Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass spectrometer pumping system

For capillary columns, the usual practice is to insert the exit end of the column into the ion source. This is possible because under normal operating conditions the mass spectrometer pumping system can handle the entire effluent from the column. It is then only necessary to heat the capillary column between the GC and the MS ion source, taking care to eliminate cold spots where analyte could condense. The interface must be heated above the boiling point of the highest-boiling component of the sample. [Pg.11]

The development of the open tubular columns eliminated the need for concentrating devices, as the mass spectrometer pumping system could dispense with the entire carrier gas flow from such columns. Consequently, the column flow was passed directly into the mass... [Pg.381]

The introduction of the open-tubular columns eliminated the need for concentrating devices as the mass spectrometer pumping system could cope with the entire column eluent. Consequently, the column eluent could be passed directly into the mass spectrometer and the total sample can enter the ionization source. The first mass spectrometer used in a GC-MS tandem system was a rapid-scanning magnetic sector instrument that easily provided a resolution of one mass unit. Contemporary mass spectrometers have vastly improved resolution and the most advanced system (involving the triple quadrupole mass spectrometer) gives high in-line sensitivity, selectivity, and resolution. [Pg.732]

The coupling of LC with MS requires an appropriate interface due to the incompatibilities of the two methods. In order to introduce a conventional LC flow (0.5-1.5 ml/ min) into the mass spectrometer, one should evaporate it, producing a vapor with a volume far beyond the capacity of the mass spectrometer pump systems. The selection of the interface is the key decision and a limiting factor in the utilization of the instrument. Some years ago, LC/MS systems were state-of-the-art, space-occupying machines. [Pg.1331]

TPD is frequently used to detenuine (relative) surface coverages. The area below a TPD spectrum of a certain species is proportional to the total amount that desorbs. In this way one can detennine uptake curves that correlate gas exposure to surface coverage. If tire pumping rate of the UHV system is sufiBciently high, the mass spectrometer signal for a particular desorption product is linearly proportional to the desorption rate of the adsorbate [20, 21] ... [Pg.1863]

Vacuum system. Components associated with lowering the pressure within a mass spectrometer. A vacuum system includes not only the various pumping components but also valves, gauges, and associated electronic or other control devices the chamber in which ions are formed and detected and the vacuum envelope. [Pg.430]

Moving-belt (ribbon or wire) interface. An interface that continuously applies all, or a part of, the effluent from a liquid chromatograph to a belt (ribbon or wire) that passes through two or more orifices, with differential pumping into the mass spectrometer s vacuum system. Heat is applied to remove the solvent and to evaporate the solute into the ion source. [Pg.433]

Solvent-divert system. Used in conjunction with an interface, it permits temporary interruption of the flow from a chromatograph to a mass spectrometer by briefly opening a valve to a pumping line. Thus effluent present at a high concentration (usually solvent) does not enter the mass spectrometer ion source. [Pg.433]

Two vacuum systems are used to provide both the high vacuum needed for the mass spectrometer and the differential pumping required for the interface region. Rotary pumps are used for the interface region. The high vacuum is obtained using diffusion pumps, cryogenic pumps, or turbo pumps. [Pg.626]

A versatile Laser-SNMS instrument consists of a versatile microfocus ion gun, a sputtering ion gun, a liquid metal ion gun, a pulsed flood electron gun, a resonant laser system consisting of a pulsed Nd YAG laser pumping two dye lasers, a non-resonant laser system consisting of a high-power excimer or Nd YAG laser, a computer-controlled high-resolution sample manipulator on which samples can be cooled or heated, a video and electron imaging system, a vacuum lock for sample introduction, and a TOF mass spectrometer. [Pg.135]

For IBSCA analysis, standard HV or, better, UHV-equipment with turbomolecular pump and a residual gas pressure of less than 10 Pa is necessary. As is apparent from Fig. 4.46, the optical detection system, which consists of transfer optics, a spectrometer, and a lateral-sensitive detector, is often combined with a quadrupole mass spectrometer for analysis of secondary sputtered particles (ions or post-ionized neutrals). [Pg.242]

GCMS Analysis. The GCMS system utilized a Bendix model 2200 chromatograph connected to an AEI MS902 mass spectrometer, equipped with a high speed pumping system (9-Systems,... [Pg.337]

The DBMS setup and experimental procedures used in this study were the same as described in more detail elsewhere [Jusys et al., 2001]. Briefly, the DBMS setup consisted of two differentially pumped chambers, a Balzers QMS 112 quadrupole mass spectrometer (MS), a Pine Instruments potentiostat, and a computerized data acquisition system. [Pg.415]

The central transport chamber is an 80-cm-diameter stainless steel vessel, and is pumped by a 1000-1/s turbomolecular pump, which is backed by a small (501/s) turbomolecular pump to increase the compression ratio for hydrogen, and by a 16-m /h rotating-vane pump. UHV is obtained after a bake-out at temperatures above 100°C (measured with thermocouples at the outside surface) of the whole system for about a week. A pressure in the low 10 " -mbar range is then obtained. With a residual gas analyzer (quadrupole mass spectrometer, QMS) the partial pressures of various gases can be measured. During use of the system, the pressure in the central chamber is in the low 10 -mbar range due to loading of samples. Water vapor then is the most abundant species in the chamber. [Pg.22]

The partial pressures of the stable neutral molecules in the discharge (silane, hydrogen, disilane, trisilane) can be measured by a quadrupole mass spectrometer (QMS). The QMS usually is mounted in a differentially pumped chamber, which is connected to the reactor via a small extraction port [286]. In the ASTER system a QMS is mounted on the reactor that is used for intrinsic material deposition. The QMS background pressure (after proper bake-out) is between 10 and 10 mbar. The controllable diameter in the extraction port is adjusted so that during discharge operation the background pressure never exceeds 10"" mbar. [Pg.85]

FIG. 35. Vertical cross section of the reaction chamber equipped with the mass spectrometer system. Indicated are QMF. the quadmpole mass filter ESA. the electrostatic analyzer CD, the channeltron detector DE, the detector electronics DT, the drift tube lO, the ion optics TMP, the turbomolecular pump PR, the plasma reactor and MN. the matching network. [Pg.93]

Electrospray Ionization - Mass Spectrometry (ESI-MS). The Jacobsen s Co-salen catalysts dissolved in dichloromethane were pumped to the mass spectrometer system after dilution with methanol at a flow rate of 50 pi min and 600 scans were collected in 1 min. [Pg.392]

Schematic representation of the experimental setup is shown in Fig 1.1. The electrochemical system is coupled on-line to a Quadrupole Mass Spectrometer (Balzers QMS 311 or QMG 112). Volatile substances diffusing through the PTFE membrane enter into a first chamber where a pressure between 10 1 and 10 2 mbar is maintained by means of a turbomolecular pump. In this chamber most of the gases entering in the MS (mainly solvent molecules) are eliminated, a minor part enters in a second chamber where the analyzer is placed. A second turbo molecular pump evacuates this chamber promptly and the pressure can be controlled by changing the aperture between both chambers. Depending on the type of detector used (see below) pressures in the range 10 4-10 5 mbar, (for Faraday Collector, FC), or 10 7-10 9 mbar (for Secondary Electrton Multiplier, SEM) may be established. Schematic representation of the experimental setup is shown in Fig 1.1. The electrochemical system is coupled on-line to a Quadrupole Mass Spectrometer (Balzers QMS 311 or QMG 112). Volatile substances diffusing through the PTFE membrane enter into a first chamber where a pressure between 10 1 and 10 2 mbar is maintained by means of a turbomolecular pump. In this chamber most of the gases entering in the MS (mainly solvent molecules) are eliminated, a minor part enters in a second chamber where the analyzer is placed. A second turbo molecular pump evacuates this chamber promptly and the pressure can be controlled by changing the aperture between both chambers. Depending on the type of detector used (see below) pressures in the range 10 4-10 5 mbar, (for Faraday Collector, FC), or 10 7-10 9 mbar (for Secondary Electrton Multiplier, SEM) may be established.

See other pages where Mass spectrometer pumping system is mentioned: [Pg.376]    [Pg.1470]    [Pg.976]    [Pg.376]    [Pg.1470]    [Pg.976]    [Pg.992]    [Pg.370]    [Pg.488]    [Pg.290]    [Pg.90]    [Pg.37]    [Pg.752]    [Pg.1910]    [Pg.149]    [Pg.327]    [Pg.332]    [Pg.754]    [Pg.255]    [Pg.341]    [Pg.345]    [Pg.354]    [Pg.323]    [Pg.30]    [Pg.157]    [Pg.162]    [Pg.827]    [Pg.1185]    [Pg.631]    [Pg.998]    [Pg.26]    [Pg.228]    [Pg.129]    [Pg.132]    [Pg.70]   
See also in sourсe #XX -- [ Pg.20 , Pg.22 ]




SEARCH



Pump systems

Pump, pumping system

Pumping systems

Spectrometer systems

© 2024 chempedia.info