane extraction techniques have been developed that largely prevent the introduction of water into the analytical system. Typical implementations of this form of sample introduction have been called by its developers membrane extraction with a sorbent interface (MESI),97 or membrane introduction mass spectrometry (MIMS).98 " They are based on a silicone hollow-fiber membrane that is inserted into the sample to be monitored, and the passing of a certain volume of inert gas through the membrane. Volatile compounds permeate the membrane and are swept to the adsorbent trap from which they are desorbed into the GC. This method of sample introduction is particularly suited for field and process monitoring and for dirty samples, since it prevents any nonvolatile compounds from entering the analytical system.100... [Pg.320]

The use of membrane introduction mass spectrometry (MIMS) was first reported in 1963 by Hoch and Kok for measuring oxygen and carbon dioxide in the kinetic studies of photosynthesis [46], The membrane module used in this work was a flat membrane fitted on the tip of a probe and was operated in the MIS mode. The permeated anaytes were drawn by the vacuum in the MS through a long transfer line. Similar devices were later used for the analysis of organic compounds in blood [47], Memory effects and poor reproducibility plagued these earlier systems. In 1974, the use of hollow-fiber membranes in MIMS was reported, which was also operated in the MIS mode [48], Lower detection limits were achieved thanks to the larger surface area provided by hollow fibers. However, memory effects caused by analyte condensation on the wall of the vacuum transfer line remained a problem. [Pg.217]

A consequence of MESI application is the membrane introduction mass spectrometry system (MIMS), which allows selective extraction of VOCs and direct MS analysis without chromatographic separation. The gaseous substances permeate through the highly hydrophobic membrane and are introduced directly to a mass spectrometer [115-118]. [Pg.416]

Example 7.2 A sharp filter sand has the sieve analysis shown below. The porosity of the nnstratified bed is 0.39, and that of the stratified bed is 0.42. The lowest temperatnre anticipated of the water to be filtered is 4°C. Find the head loss if the sand is to be nsed in (a) a slow-sand filter 76 cm dee ) operated at 9.33 mim d and (b) a rapid-sand filter 76 cm deep operated at 117 m /m d. [Pg.357]

Pilot plant analysis on a mixed-media hlter shows that a hltration rate of 15 mim-h is acceptable. If a surface conhguration of 5 m x 8 m is ajtpro-priate, how many hlter units will be required to process 100,000 m /d of raw water ... [Pg.384]

Another MS technique used in connection to pyrolysis is MIMS (membrane introduction mass spectrometry). MIMS is in fact a special inlet for the mass spectrometer, where a membrane (usually silicone, non-polar) lets only certain molecule types enter the Ionization chamber of the MS. This allows, for example, direct analysis of certain volatile organic compounds from air. The system makes possible the coupling of atmospheric pyrolysis to a mass spectrometer [61a] allowing direct sampling of the pyrolysate. Other parts of the mass spectrometer do not need to be changed when using MIMS. [Pg.159]

Acknowledgements The authors thank Michele R. Mims for the assistance in sample analysis. Financial support from USDA CSREES Evans-Alien Program is also greatly appreciated. [Pg.130]


See other pages where MIM analysis is mentioned: [Pg.151]    [Pg.162]    [Pg.1372]    [Pg.123]    [Pg.598]    [Pg.151]    [Pg.162]    [Pg.1372]    [Pg.123]    [Pg.598]    [Pg.281]    [Pg.271]    [Pg.120]    [Pg.240]    [Pg.255]    [Pg.153]    [Pg.390]    [Pg.213]    [Pg.217]    [Pg.267]    [Pg.6493]    [Pg.6505]    [Pg.381]    [Pg.1359]    [Pg.1359]    [Pg.1371]    [Pg.89]    [Pg.476]    [Pg.536]    [Pg.214]    [Pg.181]    [Pg.9]    [Pg.19]    [Pg.433]    [Pg.1393]   
See also in sourсe #XX -- [ Pg.162 ]




SEARCH



MIM

© 2024 chempedia.info