Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Low quality control

The calculated accuracies ranged from 86.2-122.8 % for the low quality controls and from 88-106.8 for the MQC and finally from 93.6 to 116.5 % for the high concentration quality control samples (HQC). [Pg.617]

Recommended analysis and acceptance criteria apply to each sample concentration HQC = high-quality control sample LLOQ — lower limit of quantification sample LQC = low-quality control sample MQC — middle quality control sample ULOQ — upper limit of quantification sample) %CV — percent coefficient of variation %RE — percent relative error [4]... [Pg.97]

Leaks caused by low quality control of welds after installation were detected immediately after putting into operation of failed system element. [Pg.126]

Matrix Effect and Recovery For LC-MS/ MS-based methods, the signal suppression or enhancement of the analyte due to the presence of the matrix interferences (matrix effects) in MS/MS detection should be evaluated by comparing the response (peak area) of the analyte and the IS from the extracted blank samples post-fortified with the analyte and the IS with the response of neat solutions with both the analyte and the IS at the same concentrations as above. Matrix effects should be evaluated in one pooled batch of animal matrix or in at least three different batches of human matrix, using three replicates at a minimum of three QC concentrations (e.g., low quality control [LQC], medium quality control [MQC], high quality control [HQC]) with IS at working concentration. The coefficient of variation (CV%) of the matrix effect variability should be <15% at each concentration level and between the three (LQC, MQC, and HQC) concentration levels. [Pg.177]

Low Quality Control (LQC) QC samples prepared approximately three times the concentration of the LLOQ. [Pg.187]

Quality Control. The spectrometer is the most suitable instmment for determining most low level residual impurities. ASTM E414 is the standard method for the measurement of impurities in copper by the briquette dc-arc technique (65). In this method, the sample in the form of chips, drillings, or powder is briquetted and excited in a d-c arc opposite a high purity copper rod. Impurities in the ranges noted can be measured ... [Pg.211]

The method was validated in accordance to the guidelines of the international conference on harmonization (ICH). Data with respect to accuracy, within- and between run precision, recovery, detection and quantitation limits were reported and found to be within the accepted international criteria. Neither endogeneous substances nor the commonly used dmgs were found to interfere with the retention times of the analytes. Standard solutions of the dmg and quality control preparations at high and low level concentrations were demonstrated to be stable at room temperature and/or -20°C for long and short periods of time. [Pg.395]

Shafts are made of material ranging from medium carbon to low alloy steel and are usually heat treated. Shafts were originally made of forgings for the compressors in process service. But because of the availability ot high quality material, hot rolled bar stock has been used for shafts up to S inches in diameter. Bar stock shafts are given the same heat treatment and quality control as forgings. Many of the process users prefer a low alloy, chrome-moly-nickel material for shafting, particularly for compressors in critical service. [Pg.197]

Whilst nothing can improve upon the disadvantage of low molar absorption coefficients, instrumental designs and improvements with ratio recording and FT-IR instruments have virtually overcome the accuracy and instrumental limitations referred to in (b) and (c) above. As a result, quantitative infrared procedures are now much more widely used and are frequently applied in quality control and materials investigations. Applications fall into several distinct groups ... [Pg.752]

Most packaged foods require a barrier against gases, flavors, or odors to maintain product quality and provide acceptable shelf life. Baked foods usually need moisture protection, while fresh meats and vegetables require low or controlled exposure to oxygen... [Pg.238]

This test shows the ability of a material to resist the action of an arc of high voltage and low current close to the surface of the insulation in tending to form a conducting path therein. The arc resistance data are of relative value only for distinguishing materials of nearly identical composition, such as for quality control, development, or identification. [Pg.327]

The Ml test equipment is easy to operate, provides repeatable results, and low cost to operate. It is widely used for quality control and for distinguishing between members of a single family of plastics. Specifically, this MI makes a single-point test that provides information on resistance to flow at only a single shear rate. Because variations in branching or MWD can alter the shape of the viscosity curve, the MI may give a false ranking of plastics in terms of their shear rate resistance to flow. To overcome this problem, extrusion rates are sometimes measured for two loads, or other modifications are made. [Pg.450]

This reasonably low viscosity premix can be prepared in rather large batches which are sampled for quality control of ingredient proportions. [Pg.898]

A quality control laboratory had a certain model of HPLC in operation. One of the products that was routinely run on the instrument contained two compounds, A and B, that were quantitated in one run at the same detector wavelength setting. At an injection volume of 20 /tL, both compounds showed linear response. The relatively low absorption for compound B resulted in an uncertainty that was just tolerable, but an improvement was sought. [Pg.277]

Solution viscosity measurements have sometimes been utilized as qioality control tests for this polymer. Chromatographs of three samples that showed Identical intrinsic viscosities (0.8 g/dl) in toluene are shown in Figure 9. These chromatographs indicate that the identical viscosities are the result of different combinations of high and low MW components. These three polymer samples probably have significantly different physical properties and if viscosity measurments alone are utilized for quality control purposes, they may be quite misleading. [Pg.263]

However, it has been established that an intense control of certain variables may improve the execution of a hydraulic fracturing job and the success of a stimulation. Therefore an intense quality control is recommended [552,553]. Such a program includes monitoring the breaker performance at low temperatures and measuring the sensitivity of fracturing fluids to variations in crosslinker loading, temperature stabilizers, and other additives at higher temperatures. [Pg.238]

When developing or routinely using an analytical method, quality control (QC) fortifications can be added to each sample at critical points in the procedure to ensure that sensitive steps in the method were conducted properly and to pinpoint where problems occurred if results are less than satisfactory. For example, if the QC fortification samples for detection and cleanup were to show acceptable results in a batch of samples, but the extraction QC spike gave low recovery and/or high variability, then the analyst could modify instrument conditions or altering cleanup parameters immediately. Likewise, if the QC spike added just before analysis gives poor results, then instrument maintenance could be done and the samples merely re-analyzed rather than re-extracted. [Pg.754]

Cortes et al. [18] have quantitatively determined polymer additives in a polycarbonate homopolymer and an ABS terpolymer. In that case, a multidimensional system consisting of a microcolumn SEC was coupled on-line to either capillary GC or a conventional LC system. The results show losses of certain additives when using the conventional precipitation approach. An at-column GC procedure has been developed for rapid determination (27 min) of high-MW additives (ca. 1200Da) at low concentrations (lOOppm) in 500- xL SEC fractions in DCM for on-line quality control (RSD of 2-7%) [36], Also, SEC-NPLC has been used for the analysis of additives in dissolution of polymeric... [Pg.695]

Workplace safety has been taken care of by the reworking of some classes of additives into more environmentally acceptable forms. Some trends are the increased use of additive concentrates or masterbatches and the replacement of powder versions by uniform pellets or pastilles which release less dust and flow more easily. Moreover, the current move to multicomponent formulations of stabilisers and processing aids in a low- or nondusting product also takes away the risk of operator error, aids quality control, ISO protocols and good housekeeping. An additional benefit is more homogeneous incorporation of the additives in the polymeric matrix. [Pg.725]


See other pages where Low quality control is mentioned: [Pg.617]    [Pg.403]    [Pg.617]    [Pg.403]    [Pg.1642]    [Pg.153]    [Pg.258]    [Pg.202]    [Pg.954]    [Pg.43]    [Pg.220]    [Pg.264]    [Pg.512]    [Pg.164]    [Pg.538]    [Pg.302]    [Pg.318]    [Pg.168]    [Pg.287]    [Pg.87]    [Pg.308]    [Pg.429]    [Pg.207]    [Pg.39]    [Pg.50]    [Pg.186]    [Pg.2]    [Pg.434]    [Pg.473]    [Pg.51]    [Pg.57]    [Pg.323]    [Pg.430]    [Pg.237]   
See also in sourсe #XX -- [ Pg.177 , Pg.187 ]




SEARCH



Low-Field NMR Magnetisation Relaxation Experiments for Quality Control Purposes

© 2024 chempedia.info